| | Class 3 Device Recall |  |
| Date Initiated by Firm | January 27, 2006 |
| Date Posted | February 28, 2006 |
| Recall Status1 |
Terminated 3 on October 19, 2006 |
| Recall Number | Z-0577-06 |
| Recall Event ID |
34522 |
| 510(K)Number | K022746 |
| Product Classification |
Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph - Product Code CHL
|
| Product | Stat Profile Critical Care PLUS (CCX) Analyzer
Catalog Number: 37413 |
| Code Information |
All S/N''s |
| FEI Number |
1219029
|
Recalling Firm/ Manufacturer |
Nova Biomedical Corporation 200 Prospect St Waltham MA 02453-3457
|
| For Additional Information Contact | Paul W. MacDonald 781-894-0800 |
Manufacturer Reason for Recall | Analyzers may fail to perform scheduled automatic two-point calibrations and Quality Control runs |
FDA Determined Cause 2 | Other |
| Action | Nova Biomedical Issued on 1/27/2006 an Advisory Notice to Stat Profile CCX Customers to report the malfunction and request the Good Laboratory Practices (GLP) be used and review be made to assure the calibration status and QC is being performed. A new software version is scheduled to be released to correct the error. |
| Quantity in Commerce | 408 units |
| Distribution | Nationwide
Foreign: Canada, UAE, Taiwan, Singapore, Budapest, UK, France, Malaysia, France, Mexico, Korea, Finland, Germany, Turkey, Ireland, Italy, Austria, Australia, Netherlands, Venezuela, Japan, Greece, Thailand, Israel, Spain, China, Hong Kong, Argentina, Czech Republic, Croatia,Trinidad WI |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CHL
|
|
|
|