Date Initiated by Firm |
November 03, 2005 |
Date Posted |
May 09, 2006 |
Recall Status1 |
Terminated 3 on June 20, 2006 |
Recall Number |
Z-0855-06 |
Recall Event ID |
34776 |
510(K)Number |
K050810
|
Product Classification |
computed radiography system digitizer - Product Code MQB
|
Product |
CR DX-S, DX-S, Computed radiography system (Digitizer) |
Code Information |
CR DX-S: Units with Serial numbers SN1001-SN1079, Software versions: STR1102B and below. |
Recalling Firm/ Manufacturer |
AGFA Corp. 10 S Academy St Greenville SC 29601-2632
|
For Additional Information Contact |
Ray B. Myers Ph.D 864-421-1827
|
Manufacturer Reason for Recall |
Corrupted image after system start; Corrupted image after workflow interruption; An erasure unit problem can cause system to stop.
|
FDA Determined Cause 2 |
Other |
Action |
Consignees were notified at the time of upgrading. The software was upgraded on 11/3/2005, 12/19/2005 and 12/28/2005. A Mandatory Service Bulletin, No. 11, explaining the issues was shared with the accounts. |
Quantity in Commerce |
3 units |
Distribution |
DC, MS, OR--three states |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MQB and Original Applicant = AGFA CORP.
|