| | Class 2 Device Recall Theratron Phoenix |  |
| Date Initiated by Firm | July 08, 2004 |
| Date Posted | June 27, 2006 |
| Recall Status1 |
Terminated 3 on June 27, 2006 |
| Recall Number | Z-1182-06 |
| Recall Event ID |
35617 |
| 510(K)Number | K863180 |
| Product Classification |
System, Radiation Therapy, Radionuclide - Product Code IWB
|
| Product | Theratron Phoenix Cobalt Radiotherapy |
| Code Information |
Model Theratron Phoenix |
Recalling Firm/ Manufacturer |
MDS Canada Inc. DBA MDS Nordion 447 March Road Kanata Canada
|
| For Additional Information Contact | Ross Kachaniwsky 613-592-3400 Ext. 2234 |
Manufacturer Reason for Recall | The source locking device may not fit properly through the smaller opening in the new field light shield which could prevent the source from being returned manually under emergency conditions. |
FDA Determined Cause 2 | Other |
| Action | A User Bulletin (CUB-2004-01dated 7/5/04) was issued to affected users on 8/24/04 and 8/31/04. The bulletin provided 2 options for corrective action. |
| Quantity in Commerce | 4 units |
| Distribution | The products were shipped to medical facilities in FL and KS. The product was also shipped to Kuwait, Spain, Honduras, Romania, France, Grrece, Bangladesh, Germany, Nepal, Canada, England, Thailand, Phillippines, Mexico, Columbia, India, Switzerland, Syria, Trinidad & Tobago, Ireland, Turkey, Algeria, Portugal, Poland, Iran, China, Cuba, Belgium, Hungary, Egypt, Italy, Vietnam, Sri Lanka, and Costa Rica. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = IWB
|
|
|
|