 Class 2 Device Recall Bardsee related information
Date Initiated by FirmJune 27, 2006
Date PostedNovember 02, 2006
Recall Status1 Terminated 3 on April 14, 2007
Recall NumberZ-0118-2007
Recall Event ID 35827
510(K)NumberK043489 
Product Classification guide catheter delivery sheath - Product Code DQY
ProductChannel Steerable Sheath in 8F Enpath part # 10775-003, BARD Part Number XD10775003 (US Distribution Only) and 9F Enpath part # 10775-004 (foreign distribution), BARD Part Number XD10775004; Labeled as Distributed by Bard Electrophysiology Division.
Code Information Model # 8F: Enpath Part Number 10775-003, BARD Part Number XD10775003 (Distributed in US and Internationally). 8F Lots #s W34968 (80 units), W35440 (35 units), W35522 (40 units), W35705 (17 units, sent as SAMPLES, labeled NOT FOR HUMAN USE), W35719 (320 units), W35952 (79 units), W37292 (300 units), W37567 (116 units), W38340 (298 units), W38341 (87 units), W38342 (87 units), W38974 (99 units), W38975 (100 units), W39174 (100 units), W40410 (100 units), W40479 (87 units), W40480 (85 units), W40481 (100 units), W40483 (87 units).  Model # 9F: Enpath Part Number 10775-004 (9F), BARD Part Number XD10775004, (Distributed Internationally (EU) only), 9F lot #s; W36392 (3 units), W36623 (5 units), W36884 (2 units), W36974 (8 units).
FEI Number 2183787
Recalling Firm/
Manufacturer
Enpath Medical, Inc.
15301 Highway 55
Plymouth MN 55447-1418
Manufacturer Reason
for Recall
Enpath has become aware that some of the Enpath Medical 8F & 9F Steerable sheath devices may have a non-conforming flushport bond that causes a leak where air may enter the flush port chamber.
FDA Determined
Cause 2
Other
ActionEnpath''s three Distributors were notified via letter or telephone on June 27th, 29th and July 6th with a description of the problem and further distribution or use of any remaining product should cease immediately. Instructed to notify their customers about the product and forward to Enpath for evaluation.
Quantity in Commerce2494 Units
DistributionCA, MA
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = DQY
-
-