 Class 2 Device Recall Allurasee related information
Date Initiated by FirmOctober 12, 2006
Date PostedNovember 14, 2006
Recall Status1 Terminated 3 on September 24, 2010
Recall NumberZ-0165-2007
Recall Event ID 36573
510(K)NumberK031333 
Product Classification x-ray and fluoroscopy - Product Code IZI
ProductAllura X-per FD 10/10 x-ray, and fluoroscopy
Code Information 127, 113, 117, 112, 131, 128, 90, 17, 35, 137, 93, 51, 95, 27, 6, 47, 98, 22, 49, 39, 103, 120, 21, 40, 70, 101, 69, 94, 89, 23, 33, 31, 30, 29, 28, 32, 100, 63, 61 Additional units are identified with site numbers: 533397, 531926, 104950, 104953, 103283, 103281, 530653, 104497
FEI Number 1000524572
Recalling Firm/
Manufacturer
Philips Medical Systems North America Co. Phillips
22100 Bothell Everett Hwy
Bothell WA 98021-8431
For Additional Information ContactSarah Baxter
425-487-7665
Manufacturer Reason
for Recall
Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.
FDA Determined
Cause 2
Other
ActionOn 10/12/06 the firm issued a letter to their consignees advising of the problem and how to avoid the problem by using specific image formats. If the system should lock up, consignees are to switch the system off and then restart. The firm will be conducting software upgrades to the devices once software is ready (estimated time for FD 10/10 is November 2006 and estimated time for FD 20/10 is January 2007). Later, the firm identified 8 additional units and issued a letter on 12/11/06 to affected customers.
Quantity in Commerce47 devices
DistributionNationwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = IZI
-
-