| Code Information |
Serial numbers affected: 17, 21, 22, 39, 44, 48, 49, 52, 54, 60, 81, 83, 93, 107, 111, 118, 121, 122, 135, 141, 144, 145, 153, 159, 160, 163, 165, 166, 167, 168, 170, 175, 176, 178, 179, 183, 184, 186, 187, 190, 191, 195, 196, 202, 203, 207, 211, 212, 213, 214, 220, 222, 225, 230, 231, 232, 234, 239, 246, 248, 249, 250, 251, 252, 254, 255, 262, 263, 265, 269, 270, 272, 277, 278, 283, 285, 286, 287, 289, 294, 301, 302, 303, 304, 306, 308, 310, 314, 315, 318, 321, 326, 327, 332, 337, 340, 341, 343, 347, 361, 362, 366, 369, 376, 379, 384, 387, 388, 389, 396, 404, 423, 425, 426, 435, 439, 445, 458, 467, 468, 469, 470, 471, 472, 479, 490, 494, 498, 504, 512, 514, 515, 518, 520, 521, 523, 528, 532, 534, 535, 543, 545, 553, 554, 559, 560, 579, 582, 586, 587, 588, 591, 592, 593, 595, 596, 599, 601, 602, 604, 606, 609, 611, 612, 615, 616, 618, 622, 623, 625, 626, 627, 628, 632, 634, 637, 638, 641, 642, 647, 649, 651, 654, 655, 658, 661, 663, 664, 667, 668, 670, 671, 672, 673, 674, 675, 676, 677, 678, 679, 680, 681, 683, 685, 686, 689, 691, 692, 694, 700, 701, 702, 705, 706, 707, 709, 716, 718, 722, 723, 730, 731, 737, 738, 741, 742, 744, 746, 750, 751, 752, 753, 754, 759, 765, 766, 770, 771, 773, 774, 775, 776, 777, 778, 779, 780, 781, 782, 785, 786, 788, 789, 790, 796, 797, 800, 804, 806, 808, 810, 813, 816, 821, 822, 823, 825, 826, 827, 828, 829, 830, 840, 841, 842, 843, 844, 845, 847, 848, 849, 850, 851, 852, 853, 857, 860, 861, 862, 863, 878, 882, 882, 886, 888, 892, 893, 894, 895, 896, 903, 904, 905, 906, 907, 908, 910, 911, 912, 913, 914, 919, 920, 921, 923, 924, 927, 929, 930, 933, 934, 939, 942, 943, 953, 959, 960, 961, 962, 966, 968, 970, 971, 972, 976, 984, 991, 992, 993, 994, 1007, 1009, 1015, 1032, 1038, 1044, 1045, 1046, 1056, 1075, 1076, 1081, 1085, 1090, 1092, 1093, 1095, 1096, 1098, 1103, 1104, 1105, 1107, 1108, 1114, 1115, 1116, 1118, 1119, 1124, 1126, 1127, 1128, 1132, 1138, 1139, 1140, 1141, 1145, 1147, 1150, 1153, 1161, 1166, 1167, 1168, 1169, 1170, 1174, 1176, 1177, 1178, 1179, 1183, 1187, 1188, 1189, 1190, 1195, 1196, 1201, 1202, 1205, 1206, 1207, 1209, 1211, 1221, 1223, 1227, 1228, 1229, 1235, 1236, 1237, 1240, 1241, 1242, 1252, 1253, 1256, 1258, 1259, 1267, 1268, 1269, 1273, 1274, 1276, 1277, 1279, 1291, 1293, 1294, 1297, 1298, 1300, 1302, 1318, 1321, 1326, 1331, 1338, 1364, 1374, 18165, 43041, 43048, 43051, 43057, 43068, 43070, 43076, 43080, 43081, 43090, 43093, 1674122, 521, 9A, and CG22N41. |