• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Impulse Angiographic Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Impulse Angiographic Cathetersee related information
Date Initiated by FirmNovember 20, 2007
Date PostedJanuary 31, 2008
Recall Status1 Terminated 3 on October 23, 2008
Recall NumberZ-0935-2008
Recall Event ID 45903
510(K)NumberK000956 K964859 K974559 K992123 
Product Classification Intravascular Diagnostic Catheter - Product Code DQO
ProductBoston Scientific, Impulse Angiographic Catheter, Sterile EO. Catheters are single lumen catheters with a single proximal hub for attachment to a dye delivery system. Catalog No. 16599-301. 6F MULTI PIT 124 (Single Pack). Distal Curve FL4, FR4, PIG 145. 100 and 110 cm Length. Made in Mexico: Parque industrial Le Mesa, Tijuana, Baja California 22050 MEX. Boston Scientific Corporation, Natick, MA 01760-1537 USA.
Code Information Catalog No./outer box Material No. (UPN) /inner pouch UPN /Lot (Batch) No.  16599-301/H749165993010/H749165993010/0020032538 and  16599-301/H749165993010/H749165993010/002003517
Recalling Firm/
Manufacturer
Boston Scientific
1 Scimed Pl
Maple Grove MN 55311-1565
For Additional Information Contact
763-494-1700
Manufacturer Reason
for Recall
Flash may be protruding from the lumen of the catheter shaft. If this flash is present in the catheter, there is the potential of an embolization resulting from the flash detaching from the device during the procedure.
FDA Determined
Cause 2
Release of Material/Component prior to receiving test results
ActionThe firm sent an Urgent Medical Device Recall letter dated 11/20/2007, to all affected hospitals. The letter describes the defect with the catheters and Product Table information including UPNs, Catalog Numbers, Lot/Batch Numbers and Product Descriptions. The firm requests that further distribution or use of any remaining product affected by the recall should cease immediately. ****IMPORTANT INFORMATION: If it is a practice of the hospital to remove product from the outer carton and store on shelves in the inner-pouch only. Users are to review the Product Table carefully and consider both the inner and outer packaging product codes when search for affected/recalled product.****
Quantity in Commerce211 single units
DistributionNationwide Distribution including states of AL, CA, FL, GA, KS, KY, LA, MN, NV, OH, OK, PA, SD, TN, TX, and WI.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DQO
510(K)s with Product Code = DQO
-
-