Date Initiated by Firm | December 05, 2007 |
Date Posted | July 02, 2008 |
Recall Status1 |
Terminated 3 on January 11, 2012 |
Recall Number | Z-1256-2008 |
Recall Event ID |
46679 |
510(K)Number | K053404 |
Product Classification |
Biliary catheter and accessories - Product Code FGE
|
Product | Edwards LifeStent FlexStar Biliary Stent System; Biliary catheter and accessories; Model Number: EX090603CD; Edwards Lifesciences LLC, Irvine, CA 92614. |
Code Information |
Lot Number: FR6H0173 (Exp 2008/08). |
Recalling Firm/ Manufacturer |
Edwards Lifesciences, Llc 1 Edwards Way Irvine CA 92614-5688
|
For Additional Information Contact | Walter Wiegand 949-250-2443 |
Manufacturer Reason for Recall | Incorrect Expiration Date: Products were labeled with an incorrect shelf-life expiration date. Therefore, the sterility of the device may be compromised. |
FDA Determined Cause 2 | Packaging process control |
Action | Consignees were notified by an Urgent Product Recall letter sent on 12/5/07. The letter instructed users to return any affected product to the recalling firm for replacement products. The letter also advised users that the likelihood of patient complications is extremely low for products that have already been implanted in patients. Users were requested to return a confirmation form after checking the inventory for any affected products. For additional information, contact 1-949-250-3779. |
Quantity in Commerce | Lot # FR6H0173 - 1 UNIT |
Distribution | Nationwide Distribution |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = FGE
|