|
Class 2 Device Recall Baxa ExactaMix 600 Pharmacy compounding system |
 |
Date Initiated by Firm |
February 11, 2008 |
Date Posted |
August 19, 2008 |
Recall Status1 |
Terminated 3 on March 16, 2012 |
Recall Number |
Z-1371-2008 |
Recall Event ID |
46771 |
Product Classification |
Automated pharmacy compounding system - Product Code NEP
|
Product |
Exacta-Mix 600 Pharmacy Compounding System is used in the pharmacy to compound multiple source ingredients into on final solution based on user-specified sequences and volumes. |
Code Information |
Model 071B-RS232, all serial numbers |
Recalling Firm/ Manufacturer |
Baxa Corporation 14445 Grasslands Dr Englewood CO 80112-7062
|
For Additional Information Contact |
303-617-2242
|
Manufacturer Reason for Recall |
Flush solution from TPN compounder may be added to patient's TPN bag.
|
FDA Determined Cause 2 |
Software design |
Action |
Consignees were notified by letter (Exacta-Mix Compounder Software Safety Alert) on 02/11/2008 advising them of the necessity to install a waste bag prior to a machine flush. |
Quantity in Commerce |
221 units |
Distribution |
Nationwide, including military facilities in CA, FL, HI, MS, NC, TX, VA and VA facilities in NY, OR, SC, UT. Foreign distribution to Australia, Canada, China and New Zealand. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|