 Class 2 Device Recall Zimmer Surgivac Wound Drainage Devicesee related information
Date Initiated by FirmJanuary 29, 2008
Date PostedSeptember 19, 2008
Recall Status1 Terminated 3 on November 12, 2009
Recall NumberZ-1504-2008
Recall Event ID 47233
Product Classification non-powered, portable suction appartus for a single patient - Product Code GCY
ProductSurgivac Wound Drainage Device, 400 ml Surgivac Component Kit, sterile, latex free, Rx only; Zimmer, Swindon, Wiltshire, United Kingdom; Zimmer, Dover, Ohio; REF 00-5300-004-10.
Code Information All lots manufactured on or before 1/18/08. Product has a 5 year expiration period.
FEI Number 1000220733
Recalling Firm/
Manufacturer
Zimmer Inc.
345 E Main St
Warsaw IN 46580-2746
For Additional Information Contact
330-365-3793
Manufacturer Reason
for Recall
The sterility of the device may be compromised due to the possible presence of slits or pinholes in the packaging.
FDA Determined
Cause 2
Packaging process control
ActionConsignees were notified via recall letter dated 1/29/08 to locate and return the affected product. The recall action was expanded to additional products/lots via recall letter dated 3/17/08. Custom kit packers were instructed to notify their customers.
DistributionNationwide, Australia, Canada, Chile, China, Czech Republic, Dubai, Ecuador, France, Germany, Greece, Hungary, Italy, Japan, Korea, Poland, Romania, Russia, Saudi Arabia, Serbia & Montenegro, Singapore, Slovakia, South Africa, Spain, and Switzerland.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-