| | Class 2 Device Recall 740 Series Multiparameter Monitor |  |
| Date Initiated by Firm | June 24, 2008 |
| Date Posted | September 16, 2008 |
| Recall Status1 |
Terminated 3 on August 05, 2010 |
| Recall Number | Z-2203-2008 |
| Recall Event ID |
48666 |
| 510(K)Number | K033048 |
| Product Classification |
Patient Physiological Monitor with Arrhythmia Alarms - Product Code MHX
|
| Product | 740 Series Multiparameter Monitor Software Version 3.0
Model Numbers: 740-1, 740M-1, 740-2T, 740M-2T, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS, 740-3NL, 740M-3NL, 740-3NN, and 740M-3NN
CAS Medical Systems, Inc., Branford, Connecticut 06405
Indicated for use for non-invasive monitoring of blood pressure, oxygen saturation, pulse and temperature of adults, pediatric and neonatal patients, in the care of health professionals. |
| Code Information |
Serial Numbers: 0809834-0823035 |
| FEI Number |
2244861
|
Recalling Firm/ Manufacturer |
CAS Medical Systems, Inc. 44 E Industrial Rd Branford CT 06405-6507
|
| For Additional Information Contact | SAME 203-488-6056 |
Manufacturer Reason for Recall | Software: Audible alarm may be silenced if changed from the factory setting |
FDA Determined Cause 2 | Software change control |
| Action | CASMED notified customers by a Product Correction Notification letter on 6/24/08 to advise of problem of the audible alarm and a software upgrade to correct the problem. The firm requests that the alarm volume not be adjusted from its factory set value to ensure that alarms sound correctly. Also, the letter states that if the volume needs to be adjusted, that customers needs to be sure that they only select two clicks above, or two clicks below the mid-point, and that there is a corresponding beep when you make your selection. Customers are asked to review the attached list of serial numbers that will need to be updated with new software and to then email the a contact at the firm once updates have been completed. All updates must be completed by August 31, 2008. |
| Quantity in Commerce | 820 units |
| Distribution | Worldwide Distribution --- including USA and countries of Canada, Jordan, UK, France, and Taipei. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MHX
|
|
|
|