| | Class 2 Device Recall Zimmer NEXGEN Tibial Broach Impactor |  |
| Date Initiated by Firm | June 30, 2008 |
| Date Posted | September 20, 2008 |
| Recall Status1 |
Terminated 3 on November 12, 2009 |
| Recall Number | Z-2301-2008 |
| Recall Event ID |
48717 |
| Product Classification |
Broach Impactor - Product Code HWA
|
| Product | Zimmer NEXGEN Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor (a pounding instrument used during knee surgery), nonsterile, Zimmer, Warsaw, Indiana; REF 5951-90.. |
| Code Information |
All units. |
Recalling Firm/ Manufacturer |
Zimmer Inc. 345 E Main St Warsaw IN 46580-2746
|
| For Additional Information Contact | 574-372-4487 |
Manufacturer Reason for Recall | The instrument may fracture during use, resulting in metal fragments being left in the patient post-surgery, which could cause implant failure. |
FDA Determined Cause 2 | Device Design |
| Action | Consignees were notified via letter dated 6/30/08. The letter described the issue, provided instructions for the return of the product and asked customers to complete a reply back form. For assistance, contact Zimmer at 1-574-372-4487. |
| Quantity in Commerce | 1439 |
| Distribution | Nationwide, Australia, China, England, France, Germany, India, Italy, Japan, Korea, Singapore and Taiwan. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|