| | Class 3 Device Recall Micro Plate EIA Oral Fluid Negative Control |  |
| Date Initiated by Firm | September 08, 2008 |
| Date Posted | December 23, 2008 |
| Recall Status1 |
Terminated 3 on May 05, 2009 |
| Recall Number | Z-0478-2009 |
| Recall Event ID |
49557 |
| 510(K)Number | K002375 |
| Product Classification |
Cannabinoid Test System - Product Code LDJ
|
| Product | Micro Plate EIA Oral Fluid Negative Control which is a component of the Intercept Micro Plate EIA kit. |
| Code Information |
Product Number 61452. Lot numbers 6604827, 6604834, 6606583, 5506589, and 6606034 |
| FEI Number |
2528909
|
Recalling Firm/ Manufacturer |
OraSure Technologies, Inc. 1745 Eaton Ave Bethlehem PA 18018-1769
|
| For Additional Information Contact | Jackie Ferro 610-882-1820 |
Manufacturer Reason for Recall | Results, false-positive test: low % displacement result of 36.0% (spec 39-60%). For further information, please contact the firm at 610-882-1820. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | The recalling firm issued a letter entitled "URGENT PRODUCT RECALL NOTICE" dated 9/8/08 to their customers to inform them of the problem and the need to return the product. |
| Quantity in Commerce | 2198 units |
| Distribution | Worldwide Distribution --- USA including states of AZ, CA, CT, DE, FL, GA, ID, KS, KY, MA, MD, MI, NC, NH, NV, NY, PA, VA, and WY, and countries of United Kingdom, Italy, Germany, and Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|