| | Class 2 Device Recall Fisher & Paykel Healthcare RT240 Adult Breathing Circuit Kit. |  |
| Date Initiated by Firm | October 22, 2008 |
| Date Posted | December 10, 2008 |
| Recall Status1 |
Terminated 3 on March 26, 2012 |
| Recall Number | Z-0414-2009 |
| Recall Event ID |
50105 |
| 510(K)Number | K983112 |
| Product Classification |
Respiratory Gas Humidifier (Direct Patient Interface). - Product Code BTT
|
| Product | Fisher & Paykel Healthcare RT240 Adult Breathing Circuit Kit.
lntended Use: The dual heated adult respiratory ventilator circuits are intended as conduits of breathing gas for ventilation of patients, and to maintain the temperature of the humidified gas, to reduce condensation. |
| Code Information |
Lot Number: All lot numbers up to and including Lot 080627. |
Recalling Firm/ Manufacturer |
Fisher & Paykel Healthcare Inc 5900 Skylab Rd Huntington Beach CA 92647-2061
|
| For Additional Information Contact | 949-453-4000 |
Manufacturer Reason for Recall | The recall was initiated because certain lot dates of the RT240 Adult Breathing Circuit Kits manufactured on or before June 27, 2008 include a heated breathing circuit that may be more susceptible to damage when used in excess of the specified seven-day maximum duration of use, which may increase the risk of malfunction or fire. |
FDA Determined Cause 2 | Process change control |
| Action | Fisher & Paykel Healthcare, Inc. (FPH) issued an URGENT-PRODUCT RECALL letter (dated October 8, 2008) including a Customer recall letter for consignees/customers to Cardinal Health. Cardinal Health will, in turn, send the Customer recall letter dated October 10, 2008 to its consignees/customers. Fisher & Paykel Healthcare (FPH) also issued a similar set of recall notification letters to Norco on October 15, 2008.
Recipients of the recall letter must perform the following:
1) Check all Cardinal Health stock of the product for the affected lots, 2) Immediately quarantine the affected product, 3) Complete a Device Recall Response Sheet, and 4) Fax or e-mail the completed Device Recall Response Sheet.
Fisher & Paykel Healthcare, Inc.'s "Dear Customer Letter" for Cardinal Health and Norco instructs consignees to complete a customer response (i.e. fax enclosed Customer Acknowledgement Form) to FPH. The customer letter also provides instructions for product return.
Fisher & Paykel Healthcare, Inc. recall communication instructions to Cardinal Health and Norco was for all affected product to be returned and quarantined by the distributors. The distributors will return any affected stock to Fisher & Paykel Healthcare, Inc. for documented destruction. |
| Quantity in Commerce | 229,320 units |
| Distribution | Worldwide Distribution including US and country of Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = BTT
|
|
|
|