|
Class 2 Device Recall Integris BH5000 |
|
Date Initiated by Firm |
October 23, 2008 |
Date Posted |
September 22, 2010 |
Recall Status1 |
Terminated 3 on August 10, 2011 |
Recall Number |
Z-2473-2010 |
Recall Event ID |
50114 |
510(K)Number |
K984545
|
Product Classification |
System, x-ray, angiographic - Product Code IZI
|
Product |
Philips Integris BH5000 Angiographic X-ray system. This system with Velara Generator can be identified as type number 9890 000 6501 * (number located on the generator cabinet). Intended use for Angiographic X-ray system with and without a solid state X-ray imaging device for cardiovascular and vascular diagnostic and interventional procedures. |
Code Information |
Devices are identified with Site Numbers: 103581, 105399, 104386, 105184, 105585, 104021, 105581, 505120, 104020, 504733, 505067, 104068, 504868, 504921, 504674, 505068, 504903, 104628, 105654, 505530, 505703, 505380, 504655, 104926, 504779, 506026, 505611, 506069, 506119, 504742, 104497, 506008, 505903, 506655, 103921, 531926, 519136, 505356, 104950, 519118, 103526, 104953, 104953, 519730, 520949, 520275, 104959, 76992, 520582, 519981, 521525, 519196, 521325, 521149, 522456, 506509, 520688, 44147585, 530653, 531939, 530654, 522136, 520666, 530660, 520964, 532383, 531649, 531799, 532218, 532152, 520756, 533290, 532240, 530396, 532241, 534131, 531758, 534786, 535761, 535775, 519197, 548140, 537175, 533288, 535345, 537593, 538172, and 534156. |
Recalling Firm/ Manufacturer |
Philips Medical Systems North America Co. Phillips 22100 Bothell Everett Hwy Bothell WA 98021-8431
|
For Additional Information Contact |
Philips Call Center 800-722-9377
|
Manufacturer Reason for Recall |
A high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
|
FDA Determined Cause 2 |
Other |
Action |
The firm, Philips, sent "URGENT-Device Correction" letters dated November 10, 2008, to their customers and informed them about the reliability issue in the Allura XPER and Integris X-ray systems with the High Voltage Velera (HV) Generator. This reliability issue can lead to a system error causing intermittent loss of fluro and/or exposure acquisition, potentially resulting in the loss of the X-ray image on the display.
The customers are advised if they experience High Voltage Generator reliability issue and the HV Generator stops working generating an O3EA error, a cold restart is required and reboot the system when the X-Ray image disappears.
The firm will implement multiple mandatory Field Change Orders (FCOs) to solve the Generator reliability by updating the software for each X-ray system.
The customers could contact Philips Care Center at 800-722-9377, #5, #4, #1 for any questions regarding this recall or 425-487-7469. |
Quantity in Commerce |
88 units in US |
Distribution |
Worldwide distribution. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IZI and Original Applicant = PHILIPS MEDICAL SYSTEMS NORTH AMERICA, INC.
|
|
|
|