Class 2 Device Recall Low Profile U Joint Driver for the SynFix Mini Open System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Low Profile U Joint Driver for the SynFix Mini Open Systemsee related information
Date Initiated by FirmNovember 20, 2008
Date PostedJanuary 30, 2009
Recall Status1 Terminated 3 on October 27, 2009
Recall NumberZ-0931-2009
Recall Event ID 50523
510(K)NumberK072253 
Product Classification Lumbar Intervertebral Fusion Device with Bone Graft - Product Code MAX
ProductLow Profile U-Joint Driver for the SynFix Mini-Open System, Catalog number 03.802.231. The Synthes SynFix LR is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). The interior of the spacer component of the SynFix LR can be packed with autograft.
Code Information Lot numbers: 2317363, 2331806, 2331810, 2387975, 2387976, 2401982, 5637286, 5644605, 5660377, 5664629, 5826243, and 5826244.
FEI Number 3005180112
Recalling Firm/
Manufacturer
Synthes Spine
1302 Wrights Ln E
West Chester PA 19380-3417
For Additional Information ContactLynne Lewis BSN RN
610-719-5000
Manufacturer Reason
for Recall
screwdriver tip may break off and lodge in the screw head
FDA Determined
Cause 2
Device Design
ActionThe recalling firm emailed all Spine Sales Consultants on 11/20/08. They were instructed to immediately examine their stock and discuss appropriate technique with surgeon users. In addition, a Medical Device Correction letter dated 11/19/08 was issued via USPS certified mail to the medical facilities. This initial letter stated the issue and requested the users take the following actions: discard all SynFix Technique Guides and replace with the updated literature contained in the mailing packet with the letter, and is identified by code J7022-E (Rev. E); to ensure receipt and understanding of the information, the Verification Section of the letter needs to be completed and returned to Synthes by either fax: 610-719-5120 or scan/e-mail: lewis.lynne@synthes.com. An Urgent: Medical Device Recall letter, dated 1/16/2009, was also sent out. This latter letter requested that users examine their stock immediately and remove any of the affected devices from their inventory. Customers are also to call Synthes at 1-800-479-6329 to obtain a Return Authorization Number; complete the Verification Section; and return the letter with the recalled product to the address listed. There are also actions for those who do not have the affected product. Questions should be directed to 800-620-7025 x 5453 or 610-719-5453.
Quantity in Commerce381 units
DistributionClass II Recall - Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA ,SC, TN, TX, UT, VA, WA, and WI, and countries of Canada and Switzerland.
Total Product Life CycleTPLC Device Report