| Date Initiated by Firm | May 14, 2009 |
| Date Posted | September 04, 2009 |
| Recall Status1 |
Terminated 3 on January 20, 2010 |
| Recall Number | Z-1827-2009 |
| Recall Event ID |
52163 |
| 510(K)Number | K041585 K060258 |
| Product Classification |
Infusion pump for pain medication. - Product Code MEB
|
| Product | Stryker 100 mL AutoFuser PainPump, dual site continuous, 2 mL/hr, sterile, REF 0531-200-200. |
| Code Information |
Lots 08142756094 through 09029881575. |
Recalling Firm/ Manufacturer |
Stryker Instruments Div. of Stryker Corporation 4100 E. Milham Ave. Kalamazoo MI 49001
|
| For Additional Information Contact | 800-800-4236 Ext. 3808 |
Manufacturer Reason for Recall | Failure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation. |
FDA Determined Cause 2 | Process control |
| Action | Stryker sent all consignees a Medical Device Recall Notification letter dated 5/14/09 and instructed to locate and destroy the recalled products and to complete and return the business reply card. |
| Quantity in Commerce | 28,755 of all pumps |
| Distribution | Nationwide and Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MEB 510(K)s with Product Code = MEB
|