Class 3 Device Recall Bone Screw

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 3 Device Recall Bone Screwsee related information
Date Initiated by FirmApril 02, 2009
Date PostedSeptember 22, 2009
Recall Status1 Terminated 3 on September 23, 2009
Recall NumberZ-2310-2009
Recall Event ID 51730
510(K)NumberK993956 
Product Classification Spinal intervertebral body fixation - Product Code MQP
ProductBone Screw 6MM (D) x 35MM (L) Telescopic Plate Spacer, Thoracolumbar Spinal System Made in USA, Biomet Spine, Parsippany, NJ 07054 The screw devices are an internal fixation device intended to aid in the alignment and stabilization of fractures to the skeletal system until healing has occurred. Implants used in this application include non sterile and sterile bone screws.
Code Information Catalog number 1300-0635, Packaging lot: 691780; Unpackaged lot: 605339.
FEI Number 2242816
Recalling Firm/
Manufacturer
EBI, L.P.
100 Interpace Pkwy
Parsippany NJ 07054-1149
For Additional Information ContactWilliam Hennig
973-299-9300 Ext. 1557
Manufacturer Reason
for Recall
Some devices could have been made with raw material lots that may have inclusions, and could result in decreased strength of the device.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionThe firm sent Urgent Recall Notices to all direct accounts, hospital Risk Management and Physicians by first class mail. Letters are dated March 29, 2009, April 2, 2009 and April 16, 2009 respectively. The letter describes the problem and that that there is a possible risk of a device failure. An additional e-mail was sent on May 22, 2009 to 5 customers due to an expansion of the recall to include 4 additional items. On June 19, 2009, the firm again expanded the recall to include 2 additional products and notified customers. Questions should be directed to William Hennig, at 1.973.299.9300. ext. 1557.
Quantity in Commerce24
DistributionWorldwide - US -NY, NJ, MI, MD, FL, GA, PR, MN, SD, OK, TX,, LA, CA, PA, OH, NV, HI, ME, NC, KS, UT, CO, TN, WV, VA, and WA. Foreign: Brazil, Colombia, Denmark, Germany, Finland, Greece, Italy, Japan, The Netherlands, Switzerland, Sweden, Portugal, Spain, United Kingdom, Saudi Arabia, and Trinidad.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MQP
-
-