Date Initiated by FirmFebruary 12, 2007
Date PostedAugust 26, 2009
Recall Status1 Terminated 3 on October 15, 2009
Recall NumberZ-1917-2009
Recall Event ID 52641
Product Classification orthopedic manual surgical instrument - Product Code LXH
ProductTriathlon MIS 4:1 Cutting Blocks - Size 1-8; Catalog Numbers: 6541-5-701, 6541-5-702, 6541-5-703, 6541-5-704, 6541-5-705, 6541-5-706, 6541-5-707, & 6541-5-708; Triathlon MIS Instruments; Non Sterile: Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 The Triathlon AR 4:1 Cutting Blocks and the Triathlon MIS 4:1 Cutting Blocks are used to prepare the Femur (bone cuts) for the Triathlon femoral component.
Code Information Catalog numbers: 6541-5-701, 6541-5-702, 6541-5-703, 6541-5-704, 6541-5-705, 6541-5-7-6, 6541-5-707, 6541-5-708. All lot codes.
FEI Number 3003070421
Recalling Firm/
Manufacturer
Stryker Howmedica Osteonics Corp.
325 Corporate Dr
Mahwah NJ 07430
For Additional Information ContactRita Intorella
201-831-5000
Manufacturer Reason
for Recall
Triathlon Cutting Blocks did not assemble to the Triathlon Impactor/Extractor Handle.
FDA Determined
Cause 2
Process control
ActionImportant Product Correction letters were sent to all branches/agencies on February 12, 2008 by Federal Express. The letter asked consignees to examine their inventory and hospital locations to identify the affected product. Customers were asked to perform a functional inspection by assembling the Triathlon Impactor/Extractor Handle to all of their MIS Cutting Blocks. If the cutting block does not assemble, customers are to list the affected lot codes on the Attached Product Accountability Form and fax a copy to (201) 831-6069. The letter states that product should not be returned until further instructions are received. Questions should be directed to (201) 831-5825.
Quantity in Commerce690 - both varieties
DistributionWorldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-