• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 3 Device Recall Stryker Orthopaedics Bone Screw

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 3 Device Recall Stryker Orthopaedics Bone Screw see related information
Date Initiated by Firm June 26, 2009
Date Posted September 14, 2009
Recall Status1 Terminated 3 on July 27, 2010
Recall Number Z-2118-2009
Recall Event ID 52713
510(K)Number K873251  
Product Classification Screw, fixation, bone - Product Code HWC
Product Stryker Orthopaedics Cancellous Bone Screw;
Dia 6.5 mm; Lnth: 20 mm; Sterile
Made in the USA;
Stryker Osteonics Corp.
325 Corporate Drive
Mahwah, NJ 07430.

The bone screws are optional fixtures that may be used during knee and hip replacement procedures.
Code Information Catalog number: 2030-6520-1; 6.5 CANCELLOUS BONE SCREW 20mm AH5MKE 16-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm AH1MKE 16-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm AH3MKE 17-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm H8EMKE 18-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm H8AMKE 18-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm H8KMKE 18-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm H8DMKE 21-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm JE6MKE 21-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm H88MKE 22-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm H8JMKE 22-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm JE7MKE 22-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm JE2MKE 22-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm JE4MKE 22-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm JE3MKE 22-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm H8HMKE 23-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm JE5MKE 23-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm LL9MKE 23-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm M1YMKE 23-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm MHMMKE 23-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm LLDMKE 23-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm MK4MKE 23-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm M20MKE 24-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm MHEMKE 24-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm MHDMKE 24-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm MN9MKE 24-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm MNAMKE 24-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm NMRMKE 24-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm MN8MKE 25-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm MNDMKE 25-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm MK3MKE 25-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm MK5MKE 25-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm NMVMKE 25-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm NMWMKE 25-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm NMTMKE 25-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm MK6MKE 25-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm NMPMKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm PY7MKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm R8DMKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm PYDMKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm R44MKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm T14MKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm T15MKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm R8AMKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm R8EMKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm T11MKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm T17MKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm T1DMKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm T13MKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm T12MKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm T18MKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm T1AMKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm T19MKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm T16MKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 20mm 03JMLE 06-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 252MLE 07-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 257MLE 07-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 250MLE 08-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 258MLE 08-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 2Y0MLE 08-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 2Y1MLE 08-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 2Y2MLE 08-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 2XXMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 2XYMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 346MLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 34AMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 348MLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 52JMLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 52HMLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 52RMLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 52MMLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 52PMLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 52LMLE 13-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 7H5MLE 13-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 7H4MLE 13-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 347MLE 14-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 52KMLE 14-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 52NMLE 14-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 349MLE 14-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 7H7MLE 14-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 7VJMLE 14-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 7VHMLE 14-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 7VEMLE 14-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 7VKMLE 15-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 97MMLE 15-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 7VMMLE 15-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm 7VLMLE 16-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm H1DMLE 21-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm J4EMLE 21-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm J4DMLE 21-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm J4AMLE 21-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm H8YMLE 22-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm H1EMLE 22-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm H93MLE 22-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm J49MLE 22-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm HP4MLE 22-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm HP5MLE 22-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm H1HMLE 23-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm MYNMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm MYRMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P21MLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm MY1MLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm MXYMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P0JMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P0LMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P0MMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P8HMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P8PMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P8KMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P0EMLE 29-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P1PMLE 29-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P0KMLE 29-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm MYPMLE 30-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P8JMLE 30-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm P20MLE 30-Jul-2013 6.5 CANCELLOUS BONE SCREW 20mm EA6MME 19-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm A46MME 20-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EA7MME 20-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EK7MME 20-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EMDMME 20-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EMMMME 20-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EKEMME 22-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EKHMME 22-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EKKMME 22-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EKJMME 22-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EMKMME 22-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EKLMME 22-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EMAMME 22-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm EMJMME 22-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm KXKMME 25-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm KXPMME 25-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm KXRMME 25-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm KXMMME 25-Aug-2013 6.5 CANCELLOUS BONE SCREW 20mm 1AMMNE 07-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 1APMNE 07-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 1ARMNE 07-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 1ANMNE 07-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 14WMNE 07-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 14XMNE 07-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 25KMNE 09-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 2APMNE 09-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4L6MNE 09-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4L9MNE 09-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 25MMNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 1ALMNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 2AMMNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 25NMNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 2ANMNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 2ARMNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4PPMNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4PRMNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4PVMNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4PTMNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4PWMNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4W8MNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 25LMNE 11-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4PNMNE 11-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4PXMNE 11-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4LAMNE 11-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4W9MNE 11-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 4WAMNE 11-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 61XMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 620MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 621MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 622MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 61YMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 6N8MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm 6N9MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm ADJMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm ADMMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm ADLMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm ADNMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm ADKMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm D03MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm D04MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm D05MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm D06MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm E6YMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm E72MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm E70MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm E73MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm EA3MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm EA5MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 20mm MER502 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER5V0 16-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER70T 16-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER70X 16-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER70V 16-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER70Y 16-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER712 16-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER711 16-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER713 16-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER8DX 16-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER8DY 16-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER503 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER506 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER504 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER501 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER5V1 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER5V2 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER5V3 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER5TY 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER710 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER8E0 19-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER8DV 19-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER8DW 19-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER8DT 19-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER98Y 19-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER990 19-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER98X 19-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER9VP 19-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER9VX 19-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERAPT 19-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER98W 20-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER9VN 20-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER9VV 20-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER9VT 20-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MER9VR 20-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERAPW 20-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERAPY 20-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERAPX 20-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERDD0 20-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERAPV 23-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERAR0 23-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERDAY 23-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERJTX 30-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERK8R 30-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERK8T 30-Nov-2013 6.5 CANCELLOUS BONE SCREW 20mm MERJTW 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERK8V 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERK8N 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERKWR 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERKWT 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERM8K 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERM8N 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERN15 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERN17 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERN51 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERN55 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERK8P 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERJTY 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERK8M 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERKWW 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MER505 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERM8P 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERM8L 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERN18 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET1H5 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET20W 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET20T 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET1H6 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET27E 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERKWV 08-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERM8M 08-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERN16 08-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERN52 08-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MERN54 08-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET20P 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET20R 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET20V 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET27D 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET27H 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET27J 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET486 12-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET11D 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET1H4 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET1H7 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET1H8 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET485 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET6M1 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET720 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET62P 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET5N0 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET6M3 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET7N8 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET7ND 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET7NA 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET71Y 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET7N7 18-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET81L 18-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET9X1 18-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET9X2 18-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET9X5 18-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METEE6 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METEE7 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METHKT 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET7N9 22-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET81M 22-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET6N8 22-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET6N7 22-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METKE9 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METKEE 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METKEA 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METL9V 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METKEL 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METL9W 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METKM6 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METKM3 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METJ6H 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METJ6M 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METHKR 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METKE8 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METL6E 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METL9X 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METKM4 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METKM5 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET9X3 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET9X6 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm MET9X4 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 20mm METN6D 04-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METJ6J 06-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METJ6K 06-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METP7L 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METNKL 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METPA6 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METNHJ 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METP1W 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METPE0 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METP0M 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METP7D 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METP7M 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METP7N 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METPA4 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METPKH 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METPMH 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METPA5 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METPTK 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METPTT 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm METPTP 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA02H 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA04E 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA29R 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA2N7 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA35E 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA39E 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA39J 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA39H 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA3DN 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA3DP 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA3KD 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA3M3 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA3M7 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA3M4 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA4J9 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA3M5 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA4J8 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA4JA 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA53L 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 20mm MHA29P 25-Jan-2014 
Recalling Firm/
Manufacturer
Stryker Howmedica Osteonics Corp.
325 Corporate Drive
Mahwah NJ 07430
For Additional Information Contact Colleen O'Meara
201-972-2100
Manufacturer Reason
for Recall
Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in ASTM standard for titanium surgical implants.
FDA Determined
Cause 2
Nonconforming Material/Component
Action Stryker Orthopaedics issued "Urgent Product Recall" letters via Federal Express priority overnight beginning June 26, 2009 to Branches, Hospital Risk Management and Chief of Orthopaedics and Surgeons. All users are requested to examine their inventory, retrieve all affected product and return for reconciliation to Stryker Orthopaedics. Users are also requested to complete the enclosed Product Accountability Form and fax with necessary information to the firm at 1-201-831-6069. For further information, contact Stryker Orthopaedics at 1-201-831-5970.
Quantity in Commerce 51,113 total - all sizes.
Distribution Worldwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = HWC and Original Applicant = OSTEONICS CORP.
-
-