| Date Initiated by Firm | June 26, 2009 |
| Date Posted | September 14, 2009 |
| Recall Status1 |
Terminated 3 on July 27, 2010 |
| Recall Number | Z-2119-2009 |
| Recall Event ID |
52713 |
| 510(K)Number | K873251 |
| Product Classification |
Screw, fixation, bone - Product Code HWC
|
| Product | Stryker Orthopaedics Cancellous Bone Screw;
Dia 6.5 mm; Lnth: 25 mm; Sterile;
Made in the USA;
Stryker Osteonics Corp.
325 Corporate Drive
Mahwah, NJ 07430.
The bone screws are optional fixtures that may be used during knee and hip replacement procedures. |
| Code Information |
Catalog number: 2030-6525-1; 6.5 CANCELLOUS BONE SCREW 25mm T29MKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm V0EMKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm V0JMKE 08-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm KYRMLE 23-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LP2MLE 25-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm MPWMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm MHDMKE 24-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm HVLMKE 21-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm HVMMKE 21-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm HVKMKE 21-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm HVPMKE 21-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm HVNMKE 22-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm NNNMKE 25-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm NNPMKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm NNRMKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm R6VMKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm R6WMKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm R72MKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm R70MKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm T26MKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm T28MKE 27-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm R6TMKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm R75MKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm R73MKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm R6XMKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm R74MKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm R6YMKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm T25MKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm T27MKE 29-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm T2AMKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm VDVMKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm V5AMKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm V0HMKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm V13MKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm V5LMKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm V14MKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm VE0MKE 30-Jun-2013 6.5 CANCELLOUS BONE SCREW 25mm VDRMKE 01-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm VDXMKE 01-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm VDWMKE 01-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm WDDMKE 01-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm WDAMKE 01-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm WDEMKE 01-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm WDNMKE 02-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm WDHMKE 02-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm WD9MKE 02-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 0PRMLE 02-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 0PPMLE 06-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 0PLMLE 06-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 0PTMLE 06-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 0PNMLE 06-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 0PMMLE 07-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 1K4MLE 07-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 1K2MLE 07-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 0PWMLE 07-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 1J6MLE 07-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 0PVMLE 07-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 1K6MLE 07-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 1JMMLE 08-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 1JXMLE 08-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 1JVMLE 08-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 1JPMLE 08-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 39HMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 39JMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 39LMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 39KMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 39EMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 39DMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 39RMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 39PMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 39MMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 39NMLE 09-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 4P9MLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 4PDMLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 4PEMLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 4PHMLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 4PJMLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 4PAMLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm 4PKMLE 11-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D0TMLE 18-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D0WMLE 18-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D0VMLE 18-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D11MLE 18-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D12MLE 18-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D0XMLE 21-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D10MLE 21-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D0YMLE 21-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D16MLE 21-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D15MLE 21-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D17MLE 21-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm D13MLE 22-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm KYPMLE 25-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LPKMLE 25-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm KYTMLE 25-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LPDMLE 25-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm KYNMLE 25-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LP4MLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LPHMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LWRMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LWVMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LWPMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LWYMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LPJMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LWXMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LX1MLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LP3MLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LP5MLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LWTMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm P30MLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm LX0MLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm P2YMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm P31MLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm P1DMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm P33MLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm PPKMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm P32MLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm PPJMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm P1EMLE 28-Jul-2013 6.5 CANCELLOUS BONE SCREW 25mm A4YMME 17-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm D85MME 17-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm A52MME 17-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm A50MME 17-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm A53MME 17-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm A51MME 19-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm D88MME 19-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm A4WMME 19-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm D83MME 19-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm D86MME 19-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm A54MME 19-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm A4TMME 19-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm A4XMME 19-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm D89MME 22-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm D8AMME 22-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm KXDMME 24-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm KXEMME 24-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm KXHMME 24-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm KXJMME 24-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm PL5MME 27-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm PL6MME 27-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm PL7MME 27-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm NTJMME 27-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm NTDMME 27-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm NTNMME 27-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm R3WMME 27-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm R4YMME 27-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm PN0MME 28-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm PMWMME 28-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm PL4MME 28-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm NTEMME 28-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm NTHMME 28-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm NTKMME 28-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm NTLMME 28-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm NTMMME 28-Aug-2013 6.5 CANCELLOUS BONE SCREW 25mm 0K3MNE 09-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 0K5MNE 09-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 0K4MNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 572MNE 10-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 5V1MNE 11-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 5V2MNE 11-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 56YMNE 11-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 56VMNE 11-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 574MNE 11-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 4RWMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 4T2MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 4T3MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 4T1MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 4RXMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 4RYMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 4T4MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 4T6MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 7E9MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 4T5MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 7EAMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 7EDMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 7EEMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 7ELMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 7EKMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 7EMMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 97YMNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9M5MNE 16-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 7EHMNE 17-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 7EJMNE 17-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9MWMNE 17-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9K8MNE 17-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9MVMNE 17-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9MRMNE 17-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9TLMNE 17-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 7ENMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 97XMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 980MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9MTMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 981MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9M8MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9TKMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9TMMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9TNMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9TPMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9XTMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9Y3MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9Y0MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9XYMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9XRMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9XVMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9XWMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9Y1MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9Y2MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm 9XXMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm E2HMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm E2RMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm E2KMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm E9KMNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm EV0MNE 21-Sep-2013 6.5 CANCELLOUS BONE SCREW 25mm MER2R6 11-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER2RA 11-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER2R7 11-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER2RD 11-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER2RH 11-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER2RJ 11-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER2RE 11-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER3EV 11-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER2R9 12-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER2R8 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER2RK 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER3EY 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER3EN 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER3ER 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER3EP 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER3EW 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER3H0 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER3H1 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4L5 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4L6 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4W9 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4WA 13-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER3EX 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER3H2 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4L3 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4L7 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4W7 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4WD 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4W8 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4W4 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4W5 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4WE 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER5VD 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER5VK 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER5VJ 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER5VH 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER5VE 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4W6 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4WH 18-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MERED3 23-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MERED9 23-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MERED6 23-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MERED7 23-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MERED8 23-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MERJJV 30-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MERJJX 30-Nov-2013 6.5 CANCELLOUS BONE SCREW 25mm MERK64 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERK66 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERLYE 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERK67 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERM02 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERM01 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERM00 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERM05 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERP39 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERJJN 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERJJP 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERK61 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERK5Y 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERJK0 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERJJT 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERJK1 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERK60 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERK65 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERLYD 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERK63 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERK62 04-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERK68 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERM06 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERM08 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERP37 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERP3A 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERP3D 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERP3H 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERP3J 05-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERP34 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET10K 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET10L 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET10P 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET0X6 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET10W 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET10T 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET10M 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET0X9 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET0XE 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET1PA 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET1PM 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET1PP 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET33H 07-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERM04 08-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET10R 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET10V 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET1PD 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET1PH 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET1PJ 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET1PK 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET1PE 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET1PL 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET1PN 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET339 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET33A 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET33D 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET33J 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET33E 10-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MER4L4 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERM07 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERP35 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERP36 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERP38 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MERP3E 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET10N 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET10X 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET0X7 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET0XD 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET0XA 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET0X8 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET336 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET338 14-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET9V1 18-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET9V7 18-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET9V9 18-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET9V3 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METAMD 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METAME 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METD34 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METD4L 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METDAA 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METDAE 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METDAJ 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METDAD 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METDA7 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METDAK 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METAM5 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METE93 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METDA9 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METDAH 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METE92 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METE94 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METE95 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METE96 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METE97 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METE98 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METE99 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METE9A 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METE9D 21-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET334 22-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET337 22-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET335 22-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METALY 22-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METHAM 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METHAV 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METHAT 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METJL2 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METJL3 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METJL6 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METL1T 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METL1X 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METKKK 28-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METHAL 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METHAP 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METL1W 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METL22 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METL1Y 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METL20 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METL23 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METL24 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METLYL 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METM1D 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METALW 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METALX 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METAM2 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METAM0 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METAMM 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET9V0 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET9V2 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm MET9V5 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METAMJ 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METD4K 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METDA6 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METDA8 30-Dec-2013 6.5 CANCELLOUS BONE SCREW 25mm METM6D 04-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METM6E 04-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METHAN 06-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METHAW 06-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METJKW 06-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METJKY 06-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METJKX 06-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METJL0 06-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METJL1 06-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METJL5 06-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METPMP 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METPN4 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METPN7 07-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METPMR 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METPN1 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METPMX 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METPMW 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METPMT 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METPN5 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm METPN6 11-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA3LL 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA3LN 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA4JE 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA4JH 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA54X 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA54Y 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA3LT 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA4JJ 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA54W 14-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA3LW 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA65E 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA65H 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA65K 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA65L 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA5LR 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA5LV 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA5M0 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA5LX 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA5LW 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA5LT 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA5M1 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA65A 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA65D 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA65M 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA65J 18-Jan-2014 6.5 CANCELLOUS BONE SCREW 25mm MHA75D 18-Jan-2014 |
Recalling Firm/ Manufacturer |
Stryker Howmedica Osteonics Corp. 325 Corporate Drive Mahwah NJ 07430
|
| For Additional Information Contact | Colleen O'Meara 201-972-2100 |
Manufacturer Reason for Recall | Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in ASTM standard for titanium surgical implants. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Stryker Orthopaedics issued "Urgent Product Recall" letters via Federal Express priority overnight beginning June 26, 2009 to Branches, Hospital Risk Management and Chief of Orthopaedics and Surgeons. All users are requested to examine their inventory, retrieve all affected product and return for reconciliation to Stryker Orthopaedics. Users are also requested to complete the enclosed Product Accountability Form and fax with necessary information to the firm at 1-201-831-6069.
For further information, contact Stryker Orthopaedics at 1-201-831-5970. |
| Quantity in Commerce | 51,113 total - all sizes. |
| Distribution | Worldwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = HWC
|