Date Initiated by Firm |
October 23, 2009 |
Date Posted |
November 17, 2009 |
Recall Status1 |
Terminated 3 on June 09, 2010 |
Recall Number |
Z-0307-2010 |
Recall Event ID |
53605 |
510(K)Number |
K040509 K063441 K063431 K032061 K042858
|
Product Classification |
Blood access device and accessories - Product Code MSD
|
Product |
28CM DECATHLON EXCHANGE KIT Model: DX23SH28-XP
DECATHLON MINI KIT 15.5 Fr polyurethane twin lumen hemodialysis catheter with separated tips, cuff and side holes for chronic hemodialysis and apheresis. Kit contains catheter, 18 gauge introducer needle, 16 Fr valved peel-away sheath dilator, tunneling stylet, (2) injection sealing caps, and warning label. Latex free. |
Code Information |
All lot codes |
Recalling Firm/ Manufacturer |
Spire Biomedical, Inc. One Patriots Park Bedford MA 01730-2343
|
For Additional Information Contact |
Edward Rule 617-275-6000 Ext. 274
|
Manufacturer Reason for Recall |
There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential clinical complication is air embolism, particularly for the segment of the patient population that develops negative central venous pressure during normal respira
|
FDA Determined Cause 2 |
Process control |
Action |
Spire Medical issued an "Urgent device Recall" letter dated 10/22/09 by FEDEX to end users. Users are requested to return product. For questions, contact Spire Biomedical Customer Service at 1-877-876-8635. |
Quantity in Commerce |
97 units |
Distribution |
Worldwide distribution - Nationwide, Government. Foreign: Canada, France. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MSD and Original Applicant = SPIRE BIOMEDICAL INC 510(K)s with Product Code = MSD and Original Applicant = SPIRE BIOMEDICAL, INC.
|