| | Class 2 Device Recall Dimension LIP Flex Reagent Cartridge |  |
| Date Initiated by Firm | March 10, 2010 |
| Date Posted | December 10, 2010 |
| Recall Status1 |
Terminated 3 on August 04, 2011 |
| Recall Number | Z-0588-2011 |
| Recall Event ID |
55571 |
| 510(K)Number | K952816 |
| Product Classification |
Lipase-esterase, enzymatic, photometric, lipase - Product Code CHI
|
| Product | Dimension LIP Flex Reagent Cartridge, Catalog# DF55A |
| Code Information |
Lot Number: DE0320, FC0306, GB0342, EA0334, FB0278, DA0293, GB0286, FA0355, EA0362, EB0362. |
| FEI Number |
2517506
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 500 GBC Drive, Mailstop 514 PO BOX 6101 Newark DE 19714-6101
|
| For Additional Information Contact | Siemens Technical Solutions Center 800-441-9250 |
Manufacturer Reason for Recall | Open Well instability: After calibration, QC shifts high (out of range) after the Flex Well is hydrated. Shifts of up to 20% have been observed in both QC and patient samplesas soon as two hours post hydration. |
FDA Determined Cause 2 | Other |
| Action | Recalling firm issued "Urgent Field Safety Notice", dated March 2010, informing users of the problem and to transition to the revised Lipasae method. |
| Quantity in Commerce | 21171 |
| Distribution | Nationwide Distribution throughout the US. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CHI
|
|
|
|