| | Class 2 Device Recall Dimension Vista System |  |
| Date Initiated by Firm | May 05, 2010 |
| Date Posted | September 15, 2010 |
| Recall Status1 |
Terminated 3 on October 25, 2010 |
| Recall Number | Z-2423-2010 |
| Recall Event ID |
55763 |
| 510(K)Number | K051087 |
| Product Classification |
in vitro diagnostics - Product Code JJE
|
| Product | Siemens Healthcare Diagnostics, Inc., Dimension Vista Intelligent Laboratory System with Alpha Pinnacle 10kVA UPS External Battery Pack. Part numbers: 070200001, 070200011, 070200021; Dimension Vista 1500 and 3000T instruments. |
| Code Information |
Part numbers: 070200001, 070200011, 070200021; Dimension Vista 1500 and 3000T instruments within the serial number range of B01060900055 through B01090100414R0, excluding B01081100388R0, B01081100389R0, B01081200396R0, B01081200402R0, B01081200403R0, B01081200404R0, B01081200406R0, B01081200408R0, B01081200410R0, B01081200411R0, B01090100412R0, B01090100413R0. |
| FEI Number |
2517506
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc.
500 GBC Drive, Mailstop 514
PO BOX 6101
Newark
DE
19714-6101
|
| For Additional Information Contact | 302-631-6311 |
Manufacturer Reason for Recall | Defective batteries used in the Alpha Pinnacle P10 UPS External Battery Pack may overheat and cause a strong sulfuric odor to be released and a corrosive sulfuric acid solution leak to occur. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Siemens issued an Urgent Field Safety Notice in May 2010 to customers with information on what to do if the battery leaks, overheats, or vents an odor. Firm also issued a Support Bulletin to field service personnel to inspect and replace affected units. |
| Quantity in Commerce | 228 units |
| Distribution | Worldwide distribution: USA, Australia, Austria, Belgium, Canada, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Portugal, Slovenia, Spain, and Switzerland. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|