| Class 2 Device Recall Computed Tomography XRay System MX 8000 IDT 16 | |
Date Initiated by Firm | July 27, 2010 |
Date Posted | October 22, 2010 |
Recall Status1 |
Terminated 3 on September 21, 2012 |
Recall Number | Z-0133-2011 |
Recall Event ID |
56508 |
510(K)Number | K012009 |
Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
Product | Computed Tomography X-Ray System MX 8000 IDT 16; Model # 728120; Serial #'s: 2513, 2518, 3147, 3217, 3014, 3050, 3068, 3188, 3091, 3051, 3099, 3126, 3183, 3303, 2539 & 3056.
Produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. |
Code Information |
Model # 728120; Serial #'s: 2513, 2518, 3147, 3217, 3014, 3050, 3068, 3188, 3091, 3051, 3099, 3126, 3183, 3303, 2539 & 3056. |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Road Cleveland OH 44143-2131
|
For Additional Information Contact | 440-483-7000 |
Manufacturer Reason for Recall | Philips Healthcare has decided to recall the Computed Tomography X-Ray System MX 8000 IDT due to the fact that the software upgrade for the MX 8000 IDT 16 v4.2.1 may not provide the optional collimation of 4x4.5 (18mm). If the customer does not verify the increment and thickness coincide with the newly selected collimation suggested by the conversion report, there is potential for a 6mm gap to ap |
FDA Determined Cause 2 | Software change control |
Action | Philips Medical Systems (Cleveland), Inc. sent an URGENT - Field Safety Notice dated July 29, 2010 and August 23, 2010 to all customers. The notices identified the product, the problem, and the steps to be taken.
Customers were instructed to locate and correct the error.
Field Service Engineers were to visit each site to update the software
For further information customers were to contact their local Philips Customer Care Center at 1-800-722-9377 |
Quantity in Commerce | 16 units |
Distribution | Product was sold in the following states: CA, DE, IN,OR, MA, NV, TN. Product was also sold in Canada and Australia |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|