| | Class 2 Device Recall Welch Allyn PIC 50 Portable Intensive Care System |  |
| Date Initiated by Firm | September 17, 2010 |
| Date Posted | January 24, 2011 |
| Recall Status1 |
Terminated 3 on July 25, 2011 |
| Recall Number | Z-0937-2011 |
| Recall Event ID |
56804 |
| 510(K)Number | K974034 |
| Product Classification |
External Transcutaneous Cardiac Pacemaker (Non-Invasive) - Product Code DRO
|
| Product | Welch Allyn PIC 50 Portable Intensive Care System. Model 971081 series.
Designed for use in and out of the hospital. |
| Code Information |
Devices are identified with Serial Numbers: 1005, 1011, 1012, 1013, 1014, 2355, 80001, 80003, 80007, 80008, 80009, 80012, 80013, 80018, 80022, 80025, 80031, 80032, 80033, 80035, 80042, 80045, 80050, 80059, 80076, 80085, 80117, 80228, 80276, 80281, 80282, 80283, 80285, 80286, 80288, 80289, 80290, 80296, 80297, 80298, 80299, 80301, 80314, 80315, 80316, 80317, 80318, 80319, 80347, 80348, 80349, 80350, 80351, 80352, 80353, 80354, 80355, 80356, 80357, 80358, 80359, 80360, 80361, 80362, 80363, 80364, 80365, 80366, 80367, 80368, 80369, 80370, 80371, 80372, 80373, 80374, 80375, 80376, 80378, 80379, 80380, 80381, 80383, 80384, 80385, 80386, 80387, 80388, 80389, 80390, 80391, 80392, 80393, 80394, 80395, 80405, 80406, 80407, 80408, 80419, 80420, 80423, 80424, 80425, 80426, 80427, 80428, 80429, 80430, 80431, 80432, 80433, 80434, 80435, 80436, 80437, 80438, 80439, 80440, 80441, 80443, 80444, 80445, 80446, 80447, 80448, 80449, 80450, 80451, 80453, 80454, 80455, 80456, 80458, 80459, 80460, 80461, 80462, 80463, 80464, 80465, 80466, 80467, 80468, 80475, 80476, 80478, 80487, 80488, 80489, 80490, 80491, 80492, 80493, 80495, 80496, 80497, 80498, 80499, 80500, 80501, 80502, 80503, 80504, 80505, 80506, 80507, 80508, 80509, 80510, 80511, 80512, 80513, 80514, 80515, 80516, 80517, 80518, 80519, 80520, 80521, 80523, 80524, 80525, 80531, 80532, 80533, 80534, 80535, 80536, 80539, 80540, 80542, 80543, 80554, 80555, 80556, 80557, 80558, 80559, 80560, 80564, 80569, 80571, 80572, 80574, 80575, 80576, 80577, 80578, 80579, 80580, 80581, 80582, 80583, 80584, 80585, 80586, 80587, 80589, 80590, 80591, 80592, 80593, 80594, 80596, 80597, 80600, 80602, 80603, 80604, 80605, 80606, 80607, 80608, 80609, 80611, 80616, 80617, 80618, 80619, 80620, 80621, 80622, 80624, 80625, 80627, 80628, 80629, 80632, 80637, 80638, 80639, 80640, 80641, 80642, 80643, 80644, 80645, 80646, 80647, 80648, 80649, 80650, 80651, 80652, 80653, 80654, 80655, 80656, 80657, 80658, 80659, 80660, 80662, 80663, 80664, 80665, 80666, 80667, 80672, 80673, 80674, 80675, 80676, 80677, 80678, 80679, 80680, 80681, 80682, 80683, 80684, 80685, 80686, 80687, 80689, 80690, 80691, 80692, 80693, 80694, 80695, 80696, 80697, 80699, 80702, 80703, 80704, 80705, 80706, 80707, 80711, 80713, 80714, 80715, 80716, 80717, 80718, 80719, 80720, 80721, 80722, 80723, 80724, 80725, 80726, 80727, 80728, 80729, 80730, 80731, 80755, 80756, 80757, 80758, 80759, 80761, 80762, 80763, 80764, 80765, 80766, 80767, 80768, 80769, 80770, 80771, 80778, 80779, 80781, 80782, 80783, 80784, 80785, 80786, 80787, 80788, 80789, 80790, 80791, 80792, 80794, 80795, 80796, 80797, 80798, 80799, 80800, 80801, 80802, 80803, 80807, 80808, 80809, 80810, 80822, 80823, 80824, 80826, 80827, 80838, 80839, 80840, 80841, 80842, 80843, 80844, 80845, 80847, 80848, 80849, 80850, 80851, 80852, 80855, 80856, 80857, 80858, 80859, 80860, 80861, 80862, 80863, 80864, 80865, 80867, 80868, 80869, 80870, 80871, 80872, 80873, 80874, 80875, 80876, 80877, 80878, 80879, 80880, 80881, 80882, 80883, 80884, 80885, 80886, 80887, 80891, 80913, 80914, 80915, 80916, 80917, 80919, 80920, 80921, 80922, 80923, 80925, 80926, 80927, 80928, 80929, 80930, 80931, 80932, 80934, 80935, 80936, 80937, 80939, 80940, 80941, 80942, 80943, 80944, 80945, 80946, 80947, 80948, 80949, 80950, 80951, 80960, 80961, 80962, 80963, 80964, 80965, 80966, 80967, 80968, 80969, 80971, 80972, 80973, 80974, 80975, 80976, 80977, 80978, 80979, 80980, 80982, 80983, 80984, 80985, 80987, 80988, 80989, 80990, 80991, 80992, 80993, 80994. 80995, 81001, 81003, 81004, 81005, 81006, 81007, 81008, 81009, 81010, 81011, 81012, 81013, 81014, 81015, 81016, 81017, 81018, 81019, 81020, 81021, 81022, 81023, 81024, 81025, 81026, 81027, 81028, 81029, 81030, 81031, 81032, 81033, 81034, 81035, 81036, 81037, 81038, 81039, 81040, 81041, 81042, 81043, 81044, 81045, 81046, 81047, 81047, 81049, 81050, 81051, 81052, 81053, 81054, 81055, 81056, 81057, 81058, 81059, 81060, 81061, 81062, 81063, 81064, 81065, 81066, 81067, 81068, 81069, 81070, 81071, 81072, 81073, 81074, 81075, 81076, 81077, 81078, 81079, 81080, 81081, 81082, 81083, 81084, 81085, 81086, 81087, 81088, 81089, 81090, 81092, 81093, 81094, 81095, 81096, 81097, 81098, 81099, 81102, 81103, 81105, 81107, 81108, 81109, 81116, 81118, 81121, 81122, 81123, 81124, 81125, 81126, 81127, 81128, 81129, 81130, 81131, 81132, 81133, 81134, 81135, 81136, 81137, 81138, 81139, 81140, 81141, 81142, 81143, 81144, 81145, 81146, 81147, 81148, 81149, 81150, 81151, 81152, 81153, 81154, 81155, 81156, 81157, 81158, 81159, 81160, 81165, 81166, 81167, 81168, 81169, 81176, 81177, 81178, 81179, 81180, 81181, 81182, 81182, 81183, 81184, 81185, 81186, 81187, 81188, 81189, 81190, 81191, 81192, 81193, 81194, 81195, 81196, 81197, 81198, 81199, 81200, 81206, 81216, 81217, 81218, 81219, 81220, 81221, 81222, 81223, 81227, 81228, 81229, 81230, 81231, 81232, 81233, 81234, 81235, 81236, 81237, 81238, 81239, 81240, 81241, 81242, 81243, 81244, 81245, 81246, 81247, 81248, 81249, 81250, 81251, 81252, 81253, 81254, 81255, 81256, 81257, 81258, 81259, 81261, 81262, 81264, 81265, 81269, 81270, 81277, 81290, 81291, 81292, 81296, 81297, 81298, 81299, 81301, 81302, 81303, 81304, 81305, 81306, 81307, 81308, 81309, 81324, 81325, 81326, 81327, 81328, 81329, 81330, 81331, 81332, 81333, 81334, 81335, 81336, 81337, 81338, 81339, 81340, 81341, 81342, 81343, 81346, 81347, 81348, 81351, 81352, 81353, 81354, 81355, 81358, 81359, 81360, 81361, 81362, 81363, 81364, 81365, 81366, 81371, 81372, 81373, 81374, 81376, 81377, 81378, 81379, 81380, 81381, 81382, 81383, 81384, 81385, 81386, 81387, 81388, 81389, 81390, 81391, 81392, 81393, 81394, 81395, 81396, 81399, 81400, 81416, 81417, 81418, 81419, 81420, 81421, 81424, 81425, 81426, 81427, 81428, 81431, 81436, 81437, 81438, 81439, 81440, 81441, 81442, 81443, 81444, 81445, 81446, 81447, 81448, 81450, 81451, 81452, 81453, 81454, 81455, 81456, 81457, 81458, 81459, 81460, 81461, 81462, 81463, 81464, 81465, 81466, 81467, 81468, 81469, 81470, 81471, 81473, 81474, 81485, 81486, 81487, 81488, 81489, 81490, 81491, 81492, 81493, 81494, 81495, 81496, 81497, 81498, 81499, 81500, 81501, 81502, 81530, 81531, 81532, 81533, 81568, 81569, 81570, 81571, 81573, 81575, 81576, 81577, 81578, 81579, 81580, 81581, 81582, 81583, 81584, 81585, 81587, 81588, 81602, 81603, 81604, 81605, 81606, 81607, 81608, 81609, 81610, 81611, 81612, 81613, 81614, 81615, 81616, 81617, 81618, 81619, 81620, 81621, 81622, 81623, 81624, 81625, 81626, 81627, 81628, 81630, 81664, 81665, 81666, 81667, 81668, 81669, 81670, 81671, 81672, 81673, 81674, 81675, 81676, 81677, 81678, 81679, 81680, 81681, 81682, 81683, 81684, 81685, 81686, 81687, 81688, 81689, 81690, 81691, 81692, 81693, 81694, 81695, 81696, 81698, 81699, 81700, 81701, 81702, 81703, 81704, 81705, 81706, 81707, 81708, 81709, 81710, 81711, 81712, 81713, 81714, 81715, 81716, 81717, 81718, 81719, 81720, 81721, 81733, 81738, 81739, 81743, 81744, 81745, 81746, 81747, 81748, 81749, 81750, 81751, 81752, 81757, 81761, 81762, 81763, 81764, 81765, 81766. 81767, 81768, 81769, 81770, 81771, 81772, 81773, 81774, 81775, 81776, 81776, 81777, 81778, 81779, 81780, 81781, 81802, 81803, 81804, 81805, 81806, 81807, 81822, 81823, 81824, 81825, 81826, 81827, 81828, 81829, 81830, 81831, 81832, 81833, 81834, 81835, 81836, 81837, 81838, 81839, 81840, 81841, 81842, 81843, 81844, 81845, 81846, 81849, 81850, 81851, 81852, 81853, 81854, 81855, 81856, 81857, 81858, 81860, 81861, 81862, 81863, 81864, 81865, 81866, 81867, 81868, 81869, 81884, 81885, 81886, 81892, 81893, 81894, 81895, 81896, 81897, 81918, 81919, 81920, 81921, 81922, 81923, 81924, 81925, 81926, 81927, 81928, 81929, 81930, 81931, 81932, 81933, 81934, 81935, 81936, 81937, 81938, 81939, 81940, 81941, 81942, 81943, 81944, 81944, 81945, 81946, 81947, 81948, 81949, 81950, 81951, 81952, 81953, 81954, 81955, 81957, 81983, 81985, 81986, 81996, 81997, 81998, 81999, 81999, 82000, 82001, 82002, 82003, 82004, 82005, 82006, 82007, 82008, 82009, 82010, 82011, 82012, 82013, 82014, 82015, 82016, 82017, 82018, 82019, 82020, 82021, 82022, 82023, 82023, 82024, 82024, 82025, 82027, 82028, 82029, 82029, 82030, 82031, 82032, 82033, 82034, 82034, 82035, 82035, 82036, 82036, 82037, 82038, 82039, 82043, 82044, 82061, 82062, 82063, 82064, 82065, 82066, 82067, 82068, 82069, 82070, 82071, 82072, 82073, 82074, 82075, 82076, 82077, 82078, 82079, 82080, 82084, 82097, 82098, 82099, 82100, 82101, 82102, 82103, 82104, 82105, 82106, 82107, 82108, 82109, 82110, 82111, 82122, 82123, 82124, 82124, 82125, 82126, 82127, 82127, 82128, 82128, 82133, 82136, 82137, 82138, 82139, 82140, 82141, 82142, 82143, 82150, 82151, 82152, 82153, 82154, 82155, 82156, 82157, 82158, 82159, 82160, 82161, 82162, 82163, 82164, 82165, 82166, 82167, 82168, 82169, 82172, 82173, 82174, 82175, 82176, 82177, 82178, 82179, 82180, 82181, 82182, 82183, 82184, 82185, 82186, 82187, 82188, 82189, 82190, 82191, 82192, 82193, 82194, 82195, 82196, 82197, 82198, 82199, 82200, 82201, 82202, 82203, 82204, 82234, 82245, 82252, 82253, 82254, 82255, 82256, 82257, 82258, 82259, 82260, 82261, 82262, 82263, 82264, 82265, 82266, 82267, 82268, 82269, 82270, 82271, 82272, 82273, 82283, 82283, 82284, 82285, 82286, 82286, 82307, 82308, 82309, 82310, 82311, 82312, 82313, 82314, 82315, 82316, 82317, 82318, 82319, 82320, 82321, 82322, 82323, 82324, 82325, 82326, 82327, 82328, 82329, 82330, 82340, 82341, 82342, 82343, 82344, 82345, 82346, 82347, 82348, 82371, 82405, 82406, 82407, 82408, 82409, 82410, 82411, 82412, 82413, 82414, 82415, 82416, 82417, 82418, 82419, 82420, 82422, 82423, 82424, 82425, 82426, 82429, 82430, 82460, 82461, 82462, 82463, 82464, 82465, 82466, 82467, 82468, 82469, 82470, 82471, 82472, 82473, 82474, 82475, 82476, 82477, 82478, 82479, 82480, 82481, 82482, 82483, 82484, 82485, 82486, 82487, 82488, 82489, 82490, 82498, 82499, 82500, 82501, 82502, 82503, 82504, 82505, 82506, 82507, 82508, 82509, 82510, 82511, 82512, 82513, 82514, 82515, 82516, 82517, 82518, 82519, 82520, 82521, 82522, 82523, 82524, 82525, 82526, 82527, 82528, 82529, 82530, 82531, 82531, 82531, 82532, 82533, 82534, 82535, 82536, 82538, 82538, 82539, 82540, 82541, 82542, 82543, 82544, 82545, 82546, 82547, 82548, 82549, 82550, 82551, 82552, 82553, 82555, 82555, 82555, 82556, 82557, 82557, 82558, 82558, 82559, 82559, 82560, 82560, 82561, 82561, 82562, 82563, 82564, 82565, 82565, 82568, 82568, 82569, 82570, 82612, 82612, 82613, 82613, 82614, 82614, 82615, 82615, 82616, 82616, 82617, 82617, 82618, 82619, 82619, 82620, 82620, 82621, 82621, 82622, 82622, 82623. 82623, 82624, 82624, 82625, 82625, 82626, 82627, 82628, 82629, 82630, 82631, 82654, 82655, 82656, 82659, 82660, 82661, 82662, 82663, 82664, 82665, 82666, 82667, 82668, 82669, 82670, 82671, 82672, 82673, 82689, 82690, 82773, 82774, 82775, 82776, 82777, 82778, 82779, 82780, 82781, 82782, 82783, 82784, 82785, 82786, 82787, 82788, 82789, 82790, 82791, 82792, 82793, 82794, 82795, 82796, 82797, 82798, 82799, 82800, 82801, 82802, 82803, 82804, 82805, 82806, 82807, 82808, 82809, 82810, 82811, 82812, 82813, 82827, 82828, 82829, 82830, 82852, 82856, 82857, 82858, 82859, 82859, 82906, 82936, 82937, 82958, 82959, 82964, 82965, 82966, 82967, 82969, 82972, 82973, 82974, 82975, 82977, 82984, 82985, 82986, 82987, 82988, 82989, 82990, 82991, 82992, 82995, 82996, 82998, 82999, 83001, 83002, 83005, 83008, 83009, 83010, 83012, 83013, 83033, 83034, 83035, 83036, 83037, 83038, 83060, 83061, 83062, 83062, 83062, 83063, 83063, 83064, 83064, 83065, 83065, 83066, 83066, 83067, 83067, 83068, 83068, 83070, 83071, 83083, 83083, 83084, 83084, 83085, 83085, 83086, 83087, 83088, 83088, 83089, 83089, 83090, 83090, 83093, 83145, 83153, 83154, 83155, 83156, 83157, 83158, 83159, 83160, 83161, 83162, 83163, 83164, 83165, 83166, 83167, 83168, 83169, 83170, 83171, 83172, 83198, 83199, 83200, 83201, 83202, 83203, 83204, 83205, 83206, 83207, 83208, 83209, 83210, 83211, 83212, 83213, 83214, 83215, 83216, 83217, 83218, 83219, 83220, 83221, 83222, 83223, 83224, 83225, 83228, 83230, 83233, 83237, 83255, 83288, 83289, 83290, 83291, 83368, 83369, 83370, 83400, 83401, 83402, 83403, 83404, 83405, 83406, 83407, 83414, 83415, 83416, 83417, 83418, 83420, 83421, 83422, 83423, 83424, 83425, 83426, 83428, 83440, 83441, 83442, 83443, 83499, 83500, 83501, 83502, 83503, 83504, 83505, 83506, 83507, 83508, 83509, 83510, 83511, 83512, 83513, 83514, 83515, 83516, 83517, 83578, 83579, 83580, 83633, 83644, 83645, 83654, 85009, 85010, 85011, 85012, 85013, 85026, 85027, 85028, 85029, 85032, 85033, 85034, 85035, 85036, 85037, 85038, 85039, 85040, 85041, 85042, 85043, 85044, 85045, 85046, 85047, 85050, 85051, 85052, 85073, 85074, 85075, 85076, 85077, 85078, 85079, 85093, 85094, 85110, 85111, 85143, 85144, 85145, 85146, 85147, 85148, 85149, 85150, 85151, 85152, 85153, 85154, 85155, 85156, 85157, 85187, 85188, 85189, 85190, 85191, 85192, 85193, 85194, 85195, 85196, 85197, 85198, 85199, 85200, 85201, 85226, 85227, 85228, 85229, 85234, 85235, 85236, 85237, 85238, 85244, 85245, 85247, 85248, 85249, 85255, 85329, 85332, 85333, 85334, 85337, 85338, 85340, 85341, 85342, 85343, 85344, 85345, 85346, 85347, 85348, 85349, 85350, 85364, 85365, 85368, 85369, 85381, 85388, 85389, 85391, 85392, 85393, 85394, 85395, 85396, 85397, 85398, 85402, 85403, 85404, 85410, 85411, 85412, 85413, 85414, 85420, 85421, 85422, 85423, 85424, 85425, 85426, 85427, 85428, 85429, 85430, 85431, 85432, 85433, 85441, 85505, 85506, 85507, 85508, 85509, 85510, 85511, 85512, 85513, 85514, 85515, 85517, 85518, 85519, 85520, 85522, 85523, 85524, 85525, 85526, 85527, 85528, 85529, 85530, 85552, 85554, 85555, 85556, 85557, 85587, 85588, 85589, 85602, 85603, 85604, 85604, 85605, 85606, 85607, 85619, 85620, 85621, 85622, 85623, 85630, 85645, 85660, 85686, 85728, 85729, 85730, 85731, 85732, 85733, 85734, 85755, 85756, 85757, 85761, 85762, 85763, 85764, 85765, 85766, 85771, 85771, 85772, 85772, 85773, 85773, 85774, 85774, 85775, 85775, 85776, 85776, 85777, 85778, 85778, 85779, 85779, 85780, 85780, 85781, 85782, 85783, 85784, 85785, 85787, 85788, 85789. 85790, 85791, 85792, 85793, 85810, 85812, 85813, 85814, 85815, 85816, 85817, 85818, 85819, 85820, 85821, 85822, 85823, 85825, 85826, 85827, 85828, 85829, 85830, 85831, 85832, 85833, 85834, 85835, 85836, 85837, 85838, 85856, 85857, 85858, 85859, 85860, 85861, 85862, 85863, 85864, 85865, 85866, 85867, 85868, 85869, 85870, 85871, 85872, 85877, 85878, 85879, 85880, 85881, 85882, 85883, 85884, 85888, 85889, 85890, 85891, 85892, 85893, 85894, 85895, 85896, 85897, 85898, 85899, 85900, 85901, 85902, 85931, 85932, 85933, 85945, 85955, 85964, 85967, 85968, 85969, 85970, 86017, 86018, 86024, 86025, 86026, 86027, 86028, 86029, 86055, 86056, 86057, 86103, 86104, 86118, 86141, 86142, 86143, 86144, 86145, 86146, 86147, 86149, 86150, 86222, 86223, 86224, 86225, 86226, 86227, 86228, 86230, 86245, 86256, 86257, 86258, 86261, 86262, 86263, 86264, 86265, 86266, 86267, 86268, 86281, 86282, 86283, 86289, 86290, 86291, 86292, 86293, 86295, 86296, 86299, 86305, 86306, 86314, 86315, 86320, 86321, 86322, 86323, 86324, 86325, 86326, 86327, 86328, 86334, 86335, 86358, 86359, 86385, 86387, 86388, 86389, 86390, 86391, 86392, 86393, 86394, 86395, 86396, 86397, 86398, 86399, 86400, 86411, 86412, 86419, 86420, 86489, 86490, 86491, 86495, 86496, 86497, 86498, 86499, 86533, 86534, 86535, 86536, 86537, 86538, 86539, 86540, 86541, 86542, 86543, 86544, 86545, 86546, 86562, 86563, 86564, 86575, 86599, 86600, 86640, 86641, 86642, 86643, 86644, 86650, 86651, 86663, 86664, 86665, 86666, 86667, 86668, 86669, 86676, 86677, 86678, 86683, 86684, 86696, 86711, 86712, 86713, 86714, 86715, 86716, 86717, 86718, 86767, 86768, 86769, 86770, 86771, 86778, 86819, 86835, 86836, 86837, 86838, 86840, 86848, 86851, 86871, 86872, 86873, 86874, 86875, 86876, 86877, 86878, 86879, 86880, 86881, 86882, 86883, 86973, 86999, 87000, 87001, 87002, 87003, 87004, 87005, 87007, 87008, 87009, 87010, 87011, 87012, 87012, 87013, 87014, 87015, 87027, 87029, 87030, 87309, 87310, 87327, 87328, 87329, 87330, 87331, 87332, 87333, 87334, 87335, 87336, 87337, 87338, 87339, 87340, 87341, 87342, 87343, 87344, 87345, 87347, 87348, 87349, 87350, 87351, 87352, 87403, 87404, 87413, 87414, 87415, 87416, 87417, 87418, 87420, 87572, 87573, 87574, 87575, 87584, 87588, 87589, 87590, 87591, 87592, 87593, 87594, 87595, 87596, 87597, 87598, 87599, 87600, 87601, 87602, 87603, 87604, 87605, 87606, 87607, 87632, 87633, 87667, 87668, 87669, 87670, 87671, 87683, 87684, 87685, 87686, 87687, 87688, 87689, 87690, 87691, 87692, 87693, 87694, 87695, 87696, 87697, 87698, 87699, 87700, 87701, 87702, 87703, 87704, 87705, 87706, 87707, 87708, 87728, 87729, 87730, 87731, 87732, 87733, 87734, 87735, 87736, 87737, 87738, 87739, 87740, 87741, 87742, 87776, 87777, 87778, 87779, 87780, 87781, 87782, 87783, 87789, 87790, 87792, 87793, 87794, 87795, 87796, 89003, 89004, 89005, 89006, 89007, 89008, 89009, 89010, 89011, 89012, 89013, 89014, 89015, 89016, 89020, 89021, 89022, 89023, 89026, 89027, 89028, 89029, 89030, 89033, 89034, 89035, 89054, 89056, 89057, 89058, 89059, 89060, 89061, 89062, 89063, 89064, 89065, 89073, 89074, 89075, 89076, 89077, 89078, 89081, 89082, 89083, 89084, 89085, 89086, 89087, 89088, 89089, 89090, 89091, 89092, 89093, 89094, 89095, 89096, 89097, 89098, 89099, 89100, 89101, 89102, 89103, 89104, 89119, 89120, 89121, 89122, 89123, 89124, 89125, 89126, 89127, 89142, 89143, 89144, 89145, 89146, 89147, 89148, 89149, 89150, 89151, 89152, 89153, 89154, 89155, 89156, 89157, 89158, 89159, 89160, 89161, 89162, 89163, 89164, 89165, 89166. 89186, 89187, 89188, 89189, 89190, 89191, 89192, 89193, 89194, 89197, 89198, 89199, 89219, 89220, 89221, 89222, 89223, 89224, 89225, 89232, 89233, 89234, 89235, 89236, 89237, 89238, 89240, 89251, 89252, 89253, 89254, 89256, 89299, 89300, 89301, 89302, 89303, 89304, 89305, 89306, 89307, 89308, 89331, 89332, 89334, 89336, 89337, 89349, 89350, 89351, 89352, 89353, 89354, 89355, 89356, 89357, 89358, 89359, 89366, 89366, 89367, 89368, 89399, 89400, 89402, 89403, 89404, 89405, 89406, 89407, 89412, 89413, 89414, 89420, 89480, 89502, 89505, 89506, 89507, 89508, 89509, 89510, 89511, 89511, 89512, 89512, 89513, 89513, 89534, 89535, 89536, 89562, 89563, 89564, 89605, 89606, 89610, 89646, 89647, 89648, 89686, 89687, 89688, 89689, 89718, 89719, 89720, 89761, 89762, 89793, 89815, 89816, 89817, 89818, 89819, 89820, 89821, 89822, 89863, 89881, 89882, 89922, 89923, 89924, 89927, 89928, 89929, 89930, 89931, 89932, 89933, 89934, 89935, 89936, 89937, 89938, 89939, 89940, 90002, 90003, 90004, 90005, 90006, 90007, 90008, 90009, 90010, 90011, 90012, 90013, 90014, 90015, 90020, 90030, 90031, 90032, 90033, 90034, 90035, 90036, 90037, 90047, 90048, 90049, 90050, 90051, 90052, 90053, 90054, 90055, 90056, 90057, 90058, 90059, 90060, 90063, 90064, 90066, 90067, 90074, 90080, 90081, 90164, 90165, 90166, 90209, 90210, 90211, 90212, 90213, 90271, 90272, 90273, 90308, 90309, 90320, 90321, 90322, 90364, 90365, 90366, 90402, 90403, 90404, 90405, 90406, 90407, 90408, 90409, 90410, 90411, 90412, 90413, 90414, 90415, 90416, 90417, 90503, 90504, 90505, 90506, 90507, 90508, 90510, 90511, 90516, 90517, 90518, 90519, 90533, 90534, 90562, 90570, 90571, 90572, 90573, 90637, 90700, 90712, 90724, 90725, 90726, 90731, 90732, 90733, 90734, 90735, 90736, 90737, 90738, 90739, 90740, 90741, 90742, 90743, 90744, 90745, 90746, 90747, 90748, 90749, 90750, 90751, 90752, 90753, 90754, 90755, 90756, 90757, 90758, 90759, 90760, 90761, 90762, 90763, 90764, 90765, 90766, 90796, 90797, 90840, 90888, 90889, 90890, 90891, 90900, 90901, 90934, 90935, 90960, 90961, 90962, 90963, 90964, 90969, 90970, 90971, 91079, 91129, 91140, 91141, 91142, 91143, 91144, 91145, 91146, 91147, 91148, 91149, 91150, 91151, 91152, 91154, 91155, 91156, 91199, 91200, 91201, 91202, 91233, 91234, 91235, 91238, 91290, 91295, 91446, 91447, 91495, 91498, 91499, 91500, 91503, 91505, 91506, 91507, 91508, 91509, 91536, 91569, 91570, 91571, 91572, 91575, 91576, 91577, 91578, 91579, 91580, 91581, 91582, 91583, 91585, 91587, 91588, 91589, 91590, 91591, 91592, 91593, 91594, 91595, 91598, 91599, 91600, 91601, 91602, 91603, 91604, 91606, 91607, 91608, 91608, 91609, 91610, 91610, 91611, 91612, 91613, 91615, 91616, 91618, 91619, 91621, 91623, 91624, 91626, 91631, 91632, 91633, 91636, 91637, 91638, 91639, 91640, 91641, 91642, 91643, 91651, 91652, 91653, 91654, 91656, 91657, 91658, 91660, 91661, 91662, 91664, 91665, 91666, 91669, 91670, 91671, 91672, 91675, 91678, 91679, 91681, 91723, 91735, 91736, 91741, 91742, 91743, 91744, 91748, 91749, 91750, 91751, 91752, 91753, 91754, 91755, 91756, 91757, 91758, 91759, 91774, 91775, 91776, 91777, 91778, 91889, 91890, 91891, 91892, 91892, 91893, 91893, 91894, 91895, 91896, 91923, 91924, 92021, 92024, 92105, 92106, 92107, 92108, 92110, 92111, 92113, 92114, 92115, 92116, 92117, 92118, 92119, 92120, 92121, 92122, 92123, 92124, 92125, 92126, 92127, 92128, 92129, 92130, 92131, 92132, 92133, 92134, 92135, 92135, 92136, 92137, 92138, 92139, 92140, 92141, 92142, 92143, 92144, 92145, 92146, 92147, 92148, 92149. 92150, 92151, 92152, 92153, 92154, 92155, 92156, 92157, 92158, 92159, 92160, 92161, 92162, 92195, 92219, 92220, 92224, 92225, 92226, 92227, 92228, 92229, 92230, 92235, 92236, 92237, 92238, 92239, 92242, 92243, 92246, 92247, 92253, 92254, 92257, 92258, 92259, 92259, 92260, 92260, 92261, 92262, 92263, 92264, 92265, 92316, 92317, 92318, 92319, 92320, 92321, 92322, 92323, 92324, 92325, 92326, 92327, 92328, 92329, 92330, 92331, 92332, 92333, 92334, 92335, 92336, 92337, 92339, 92340, 92340, 92341, 92342, 92343, 92344, 92345, 92346, 92348, 92364, 92364, 92365, 92365, 92366, 92367, 92368, 92369, 92371, 92371, 92372, 92373, 92424, 92439, 92440, 92462, 92463, 92477, 92478, 92479, 92480, 92522, 92538, 92539, 92540, 92541, 92542, 92548, 92549, 92550, 92551, 92552, 92557, 92558, 92559, 92565, 92566, 92567, 92568, 92568, 92569, 92570, 92570, 92571, 92572, 92572, 92573, 92573, 92574, 92575, 92576, 92577, 92578, 92579, 92580, 92582, 92583, 92584, 92585, 92586, 92587, 92588, 92589, 92590, 92591, 92592, 92593, 92594, 92595, 92596, 92597, 92598, 92599, 92600, 92601, 92602, 92603, 92604, 92605, 92606, 92607, 92608, 92658, 92659, 92659, 92660, 92662, 92663, 92667, 92668, 92669, 92670, 92671, NF160001, NF160001, NF160002, NF160002, NF160003, NF160003, NF160004, NF160004, NF160005, NF160005, NF160006, NF160006, NF160007, NF160007, NF160008, NF160008, NF160009, NF160009, NF160010, NF160010, NF160011, NF160011, NF160012, NF160013, NF160014, NF160015, NF160016, NF160017, NF160018, NF160019, NF160020, NF160021, NF160022, NF160023, NF160024, NF160025, NF160026, NF160027, NF160028, NF160029, NF160030, NF160031, NF160032, NF160033, NF160034, NF160035, NF160036, NF160037, NF160038, NF160039, NF160040, NF160041, NF160042, NF160043, NF160044, NF160045, NF160046, NF160047, NF160048, NF160049, NF160050, NF160051, NF160053, NF160054, NF160055, NF160056, NF160057, NF160058, NF160059, NF160060, NF160061, NF160062, NF160063, NF160064, NF160065, NF160066, NF160067, NF160068, NF160069, NF160070, NF160071, NF160072, NF160073, NF160074, NF160075, NF160076, NF160077, NF160078, NF160079, NF160080, NF160081, NF160082, NF160083, NF160084, NF160085, NF160086, NF160087, NF160088, NF160089, NF160090, NF160091, NF160092, NF160093, NF160094, NF160095, NF160096, NF160097, NF160098, NF160099, NF160100, NF160101, NF160102, NF160103, NF160104, NF160105, NF160106, NF160107, NF160108, NF160109, NF160110, NF160111, NF160112, NF160113, NF160114, NF160115, NF160116, NF160117, NF160118, NF160119, NF160120, NF160121, NF160122, NF160123, NF160124, NF160125, NF160126, NF160127, NF160128, NF160129, NF160130, NF160131, NF160132, NF160133, NF160134, NF160135, NF160136, NF160137, NF160138, NF160139, NF160140, NF160141, NF160142, NF160143, NF160144, NF160145, NF160146, NF160147, NF160148, NF160149, NF160150, NF160151, NF160152, NF160153, NF160154, NF160155, NF160156, NF160157, NF160158, NF160159, NF160160, NF160161, NF160162, NF160163, NF160164, NF160165, NF160166, NF160167, NF160168, NF160169, NF160170, NF160171, NF160172, NF160173, NF160174, NF160175, NF160176, NF160177, NF160178, NF160179, NF160180, NF160181, NF160182, NF160183, NF160184, NF160185, NF160186, NF160187, NF160188, NF160189, NF160190, NF160191, NF160192, NF160193, NF160194, NF160195, NF160196, NF160197, NF160198, NF160199, NF160200, NF160201, NF160202, NF160203, NF160204, NF160205, NF160206, NF160207, NF160208, NF160209, NF160210, NF160211, NF160212, NF160213, NF160214, NF160215, NF160216, NF160217, NF160218, NF160219, NF160220, NF160221, NF160222, NF160223, NF160224, NF160225, NF160226, NF160227, NF160228, NF160229, NF160230. NF160231, NF160232, NF160233, NF160234, NF160235, NF160236, NF160237, NF160238, NF160239, NF160240, NF160241, NF160242, NF160243, NF160244, NF160245, NF160246, NF160247, NF160248, NF160249, NF160250, NF160251, NF160252, NF160253, NF160254, NF160255, NF160256, NF160257, NF160258, NF160259, NF160260, NF160261, NF160262, NF160263, NF160264, NF160265, NF160266, NF160267, NF160268, NF160269, NF160270, NF160271, NF160272, NF160273, NF160274, NF160275, NF160276, NF160277, NF160278, NF160279, NF160280, NF160281, NF160282, NF160283, NF160284, NF160285, NF160286, NF160287, NF160288, NF160289, NF160290, NF160291, NF160292, NF160293, NF160294, NF160295, NF160296, NF160297, NF160298, NF160299, NF160300, NF160301, NF160302, NF160303, NF160304, NF160305, NF160306, NF160307, NF160308, NF160309, NF160310, NF160311, NF160312, NF160313, NF160314, NF160315, NF160316, NF160317, NF160318, NF160319, NF160320, NF160321, NF160322, NF160323, NF160324, NF160325, NF160326, NF160327, NF160328, NF160329, NF160330, NF160331, NF160332, NF160333, NF160334, NF160335, NF160336, NF160337, NF160338, NF160339, NF160340, NF160341, NF160342, NF160343, NF160344, NF160345, NF160346, NF160347, NF160348, NF160349, NF160350, NF160351, NF160352, NF160353, NF160354, NF160355, NF160356, NF160357, NF160358, NF160359, NF160360, NF160361, NF160362, NF160363, NF160364, NF160365, NF160366, NF160367, NF160368, NF160369, NF160370, NF160371, NF160372, NF160373, NF160374, NF160375, NF160376, NF160379, NF160380, NF160381, NF160382, NF160383, NF160384, NF160385, NF160386, NF160387, NF160388, NF160389, NF160390, NF160391, NF160392, NF160393, NF160394, NF160395, NF160396, NF160397, NF160400, NF160401, NF160402, NF160403, NF160404, NF160405, NF160406, NF160407, NF160408, NF160409, NF160410, NF160411, NF160412, NF160413, NF160414, NF160415, NF160416, NF160417, NF160418, NF160419, NF160420, NF160421, NF160422, NF160423, NF160424, NF160425, NF160426, NF160427, NF160428, NF160429, NF160430, NF160431, NF160432, NF160433, NF160434, and P1011. |
Recalling Firm/ Manufacturer |
Welch Allyn Protocol, Inc 8500 Sw Creekside Pl Beaverton OR 97008
|
| For Additional Information Contact | ZOLL Medical Corporation 503-530-7935 |
Manufacturer Reason for Recall | Potential likelihood that the Portable Intensive Care (PIC) 50 Defibrillator's screen display may fail. The loss of the ability to view the patient's Electrocardiogram (ECG) on the screen could cause the user to delay or fail to administer defibrillating shocks needed to resuscitate a patient. |
FDA Determined Cause 2 | Other |
| Action | On 9/17/10, Welch Allyn began sending out the Important Product Alert letter to all their direct consignees. On 9/24/2010, the firm sent out the letters to their distributors.
The Product Alert letter informs the consignees of a low potential likelihood that the Portable Intensive Care (PIC) 50 Defibrillator's screen display may fail.
The consignees are advised to follow the instruction in the PIC 50 user's manual requiring routine shift equipment checks daily or at each shift change to minimize the likelihood of the malfunction from occurring. The checks ensure that the PIC 50 and its component (including the screen display) are fully operation and undamaged.
The letter further states "our analysis indicates that routinely doing such checks will likely lead to discovery of this malfunction as it occurs, and thus timely maintenance and repair".
The distributors are advised to communicate this Product Alert to their customers if the devices were further distributed.
For customers in US and Canada, distributors are advised to provide the electronic Excel file with customer's name, address, serial number and part number to WA2221@Stericycle.com.
For customers in other countries, they are advised to provide information to PIC50Product Alert@welchallyn.com
Consignees can call Welch Allyn at 800-289-2500 for any questions during normal business hours. |
| Quantity in Commerce | 2002 units to US consignees and 1604 units to foreign consignees |
| Distribution | Worldwide Distribution -- US, including states: AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, and WY and countries of
Australia, Austria, Belgium, Canada, China, Columbia, Cyprus, Denmark, Egypt, Estonia, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Jordan, Kuwait, Kyrgyzstan, Latvia, Lebanon, Liechtenstein, Lithuania, Malaysia, Mexico, Oman, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Switzerland, Syrian Arab Republic, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam, and Yemen. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = DRO
|
|
|
|