Class 2 Device Recall STRYKER NAVIGATION SYSTEM OrthoMap 3D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall STRYKER NAVIGATION SYSTEM OrthoMap 3Dsee related information
Date Initiated by FirmNovember 10, 2010
Date PostedFebruary 25, 2011
Recall Status1 Terminated 3 on June 01, 2011
Recall NumberZ-1414-2011
Recall Event ID 57487
510(K)NumberK062640 
Product Classification Neurological Stereotaxic Instrument - Product Code HAW
ProductStryker Navigation System - OrthoMap 3D Module, Part number 6005-650-000. It provides precise stereostatic determination of surgical targets using stereostatic methodology. The three principle features include computer calculation of stereostatic coordinates from the diagnostic images, measurement of the stereostatic coordinates within the surgical field, high-resolution computer display or diagnostic images with stereostatic coordinates indicated. The system is comprised of software and hardware components.
Code Information Verision 1.0; all batches: K4600F3004; K4L00F3004; K6400F3004; K6K00F3004; K7D00F3004; K8700F3004; KB400F3004; KC400F3004; L1400F3004; L2500F3004; L3800F3004; L5300F3004; and L5J00F3004.
FEI Number 1811755
Recalling Firm/
Manufacturer
Stryker Instruments Division of Stryker Corporation
4100 East Milham Ave.
Kalamazoo MI 49001
For Additional Information ContactStryker Field Action Team
800-200-4263 Ext. 3808
Manufacturer Reason
for Recall
When users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate correlation results or incorrectly shown orientation of every second image series during navigation.
FDA Determined
Cause 2
Software design
ActionUrgent: Medical Device Recall Notification letters were sent on 12/3/2010, and software upgrades were to be performed in the field by the end of 2010. Specific Customer Instructions: 1) Locate the units listed in the notification. 2) Distribute this notification to all departments affected. 3) Return the enclosed business reply form to confirm receipt of this notification and identify how many affected units are currently in your inventory. 4) Once the new software is available, Stryker will dispatch a service tech in order to update the software. Note: Software upgrades for all applications will be available by the end of the year. The upgrades will be scheduled as we receive your business reply form. Note: If you currently have a Time License or Pay Per Use card, a new one will be issued to you at the time of the update. 5) If you have loaned or sold any of the units listed in this notification, please, forward a copy of this notice to the new users and advise Stryker of the new users name and their new location. 6) If you have disposed of any of the affected units and they are no longer in use, please, advise us of their obsolescence by providing us with their serial numbers. Questions regarding this recall should be directed to Stryker Field Action Team at 800-800-4236, ext 3808, 3584, or 4354 between the hours of 8 a.m. - 5 p.m. Monday - Friday.
Quantity in Commerce1,155 total
DistributionWorldwide Distribution -- USA, Canada, Mexico, Australia, Argentina, Brazil, United Kingdom, Japan, Poland, S. Korea, Hong Kong, Taiwan, India, and New Zealand.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = HAW
-
-