| | Class 2 Device Recall ARTISTE MV with Artiste Digital Linear Accelerators |  |
| Date Initiated by Firm | December 03, 2010 |
| Date Posted | February 24, 2011 |
| Recall Status1 |
Terminated 3 on June 30, 2011 |
| Recall Number | Z-1422-2011 |
| Recall Event ID |
57704 |
| 510(K)Number | K072485 |
| Product Classification |
Accelerator, linear, medical - Product Code IYE
|
| Product | Siemens brand ARTISTE MV with Artiste Digital Linear Accelerators in combination with the Flat Panel Positioner; Part No. 8139789;
Intended use: To deliver x-ray radiation for therapeutic treatment of cancer.
Product is manufactured and distributed by Siemens Medical Solutions USA, Inc., Concord, CA; and Siemens Medical Solutions AG, Roentgenstrasse, Kemnath, Germany, and Siemens Medical Solutions AG, Henkestrasse, Erlangen, Germany |
| Code Information |
Serial No. 5647 - 5718 (Specifically 5647,5651,5652,5660,5657, 5664, 5654,5686,5687,5688,5689, 5606, 5690,5700,5701,5699,5704,5705,5707,5708, 5714, 5716, and 5718) |
| FEI Number |
2937457
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc., Concord, CA
4040 Nelson Ave
Concord
CA
94520-1200
|
| For Additional Information Contact | Christine Dunbar 925-246-8407 |
Manufacturer Reason for Recall | The firm become aware of a potential issue with ARTISTE system that may result in patient injury should the positioner belt break, and the cam disc (secondary safety mechanism) is not installed and if the flat panel is over the patient. |
FDA Determined Cause 2 | Other |
| Action | The firm, Siemens, issued a "Customer Information" letter on December 3, 2010 via mail to all its direct consignees/customers. The letter described the product, problem and actions taken by the firm to the customer. Siemens arranged for an inspection of their customers systems to make sure the safety mechanism is fully functional. Siemens instructed the customers to include the Customer Information letter in their ARTISTE System Owner manual chapter 'Safety Advisory Letters' .
If you any questions, please call 925-246-8407. |
| Quantity in Commerce | 20 units |
| Distribution | Worldwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = IYE
|
|
|
|