| | Class 2 Device Recall Infusion Administration Set |  |
| Date Initiated by Firm | January 24, 2011 |
| Date Posted | February 25, 2011 |
| Recall Status1 |
Terminated 3 on September 27, 2011 |
| Recall Number | Z-1440-2011 |
| Recall Event ID |
57829 |
| 510(K)Number | K981816 |
| Product Classification |
Pump, infusion - Product Code FRN
|
| Product | Curlin Infusion Administration Set, REF 340-4114, Sterile EO, Zevex, Inc., 4314 Zevex Park Lane, Salt Lake City, UT 84123.
The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications. |
| Code Information |
***EXPANDED*** Additional Lots Added 05/12/2011: CF1017602, CRF10054001, CRF10056001, CRF10077001, CRF10079001, CRF10081001, CRF10113002, CRF10166001, CRF10167002, CRF10170001, CRF10113002. Lot Codes: CF1019715, CF1028109, CF1028110, CRF10042001, CRF10110001, CRF10121001, CRF10123001, CRF10124001, CRF10132003, CRF10133002, CRF10134001, CRF10163003, CRF10165001, CRF10167002, CRF10168001, CRF10169003, CRF1017501, CRF1017602, CRF1017702, CRF1017901, CRF1018002, D018718, D100319, D100320. |
| FEI Number |
1000117172
|
Recalling Firm/ Manufacturer |
Zevex International, Inc.
4314 Zevex Park Ln
Salt Lake City
UT
84123
|
| For Additional Information Contact | 801-264-1001 Ext. 244 |
Manufacturer Reason for Recall | Defect in tubing may trigger false Air-In-Line warning. |
FDA Determined Cause 2 | Process control |
| Action | The firm, MOOG, sent an "Expansion of Moog/Curlin Ambulatory Pump Administration Set Recall" letter dated May 2, 2011 to its customers. This letter informed the customers that this recall has been expanded to included specific lot numbers of Curlin model ambulatory infusion administration sets. The customers were instructed to follow the outlined process in the letter: remove suspect product code from inventory, contact MOOG Customer Service at 1-800-970-2337, prompt #7, contact your distributor to arrange replacement product, request a shipping return label as per the distributor's process and ship back to their location and follow the temporary solution outlined in the letter. Moog Customer Service will provide a call tag for convenient product return.
If you have any questions contact MOOG Customer Service at 1-800-970-2337.
The firm, MOOG, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated January 24, 2011 to all customers. The letter described the product, problem and actions to be taken by the customers. The customers were instructed to remove suspect product code from inventory; contact their distributor to arrange replacement product, if purchased from distributor and request shipping return label per distributors process and ship back to their location or contact MOOG Customer Service at 800-970-2337, prompt #7, if purchased directly from MOOG and MOOG Customer Service will provide a call tag for convenient product return; follow the temporary solution provided until replacements are administered; and complete and return the MEDICAL DEVICE CORRECTION REPLY FORM in the self addressed envelope provided.
If you have any questions, contact MOOG Customer Service at 800-970-2337. |
| Quantity in Commerce | 246,080 units (US), 260 units (INTL) |
| Distribution | Worldwide Distribution: USA including states of: AR, CA, FL, IA, KS, MS, NE, NY, OH, OK, TN, TX, and UT; and countries including: Australia, Croatia, France, Germany, Kuwait, New Zealand, Qatar, Saudi Arabia, South Africa, United Arab Emirates, and UK. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|