| | Class 2 Device Recall McKesson |  |
| Date Initiated by Firm | January 20, 2011 |
| Date Posted | May 25, 2011 |
| Recall Status1 |
Terminated 3 on October 25, 2012 |
| Recall Number | Z-2250-2011 |
| Recall Event ID |
58021 |
| Product Classification |
Tray, catheterization, sterile urethral, with or without catheter - Product Code FCM
|
| Product | Case Label: McKesson Medi-Pak Irrigation Tray With Thumb-Control Ring Piston Syringe Sterile Single Use Disposable Latex-Free 20 Trays Per Case Reorder No. 37-104 Marketed By McKesson Corporation McKesson Medical-Surgical, Richmond, VA 23228 Made in the U.S.A.
Tray Label: McKESSON Medi-Pak Irrigation Tray With Thumb-Control Ring Piston Syringe Sterile Single Use Disposable Latex-Free Contents: 1-Graduated Tray, 1200 cc 1 - Thumb Control Ring Piston Syringe (Catheter Tip), 60 cc 1- Graduated Solution Container, 500 cc, 1 - Underpad, 1-Sterile Alcohol Prep Pad Reorder No. 37-104 Contents STERILE In unopened, undamaged package. Marketed By McKesson Corporation McKesson Medical-Surgical, Richmond, VA 23228 Made in the U.S.A.
Intended use: The trays are irrigation trays for catheters |
| Code Information |
Lot codes: 807208, 808178, 809213, 810235, 811133, 812197, 901173, 902219, 903177, 904253, 905272, 906275, 908217, 909101, 910153, 911129, 912189, 1001116, 1002026, 1002214, 1004072, 1005140, and 1006116 |
| FEI Number |
3002695476
|
Recalling Firm/ Manufacturer |
Nurse Assist, Inc
3400 Northern Cross Blvd
Fort Worth
TX
76137-3600
|
| For Additional Information Contact | 817-231-1300 Ext. 14 |
Manufacturer Reason for Recall | The irrigation procedure trays contain Triad Group Alcohol Prep Pads recalled by Triad Group because they have the potential to be contaminated with an objectionable organism. |
FDA Determined Cause 2 | Other |
| Action | The firm, Nurse Assist Incorporated, sent a letter dated January 20, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to examine their inventory immediately; quarantine the product; identify their customers and notify them at once about the product recall (Nurse Assist will provide bright labels to affix to your affected inventory); provide their customers with labels as well, and complete and return the attached RECALL ACKNOWLEGEMENT form via fax to: 817-231-1500 or mail to: Nurse Assist, Inc., 3400 Northern Cross Blvd., Forth Worth, TX 76137.
Only the product and lots identified on the attached list are affected.
If you have any questions or to request labels, please call Nurse Assist Customer Service Monday through Friday, from 8:00 am to 5:00 pm Central Time at 1-800-649-6800. |
| Quantity in Commerce | 18,320 cases |
| Distribution | Worldwide distribution: USA (nationwide) and countries including: Canada, Dubai and Saudi Arabia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|