| | Class 2 Device Recall Water Soluble Lubricant |  |
| Date Initiated by Firm | February 07, 2011 |
| Date Posted | April 18, 2011 |
| Recall Status1 |
Terminated 3 on August 05, 2014 |
| Recall Number | Z-2013-2011 |
| Recall Event ID |
58148 |
| 510(K)Number | K871169 |
| Product Classification |
Patient Lubricant - Product Code KMJ
|
| Product | Lubricating Jelly packaged as follows:
(1) DeRoyal(R) Water Soluble Lubricant, Size: 5 Gram, Rx Only, REF 31-702, STERILE EO, Distributed by: DeRoyal - 200 DeBusk Lane, Powell, TN 37849, USA;
(2) DeRoyal(R) Water Soluble Lubricant, 1 / Pack, Size: 2 oz Rx Only, REF 31-704, STERILE EO, Distributor: DeRoyal, 200 DeBusk Lane, Powell, TN 37849, USA;
(3) DeRoyal(R) Lubricant, Size: 5 Gram, Rx Only, REF 31-938, STERILE EO, Distributed by: DeRoyal - 200 DeBusk Lane, Powell, TN 37849, USA
Lubricating body orifices. |
| Code Information |
31-702: 18040107, 18125741, 18150322, 18368014, 18605067, 18605067, 19032694, 19032694, 19136741, 19557491, 20023453, 20102637, 20190729, 20190729, 20023453, 20190729, 20510139, 20023453, 20190729, 21049840, 21049840, 21517201, 21780683, 21780683, 21517201, 21780683, 22025604, 22147169, 21517201, 21517201, 22269799, 22952133, 23474014, 23474014, 23474014, 24194532, 24194532, 24406620, 17950117, 19136741, 21517201, 21780683, 22147169, 22025604, 22595121, 23918771, 24406620, 20190729, 18040107, 18605067, 19032694, 19136741, 20190729, 18150322, 19032694, 20102637, 20190729, 20190729, 21517201, 22269799, 22269799, 24406620, 24406620, 17296856, 17950117, 18368014, 18368014, 18368014, 18605067, 19557491, 19557491, 20510139, 20634166, 21168988, 22952133, 22595121, 22898042, 22952133, 22269799, 24194532, 24194532, 24406620, 19557491, 19032694, 19032694, 20023453, 20023453, 20190729, 20510139, 20634166, 21517201, 21517201, 21517201, 22898042, 21517201, 23918771, 18040107, 18605067, 20023453, 20634166, 21049840, 21168988, 21517201, 22595121, 17950117, 17296856, 19136741, 19032694, 20190729, 20190729, 20190729, 20190729, 20190729, 20023453, 20634166, 21517201, 22025604, 21517201, 21517201, 23918771, 17950117, 18040107, 18368014, 18605067, 18605067, 19032694, 19218034, 19136741, 19557491, 20190729, 20190729, 20190729, 20510139, 21049840, 21049840, 21517201, 21780683, 22025604, 21517201, 21517201, 21517201, 21517201, 22952133, 22595121, 21517201, 23918771, 23474014, 24194532, 23474014, 24892314, 24194532, 22025604, 18040107, 18125741, 18368014, 18605067, 19032694, 19136741, 19218034, 20023453, 20102637, 20190729, 20190729, 20510139, 20634166, 21168988, 21517201, 21588464, 21517201, 20634166, 21517201, 21780683, 22025604, 22595121, 21517201, 22269799, 23474014, 24194532, 24194532, 20023453, 20190729, 20634166, 22595121, 22898042, 21517201, 23474014, 24641355 31-704: 17407430, 18395135, 18605075, 18605075, 19476391, 19476391, 19557520, 20510155, 20914037, 21049866, 21238801, 1397561, 21588481, 21746987, 22269810, 23474022, 23474022, 23750736, 23750736, 24406638, 24406638, 24406638, 17407430, 17407430, 17407430, 17950133, 18234024, 18395143, 18368031, 18368031, 18368031, 18605075, 18962021, 18962021, 18605075, 18962021, 18605075, 18962021, 18605075, 19089121, 19089121, 19476391, 19476391, 19476391, 19476391, 19792802, 19792802, 19557520, 19792802, 19557520, 19792802, 20510155, 20510155, 20510155, 20656576, 20656576, 20656576, 20914037, 20510155, 20656576, 20914037, 21049866, 20656576, 21049866, 21049866, 21238801, 21238801, 21397561, 21588481, 21588481, 21588481, 21746987, 21746987, 21746987, 22037111, 22037111, 22269810, 22531646, 22531646, 22531646, 22758218, 22758218, 22758218, 22269810, 23192682, 23192682, 23474022, 23474022, 23750736, 23474022, 23750736, 24175892, 24175892, 24175892, 24406638, 24406638, 24406638, 24764187, 17950133, 18234024, 18368031, 18368031, 18368031, 18605075, 18605075, 18962021, 18962021, 19089121, 19089121, 19476391, 19792802, 19557520, 19557520, 20510155, 20656576, 20656576, 20656576, 21049866, 21238801, 21238801, 21397561, 21397561, 21588481, 21746987, 21746987, 22037111, 22037111, 22269810, 22531646, 22758218, 22269810, 23474022, 23750736, 23750736, 23750736, 24175892, 23750736, 24175892, 24175892, 24175892, 24175892, 24406638, 24175892, 24175892, 24406638, 24406638, 24764187, 22269810, 18962021, 17407430, 18234024, 18234024, 18368031, 18395135, 18605075, 18962021, 19089121, 19089121, 19476391, 19476391, 19792802, 17407430, 18234024, 19089121, 18605075, 19476391, 20510155, 20656576, 21397561, 21588481, 21588481, 21746987, 22531646, 22758218, 22269810, 22269810, 23192682, 23474022, 23192682, 23474022, 24406638, 24764187, 18605075, 19089121, 19476391, 19557520, 19792802, 20510155, 20914037, 21049866, 21049866, 21049866, 21238801, 23750736, 24406638, 24406638, 24406638, 23192682, 18962021, 21049866, 21588481, 23750736, 17407430, 17950133, 18234024, 18234024, 20510155, 21049866, 21238801, 21746987, 22037111, 20914037, 18605075, 23474022, 22758218, 17407430, 17407430, 17407430, 17950133, 17950133, 18234024, 18368031, 18368031, 18605075, 18962021, 18962021, 19089121, 19089121, 18605075, 18962021, 18605075, 18962021, 19476391, 19476391, 19476391, 19476391, 19792802, 19792802, 19557520, 19792802, 20510155, 20510155, 19557520, 20510155, 20656576, 20656576, 20914037, 20914037, 20656576, 20656576, 21049866, 21049866, 21506318, 21746987, 21746987, 22037111, 22037111, 22269810, 22758218, 22758218, 22758218, 22269810, 22269810, 23192682, 23192682, 23192682, 23474022, 23474022, 23474022, 23474022, 23474022, 23474022, 23750736, 23750736, 24175892, 24175892, 24406638, 24406638, 24406638, 24764187, 24406638, 21397561, 22269810, 23474022, 23750736, 24175892, 24764187, 19792802, 17407430, 17950133, 18234024, 18395143, 18368031, 18605075, 18962021, 18962021, 19089121, 19089121, 18605075, 19476391, 19476391, 19792802, 19792802, 19557520, 19557520, 19792802, 20510155, 20510155, 20510155, 20510155, 20656576, 20656576, 20656576, 20914037, 20656576, 21238801, 21506318, 21397561, 21397561, 21588481, 21397561, 22037111, 22531646, 22531646, 22758218, 22269810, 22269810, 22269810, 23192682, 23192682, 23192682, 23192682, 23474022, 23474022, 23474022, 23750736, 23474022, 23750736, 23750736, 24175892, 24175892, 23750736, 24406638, 24406638, 24406638, 24406638, 24764187, 23192682, 17407430, 18605075, 18605075, 18605075, 19476391, 19476391, 20510155, 20914037, 21506318, 21588481, 22037111, 22758218, 22269810, 23474022, 23750736, 24175892, 24764187, 17950133, 18234024, 18368031, 18234024, 18605075, 19089121, 19089121, 18605075, 19476391, 1792802, 19557520, 20510155, 20510155, 20510155, 20914037, 21238801, 21238801, 21506318, 21397561, 21588481, 21746987, 21746987, 21746987, 22037111, 22269810, 22531646, 22758218, 22269810, 22758218, 22269810, 22531646, 23192682, 23192682, 23750736, 23750736, 24175892, 24175892, 24764187, 18605075, 17950133, 17950133, 18395143, 18368031, 18962021, 18605075, 19089121, 18605075, 19476391, 18605075, 19557520, 19557520, 20510155, 20510155, 20656576, 20656576, 20656576, 21238801, 21397561, 21588481, 21746987, 22269810, 22531646, 22758218, 22269810, 23192682, 23474022, 23750736, 24406638, 24406638, 24406638, 24764187, 19476391, 21049866, 22037111, 17950133, 18395143, 18368031, 18605075, 17950133, 18368031, 18605075, 19089121, 19476391, 19557520, 20510155, 20656576, 21238801, 21397561, 21588481, 21746987, 22037111, 22758218, 22758218, 23192682, 23750736, 24175892, 24406638, 18605075, 20510155, 20510155, 20656576, 21049866, 21049866, 21746987, 21746987, 22269810, 23192682, 23192682, 24406638, 17950133, 18368031, 18605075, 19792802, 20510155, 20914037, 21049866, 21397561, 21588481, 21588481, 21588481, 22269810, 23192682, 23192682, 23192682, 23474022, 23750736, 24175892, 24175892, 24406638, 24764187, 24764187, 18234024, 18234024, 18234024, 18234024, 18962021, 19089121, 19476391, 19476391, 19557520, 19557520, 20510155, 20510155, 20510155, 22758218, 23750736, 19476391, 19089121, 19476391, 19089121, 19476391, 20510155, 20510155, 20656576, 21238801, 18605075, 22531646, 22758218, 20656576, 22037111, 17407430, 17950133, 18395143, 18234024, 18368031, 18368031, 18605075, 18605075, 18605075, 18605075, 18605075, 18962021, 19089121, 18962021, 18605075, 18605075, 19476391, 19476391, 19089121, 19792802, 19557520, 19557520, 20510155, 20510155, 20510155, 20510155, 20914037, 20914037, 18234024, 20656576, 20914037, 21049866, 21049866, 20656576, 21049866, 21238801, 21238801, 21397561, 21397561, 21397561, 21746987, 21746987, 21746987, 22037111, 22269810, 22531646, 22531646, 22531646, 22758218, 22758218, 22269810, 22269810, 22269810, 23192682, 23192682, 23474022, 23474022, 23474022, 23750736, 23750736, 23750736, 23750736, 23750736, 24175892, 24175892, 24406638, 24406638, 24406638, 24406638, 24764187, 17407430, 23192682, 23474022, 24175892, 24175892, 21397561, 21588481, 21746987, 22037111, 22037111, 22269810, 22269810, 22269810, 22531646, 22269810, 23474022, 24406638, 24764187, 24406638, 24764187, 17407430, 18234024, 18368031, 18395135, 18395135, 18605075, 18605075, 18962021, 18962021, 18962021, 18962021, 19089121, 19089121, 19476391, 19476391, 19792802, 20510155, 20510155, 20656576, 20656576, 20656576, 21238801, 21397561, 21506318, 17950133, 17950133, 17950133, 18234024, 18368031, 18605075, 18605075, 18605075, 18962021, 18962021, 19089121, 19089121, 18605075, 19476391, 19476391, 19792802, 19557520, 19557520, 20510155, 20510155, 20510155, 20656576, 20914037, 20914037, 24406638, 24764187, 18962021, 19476391, 19792802, 19792802, 20510155, 20656576, 21588481, 21746987, 22037111, 22531646, 23192682, 23474022, 23750736, 18234024, 19476391, 21238801, 22037111, 23192682, 24175892, 21397561, 21746987, 22531646, 22269810, 23474022, 24175892, 21049866 31-938: 18368057, 17070224, 17407448, 17407448, 18368057, 20656584, 21588499, 22147193, 20510180, 21747007, 24194559, 18500214, 20656584, 22269836, 23750744, 21747007, 24046927, 24194559, 24406646, 20069646, 21672797, 21672797, 18368057, 24406646, 17407448, 17407448, 18368057, 18500214, 18500214, 18500214, 19089148, 19089148, 19089148, 19089148, 19089148, 18500214, 20069646, 19940879, 20069646, 20510180, 20656584, 20656584, 20656584, 20069646, 20510180, 21672797, 21747007, 21747007, 21907509, 22147193, 22269836, 21907509, 22269836, 22853565, 21747007, 21907509, 23382419, 23750744, 24046927, 24194559, 24406646, 24194559, 17407448, 18500214, 18500214, 19089148, 20069646, 19940879, 20656584, 21672797, 22269836, 23750744, 19089148, 20656584, and 22269836. |
| FEI Number |
1000208624
|
Recalling Firm/ Manufacturer |
DeRoyal Processing Center 1703 Highway 33 S New Tazewell TN 37825-5043
|
| For Additional Information Contact | Mr. Tracy Edmundson 865-362-2334 |
Manufacturer Reason for Recall | Possible contamination due to sterilization issue. Inspection of contract manufacturer found cGMP deficiencies. |
FDA Determined Cause 2 | Process control |
| Action | The firm notified their consignees by a Re: DeRoyal Recall Of Triad Group Lubricating Jelly letter, dated Friday, March 18,2011. The letter discussed the reason for recall and the affected products. The letter asked customers to identify inventory using the attached spreadsheet and quarantine to prevent further use. Customers are to complete the spreadsheet regardless if they have effected product or not. The completed form can be faxed to 865-362-3716 or emailed to recalls@deroyal.com. All affected products should be return to DeRoyal at the address provided. Distributors are to forward a copy of the notice to their end-users. If there are questions or assistance is needed with the recall, Stephanie DeBusk should be contacted at 865-362-1037 or Sharon Cook at 865-362-6157. |
| Quantity in Commerce | 323,440 units |
| Distribution | Nationwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KMJ
|
|
|
|