 Class 2 Device Recall Catheter Insertion Kitsee related information
Date Initiated by FirmMarch 24, 2011
Date PostedMay 03, 2011
Recall Status1 Terminated 3 on February 28, 2012
Recall NumberZ-2133-2011
Recall Event ID 58278
Product Classification Catheter care tray - Product Code OHR
ProductCatheter Insertion Kit (containing, Under pad, PVP Pad, BZK Prep Pad, Sterile Lubricating Jelly, Vinyl Gloves and Poly Zipper Bag). Intended usage: Kit components to be used during Intermittent Urinary Catheterizations
Code Information Catalog Reference number: 6100A, Lot numbers: 8315, 9317, 0317, 9G176, 0C046, 0G226, 0K186, 1A316, 8316, 9318, 0318, 9G206, 0C086, 0G236, 0K196, 1B016, 8317, 9320, 0319, 9G236, 0C096, 0G246, 0K206, 1B026, 8319, 9321, 0320, 9G286, 0C106, 0G296, 0K226, 1B036, 8320, 9324, 0321, 9G296, 0C186, 0G306, 0K236, 1B046, 8321, 9325, 0322, 9H066, 0C196, 0H216, 0K246, 1B056, 8324, 9326, 0323, 9H076, 0C206, 0H226, 0K256, 1B076, 8325, 9327, 0324, 9H086, 0C226, 0H266, 0K286, 1B186, 8326, 9328, 0325, 9H126, 0C236, 0H276, 0K296, 1B196, 8327, 9329, 0326, 9H176, 0C256, 0H286, 0K306, 1B176, 8329, 9340, 0327, 9H236, 0C266, 0I166, 0L036, 1B196, 8340, 0301, 0328, 9H246, 0C276, 0I176, 0L046, 1B216, 8341, 0302, 0329, 9H256, 0D076, 0I186, 0L066, 1B226, 8342, 0303, 0330, 0A186, 0D086, 0I206, 0L076, 1B236, 9301, 0304, 0331, 0A196, 0D096,0I256, 0L086, 1C016, 9302, 0305, 0332, 0A206, 0D106, 0I276, 0L096, 1C026, 9304, 0306, 1302, 0A216, 0D126, 0J236, 0L226, 1C036, 9305, 0307, 1303, 0A226, 0E086, 0J256, A196 and 1C046.
FEI Number 3007829657
Recalling Firm/
Manufacturer
Apogee Medical, Inc
90 Weathers St.
Youngsville NC 27596-7801
For Additional Information ContactDiane N. Peper
919-570-9605 Ext. 224
Manufacturer Reason
for Recall
Kits contain PVP (Providone Iodine) prep pads that are being recalled by the Triad Group due to possible micro contamination, Elizabethkingia meningoseptica.
FDA Determined
Cause 2
Other
ActionThe firm, Apogee Medical LLC, sent an "Urgent Medical Device Recall" letter dated March 23, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to review their inventory; complete and return the enclosed Product Recall Response Form as soon as possible via fax to (919) 570-9611, mail to: Apogee Medical LLC, Youngsville Commerce Center, 90 Weathers Street, Youngsville, MC 27596, or e-mail to: dpeper@apogeemed.com; cease further distribution and return any product that is the subject of this recall to Apogee Medical LLC. All returned product must be labeled with a returned goods authorization label that has been provided with the letter. Apogee will replace returned product upon receipt of the Product Recall Response Form and identified product. In addition, if the customer distributed the product, they are to identify their customer(s) and notify them at once of this product recall. The notification should include a copy of the this recall notification letter and should instruct the customer to return the recalled product so that it can be returned to Apogee. Should you have any questions regarding this notification, please call the Manager Regulatory Affairs/Quality Assurance at (919) 435-5409.
Quantity in Commerce5,503,430 units
DistributionNationwide distribution: USA including states of: AL, AZ, CA, FL, GA, IL, IN, KS, LA, MA, MO, NJ, NY, OH, OK, TN, TX, VA and WA.
Total Product Life CycleTPLC Device Report