| | Class 2 Device Recall VARIAN HIGH ENERGY LINEAR ACCELERATOR |  |
| Date Initiated by Firm | March 28, 2011 |
| Date Posted | May 13, 2011 |
| Recall Status1 |
Terminated 3 on June 07, 2013 |
| Recall Number | Z-2195-2011 |
| Recall Event ID |
58438 |
| 510(K)Number | K100890 |
| Product Classification |
Accelerator, linear, medical - Product Code IYE
|
| Product | Varian brand Clinac, Medical Linear Accelerator, All Varian Clinac with Model Numbers: H14, H18, H26, H27, H28, H29, Hcx; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA
The Trilogy Radiotherapy Delivery System is a radiation therapy accelerator intended to deliver megavoltage x-ray treatments for conventional radiotherapy (three dimensional conformal radiotherapy and intensity modulated radiotherapy) and stereotactic radiosurgery and radiotherapy. Stereotactic treatments are intended for therapy of lesions, e.g., arteriovenous malformations, primary tumors and metastases. |
| Code Information |
LIST OF SERIAL NUMBERS: H140001, H140048, H140095, H140141, H140232, H140736, H140781, H140826, H140871, H180004, H180108, H140002, H140049, H140096, H140142, H140233, H140737, H140782, H140827, H140872, H180007, H180115, H140003, H140051 ,H140097 ,H140143 ,H140234 ,H140738 ,H140783 ,H140828 ,H140873 ,H180008 ,H180116 ,H140004 ,H140052 ,H140098 ,H140144 ,H140235 ,H140739 ,H140784 ,H140829 ,H140874 ,H180016 ,H180117 ,H140005 ,H140053 ,H140099 ,H140145 ,H140236 ,H140740 ,H140785 ,H140830 ,H140875 ,H180017 ,H180119 ,H140006 ,H140054 ,H140100 ,H140146 ,H140237 ,H140741 ,H140786 ,H140831 ,H140876 ,H180019 ,H180122 ,H140007 ,H140055 ,H140101 ,H140147 ,H140238 ,H140742 ,H140787 ,H140832 ,H140877 ,H180025 ,H180124 ,H140008 ,H140056 ,H140102 ,H140148 ,H140239 ,H140743 ,H140788 ,H140833 ,H140878 ,H180026 ,H180128 ,H140009 ,H140057 ,H140103 ,H140149 ,H140240 ,H140744 ,H140789 ,H140834 ,H140879 ,H180028 ,H180129 ,H140010 ,H140058 ,H140105 ,H140150 ,H140700 ,H140745 ,H140790 ,H140835 ,H140881 ,H180034 ,H180131 ,H140011 ,H140059 ,H140106 ,H140151 ,H140701 ,H140746 ,H140791 ,H140836 ,H140882 ,H180040 ,H180132 ,H140012 ,H140060 ,H140107 ,H140152 ,H140702 ,H140747 ,H140792 ,H140837 ,H140883 ,H180043 ,H180133 ,H140014 ,H140061 ,H140108 ,H140153 ,H140703 ,H140748 ,H140793 ,H140838 ,H140884 ,H180044 ,H180135 ,H140015 ,H140062 ,H140109 ,H140154 ,H140704 ,H140749 ,H140794 ,H140839 ,H140886 ,H180045 ,H180143 ,H140016 ,H140063 ,H140110 ,H140201 ,H140705 ,H140750 ,H140795 ,H140840 ,H140887 ,H180048 ,H180144 ,H140017 ,H140064 ,H140111 ,H140202 ,H140706 ,H140751 ,H140796 ,H140841 ,H140889 ,H180049 ,H180147 ,H140018 ,H140065 ,H140112 ,H140203 ,H140707 ,H140752 ,H140797 ,H140842 ,H140890 ,H180050 ,H180148 ,H140019 ,H140066 ,H140113 ,H140204 ,H140708 ,H140753 ,H140798 ,H140843 ,H140891 ,H180052 ,H180151 ,H140021 ,H140067 ,H140114 ,H140205 ,H140709 ,H140754 ,H140799 ,H140844 ,H140892 ,H180053 ,H180154 ,H140022 ,H140068 ,H140115 ,H140206 ,H140710 ,H140755 ,H140800 ,H140845 ,H140893 ,H180056 ,H180157 ,H140023 ,H140069 ,H140116 ,H140207 ,H140711 ,H140756 ,H140801 ,H140846 ,H140895 ,H180058 ,H180158 ,H140024 ,H140070 ,H140117 ,H140208 ,H140712 ,H140757 ,H140802 ,H140847 ,H140896 ,H180059 ,H180159 ,H140025 ,H140071 ,H140118 ,H140209 ,H140713 ,H140758 ,H140803 ,H140848 ,H140897 ,H180062 ,H180160 ,H140026 ,H140072 ,H140119 ,H140210 ,H140714 ,H140759 ,H140804 ,H140849 ,H140898 ,H180065 ,H180161 ,H140027 ,H140073 ,H140120 ,H140211 ,H140715 ,H140760 ,H140805 ,H140850 ,H140899 ,H180066 ,H180164 ,H140028 ,H140074 ,H140121 ,H140212 ,H140716 ,H140761 ,H140806 ,H140851 ,H140900 ,H180067 ,H180166 ,H140029 ,H140075 ,H140122 ,H140213 ,H140717 ,H140762 ,H140807 ,H140852 ,H140901 ,H180069 ,H180167 ,H140030 ,H140076 ,H140123 ,H140214 ,H140718 ,H140763 ,H140808 ,H140853 ,H140902 ,H180073 ,H180168 ,H140031 ,H140077 ,H140124 ,H140215 ,H140719 ,H140764 ,H140809 ,H140854 ,H140903 ,H180075 ,H180170 ,H140032 ,H140078 ,H140125 ,H140216 ,H140720 ,H140765 ,H140810 ,H140855 ,H140904 ,H180077 ,H180171 ,H140033 ,H140079 ,H140126 ,H140217 ,H140721 ,H140766 ,H140811 ,H140856 ,H140905 ,H180079 ,H180173 ,H140034 ,H140080 ,H140127 ,H140218 ,H140722 ,H140767 ,H140812 ,H140857 ,H140906 ,H180080 ,H180175 ,H140035 ,H140082 ,H140128 ,H140219 ,H140723 ,H140768 ,H140813 ,H140858 ,H140907 ,H180082 ,H180178 ,H140036 ,H140083 ,H140129 ,H140220 ,H140724 ,H140769 ,H140814 ,H140859 ,H140908 ,H180083 ,H180179 ,H140037 ,H140084 ,H140130 ,H140221 ,H140725 ,H140770 ,H140815 ,H140860 ,H140909 ,H180087 ,H180180 ,H140038 ,H140085 ,H140131 ,H140222 ,H140726 ,H140771 ,H140816 ,H140861 ,H140910 ,H180088 ,H180181 ,H140039 ,H140086 ,H140132 ,H140223 ,H140727 ,H140772 ,H140817 ,H140862 ,H140911 ,H180089 ,H180182 ,H140040 ,H140087 ,H140133 ,H140224 ,H140728 ,H140773 ,H140818 ,H140863 ,H140912 ,H180093 ,H180185 ,H140041 ,H140088 ,H140134 ,H140225 ,H140729 ,H140774 ,H140819 ,H140864 ,H140913 ,H180098 ,H180186 ,H140042 ,H140089 ,H140135 ,H140226 ,H140730 ,H140775 ,H140820 ,H140865 ,H140914 ,H180099 ,H180188 ,H140043 ,H140090 ,H140136 ,H140227 ,H140731 ,H140776 ,H140821 ,H140866 ,H140915 ,H180102 ,H180192 ,H140044 ,H140091 ,H140137 ,H140228 ,H140732 ,H140777 ,H140822 ,H140867 ,H140916 ,H180103 ,H180193 ,H140045 ,H140092 ,H140138 ,H140229 ,H140733 ,H140778 ,H140823 ,H140868 ,H140917 ,H180104 ,H180195 ,H140046 ,H140093 ,H140139 ,H140230 ,H140734 ,H140779 ,H140824 ,H140869 ,H140919 ,H180106 ,H180197 ,H140047 ,H140094 ,H140140 ,H140231 ,H140735 ,H140780 ,H140825 ,H140870 ,H180003 ,H180107 ,H180199 ,H180201 ,H180283 ,H180414 ,H180476 ,H180538 ,H180590 ,H180639 ,H180689 ,H180739 ,H180788 ,H180839 ,H180203 ,H180284 ,H180415 ,H180477 ,H180539 ,H180591 ,H180640 ,H180690 ,H180740 ,H180790 ,H180840 ,H180204 ,H180285 ,H180417 ,H180478 ,H180540 ,H180592 ,H180641 ,H180691 ,H180741 ,H180791 ,H180841 ,H180206 ,H180287 ,H180419 ,H180479 ,H180541 ,H180593 ,H180642 ,H180692 ,H180742 ,H180792 ,H180842 ,H180208 ,H180289 ,H180420 ,H180480 ,H180542 ,H180594 ,H180643 ,H180693 ,H180743 ,H180793 ,H180843 ,H180209 ,H180290 ,H180422 ,H180482 ,H180543 ,H180595 ,H180644 ,H180694 ,H180744 ,H180794 ,H180844 ,H180210 ,H180291 ,H180424 ,H180484 ,H180544 ,H180596 ,H180645 ,H180695 ,H180745 ,H180795 ,H180845 ,H180214 ,H180292 ,H180425 ,H180485 ,H180545 ,H180597 ,H180646 ,H180696 ,H180746 ,H180796 ,H180846 ,H180215 ,H180293 ,H180426 ,H180486 ,H180547 ,H180598 ,H180647 ,H180697 ,H180747 ,H180798 ,H180847 ,H180217 ,H180294 ,H180427 ,H180487 ,H180548 ,H180599 ,H180648 ,H180698 ,H180748 ,H180799 ,H180848 ,H180218 ,H180295 ,H180430 ,H180489 ,H180549 ,H180600 ,H180649 ,H180699 ,H180749 ,H180800 ,H180849 ,H180220 ,H180296 ,H180432 ,H180490 ,H180550 ,H180601 ,H180650 ,H180700 ,H180750 ,H180801 ,H180850 ,H180221 ,H180298 ,H180433 ,H180491 ,H180551 ,H180602 ,H180651 ,H180701 ,H180751 ,H180802 ,H180851 ,H180222 ,H180299 ,H180436 ,H180492 ,H180552 ,H180603 ,H180652 ,H180702 ,H180752 ,H180803 ,H180852 ,H180225 ,H180303 ,H180437 ,H180493 ,H180553 ,H180604 ,H180653 ,H180703 ,H180753 ,H180804 ,H180853 ,H180226 ,H180304 ,H180440 ,H180494 ,H180554 ,H180605 ,H180654 ,H180704 ,H180754 ,H180805 ,H180854 ,H180228 ,H180305 ,H180441 ,H180495 ,H180555 ,H180606 ,H180655 ,H180705 ,H180755 ,H180806 ,H180855 ,H180229 ,H180307 ,H180442 ,H180496 ,H180556 ,H180607 ,H180656 ,H180706 ,H180756 ,H180807 ,H180856 ,H180230 ,H180309 ,H180443 ,H180498 ,H180557 ,H180608 ,H180657 ,H180707 ,H180757 ,H180808 ,H180857 ,H180233 ,H180310 ,H180444 ,H180499 ,H180559 ,H180610 ,H180658 ,H180708 ,H180758 ,H180809 ,H180858 ,H180235 ,H180311 ,H180445 ,H180501 ,H180560 ,H180611 ,H180659 ,H180709 ,H180759 ,H180810 ,H180859 ,H180236 ,H180312 ,H180446 ,H180502 ,H180561 ,H180612 ,H180660 ,H180710 ,H180760 ,H180811 ,H180860 ,H180237 ,H180314 ,H180447 ,H180505 ,H180562 ,H180613 ,H180661 ,H180711 ,H180761 ,H180812 ,H180861 ,H180238 ,H180315 ,H180449 ,H180507 ,H180563 ,H180614 ,H180662 ,H180713 ,H180762 ,H180813 ,H180862 ,H180239 ,H180318 ,H180450 ,H180508 ,H180564 ,H180615 ,H180663 ,H180714 ,H180763 ,H180814 ,H180863 ,H180240 ,H180319 ,H180451 ,H180509 ,H180565 ,H180616 ,H180664 ,H180715 ,H180764 ,H180815 ,H180864 ,H180242 ,H180321 ,H180452 ,H180510 ,H180566 ,H180617 ,H180665 ,H180716 ,H180765 ,H180816 ,H180865 ,H180243 ,H180322 ,H180453 ,H180511 ,H180567 ,H180618 ,H180667 ,H180717 ,H180766 ,H180817 ,H180866 ,H180245 ,H180323 ,H180454 ,H180513 ,H180569 ,H180619 ,H180668 ,H180718 ,H180767 ,H180818 ,H180867 ,H180246 ,H180324 ,H180455 ,H180514 ,H180570 ,H180620 ,H180669 ,H180719 ,H180768 ,H180819 ,H180868 ,H180249 ,H180325 ,H180456 ,H180515 ,H180571 ,H180621 ,H180670 ,H180720 ,H180769 ,H180820 ,H180869 ,H180250 ,H180326 ,H180457 ,H180518 ,H180572 ,H180622 ,H180671 ,H180721 ,H180770 ,H180821 ,H180870 ,H180251 ,H180327 ,H180458 ,H180519 ,H180573 ,H180623 ,H180672 ,H180722 ,H180771 ,H180822 ,H180871 ,H180254 ,H180328 ,H180459 ,H180520 ,H180574 ,H180624 ,H180673 ,H180723 ,H180772 ,H180823 ,H180872 ,H180255 ,H180329 ,H180460 ,H180521 ,H180575 ,H180625 ,H180674 ,H180724 ,H180773 ,H180824 ,H180873 ,H180256 ,H180330 ,H180461 ,H180522 ,H180576 ,H180626 ,H180675 ,H180725 ,H180774 ,H180825 ,H180874 ,H180257 ,H180401 ,H180462 ,H180524 ,H180577 ,H180627 ,H180676 ,H180726 ,H180775 ,H180826 ,H180875 ,H180258 ,H180402 ,H180463 ,H180525 ,H180578 ,H180628 ,H180678 ,H180727 ,H180776 ,H180827 ,H180876 ,H180260 ,H180403 ,H180466 ,H180527 ,H180579 ,H180629 ,H180679 ,H180728 ,H180777 ,H180828 ,H180877 ,H180262 ,H180404 ,H180467 ,H180528 ,H180580 ,H180630 ,H180680 ,H180729 ,H180778 ,H180829 ,H180878 ,H180264 ,H180405 ,H180468 ,H180529 ,H180581 ,H180631 ,H180681 ,H180730 ,H180779 ,H180830 ,H180879 ,H180267 ,H180406 ,H180469 ,H180530 ,H180582 ,H180632 ,H180682 ,H180731 ,H180780 ,H180831 ,H180880 ,H180268 ,H180407 ,H180470 ,H180531 ,H180583 ,H180633 ,H180683 ,H180733 ,H180781 ,H180832 ,H180881 ,H180270 ,H180409 ,H180471 ,H180532 ,H180584 ,H180634 ,H180684 ,H180734 ,H180782 ,H180833 ,H180882 ,H180271 ,H180410 ,H180472 ,H180533 ,H180585 ,H180635 ,H180685 ,H180735 ,H180783 ,H180834 ,H180883 ,H180272 ,H180411 ,H180473 ,H180534 ,H180586 ,H180636 ,H180686 ,H180736 ,H180784 ,H180836 ,H180884 ,H180273 ,H180412 ,H180474 ,H180535 ,H180587 ,H180637 ,H180687 ,H180737 ,H180785 ,H180837 ,H180885 ,H180275 ,H180413 ,H180475 ,H180536 ,H180588 ,H180638 ,H180688 ,H180738 ,H180786 ,H180838 ,H180886 ,H180887 ,H181034 ,H181083 ,H181137 ,H181196 ,H270216 ,H270324 ,H270420 ,H270495 ,H270624 ,H270698 ,H180888 ,H181035 ,H181084 ,H181138 ,H182000 ,H270219 ,H270329 ,H270421 ,H270498 ,H270625 ,H270699 ,H180889 ,H181036 ,H181085 ,H181139 ,H182002 ,H270226 ,H270332 ,H270425 ,H270499 ,H270626 ,H270700 ,H180890 ,H181037 ,H181086 ,H181140 ,H182006 ,H270227 ,H270335 ,H270426 ,H270500 ,H270627 ,H270702 ,H180891 ,H181038 ,H181087 ,H181141 ,H182007 ,H270228 ,H270337 ,H270427 ,H270501 ,H270631 ,H270703 ,H180892 ,H181039 ,H181088 ,H181142 ,H182009 ,H270229 ,H270338 ,H270428 ,H270502 ,H270632 ,H270705 ,H180894 ,H181040 ,H181089 ,H181143 ,H182010 ,H270230 ,H270341 ,H270430 ,H270503 ,H270633 ,H270708 ,H180895 ,H181041 ,H181090 ,H181145 ,H182012 ,H270232 ,H270342 ,H270431 ,H270504 ,H270634 ,H270709 ,H180896 ,H181042 ,H181091 ,H181146 ,H270004 ,H270233 ,H270343 ,H270433 ,H270505 ,H270635 ,H270710 ,H180897 ,H181043 ,H181092 ,H181147 ,H270006 ,H270235 ,H270344 ,H270435 ,H270506 ,H270637 ,H270711 ,H180898 ,H181044 ,H181093 ,H181148 ,H270009 ,H270237 ,H270345 ,H270436 ,H270507 ,H270639 ,H270712 ,H180899 ,H181045 ,H181094 ,H181149 ,H270010 ,H270240 ,H270346 ,H270437 ,H270508 ,H270641 ,H270713 ,H180901 ,H181046 ,H181095 ,H181151 ,H270021 ,H270242 ,H270349 ,H270438 ,H270510 ,H270642 ,H270714 ,H180907 ,H181047 ,H181096 ,H181152 ,H270026 ,H270244 ,H270351 ,H270439 ,H270511 ,H270643 ,H270715 ,H181000 ,H181048 ,H181097 ,H181153 ,H270027 ,H270245 ,H270352 ,H270443 ,H270512 ,H270644 ,H270716 ,H181001 ,H181049 ,H181098 ,H181154 ,H270029 ,H270251 ,H270353 ,H270445 ,H270513 ,H270645 ,H270717 ,H181002 ,H181050 ,H181100 ,H181155 ,H270030 ,H270252 ,H270355 ,H270446 ,H270514 ,H270646 ,H270718 ,H181003 ,H181051 ,H181101 ,H181156 ,H270031 ,H270254 ,H270356 ,H270447 ,H270516 ,H270647 ,H270719 ,H181004 ,H181052 ,H181102 ,H181157 ,H270038 ,H270255 ,H270360 ,H270448 ,H270518 ,H270648 ,H270721 ,H181005 ,H181053 ,H181103 ,H181158 ,H270045 ,H270256 ,H270361 ,H270449 ,H270520 ,H270649 ,H270722 ,H181006 ,H181055 ,H181104 ,H181159 ,H270047 ,H270260 ,H270364 ,H270450 ,H270522 ,H270651 ,H270723 ,H181007 ,H181056 ,H181106 ,H181160 ,H270068 ,H270261 ,H270365 ,H270451 ,H270525 ,H270652 ,H270725 ,H181008 ,H181057 ,H181107 ,H181161 ,H270073 ,H270264 ,H270370 ,H270452 ,H270526 ,H270653 ,H270726 ,H181009 ,H181058 ,H181108 ,H181162 ,H270074 ,H270267 ,H270371 ,H270454 ,H270529 ,H270654 ,H270727 ,H181010 ,H181059 ,H181109 ,H181163 ,H270075 ,H270270 ,H270373 ,H270455 ,H270530 ,H270656 ,H270728 ,H181011 ,H181060 ,H181110 ,H181164 ,H270079 ,H270271 ,H270375 ,H270457 ,H270532 ,H270658 ,H270729 ,H181012 ,H181061 ,H181112 ,H181165 ,H270084 ,H270273 ,H270377 ,H270458 ,H270534 ,H270660 ,H270730 ,H181013 ,H181062 ,H181113 ,H181166 ,H270088 ,H270274 ,H270378 ,H270459 ,H270535 ,H270661 ,H270732 ,H181014 ,H181063 ,H181114 ,H181167 ,H270094 ,H270277 ,H270380 ,H270461 ,H270539 ,H270662 ,H270733 ,H181015 ,H181064 ,H181115 ,H181168 ,H270096 ,H270280 ,H270384 ,H270462 ,H270540 ,H270663 ,H270735 ,H181016 ,H181065 ,H181116 ,H181170 ,H270101 ,H270281 ,H270385 ,H270463 ,H270542 ,H270664 ,H270736 ,H181017 ,H181066 ,H181117 ,H181171 ,H270106 ,H270282 ,H270386 ,H270464 ,H270544 ,H270665 ,H270737 ,H181018 ,H181067 ,H181118 ,H181173 ,H270107 ,H270284 ,H270391 ,H270466 ,H270546 ,H270667 ,H270739 ,H181019 ,H181068 ,H181119 ,H181174 ,H270110 ,H270285 ,H270392 ,H270468 ,H270547 ,H270669 ,H270740 ,H181020 ,H181069 ,H181120 ,H181175 ,H270111 ,H270288 ,H270395 ,H270470 ,H270550 ,H270670 ,H270741 ,H181021 ,H181070 ,H181123 ,H181176 ,H270118 ,H270290 ,H270396 ,H270471 ,H270601 ,H270671 ,H270743 ,H181022 ,H181071 ,H181124 ,H181177 ,H270126 ,H270293 ,H270397 ,H270473 ,H270602 ,H270673 ,H270746 ,H181023 ,H181072 ,H181125 ,H181178 ,H270128 ,H270297 ,H270398 ,H270474 ,H270603 ,H270675 ,H270749 ,H181024 ,H181073 ,H181126 ,H181179 ,H270134 ,H270302 ,H270399 ,H270476 ,H270605 ,H270680 ,H270751 ,H181025 ,H181074 ,H181127 ,H181180 ,H270136 ,H270303 ,H270400 ,H270477 ,H270606 ,H270681 ,H270752 ,H181026 ,H181075 ,H181128 ,H181181 ,H270138 ,H270307 ,H270401 ,H270479 ,H270607 ,H270682 ,H270755 ,H181027 ,H181076 ,H181129 ,H181182 ,H270141 ,H270310 ,H270404 ,H270482 ,H270613 ,H270684 ,H270757 ,H181028 ,H181077 ,H181130 ,H181183 ,H270144 ,H270311 ,H270409 ,H270483 ,H270615 ,H270685 ,H270759 ,H181029 ,H181078 ,H181132 ,H181184 ,H270201 ,H270314 ,H270410 ,H270484 ,H270616 ,H270686 ,H270760 ,H181030 ,H181079 ,H181133 ,H181185 ,H270209 ,H270316 ,H270412 ,H270486 ,H270617 ,H270690 ,H270761 ,H181031 ,H181080 ,H181134 ,H181187 ,H270211 ,H270320 ,H270416 ,H270487 ,H270618 ,H270692 ,H270762 ,H181032 ,H181081 ,H181135 ,H181191 ,H270214 ,H270321 ,H270418 ,H270489 ,H270622 ,H270693 ,H270763 ,H181033 ,H181082 ,H181136 ,H181195 ,H270215 ,H270323 ,H270419 ,H270490 ,H270623 ,H270697 ,H270766 ,H270767 ,H270841 ,H270897 ,H271066 ,H271123 ,H271185 ,H271236 ,H271296 ,H271348 ,H271402 ,H271451 ,H270768 ,H270842 ,H270898 ,H271067 ,H271124 ,H271186 ,H271238 ,H271297 ,H271349 ,H271403 ,H271453 ,H270769 ,H270844 ,H270899 ,H271068 ,H271125 ,H271188 ,H271239 ,H271298 ,H271350 ,H271404 ,H271454 ,H270777 ,H270845 ,H271000 ,H271070 ,H271126 ,H271189 ,H271241 ,H271299 ,H271351 ,H271406 ,H271455 ,H270778 ,H270846 ,H271002 ,H271071 ,H271127 ,H271190 ,H271242 ,H271300 ,H271353 ,H271407 ,H271456 ,H270779 ,H270847 ,H271004 ,H271073 ,H271130 ,H271191 ,H271243 ,H271301 ,H271355 ,H271408 ,H271457 ,H270781 ,H270848 ,H271005 ,H271075 ,H271131 ,H271192 ,H271245 ,H271302 ,H271356 ,H271409 ,H271458 ,H270782 ,H270849 ,H271006 ,H271076 ,H271132 ,H271193 ,H271247 ,H271303 ,H271357 ,H271410 ,H271459 ,H270783 ,H270850 ,H271008 ,H271079 ,H271135 ,H271194 ,H271248 ,H271304 ,H271358 ,H271411 ,H271460 ,H270785 ,H270852 ,H271009 ,H271080 ,H271136 ,H271195 ,H271249 ,H271305 ,H271359 ,H271412 ,H271461 ,H270786 ,H270854 ,H271013 ,H271082 ,H271137 ,H271196 ,H271250 ,H271306 ,H271360 ,H271413 ,H271462 ,H270787 ,H270855 ,H271015 ,H271083 ,H271138 ,H271197 ,H271251 ,H271307 ,H271361 ,H271414 ,H271463 ,H270788 ,H270857 ,H271016 ,H271084 ,H271139 ,H271198 ,H271253 ,H271308 ,H271362 ,H271415 ,H271464 ,H270789 ,H270858 ,H271017 ,H271085 ,H271141 ,H271199 ,H271256 ,H271309 ,H271363 ,H271416 ,H271465 ,H270790 ,H270859 ,H271018 ,H271086 ,H271142 ,H271200 ,H271257 ,H271311 ,H271364 ,H271417 ,H271466 ,H270791 ,H270861 ,H271019 ,H271087 ,H271143 ,H271201 ,H271258 ,H271312 ,H271365 ,H271418 ,H271467 ,H270793 ,H270862 ,H271020 ,H271088 ,H271144 ,H271202 ,H271259 ,H271314 ,H271367 ,H271419 ,H271468 ,H270794 ,H270863 ,H271021 ,H271089 ,H271145 ,H271204 ,H271260 ,H271315 ,H271369 ,H271420 ,H271470 ,H270795 ,H270864 ,H271023 ,H271091 ,H271149 ,H271205 ,H271263 ,H271316 ,H271370 ,H271421 ,H271471 ,H270796 ,H270865 ,H271024 ,H271092 ,H271151 ,H271206 ,H271264 ,H271317 ,H271371 ,H271422 ,H271472 ,H270797 ,H270866 ,H271025 ,H271093 ,H271152 ,H271207 ,H271265 ,H271318 ,H271372 ,H271423 ,H271473 ,H270799 ,H270867 ,H271026 ,H271094 ,H271153 ,H271208 ,H271266 ,H271319 ,H271373 ,H271424 ,H271474 ,H270800 ,H270868 ,H271027 ,H271095 ,H271154 ,H271209 ,H271267 ,H271320 ,H271374 ,H271425 ,H271475 ,H270803 ,H270869 ,H271029 ,H271096 ,H271155 ,H271210 ,H271268 ,H271321 ,H271375 ,H271426 ,H271476 ,H270804 ,H270870 ,H271031 ,H271097 ,H271157 ,H271211 ,H271269 ,H271322 ,H271376 ,H271427 ,H271477 ,H270805 ,H270871 ,H271032 ,H271098 ,H271159 ,H271212 ,H271271 ,H271323 ,H271377 ,H271428 ,H271478 ,H270806 ,H270872 ,H271033 ,H271099 ,H271160 ,H271213 ,H271272 ,H271324 ,H271378 ,H271429 ,H271479 ,H270808 ,H270873 ,H271034 ,H271100 ,H271161 ,H271214 ,H271273 ,H271325 ,H271379 ,H271430 ,H271480 ,H270812 ,H270875 ,H271035 ,H271101 ,H271162 ,H271215 ,H271274 ,H271327 ,H271380 ,H271431 ,H271481 ,H270813 ,H270877 ,H271036 ,H271102 ,H271163 ,H271216 ,H271275 ,H271328 ,H271381 ,H271432 ,H271483 ,H270814 ,H270878 ,H271037 ,H271103 ,H271164 ,H271217 ,H271276 ,H271329 ,H271382 ,H271433 ,H271484 ,H270817 ,H270879 ,H271038 ,H271104 ,H271165 ,H271218 ,H271278 ,H271330 ,H271383 ,H271434 ,H271485 ,H270819 ,H270880 ,H271039 ,H271105 ,H271166 ,H271219 ,H271280 ,H271331 ,H271384 ,H271435 ,H271486 ,H270820 ,H270881 ,H271041 ,H271106 ,H271167 ,H271220 ,H271281 ,H271332 ,H271385 ,H271436 ,H271487 ,H270821 ,H270882 ,H271042 ,H271107 ,H271168 ,H271222 ,H271282 ,H271333 ,H271386 ,H271437 ,H271488 ,H270822 ,H270883 ,H271043 ,H271108 ,H271169 ,H271223 ,H271283 ,H271334 ,H271387 ,H271438 ,H271489 ,H270823 ,H270884 ,H271044 ,H271109 ,H271170 ,H271224 ,H271284 ,H271335 ,H271388 ,H271439 ,H271490 ,H270825 ,H270885 ,H271046 ,H271110 ,H271171 ,H271225 ,H271285 ,H271336 ,H271389 ,H271440 ,H271492 ,H270826 ,H270886 ,H271051 ,H271111 ,H271172 ,H271226 ,H271286 ,H271337 ,H271390 ,H271441 ,H271493 ,H270827 ,H270888 ,H271052 ,H271112 ,H271174 ,H271227 ,H271287 ,H271338 ,H271391 ,H271442 ,H271494 ,H270828 ,H270889 ,H271055 ,H271113 ,H271175 ,H271228 ,H271288 ,H271339 ,H271392 ,H271443 ,H271495 ,H270829 ,H270890 ,H271056 ,H271114 ,H271176 ,H271229 ,H271289 ,H271340 ,H271393 ,H271444 ,H271496 ,H270833 ,H270891 ,H271057 ,H271115 ,H271177 ,H271230 ,H271290 ,H271342 ,H271394 ,H271445 ,H271497 ,H270835 ,H270892 ,H271059 ,H271116 ,H271178 ,H271231 ,H271291 ,H271343 ,H271395 ,H271446 ,H271498 ,H270836 ,H270893 ,H271061 ,H271117 ,H271179 ,H271232 ,H271292 ,H271344 ,H271397 ,H271447 ,H271499 ,H270837 ,H270894 ,H271062 ,H271118 ,H271182 ,H271233 ,H271293 ,H271345 ,H271398 ,H271448 ,H271500 ,H270838 ,H270895 ,H271063 ,H271120 ,H271183 ,H271234 ,H271294 ,H271346 ,H271399 ,H271449 ,H271501 ,H270840 ,H270896 ,H271065 ,H271121 ,H271184 ,H271235 ,H271295 ,H271347 ,H271401 ,H271450 ,H271502 ,H271503 ,H271552 ,H271603 ,H271652 ,H271701 ,H271749 ,H271797 ,H271845 ,H271893 ,H271942 ,H271991 ,H271504 ,H271553 ,H271604 ,H271653 ,H271702 ,H271750 ,H271798 ,H271846 ,H271894 ,H271943 ,H271992 ,H271505 ,H271554 ,H271605 ,H271654 ,H271703 ,H271751 ,H271799 ,H271847 ,H271895 ,H271944 ,H271993 ,H271506 ,H271555 ,H271606 ,H271655 ,H271704 ,H271752 ,H271800 ,H271848 ,H271896 ,H271945 ,H271994 ,H271507 ,H271557 ,H271607 ,H271656 ,H271705 ,H271753 ,H271801 ,H271849 ,H271897 ,H271946 ,H271995 ,H271508 ,H271559 ,H271608 ,H271657 ,H271706 ,H271754 ,H271802 ,H271850 ,H271898 ,H271947 ,H271996 ,H271509 ,H271560 ,H271609 ,H271658 ,H271707 ,H271755 ,H271803 ,H271851 ,H271899 ,H271948 ,H271997 ,H271510 ,H271561 ,H271610 ,H271659 ,H271708 ,H271756 ,H271804 ,H271852 ,H271901 ,H271949 ,H271998 ,H271511 ,H271562 ,H271611 ,H271660 ,H271709 ,H271757 ,H271805 ,H271853 ,H271902 ,H271950 ,H271999 ,H271512 ,H271563 ,H271612 ,H271661 ,H271710 ,H271758 ,H271806 ,H271854 ,H271903 ,H271951 ,H272000 ,H271513 ,H271564 ,H271613 ,H271662 ,H271711 ,H271759 ,H271807 ,H271855 ,H271904 ,H271952 ,H272001 ,H271514 ,H271565 ,H271614 ,H271663 ,H271712 ,H271760 ,H271808 ,H271856 ,H271905 ,H271953 ,H272002 ,H271515 ,H271566 ,H271615 ,H271664 ,H271713 ,H271761 ,H271809 ,H271857 ,H271906 ,H271954 ,H272003 ,H271516 ,H271567 ,H271616 ,H271665 ,H271714 ,H271762 ,H271810 ,H271858 ,H271907 ,H271955 ,H272004 ,H271517 ,H271568 ,H271617 ,H271666 ,H271715 ,H271763 ,H271811 ,H271859 ,H271908 ,H271956 ,H272005 ,H271518 ,H271569 ,H271618 ,H271667 ,H271716 ,H271764 ,H271812 ,H271860 ,H271909 ,H271957 ,H272006 ,H271519 ,H271570 ,H271619 ,H271668 ,H271717 ,H271765 ,H271813 ,H271861 ,H271910 ,H271958 ,H272007 ,H271520 ,H271571 ,H271620 ,H271669 ,H271718 ,H271766 ,H271814 ,H271862 ,H271911 ,H271959 ,H272008 ,H271521 ,H271572 ,H271621 ,H271670 ,H271719 ,H271767 ,H271815 ,H271863 ,H271912 ,H271960 ,H272009 ,H271522 ,H271573 ,H271622 ,H271671 ,H271720 ,H271768 ,H271816 ,H271864 ,H271913 ,H271961 ,H272010 ,H271523 ,H271574 ,H271623 ,H271672 ,H271721 ,H271769 ,H271817 ,H271865 ,H271914 ,H271962 ,H272011 ,H271525 ,H271575 ,H271624 ,H271673 ,H271722 ,H271770 ,H271818 ,H271866 ,H271915 ,H271963 ,H272012 ,H271526 ,H271576 ,H271625 ,H271674 ,H271723 ,H271771 ,H271819 ,H271867 ,H271916 ,H271964 ,H272013 ,H271527 ,H271577 ,H271626 ,H271675 ,H271724 ,H271772 ,H271820 ,H271868 ,H271917 ,H271965 ,H272014 ,H271528 ,H271578 ,H271627 ,H271676 ,H271725 ,H271773 ,H271821 ,H271869 ,H271918 ,H271966 ,H272015 ,H271529 ,H271580 ,H271628 ,H271677 ,H271726 ,H271774 ,H271822 ,H271870 ,H271919 ,H271967 ,H272016 ,H271530 ,H271581 ,H271629 ,H271678 ,H271727 ,H271775 ,H271823 ,H271871 ,H271920 ,H271968 ,H272017 ,H271531 ,H271582 ,H271630 ,H271679 ,H271728 ,H271776 ,H271824 ,H271872 ,H271921 ,H271969 ,H272018 ,H271532 ,H271583 ,H271631 ,H271680 ,H271729 ,H271777 ,H271825 ,H271873 ,H271922 ,H271970 ,H272019 ,H271533 ,H271584 ,H271632 ,H271681 ,H271730 ,H271778 ,H271826 ,H271874 ,H271923 ,H271971 ,H272020 ,H271534 ,H271585 ,H271634 ,H271682 ,H271731 ,H271779 ,H271827 ,H271875 ,H271924 ,H271972 ,H272021 ,H271535 ,H271586 ,H271635 ,H271683 ,H271732 ,H271780 ,H271828 ,H271876 ,H271925 ,H271973 ,H272022 ,H271536 ,H271587 ,H271636 ,H271684 ,H271733 ,H271781 ,H271829 ,H271877 ,H271926 ,H271974 ,H272023 ,H271537 ,H271588 ,H271637 ,H271685 ,H271734 ,H271782 ,H271830 ,H271878 ,H271927 ,H271975 ,H272024 ,H271538 ,H271589 ,H271638 ,H271686 ,H271735 ,H271783 ,H271831 ,H271879 ,H271928 ,H271976 ,H272025 ,H271539 ,H271590 ,H271639 ,H271687 ,H271736 ,H271784 ,H271832 ,H271880 ,H271929 ,H271977 ,H272026 ,H271540 ,H271591 ,H271640 ,H271688 ,H271737 ,H271785 ,H271833 ,H271881 ,H271930 ,H271978 ,H272027 ,H271541 ,H271592 ,H271641 ,H271689 ,H271738 ,H271786 ,H271834 ,H271882 ,H271931 ,H271980 ,H272028 ,H271542 ,H271593 ,H271642 ,H271691 ,H271739 ,H271787 ,H271835 ,H271883 ,H271932 ,H271981 ,H272029 ,H271543 ,H271594 ,H271643 ,H271692 ,H271740 ,H271788 ,H271836 ,H271884 ,H271933 ,H271982 ,H272030 ,H271544 ,H271595 ,H271644 ,H271693 ,H271741 ,H271789 ,H271837 ,H271885 ,H271934 ,H271983 ,H272031 ,H271545 ,H271596 ,H271645 ,H271694 ,H271742 ,H271790 ,H271838 ,H271886 ,H271935 ,H271984 ,H272032 ,H271546 ,H271597 ,H271646 ,H271695 ,H271743 ,H271791 ,H271839 ,H271887 ,H271936 ,H271985 ,H272033 ,H271547 ,H271598 ,H271647 ,H271696 ,H271744 ,H271792 ,H271840 ,H271888 ,H271937 ,H271986 ,H272034 ,H271548 ,H271599 ,H271648 ,H271697 ,H271745 ,H271793 ,H271841 ,H271889 ,H271938 ,H271987 ,H272035 ,H271549 ,H271600 ,H271649 ,H271698 ,H271746 ,H271794 ,H271842 ,H271890 ,H271939 ,H271988 ,H272036 ,H271550 ,H271601 ,H271650 ,H271699 ,H271747 ,H271795 ,H271843 ,H271891 ,H271940 ,H271989 ,H272038 ,H271551 ,H271602 ,H271651 ,H271700 ,H271748 ,H271796 ,H271844 ,H271892 ,H271941 ,H271990 ,H272039 ,H272040 ,H272090 ,H272138 ,H272186 ,H272234 ,H272284 ,H272332 ,H272380 ,H272428 ,H272477 ,H272829 ,H272041 ,H272091 ,H272139 ,H272187 ,H272235 ,H272285 ,H272333 ,H272381 ,H272429 ,H272478 ,H272830 ,H272043 ,H272092 ,H272140 ,H272188 ,H272236 ,H272286 ,H272334 ,H272382 ,H272430 ,H272479 ,H272831 ,H272044 ,H272093 ,H272141 ,H272189 ,H272237 ,H272287 ,H272335 ,H272383 ,H272431 ,H272480 ,H272833 ,H272045 ,H272094 ,H272142 ,H272190 ,H272238 ,H272288 ,H272336 ,H272384 ,H272432 ,H272482 ,H272834 ,H272046 ,H272095 ,H272143 ,H272191 ,H272239 ,H272289 ,H272337 ,H272385 ,H272433 ,H272483 ,H272836 ,H272047 ,H272096 ,H272144 ,H272192 ,H272240 ,H272290 ,H272338 ,H272386 ,H272434 ,H272484 ,H272837 ,H272048 ,H272097 ,H272145 ,H272193 ,H272241 ,H272291 ,H272339 ,H272387 ,H272435 ,H272488 ,H272838 ,H272049 ,H272098 ,H272146 ,H272194 ,H272242 ,H272292 ,H272340 ,H272388 ,H272436 ,H272489 ,H272839 ,H272050 ,H272099 ,H272147 ,H272195 ,H272243 ,H272293 ,H272341 ,H272389 ,H272437 ,H272490 ,H272840 ,H272051 ,H272100 ,H272148 ,H272196 ,H272244 ,H272294 ,H272342 ,H272390 ,H272438 ,H272491 ,H272841 ,H272052 ,H272101 ,H272149 ,H272197 ,H272245 ,H272295 ,H272343 ,H272391 ,H272439 ,H272492 ,H272842 ,H272053 ,H272102 ,H272150 ,H272198 ,H272246 ,H272296 ,H272344 ,H272392 ,H272440 ,H272493 ,H272843 ,H272054 ,H272103 ,H272151 ,H272199 ,H272247 ,H272297 ,H272345 ,H272393 ,H272441 ,H272494 ,H272844 ,H272055 ,H272104 ,H272152 ,H272200 ,H272248 ,H272298 ,H272346 ,H272394 ,H272442 ,H272495 ,H272845 ,H272056 ,H272105 ,H272153 ,H272201 ,H272249 ,H272299 ,H272347 ,H272395 ,H272443 ,H272496 ,H272846 ,H272057 ,H272106 ,H272154 ,H272202 ,H272250 ,H272300 ,H272348 ,H272396 ,H272444 ,H272497 ,H272847 ,H272058 ,H272107 ,H272155 ,H272203 ,H272251 ,H272301 ,H272349 ,H272397 ,H272445 ,H272498 ,H272848 ,H272059 ,H272108 ,H272156 ,H272204 ,H272253 ,H272302 ,H272350 ,H272398 ,H272446 ,H272500 ,H272849 ,H272060 ,H272109 ,H272157 ,H272205 ,H272254 ,H272303 ,H272351 ,H272399 ,H272447 ,H272800 ,H272850 ,H272061 ,H272110 ,H272158 ,H272206 ,H272255 ,H272304 ,H272352 ,H272400 ,H272448 ,H272801 ,H272851 ,H272062 ,H272111 ,H272159 ,H272207 ,H272256 ,H272305 ,H272353 ,H272401 ,H272449 ,H272802 ,H272852 ,H272064 ,H272112 ,H272160 ,H272208 ,H272258 ,H272306 ,H272354 ,H272402 ,H272450 ,H272803 ,H272853 ,H272065 ,H272113 ,H272161 ,H272209 ,H272259 ,H272307 ,H272355 ,H272403 ,H272451 ,H272804 ,H272854 ,H272066 ,H272114 ,H272162 ,H272210 ,H272260 ,H272308 ,H272356 ,H272404 ,H272452 ,H272805 ,H272855 ,H272067 ,H272115 ,H272163 ,H272211 ,H272261 ,H272309 ,H272357 ,H272405 ,H272453 ,H272806 ,H272856 ,H272068 ,H272116 ,H272164 ,H272212 ,H272262 ,H272310 ,H272358 ,H272406 ,H272454 ,H272807 ,H272857 ,H272069 ,H272117 ,H272165 ,H272213 ,H272263 ,H272311 ,H272359 ,H272407 ,H272455 ,H272808 ,H272858 ,H272070 ,H272118 ,H272166 ,H272214 ,H272264 ,H272312 ,H272360 ,H272408 ,H272456 ,H272809 ,H272859 ,H272071 ,H272119 ,H272167 ,H272215 ,H272265 ,H272313 ,H272361 ,H272409 ,H272457 ,H272810 ,H272860 ,H272072 ,H272120 ,H272168 ,H272216 ,H272266 ,H272314 ,H272362 ,H272410 ,H272458 ,H272811 ,H272861 ,H272073 ,H272121 ,H272169 ,H272217 ,H272267 ,H272315 ,H272363 ,H272411 ,H272459 ,H272812 ,H272862 ,H272074 ,H272122 ,H272170 ,H272218 ,H272268 ,H272316 ,H272364 ,H272412 ,H272460 ,H272813 ,H272863 ,H272075 ,H272123 ,H272171 ,H272219 ,H272269 ,H272317 ,H272365 ,H272413 ,H272461 ,H272814 ,H272864 ,H272076 ,H272124 ,H272172 ,H272220 ,H272270 ,H272318 ,H272366 ,H272414 ,H272462 ,H272815 ,H272865 ,H272077 ,H272125 ,H272173 ,H272221 ,H272271 ,H272319 ,H272367 ,H272415 ,H272463 ,H272816 ,H272866 ,H272078 ,H272126 ,H272174 ,H272222 ,H272272 ,H272320 ,H272368 ,H272416 ,H272464 ,H272817 ,H272867 ,H272079 ,H272127 ,H272175 ,H272223 ,H272273 ,H272321 ,H272369 ,H272417 ,H272465 ,H272818 ,H272868 ,H272080 ,H272128 ,H272176 ,H272224 ,H272274 ,H272322 ,H272370 ,H272418 ,H272466 ,H272819 ,H272869 ,H272081 ,H272129 ,H272177 ,H272225 ,H272275 ,H272323 ,H272371 ,H272419 ,H272467 ,H272820 ,H272870 ,H272082 ,H272130 ,H272178 ,H272226 ,H272276 ,H272324 ,H272372 ,H272420 ,H272468 ,H272821 ,H272871 ,H272083 ,H272131 ,H272179 ,H272227 ,H272277 ,H272325 ,H272373 ,H272421 ,H272469 ,H272822 ,H272872 ,H272084 ,H272132 ,H272180 ,H272228 ,H272278 ,H272326 ,H272374 ,H272422 ,H272470 ,H272823 ,H272873 ,H272085 ,H272133 ,H272181 ,H272229 ,H272279 ,H272327 ,H272375 ,H272423 ,H272471 ,H272824 ,H272875 ,H272086 ,H272134 ,H272182 ,H272230 ,H272280 ,H272328 ,H272376 ,H272424 ,H272472 ,H272825 ,H272876 ,H272087 ,H272135 ,H272183 ,H272231 ,H272281 ,H272329 ,H272377 ,H272425 ,H272473 ,H272826 ,H272877 ,H272088 ,H272136 ,H272184 ,H272232 ,H272282 ,H272330 ,H272378 ,H272426 ,H272475 ,H272827 ,H272878 ,H272089 ,H272137 ,H272185 ,H272233 ,H272283 ,H272331 ,H272379 ,H272427 ,H272476 ,H272828 ,H272879 ,H272880 ,H272928 ,H272977 ,H273028 ,H273079 ,H273128 ,H273183 ,H290119 ,H290194 ,H290244 ,H290297 ,H272881 ,H272929 ,H272978 ,H273029 ,H273080 ,H273130 ,H273184 ,H290121 ,H290195 ,H290246 ,H290299 ,H272882 ,H272930 ,H272979 ,H273030 ,H273081 ,H273131 ,H273185 ,H290122 ,H290196 ,H290247 ,H290301 ,H272883 ,H272931 ,H272980 ,H273031 ,H273082 ,H273132 ,H273186 ,H290123 ,H290197 ,H290248 ,H290302 ,H272884 ,H272932 ,H272981 ,H273032 ,H273083 ,H273133 ,H273187 ,H290125 ,H290198 ,H290249 ,H290304 ,H272885 ,H272933 ,H272982 ,H273033 ,H273084 ,H273134 ,H273189 ,H290128 ,H290199 ,H290250 ,H290305 ,H272886 ,H272934 ,H272983 ,H273034 ,H273085 ,H273135 ,H273190 ,H290130 ,H290200 ,H290253 ,H290306 ,H272887 ,H272935 ,H272984 ,H273035 ,H273086 ,H273136 ,H273191 ,H290132 ,H290201 ,H290254 ,H290307 ,H272888 ,H272936 ,H272985 ,H273036 ,H273087 ,H273137 ,H273193 ,H290135 ,H290202 ,H290255 ,H290308 ,H272889 ,H272937 ,H272986 ,H273037 ,H273088 ,H273138 ,H273194 ,H290137 ,H290203 ,H290256 ,H290309 ,H272890 ,H272938 ,H272988 ,H273038 ,H273089 ,H273140 ,H273195 ,H290138 ,H290204 ,H290257 ,H290310 ,H272891 ,H272939 ,H272989 ,H273039 ,H273090 ,H273141 ,H273196 ,H290139 ,H290205 ,H290258 ,H290311 ,H272892 ,H272940 ,H272990 ,H273040 ,H273091 ,H273142 ,H273197 ,H290141 ,H290206 ,H290259 ,H290312 ,H272893 ,H272941 ,H272991 ,H273041 ,H273092 ,H273143 ,H273198 ,H290142 ,H290208 ,H290260 ,H290313 ,H272894 ,H272942 ,H272992 ,H273042 ,H273093 ,H273144 ,H273199 ,H290143 ,H290209 ,H290261 ,H290314 ,H272895 ,H272943 ,H272993 ,H273043 ,H273094 ,H273145 ,H273200 ,H290144 ,H290210 ,H290262 ,H290315 ,H272896 ,H272944 ,H272994 ,H273044 ,H273095 ,H273147 ,H273201 ,H290145 ,H290211 ,H290263 ,H290316 ,H272897 ,H272946 ,H272995 ,H273045 ,H273097 ,H273148 ,H273202 ,H290146 ,H290212 ,H290264 ,H290317 ,H272898 ,H272947 ,H272996 ,H273046 ,H273098 ,H273149 ,H273203 ,H290151 ,H290213 ,H290265 ,H290318 ,H272899 ,H272948 ,H272997 ,H273047 ,H273099 ,H273150 ,H273204 ,H290152 ,H290214 ,H290266 ,H290319 ,H272900 ,H272949 ,H272998 ,H273048 ,H273100 ,H273151 ,H273206 ,H290154 ,H290215 ,H290267 ,H290320 ,H272901 ,H272950 ,H272999 ,H273049 ,H273101 ,H273152 ,H273208 ,H290155 ,H290216 ,H290268 ,H290321 ,H272902 ,H272951 ,H273000 ,H273050 ,H273102 ,H273153 ,H273209 ,H290156 ,H290217 ,H290269 ,H290322 ,H272903 ,H272952 ,H273001 ,H273051 ,H273103 ,H273155 ,H273210 ,H290157 ,H290218 ,H290271 ,H290323 ,H272904 ,H272953 ,H273002 ,H273052 ,H273104 ,H273156 ,H290001 ,H290158 ,H290219 ,H290272 ,H290324 ,H272905 ,H272954 ,H273003 ,H273053 ,H273105 ,H273157 ,H290003 ,H290159 ,H290220 ,H290273 ,H290325 ,H272906 ,H272955 ,H273004 ,H273054 ,H273106 ,H273158 ,H290012 ,H290160 ,H290221 ,H290274 ,H290326 ,H272907 ,H272956 ,H273005 ,H273055 ,H273107 ,H273159 ,H290017 ,H290161 ,H290222 ,H290275 ,H290327 ,H272908 ,H272957 ,H273006 ,H273056 ,H273108 ,H273160 ,H290021 ,H290162 ,H290223 ,H290276 ,H290328 ,H272909 ,H272958 ,H273007 ,H273057 ,H273109 ,H273162 ,H290026 ,H290164 ,H290224 ,H290277 ,H290329 ,H272910 ,H272959 ,H273008 ,H273058 ,H273110 ,H273163 ,H290027 ,H290165 ,H290225 ,H290278 ,H290330 ,H272911 ,H272960 ,H273010 ,H273059 ,H273111 ,H273164 ,H290030 ,H290166 ,H290226 ,H290279 ,H290331 ,H272912 ,H272961 ,H273011 ,H273060 ,H273112 ,H273165 ,H290031 ,H290168 ,H290227 ,H290280 ,H290332 ,H272913 ,H272962 ,H273012 ,H273061 ,H273113 ,H273166 ,H290032 ,H290171 ,H290228 ,H290281 ,H290333 ,H272914 ,H272963 ,H273013 ,H273062 ,H273114 ,H273167 ,H290033 ,H290173 ,H290229 ,H290282 ,H290334 ,H272915 ,H272964 ,H273014 ,H273063 ,H273115 ,H273170 ,H290034 ,H290175 ,H290230 ,H290283 ,H290335 ,H272916 ,H272965 ,H273016 ,H273065 ,H273116 ,H273171 ,H290100 ,H290176 ,H290231 ,H290284 ,H290336 ,H272917 ,H272966 ,H273017 ,H273066 ,H273117 ,H273172 ,H290103 ,H290180 ,H290232 ,H290285 ,H290337 ,H272918 ,H272967 ,H273018 ,H273067 ,H273118 ,H273173 ,H290104 ,H290181 ,H290233 ,H290286 ,H290338 ,H272919 ,H272968 ,H273019 ,H273068 ,H273119 ,H273174 ,H290105 ,H290182 ,H290234 ,H290287 ,H290339 ,H272920 ,H272969 ,H273020 ,H273069 ,H273120 ,H273175 ,H290108 ,H290183 ,H290235 ,H290288 ,H290340 ,H272921 ,H272970 ,H273021 ,H273070 ,H273121 ,H273176 ,H290109 ,H290185 ,H290236 ,H290289 ,H290341 ,H272922 ,H272971 ,H273022 ,H273071 ,H273122 ,H273177 ,H290110 ,H290186 ,H290237 ,H290290 ,H290342 ,H272923 ,H272972 ,H273023 ,H273072 ,H273123 ,H273178 ,H290111 ,H290187 ,H290239 ,H290291 ,H290343 ,H272924 ,H272973 ,H273024 ,H273074 ,H273124 ,H273179 ,H290113 ,H290189 ,H290240 ,H290292 ,H290344 ,H272925 ,H272974 ,H273025 ,H273075 ,H273125 ,H273180 ,H290114 ,H290191 ,H290241 ,H290293 ,H290345 ,H272926 ,H272975 ,H273026 ,H273077 ,H273126 ,H273181 ,H290116 ,H290192 ,H290242 ,H290295 ,H290346 ,H272927 ,H272976 ,H273027 ,H273078 ,H273127 ,H273182 ,H290118 ,H290193 ,H290243 ,H290296 ,H290347 ,H290348 ,H290398 ,H290523 ,H291009 ,H291059 ,H291108 ,H291157 ,H291205 ,H291253 ,H293324 ,H293372 ,H290349 ,H290399 ,H290524 ,H291010 ,H291060 ,H291109 ,H291158 ,H291206 ,H291254 ,H293325 ,H293373 ,H290350 ,H290400 ,H290525 ,H291011 ,H291061 ,H291111 ,H291159 ,H291207 ,H291255 ,H293326 ,H293374 ,H290351 ,H290401 ,H290526 ,H291012 ,H291062 ,H291112 ,H291160 ,H291208 ,H291256 ,H293327 ,H293375 ,H290352 ,H290402 ,H290527 ,H291013 ,H291063 ,H291113 ,H291161 ,H291209 ,H291257 ,H293328 ,H293376 ,H290353 ,H290403 ,H290528 ,H291014 ,H291064 ,H291114 ,H291162 ,H291210 ,H291258 ,H293329 ,H293377 ,H290354 ,H290404 ,H290529 ,H291015 ,H291065 ,H291115 ,H291163 ,H291211 ,H291259 ,H293330 ,H293378 ,H290355 ,H290405 ,H290530 ,H291016 ,H291066 ,H291116 ,H291164 ,H291212 ,H291260 ,H293331 ,H293379 ,H290356 ,H290406 ,H290531 ,H291017 ,H291067 ,H291117 ,H291165 ,H291213 ,H291261 ,H293332 ,H293380 ,H290357 ,H290408 ,H290532 ,H291018 ,H291068 ,H291118 ,H291166 ,H291214 ,H291262 ,H293333 ,H293381 ,H290358 ,H290409 ,H290533 ,H291019 ,H291069 ,H291119 ,H291167 ,H291215 ,H291263 ,H293334 ,H293382 ,H290359 ,H290410 ,H290534 ,H291020 ,H291070 ,H291120 ,H291168 ,H291216 ,H291264 ,H293335 ,H293383 ,H290360 ,H290411 ,H290535 ,H291021 ,H291071 ,H291121 ,H291169 ,H291217 ,H291265 ,H293336 ,H293384 ,H290362 ,H290412 ,H290536 ,H291022 ,H291072 ,H291122 ,H291170 ,H291218 ,H291266 ,H293337 ,H293385 ,H290363 ,H290413 ,H290537 ,H291023 ,H291073 ,H291123 ,H291171 ,H291219 ,H291267 ,H293338 ,H293386 ,H290364 ,H290414 ,H290538 ,H291024 ,H291074 ,H291124 ,H291172 ,H291220 ,H291268 ,H293339 ,H293387 ,H290365 ,H290415 ,H290539 ,H291025 ,H291075 ,H291125 ,H291173 ,H291221 ,H291269 ,H293340 ,H293388 ,H290366 ,H290416 ,H290540 ,H291026 ,H291076 ,H291126 ,H291174 ,H291222 ,H291270 ,H293341 ,H293389 ,H290367 ,H290417 ,H290541 ,H291027 ,H291077 ,H291127 ,H291175 ,H291223 ,H291271 ,H293342 ,H293390 ,H290368 ,H290422 ,H290542 ,H291028 ,H291078 ,H291128 ,H291176 ,H291224 ,H291272 ,H293343 ,H293391 ,H290369 ,H290423 ,H290543 ,H291029 ,H291079 ,H291129 ,H291177 ,H291225 ,H291273 ,H293344 ,H293393 ,H290370 ,H290424 ,H290544 ,H291030 ,H291080 ,H291130 ,H291178 ,H291226 ,H292832 ,H293345 ,H293394 ,H290371 ,H290425 ,H290545 ,H291031 ,H291081 ,H291131 ,H291179 ,H291227 ,H293009 ,H293346 ,H293395 ,H290372 ,H290426 ,H290546 ,H291032 ,H291082 ,H291132 ,H291180 ,H291228 ,H293207 ,H293347 ,H293396 ,H290373 ,H290427 ,H290547 ,H291034 ,H291083 ,H291133 ,H291181 ,H291229 ,H293300 ,H293348 ,H293397 ,H290374 ,H290500 ,H290548 ,H291035 ,H291084 ,H291134 ,H291182 ,H291230 ,H293301 ,H293349 ,H293398 ,H290375 ,H290501 ,H290549 ,H291036 ,H291085 ,H291135 ,H291183 ,H291231 ,H293302 ,H293350 ,H293399 ,H290376 ,H290502 ,H290550 ,H291037 ,H291086 ,H291136 ,H291184 ,H291232 ,H293303 ,H293351 ,H293401 ,H290377 ,H290503 ,H290551 ,H291038 ,H291087 ,H291137 ,H291185 ,H291233 ,H293304 ,H293352 ,H293402 ,H290378 ,H290504 ,H290552 ,H291039 ,H291088 ,H291138 ,H291186 ,H291234 ,H293305 ,H293353 ,H293403 ,H290379 ,H290505 ,H290553 ,H291040 ,H291089 ,H291139 ,H291187 ,H291235 ,H293306 ,H293354 ,H293404 ,H290380 ,H290506 ,H290554 ,H291041 ,H291090 ,H291140 ,H291188 ,H291236 ,H293307 ,H293355 ,H293405 ,H290381 ,H290507 ,H290555 ,H291042 ,H291091 ,H291141 ,H291189 ,H291237 ,H293308 ,H293356 ,H293406 ,H290382 ,H290508 ,H290556 ,H291043 ,H291092 ,H291142 ,H291190 ,H291238 ,H293309 ,H293357 ,H293407 ,H290383 ,H290509 ,H290557 ,H291044 ,H291093 ,H291143 ,H291191 ,H291239 ,H293310 ,H293358 ,H293408 ,H290384 ,H290510 ,H290558 ,H291045 ,H291094 ,H291144 ,H291192 ,H291240 ,H293311 ,H293359 ,H293409 ,H290385 ,H290511 ,H290559 ,H291046 ,H291095 ,H291145 ,H291193 ,H291241 ,H293312 ,H293360 ,H293410 ,H290386 ,H290512 ,H290560 ,H291047 ,H291096 ,H291146 ,H291194 ,H291242 ,H293313 ,H293361 ,H293411 ,H290387 ,H290513 ,H290561 ,H291048 ,H291097 ,H291147 ,H291195 ,H291243 ,H293314 ,H293362 ,H293412 ,H290388 ,H290514 ,H290600 ,H291049 ,H291098 ,H291148 ,H291196 ,H291244 ,H293315 ,H293363 ,H293413 ,H290389 ,H290515 ,H291001 ,H291050 ,H291099 ,H291149 ,H291197 ,H291245 ,H293316 ,H293364 ,H293414 ,H290390 ,H290516 ,H291002 ,H291051 ,H291101 ,H291150 ,H291198 ,H291246 ,H293317 ,H293365 ,H293415 ,H290392 ,H290517 ,H291003 ,H291052 ,H291102 ,H291151 ,H291199 ,H291247 ,H293318 ,H293366 ,H293416 ,H290393 ,H290518 ,H291004 ,H291053 ,H291103 ,H291152 ,H291200 ,H291248 ,H293319 ,H293367 ,H293417 ,H290394 ,H290519 ,H291005 ,H291054 ,H291104 ,H291153 ,H291201 ,H291249 ,H293320 ,H293368 ,H293418 ,H290395 ,H290520 ,H291006 ,H291055 ,H291105 ,H291154 ,H291202 ,H291250 ,H293321 ,H293369 ,H293419 ,H290396 ,H290521 ,H291007 ,H291056 ,H291106 ,H291155 ,H291203 ,H291251 ,H293322 ,H293370 ,H293420 ,H290397 ,H290522 ,H291008 ,H291057 ,H291107 ,H291156 ,H291204 ,H291252 ,H293323 ,H293371 ,H293421 ,H293422 ,H293470 ,H293518 ,H293568 ,H293618 ,H293666 ,H293716 ,H293766 ,H293814 ,H293863 ,H293911 ,H293423 ,H293471 ,H293519 ,H293569 ,H293619 ,H293667 ,H293717 ,H293767 ,H293815 ,H293864 ,H293912 ,H293424 ,H293472 ,H293520 ,H293570 ,H293620 ,H293668 ,H293718 ,H293768 ,H293816 ,H293865 ,H293913 ,H293425 ,H293473 ,H293521 ,H293571 ,H293621 ,H293669 ,H293719 ,H293769 ,H293817 ,H293866 ,H293914 ,H293426 ,H293474 ,H293522 ,H293572 ,H293622 ,H293670 ,H293720 ,H293770 ,H293818 ,H293867 ,H293915 ,H293427 ,H293475 ,H293523 ,H293573 ,H293623 ,H293671 ,H293721 ,H293771 ,H293819 ,H293868 ,H293916 ,H293428 ,H293476 ,H293524 ,H293574 ,H293624 ,H293672 ,H293723 ,H293772 ,H293820 ,H293869 ,H293917 ,H293429 ,H293477 ,H293525 ,H293575 ,H293625 ,H293673 ,H293724 ,H293773 ,H293821 ,H293870 ,H293918 ,H293430 ,H293478 ,H293527 ,H293576 ,H293626 ,H293674 ,H293725 ,H293774 ,H293822 ,H293871 ,H293919 ,H293431 ,H293479 ,H293528 ,H293577 ,H293627 ,H293675 ,H293726 ,H293775 ,H293823 ,H293872 ,H293920 ,H293432 ,H293480 ,H293529 ,H293578 ,H293628 ,H293676 ,H293727 ,H293776 ,H293824 ,H293873 ,H293921 ,H293433 ,H293481 ,H293530 ,H293579 ,H293629 ,H293677 ,H293728 ,H293777 ,H293825 ,H293874 ,H293922 ,H293434 ,H293482 ,H293531 ,H293581 ,H293630 ,H293678 ,H293729 ,H293778 ,H293826 ,H293875 ,H293923 ,H293435 ,H293483 ,H293532 ,H293583 ,H293631 ,H293679 ,H293730 ,H293779 ,H293827 ,H293876 ,H293924 ,H293436 ,H293484 ,H293533 ,H293584 ,H293632 ,H293681 ,H293731 ,H293780 ,H293828 ,H293877 ,H293925 ,H293437 ,H293485 ,H293534 ,H293585 ,H293633 ,H293682 ,H293732 ,H293781 ,H293829 ,H293878 ,H293926 ,H293438 ,H293486 ,H293535 ,H293586 ,H293634 ,H293683 ,H293733 ,H293782 ,H293830 ,H293879 ,H293928 ,H293439 ,H293487 ,H293536 ,H293587 ,H293635 ,H293684 ,H293734 ,H293783 ,H293831 ,H293880 ,H293929 ,H293440 ,H293488 ,H293537 ,H293588 ,H293636 ,H293685 ,H293735 ,H293784 ,H293832 ,H293881 ,H293930 ,H293441 ,H293489 ,H293538 ,H293589 ,H293637 ,H293686 ,H293736 ,H293785 ,H293833 ,H293882 ,H293931 ,H293442 ,H293490 ,H293539 ,H293590 ,H293638 ,H293687 ,H293737 ,H293786 ,H293834 ,H293883 ,H293932 ,H293443 ,H293491 ,H293540 ,H293591 ,H293639 ,H293688 ,H293738 ,H293787 ,H293835 ,H293884 ,H293933 ,H293444 ,H293492 ,H293541 ,H293592 ,H293640 ,H293689 ,H293739 ,H293788 ,H293836 ,H293885 ,H293934 ,H293445 ,H293493 ,H293542 ,H293593 ,H293641 ,H293690 ,H293740 ,H293789 ,H293837 ,H293886 ,H293935 ,H293446 ,H293494 ,H293543 ,H293594 ,H293642 ,H293691 ,H293741 ,H293790 ,H293838 ,H293887 ,H293936 ,H293447 ,H293495 ,H293544 ,H293595 ,H293643 ,H293692 ,H293742 ,H293791 ,H293839 ,H293888 ,H293938 ,H293448 ,H293496 ,H293545 ,H293596 ,H293644 ,H293693 ,H293743 ,H293792 ,H293840 ,H293889 ,H293940 ,H293449 ,H293497 ,H293546 ,H293597 ,H293645 ,H293694 ,H293744 ,H293793 ,H293841 ,H293890 ,H293941 ,H293450 ,H293498 ,H293548 ,H293598 ,H293646 ,H293695 ,H293745 ,H293794 ,H293842 ,H293891 ,H293942 ,H293451 ,H293499 ,H293549 ,H293599 ,H293647 ,H293696 ,H293746 ,H293795 ,H293843 ,H293892 ,H293943 ,H293452 ,H293500 ,H293550 ,H293600 ,H293648 ,H293697 ,H293747 ,H293796 ,H293844 ,H293893 ,H293944 ,H293453 ,H293501 ,H293551 ,H293601 ,H293649 ,H293698 ,H293748 ,H293797 ,H293845 ,H293894 ,H293946 ,H293454 ,H293502 ,H293552 ,H293602 ,H293650 ,H293699 ,H293749 ,H293798 ,H293846 ,H293895 ,H293947 ,H293455 ,H293503 ,H293553 ,H293603 ,H293651 ,H293700 ,H293750 ,H293799 ,H293847 ,H293896 ,H293948 ,H293456 ,H293504 ,H293554 ,H293604 ,H293652 ,H293701 ,H293751 ,H293800 ,H293848 ,H293897 ,H293949 ,H293457 ,H293505 ,H293555 ,H293605 ,H293653 ,H293702 ,H293752 ,H293801 ,H293849 ,H293898 ,H293950 ,H293458 ,H293506 ,H293556 ,H293606 ,H293654 ,H293703 ,H293753 ,H293802 ,H293850 ,H293899 ,H293951 ,H293459 ,H293507 ,H293557 ,H293607 ,H293655 ,H293704 ,H293754 ,H293803 ,H293851 ,H293900 ,H293952 ,H293460 ,H293508 ,H293558 ,H293608 ,H293656 ,H293705 ,H293755 ,H293804 ,H293852 ,H293901 ,H293953 ,H293461 ,H293509 ,H293559 ,H293609 ,H293657 ,H293706 ,H293756 ,H293805 ,H293854 ,H293902 ,H293954 ,H293462 ,H293510 ,H293560 ,H293610 ,H293658 ,H293707 ,H293758 ,H293806 ,H293855 ,H293903 ,H293955 ,H293463 ,H293511 ,H293561 ,H293611 ,H293659 ,H293708 ,H293759 ,H293807 ,H293856 ,H293904 ,H293956 ,H293464 ,H293512 ,H293562 ,H293612 ,H293660 ,H293709 ,H293760 ,H293808 ,H293857 ,H293905 ,H293957 ,H293465 ,H293513 ,H293563 ,H293613 ,H293661 ,H293710 ,H293761 ,H293809 ,H293858 ,H293906 ,H293958 ,H293466 ,H293514 ,H293564 ,H293614 ,H293662 ,H293712 ,H293762 ,H293810 ,H293859 ,H293907 ,H293959 ,H293467 ,H293515 ,H293565 ,H293615 ,H293663 ,H293713 ,H293763 ,H293811 ,H293860 ,H293908 ,H293960 ,H293468 ,H293516 ,H293566 ,H293616 ,H293664 ,H293714 ,H293764 ,H293812 ,H293861 ,H293909 ,H293961 ,H293469 ,H293517 ,H293567 ,H293617 ,H293665 ,H293715 ,H293765 ,H293813 ,H293862 ,H293910 ,H293962 ,H293963 ,H294012 ,H294061 ,H294110 ,H294161 ,H294209 ,H294260 ,H294309 ,H294357 ,H294406 ,H294458 ,H293964 ,H294013 ,H294062 ,H294111 ,H294162 ,H294210 ,H294261 ,H294310 ,H294358 ,H294407 ,H294460 ,H293965 ,H294014 ,H294063 ,H294112 ,H294163 ,H294211 ,H294262 ,H294311 ,H294359 ,H294408 ,H294461 ,H293966 ,H294015 ,H294064 ,H294113 ,H294164 ,H294212 ,H294263 ,H294312 ,H294360 ,H294409 ,H294462 ,H293967 ,H294016 ,H294065 ,H294114 ,H294165 ,H294213 ,H294264 ,H294313 ,H294361 ,H294410 ,H294463 ,H293968 ,H294017 ,H294066 ,H294116 ,H294166 ,H294214 ,H294265 ,H294314 ,H294362 ,H294411 ,H294464 ,H293969 ,H294018 ,H294067 ,H294117 ,H294167 ,H294215 ,H294266 ,H294315 ,H294363 ,H294412 ,H294465 ,H293970 ,H294020 ,H294068 ,H294118 ,H294168 ,H294216 ,H294267 ,H294316 ,H294364 ,H294413 ,H294466 ,H293971 ,H294021 ,H294069 ,H294119 ,H294169 ,H294217 ,H294268 ,H294317 ,H294365 ,H294414 ,H294467 ,H293972 ,H294022 ,H294070 ,H294120 ,H294170 ,H294218 ,H294269 ,H294318 ,H294366 ,H294415 ,H294468 ,H293974 ,H294023 ,H294071 ,H294121 ,H294171 ,H294219 ,H294270 ,H294319 ,H294367 ,H294416 ,H294469 ,H293975 ,H294024 ,H294072 ,H294122 ,H294172 ,H294220 ,H294271 ,H294320 ,H294368 ,H294417 ,H294470 ,H293976 ,H294025 ,H294073 ,H294123 ,H294173 ,H294221 ,H294272 ,H294321 ,H294369 ,H294418 ,H294471 ,H293977 ,H294026 ,H294074 ,H294124 ,H294174 ,H294222 ,H294273 ,H294322 ,H294370 ,H294419 ,H294472 ,H293978 ,H294027 ,H294075 ,H294125 ,H294175 ,H294223 ,H294274 ,H294323 ,H294371 ,H294420 ,H294473 ,H293979 ,H294028 ,H294076 ,H294126 ,H294176 ,H294224 ,H294275 ,H294324 ,H294372 ,H294421 ,H294474 ,H293980 ,H294029 ,H294077 ,H294127 ,H294177 ,H294225 ,H294276 ,H294325 ,H294373 ,H294422 ,H294475 ,H293981 ,H294030 ,H294078 ,H294128 ,H294178 ,H294226 ,H294277 ,H294326 ,H294374 ,H294423 ,H294476 ,H293982 ,H294031 ,H294079 ,H294129 ,H294179 ,H294227 ,H294278 ,H294327 ,H294376 ,H294424 ,H294477 ,H293983 ,H294032 ,H294080 ,H294130 ,H294180 ,H294228 ,H294279 ,H294328 ,H294377 ,H294425 ,H294478 ,H293984 ,H294033 ,H294081 ,H294131 ,H294181 ,H294229 ,H294280 ,H294329 ,H294378 ,H294426 ,H294479 ,H293985 ,H294034 ,H294082 ,H294132 ,H294182 ,H294230 ,H294281 ,H294330 ,H294379 ,H294427 ,H294480 ,H293986 ,H294035 ,H294083 ,H294133 ,H294183 ,H294231 ,H294283 ,H294331 ,H294380 ,H294428 ,H294481 ,H293987 ,H294036 ,H294084 ,H294134 ,H294184 ,H294232 ,H294284 ,H294332 ,H294381 ,H294429 ,H294482 ,H293988 ,H294037 ,H294085 ,H294135 ,H294185 ,H294233 ,H294285 ,H294333 ,H294382 ,H294430 ,H294483 ,H293989 ,H294038 ,H294086 ,H294136 ,H294186 ,H294234 ,H294286 ,H294334 ,H294383 ,H294431 ,H294484 ,H293990 ,H294039 ,H294087 ,H294137 ,H294187 ,H294235 ,H294287 ,H294335 ,H294384 ,H294432 ,H294485 ,H293991 ,H294040 ,H294088 ,H294140 ,H294188 ,H294236 ,H294288 ,H294336 ,H294385 ,H294433 ,H294486 ,H293992 ,H294041 ,H294089 ,H294141 ,H294189 ,H294237 ,H294289 ,H294337 ,H294386 ,H294434 ,H294487 ,H293993 ,H294042 ,H294090 ,H294142 ,H294190 ,H294238 ,H294290 ,H294338 ,H294387 ,H294435 ,H294488 ,H293994 ,H294043 ,H294091 ,H294143 ,H294191 ,H294239 ,H294291 ,H294339 ,H294388 ,H294436 ,H294489 ,H293995 ,H294044 ,H294092 ,H294144 ,H294192 ,H294240 ,H294292 ,H294340 ,H294389 ,H294437 ,H294490 ,H293996 ,H294045 ,H294093 ,H294145 ,H294193 ,H294241 ,H294293 ,H294341 ,H294390 ,H294438 ,H294491 ,H293997 ,H294046 ,H294094 ,H294146 ,H294194 ,H294242 ,H294294 ,H294342 ,H294391 ,H294439 ,H294492 ,H293998 ,H294047 ,H294095 ,H294147 ,H294195 ,H294243 ,H294295 ,H294343 ,H294392 ,H294440 ,H294493 ,H293999 ,H294048 ,H294097 ,H294148 ,H294196 ,H294244 ,H294296 ,H294344 ,H294393 ,H294441 ,H294494 ,H294000 ,H294049 ,H294098 ,H294149 ,H294197 ,H294245 ,H294297 ,H294345 ,H294394 ,H294442 ,H294495 ,H294001 ,H294050 ,H294099 ,H294150 ,H294198 ,H294246 ,H294298 ,H294346 ,H294395 ,H294443 ,H294496 ,H294002 ,H294051 ,H294100 ,H294151 ,H294199 ,H294247 ,H294299 ,H294347 ,H294396 ,H294444 ,H294497 ,H294003 ,H294052 ,H294101 ,H294152 ,H294200 ,H294248 ,H294300 ,H294348 ,H294397 ,H294445 ,H294498 ,H294004 ,H294053 ,H294102 ,H294153 ,H294201 ,H294250 ,H294301 ,H294349 ,H294398 ,H294446 ,H294499 ,H294005 ,H294054 ,H294103 ,H294154 ,H294202 ,H294252 ,H294302 ,H294350 ,H294399 ,H294447 ,H294500 ,H294006 ,H294055 ,H294104 ,H294155 ,H294203 ,H294253 ,H294303 ,H294351 ,H294400 ,H294448 ,H294501 ,H294007 ,H294056 ,H294105 ,H294156 ,H294204 ,H294254 ,H294304 ,H294352 ,H294401 ,H294449 ,H294502 ,H294008 ,H294057 ,H294106 ,H294157 ,H294205 ,H294255 ,H294305 ,H294353 ,H294402 ,H294450 ,H294503 ,H294009 ,H294058 ,H294107 ,H294158 ,H294206 ,H294256 ,H294306 ,H294354 ,H294403 ,H294451 ,H294504 ,H294010 ,H294059 ,H294108 ,H294159 ,H294207 ,H294257 ,H294307 ,H294355 ,H294404 ,H294452 ,H294505 ,H294011 ,H294060 ,H294109 ,H294160 ,H294208 ,H294258 ,H294308 ,H294356 ,H294405 ,H294454 ,H294506 ,H294507 ,H294555 ,H294614 ,H294666 ,H294719 ,H294775 ,H294832 ,H294901 ,H294508 ,H294556 ,H294615 ,H294667 ,H294720 ,H294777 ,H294833 ,H294902 ,H294509 ,H294564 ,H294616 ,H294668 ,H294721 ,H294778 ,H294834 ,H294904 ,H294510 ,H294565 ,H294617 ,H294670 ,H294723 ,H294779 ,H294835 ,H294906 ,H294511 ,H294567 ,H294618 ,H294671 ,H294724 ,H294780 ,H294836 ,H294907 ,H294512 ,H294568 ,H294619 ,H294673 ,H294726 ,H294781 ,H294838 ,H294908 ,H294513 ,H294569 ,H294620 ,H294674 ,H294727 ,H294782 ,H294839 ,H294910 ,H294514 ,H294570 ,H294621 ,H294675 ,H294728 ,H294783 ,H294840 ,H294914 ,H294515 ,H294571 ,H294623 ,H294676 ,H294729 ,H294784 ,H294841 ,H294915 ,H294516 ,H294572 ,H294624 ,H294677 ,H294730 ,H294786 ,H294842 ,H294917 ,H294517 ,H294573 ,H294625 ,H294678 ,H294731 ,H294787 ,H294843 ,H294918 ,H294518 ,H294574 ,H294626 ,H294679 ,H294732 ,H294790 ,H294845 ,H294919 ,H294519 ,H294575 ,H294627 ,H294680 ,H294733 ,H294792 ,H294846 ,H294921 ,H294520 ,H294576 ,H294628 ,H294681 ,H294735 ,H294793 ,H294847 ,H294922 ,H294521 ,H294577 ,H294629 ,H294682 ,H294736 ,H294794 ,H294848 ,H294923 ,H294522 ,H294579 ,H294630 ,H294683 ,H294737 ,H294795 ,H294850 ,H294924 ,H294523 ,H294580 ,H294632 ,H294684 ,H294738 ,H294796 ,H294851 ,H294926 ,H294524 ,H294581 ,H294633 ,H294685 ,H294739 ,H294797 ,H294853 ,H294928 ,H294525 ,H294582 ,H294634 ,H294686 ,H294740 ,H294798 ,H294855 ,H294929 ,H294526 ,H294583 ,H294635 ,H294687 ,H294742 ,H294799 ,H294856 ,H294930 ,H294527 ,H294584 ,H294636 ,H294688 ,H294743 ,H294800 ,H294858 ,H294932 ,H294528 ,H294585 ,H294637 ,H294689 ,H294744 ,H294801 ,H294859 ,H294933 ,H294529 ,H294586 ,H294639 ,H294690 ,H294745 ,H294802 ,H294860 ,H294934 ,H294530 ,H294587 ,H294640 ,H294692 ,H294746 ,H294803 ,H294861 ,H294936 ,H294531 ,H294588 ,H294641 ,H294693 ,H294747 ,H294804 ,H294862 ,H294937 ,H294532 ,H294589 ,H294642 ,H294694 ,H294748 ,H294805 ,H294863 ,H294945 ,H294533 ,H294590 ,H294643 ,H294695 ,H294749 ,H294806 ,H294864 ,H294950 ,H294534 ,H294591 ,H294644 ,H294696 ,H294750 ,H294807 ,H294865 ,H294951 ,H294535 ,H294592 ,H294645 ,H294697 ,H294751 ,H294808 ,H294867 ,Hcx4019 ,H294536 ,H294593 ,H294646 ,H294698 ,H294752 ,H294809 ,H294868 ,Hcx4138 ,H294537 ,H294594 ,H294647 ,H294700 ,H294753 ,H294810 ,H294871 ,Hcx4344 ,H294538 ,H294595 ,H294648 ,H294701 ,H294754 ,H294812 ,H294872 ,Hcx4400 ,H294539 ,H294597 ,H294649 ,H294702 ,H294755 ,H294813 ,H294874 ,Hcx4662 ,H294540 ,H294598 ,H294650 ,H294704 ,H294756 ,H294814 ,H294875 ,H294541 ,H294599 ,H294652 ,H294705 ,H294757 ,H294815 ,H294877 ,H294542 ,H294600 ,H294653 ,H294706 ,H294758 ,H294816 ,H294879 ,H294543 ,H294601 ,H294654 ,H294707 ,H294760 ,H294817 ,H294880 ,H294544 ,H294602 ,H294655 ,H294708 ,H294761 ,H294818 ,H294881 ,H294545 ,H294603 ,H294656 ,H294709 ,H294762 ,H294820 ,H294885 ,H294546 ,H294604 ,H294657 ,H294710 ,H294763 ,H294822 ,H294886 ,H294547 ,H294605 ,H294658 ,H294711 ,H294764 ,H294823 ,H294887 ,H294548 ,H294606 ,H294659 ,H294712 ,H294765 ,H294824 ,H294888 ,H294549 ,H294607 ,H294660 ,H294713 ,H294766 ,H294825 ,H294890 ,H294550 ,H294608 ,H294661 ,H294714 ,H294768 ,H294826 ,H294892 ,H294551 ,H294610 ,H294662 ,H294715 ,H294770 ,H294827 ,H294894 ,H294552 ,H294611 ,H294663 ,H294716 ,H294771 ,H294829 ,H294895 ,H294553 ,H294612 ,H294664 ,H294717 ,H294773 ,H294830 ,H294896 ,H294554 ,H294613 ,H294665 ,H294718 ,H294774 ,H294831 ,H294900 |
| FEI Number |
2916710
|
Recalling Firm/ Manufacturer |
Varian Medical Systems, Inc. Oncology Systems 911 Hansen Way Palo Alto CA 94304-1028
|
| For Additional Information Contact | Michael A. Pignataro 650-424-5731 |
Manufacturer Reason for Recall | Varian has identified an anomaly whereby, following prolonged use, the screw fastener holding the wedge body to the tray may fail. |
FDA Determined Cause 2 | Process control |
| Action | The firm, Varian Medical, sent an "Urgent Medical Device Correction" Letter dated March 28, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately inspect all hard wedges for missing or sheared screw fasteners, if any wedge is found to have one or more missing or sheared screw fasteners, remove the wedge from use and notify Varian immediately; inspect all hard wedges for loose screw fasteners, and retighten as necessary, and implement a program of monthly inspection of wedges for loose or missing hardware. The customers were also instructed to advise the appropriate personnel working in their Radiotherapy Department of the content of this letter.
Varian Service shall immediately begin an interim action to replace screw fasteners on all hard wedges for Clinacs installed prior to January 2004. Varian Service will contact the customers to arrange for the inspection visit.
Varian also stated that they shall continue their Investigation of this issue to further clarify the exact conditions under which the wedge screw fasteners may fail, and to analyze possible final solutions. Varian will notify the customers when final determinations are made.
If you require further clarification, please feel free to contact your local Varian Customer Support Services Manager and/or USA and Canada at 1-888-4265; Europe: +41 41 749 8844; or email: North America: support-americas@varian.com; Australia/New Zealand: support-anz@varian.com; Europe: support emea@varian.com; South East Asia: seasia.apps.helpdesk@varian.com; China/Asia: china.apps.Helpdesk@varian.com; Japan: Japan.apps helpdesk@varian.com; Latin America: soporte.al@varian.com, a d Internet: Oncology Systems-www.myvarian.com or Varian Medical Systems - www.varian.com. |
| Quantity in Commerce | 5616 units |
| Distribution | Worldwide distribution: |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = IYE
|
|
|
|