| | Class 2 Device Recall GE Healthcare, DASH 3000, 4000, 5000 Patient Monitors |  |
| Date Initiated by Firm | July 05, 2011 |
| Date Posted | August 25, 2011 |
| Recall Status1 |
Terminated 3 on September 11, 2012 |
| Recall Number | Z-3075-2011 |
| Recall Event ID |
58489 |
| 510(K)Number | K073462 |
| Product Classification |
Monitor, physiological, patient (with arrhythmia detection or alarms - Product Code MHX
|
| Product | GE Healthcare, DASH 3000, 4000, 5000 Patient Monitors
The Dash 3000/4000/5000 patient monitor is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients. The Dash 3000/4000/5000 patient monitor is designed as a bedside, portable, and intra-hospital transport monitor that can operate in all professional medical facilities including but not limited to: emergency department, operating room, post anesthesia recovery, critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care areas located in hospitals, outpatient clinics, free-standing surgical centers, and other alternate care facilities. Physiologic data includes but is not restricted to: electrocardiogram, invasive blood pressure, noninvasive blood pressure, heart rate, temperature, cardiac output, respiration, pulse oximetry, carbon dioxide, bi-spectral index, impedance cardiography, oxygen, and anesthetic agents as summarized in the operator's manual. The Dash 3000/4000/5000 patient monitor is also intended to provide physiologic data over the UNITY NETWORK TM indirectly to clinical information systems (via our Enterprise Gateway) and allow the user to access hospital data at the point-of-care. The information can be displayed, trended, stored, and printed. The Dash 3000/4000/5000 patient monitor was developed to interface with nonproprietary third party peripheral devices that support serial data outputs. |
| Code Information |
Parent SN SBG07088084GA SBG07121347GA SD009513350GA SD009513846GA SD009513847GA SD009524000GA SD009524022GA SD010127715GA SD010470400GA SBG06525985GA SBG06525988GA SBG06526380GA SBG06526381GA SBG07026902GA SBG07026903GA SD010148692GA SBG06526328GA SBG07121291GA SBG07142197GA SBG07142199GA SBG07142200GA SBG07173225GA SBG07173226GA SD009057911GA SD007346050GA SD010127681GA SD010138425GA SD010148740GA SD010429302GA SD010429303GA SD010429304GA SD010449616GA SD010449617GA SD010449618GA SD010449619GA SD010449620GA SD010449621GA SD010449622GA SD010449623GA SD010449624GA SD010449625GA SD010449626GA SD010449627GA SD010449628GA SD010449629GA SD010449630GA SD010127738GA SD010127747GA SD010127749GA SD010189722GA SD010189723GA SD010429041GA SD010429042GA SD010429043GA SD010429044GA SD010429045GA SD010429046GA SD010429047GA SD010429048GA SD010429049GA SD010429050GA SD010429051GA SD010429091GA SD010429092GA SD010429093GA SD010429094GA SD010429095GA SD010429096GA SD010429156GA SD010429157GA SD010429158GA SD010429159GA SD010429160GA SD010429161GA SD010429162GA SD010429163GA SD010481107GA SD010481108GA SD010481109GA SD010481118GA SD010512325GA SD007366627GA SD010459752GA SD010459762GA SD010459763GA SD010459764GA SD010459765GA SD010459766GA SD010459771GA SD010459772GA SD009244563GA SD010127805GA SD010138467GA SD010138484GA SD010148676GA SD010148677GA SD010148688GA SD010148689GA SD010148691GA SD010189649GA SD010189650GA SD010189655GA SD010189656GA SD010189660GA SD010189662GA SD010189663GA SD010189672GA SD010189688GA SD010189689GA SD010189690GA SD010189692GA SD010189693GA SD010189697GA SD010189699GA SD010189705GA SD010189708GA SD010189709GA SD010189713GA SBG07026904GA SBG07026905GA SBG07027014GA SBG07027015GA PA1669553 PA1669554 PA1669559 PA1669560 SBJ08062466GA SBJ08062467GA SD008087509GA SD008087518GA SD009513381GA SD009513383GA SD009119398GA SD010116608GA SD010116609GA SD010116610GA SD010116612GA PA1863389 PA1863390 PA1863391 PA1863392 PA1863393 PA1863394 PA1863395 PA1863396 PA1863397 PA1863398 PA1864326 PA1864327 PA1864328 PA1864329 PA1864330 PA1864331 PA1864332 PA1864333 PA1864334 PA1864335 PA1864336 PA1864337 PA1927648 PA1927649 PA1927650 PA1927652 PA1927653 PA1927654 PA1927655 SD009513221GA SD009513223GA SBG06056893GA SBG06026033GA SBG06505112GA SBG05260079GA SBG05320089GA SD009057919GA SD009119475GA SD009119627GA SD009119628GA SD009119629GA SD009119631GA SD009119633GA SD009119639GA SD009129871GA SD009129872GA SD010459747GA SD010459751GA SD009119394GA PA1669555 SD009119637GA SD009119730GA SD009129868GA SD010459769GA SBG06316168GA SD009119630GA SBG06441002GR SBG07077688GA SD009513320GA SD009513892GA SD009513895GA SBG06223385GA SBG06439722GA SBG06462549GA SBG06172165GA SD009513351GA SD009513352GA SD009523999GA SD009524001GA SD009524002GA SD009524003GA SD009524004GA SD009524005GA SD009524020GA SD009524023GA SD009524024GA SD009524026GA SD009223930GA SD009524178GA SD010127736GA SD008494636GA PA1669556 PA1669558 SD007251984GA SD007397784GA SBG06172209GA SBG06213085GA SBG06213087GA SBG06213089GA SBG06213092GA SBG06213093GA SBG06213094GA SBG06213096GA SBG06213201GA SBG06213366GA SBG06223368GA SBG06223373GA SBG06223374GA SBG06223376GA SBG06223379GA SBG06223380GA SBG06223381GA SBG06223383GA SBG06223421GA SBG06223422GA SBG06223415GA SD007366625GA SD007441101GA SD007441102GA SD007441107GA SD007441108GA SD007441109GA SD007441110GA SD007441111GA SD007441112GA SD007441114GA SD007441115GA SD007441116GA SD007441117GA SD007441118GA SD007441120GA SD007441121GA SD007441122GA SD007441123GA SD008138362GA SD007387286GA SD007387293GA SD007387297GA SD007387298GA SD010313863GA SD007220900GA SD007524738GA SBG06192741GA SBG06305737GA PA1674475.DEL SBG06515496GA SD007439615GA SD007439622GA SD010127739GA SD010127771GA SD010138416GA SD010138418GA SD010138429GA SD010138432GA SD010138433GA SD010138437GA SD010138443GA SD010189718GA SD010189719GA SD010189720GA SD009513848GA SBG06441269GA SD007324883GA SD007335577GA RT507101517GA SD007346139GA SD007514632GA SD007514644GA SD007514692GA SD007514702GA SD007514703GA SD007524730GA SD007524749GA SD008035428GA SD008035429GA SD008035430GA SD008035431GA SD008035432GA SD008035433GA SD008035434GA SD008035435GA SD008035436GA SBG06213358GA SBG06223371GA SBG06223372GA SBG06223388GA SBG06223392GA SBG06223393GA SBG05453020GA SD010460300GA SD010460301GA SD010502298GA SD010502299GA SD007398139GA SBG06213239GA SD009265399GA SBG07193997GA SD009119382GA SD009119669GA SD009234016GA SD009234017GA SD009234018GA SD009234019GA SD009234021GA SD009234023GA SD009234024GA SD009244555GA SD009244557GA SD009244558GA SD009244559GA SD009244560GA SD009244564GA SD009255010GA SD009255011GA SD009255014GA SBG06377825GA SBG07098594GA SD009121159GA SD010376661GA SD010376812GA SD009409691GA SD009513326GA SD009513327GA SBG06441260GA SD007283130GA SD007283147GA SD010127768GA SD010138415GA SD010138436GA SD010138438GA SD010189721GA SD009513321GA SD009513322GA SD009513323GA SD009513325GA SD010189657GA SD010189668GA SD010189669GA SD007439602GA SD008076910GA SBG06284779GA SBG06388178GA SBG06182396GR SD007439601GR SD010303447GA SD010429040GA SD008516168GA SD009121175GA SD008442085GA SBG06526117GA SD009523996GA SD009523998GA SD009524006GA SD009524025GA SD010095485GA SD010095486GA SD010116613GA SD010127708GA SD010148725GA SD010158814GA SBG06398528GA PA1669557 SD007524844GA SD010148744GA SD008077178GA SD010459754GA SD007514617GA SD008045985GA SD010459768GA SBG07173175GA SBG07173339GA SBG07173341GA SBG06056882GA SBG06223375GA SBG06264671GA SBG06264672GA SBG06264673GA SBG06264674GA SBG06264675GA SBG06264676GA SBG06264703GA SBG06264704GA SBG06264705GA SBG06264706GA SD009523997GA SD009524021GA SBG07131798GA SBG07162973GA SD009026873GA SD009026874GA SD009027110GA SD009099248GA SD009099249GA SBG06077666GA SBG06129642GA SBG06161810GA SD010127794GA PA1345863 SD009316925GA SD009368573GA SD008045990GA SD010127797GA SD010127806GA SD010189651GA SD010189653GA SD010189666GA SD010189671GA SD010189731GA SD010459746GA SD010459755GA SD010459757GA SD010459758GA SD007387385GA PA1348049 PA1348050 PA1348051 PA1348053 SBG06367354GA SBG06367356GA SBG06367361GA SBG06367363GA SBG06367364GA SBG06367365GA SBG06377820GA SBG06377823GA SBG06377826GA SBG06377829GA SBG06377831GA SBG06377832GA SD007283024GA SD007304032GA SD008076862GR SD010324795GA SD010366288GA SD010366291GA SD010366293GA SD010366295GA SD010366298GA SD008148576GA SD009119671GA SD008473919GA SBG06141092GA SBG06141093GA SBG06141094GA SBG06141095GA SBG06182342GA SBG06182343GA SBG06182345GA SBG06182346GA SBG06182347GA SBG06182348GA SBG06182349GA SBG06182351GA SBG06182352GA SBG06182353GA SBG06182355GA SBG06182357GA SBG06182358GA SBG06182359GA SD009420681GA SD009420682GA SD010158825GA SD010158826GA SD010158828GA SD010158829GA SBG06504880GA SBG06505111GA SBG06515440GA SBG06515575GA SBG06515576GA SBG06525834GA SBG06525835GA SBG07121341GA SBG07121346GA SD008462855GA SD008462856GA SD008462858GA SD008462861GA SD008463065GA SD008463066GA SD008463069GA SD008463070GA SBG06141030GA SBG06141033GA SBG06141034GA SBG06141037GA SD008473914GA SD008473920GA SD007241685GA SD007241687GA SD007241691GA SD007241694GA SD007283122GA SD007283123GA SD007283124GA SD007283125GA SD007283126GA SD007283127GA SD007283128GA SD007283129GA SD007283131GA SD007283133GA SD007283136GA SD007283137GA SD007283138GA SD007283139GA SD007283140GA SD007283142GA SD007283143GA SD007283144GA SD007283145GA SD007283146GA SD007283148GA SD007283149GA SD007283150GA SD007283151GA SD007283152GA SD007283153GA SD007283154GA SD007283155GA SD007283156GA SD007283158GA SD007387384GA SD007387387GA SD008336012GA SD009420528GA SD009420651GA SD009420652GA SD009420656GA SD009451647GA SD009451648GA SD009451649GA SD009451650GA SD009451651GA SD009451652GA SD009451654GA SD009451655GA SD009451656GA SD009451657GA SD009451658GA SD009451659GA SD009451660GA SD009451661GA SD009502925GA SD009502926GA SD009502927GA SD009502928GA SD009502929GA SD009502930GA SD009524157GA SD010127653GA SD010127654GA SD010127655GA SD010127710GA SD010127711GA SD010127712GA SD010127714GA SD010127718GA SD010127720GA SD010127721GA SD010127722GA SD010127723GA SD010127724GA SD010127725GA SD010127726GA SD010127728GA SD010127729GA SD010127730GA SD010127731GA SD010127732GA SD009119716GA SD009121199GA SD009172301GA SD009223687GA SD009223688GA SD009223689GA SD009223690GA SD009223707GA SD009223710GA SD009223711GA SD009223712GA SD009223713GA SD009223714GA SD009223716GA SD009223717GA SD009223718GA SD009223720GA SD009223721GA SD009223722GA SD009378909GA SD009378910GA SD010127645GA SD010138287GA SD010189676GA SD010189677GA SD010189678GA PA1533921 PA1533922 PA1533923 PA1533924 PA1533925 PA1533926 SBG06326308GA SBG06326309GA SBG06326310GA SBG06326311GA SBG07121296GA SBG07121297GA SBG07121299GA SBG07121300GA SBG07121302GA SBG07121303GA PA1933176 SD010148549GA SD010148553GA SD010148558GA SD010148673GA SD010148685GA SD010148686GA SD010148690GA SD010148693GA SD010148694GA PA1653267 PA1653268 PA1653269 PA1653270 PA1653271 PA1653272 PA1653273 PA1653274 PA1653275 PA1653276 PA1653537 PA1654300 PA1743077 SD009037469GA SD009037528GA SD009037536GA SD009037538GA SD009037539GA SBG06254287GA SBG06254379GA SBG06254382GA SD010314152GA SD010314155GA SD010314292GA SD010314294GA SD010314297GA PA1533890 SD007335425GA SD007346047GA SD007346048GA SD007346052GA SD007346053GA SD007346054GA SD007346055GA SD007397979GA SD007397981GA SD007397982GA SD007397983GA SD007397984GA SD007397985GA SD007397986GA SD007397987GA SD007397988GA SD007397989GA SD007397990GA SD007397991GA SD007397992GA SD007397993GA SD007397994GA SD007397995GA SD007397996GA SD007397997GA SD007397998GA SD007398000GA SD007398001GA SD007398002GA SD007398003GA SD008066784GA SD008076845GA SD008076853GA SD008076855GA SD008076857GA SD008076866GA SD007325112GA SD007335411GA SD007335413GA SD007335414GA SD007335415GA SD007335417GA SD007335421GA SD007346049GA SD007346051GA SD007397980GA SD007366658GR SD007439476GR SBG06182255GA SBG06182258GA SBG06182261GA SBG06182265GA SBG06182292GA SBG06182297GA SBG06182298GA SBG06182299GA SBG06182366GA SBG06182371GA SD010376974GA SD010377048GA SD010377049GA SD010377050GA SD010377057GA SD010387381GA PA1349672 PA1349674 PA1349675 PA1349676 PA1349677 SD009223715GA SD009254894GA SD009254895GA SD009254896GA SD010376970GA SD010377022GA SD010387378GA SD010387382GA SBG06305736GA SD008127944GA SD010377031GA SD010387380GA SD010387383GA SBG06451432GA SBG06451433GA SBG06451434GA SBG06451435GA SBG06451436GA SBG06451437GA SBG06451438GA SBG06451439GA SBG06451440GA SBG06451441GA SBG06451443GA SD008367205GA SD008367206GA SD008367208GA SD008367209GA SD008367372GA SD008367373GA SD008367374GA SD008367375GA PA1927580 PA1927581 PA1932268 PA1932269 PA1932270 PA1932271 PA1932272 PA1932273 SD009099069GA SD009129959GA PA1653256 PA1673698 PA1673699 PA1673701 PA1673702 PA1673704 PA1673710 PA1673712 PA1673714 PA1673715 PA1673716 PA1673718 PA1673719 PA1673720 PA1673721 PA1673722 PA1673724 PA1673725 PA1673726 PA1673727 PA1673728 PA1673729 PA1673730 PA1673731 PA1673732 SD007293550GA SD007366619GA SD007366620GA SD007366624GA SD007366628GA SD007366630GA SD007366632GA SD007472597GA SD009119729GA SD009223891GA SD009317056GA PA1741573 PA1741574 PA1741575 PA1741576 PA1741577 PA1741578 PA1741579 PA1743174 PA1743175 PA1743176 PA1743177 PA1743178 PA1743179 PA1743180 PA1743181 PA1743182 PA1743183 PA1743184 PA1743185 PA1743186 PA1743187 PA1743188 SD009119387GA SD009119617GA SD009223764GA SD009223897GA SD009223898GA SD009223902GA SD009223903GA SD009223904GA SD009234161GA SD009234162GA SD009234163GA SD009234164GA SD009244562GA SD009244718GA SD009244720GA SD009244721GA SD009244722GA SD009244723GA SD009244724GA SD009255005GA SD009255006GA SD009255007GA SD009255008GA SD009255013GA SD009513893GA PA1925452 PA1925453 PA1925454 PA1925455 PA1925456 PA1925457 SD009513222GA SD009513328GA PA1499698 PA1534443 PA1534445 PA1534446 PA1534447 PA1534450 PA1534452 PA1464530 PA1464531 PA1464532 PA1464533 SBG07121294GA SBG07142119GA SBG07142120GA SBG07142121GA SD010127685GA SD010127688GA SD010138312GA SD010138313GA SD008087515GA SD008097594GA SD008097595GA SD008097597GA SD008097598GA SD008097599GA SD008097875GA SD008097876GA SD008097877GA SD008097878GA SD008097879GA SD008097880GA SD008097881GA SD008097882GA SD008097885GA SD008097886GA SD008097887GA SD008168772GA SD008168774GA SD008168775GA SD008168777GA SD008168778GA SD007483097GA SD007514596GA SD008325393GA SD008367243GA SD008367244GA SD010074760GA SD010074761GA SD010074762GA SD010074763GA SD010074764GA SD010074765GA SD010074766GA SD010074814GA SD010127792GA SD010138480GA SD010148668GA SD010189686GA SD010189691GA SD010189694GA SD010189695GA SD010189696GA SD010189698GA SD010189700GA SD010189704GA SD010189706GA SD010189707GA SD010189710GA SD010189711GA SD010189712GA PA1650693 PA1650694 PA1650695 PA1650696 PA1650698 PA1650699 PA1650700 PA1668172 PA1668173 PA1668174 PA1668175 PA1668176 PA1668177 PA1668178 SD009119379GA SD009119618GA SD009223789GA SD010158818GA SD010377033GA SD010377036GA SD010377040GA SD010377043GA SD010377059GA SD008315149GA SD008325404GA SD008325407GA SD008325409GA SD008325410GA SD008325412GA SD008325413GA SD008419658GA SD008419660GA SD008421280GA SD008429834GA SD008429841GA SD008429843GA PA1347674 PA1347734 PA1347735 PA1347736 PA1347737 PA1347738 PA1347739 PA1347740 PA1347741 PA1347742 PA1347743 PA1347744 PA1347745 PA1347746 PA1347747 PA1347748 PA1347749 PA1347750 PA1347751 PA1347752 PA1347753 PA1347754 PA1347755 PA1347772 PA1347773 PA1394092 PA1394093 PA1394094 PA1394095 PA1394096 PA1394097 PA1394098 PA1394099 PA1394100 PA1394101 PA1394102 PA1394103 PA1460544 PA1460545 PA1460546 PA1460547 PA1460548 PA1462004 PA1462005 PA1462006 PA1462007 PA1462008 PA1462009 PA1462010 PA1462011 PA1462012 PA1462013 PA1462014 PA1462015 PA1462016 PA1462017 PA1462018 PA1462019 PA1462020 PA1462021 PA1462022 PA1483683 PA1483684 PA1483685 PA1483686 PA1483687 PA1483688 PA1499482 PA1563806 PA1563807 PA1563808 PA1563809 PA1563810 PA1563811 PA1563812 PA1563813 PA1563814 PA1563815 PA1563816 PA1563817 PA1607422 PA1607423 PA1607424 PA1607425 PA1607426 PA1607427 PA1607428 PA1607429 PA1607430 PA1607431 PA1607432 PA1607433 PA1650286 PA1650287 PA1651519 PA1651520 PA1651521 PA1651522 PA1651523 PA1651524 PA1651525 PA1651526 PA1653393 PA1653394 PA1653395 PA1653396 PA1653397 PA1653398 PA1653399 PA1653400 PA1653401 PA1653402 PA1654748 PA1654749 PA1654750 PA1654751 PA1654752 PA1709589 PA1709590 PA1709591 PA1709592 PA1741587 PA1741588 PA1741589 PA1741590 PA1741591 PA1741592 PA1741662 PA1741663 PA1741664 PA1742231 PA1742232 PA1742233 PA1742234 PA1742235 PA1742236 PA1742237 PA1742238 PA1742239 PA1742240 PA1742241 PA1742242 PA1763800 PA1862974 PA1862975 PA1862976 PA1862977 PA1862978 PA1862979 PA1927886 PA1927887 PA1927888 PA1927889 PA1927890 PA1927891 PA1927892 PA1927893 PA1927894 PA1927895 PA1933039 PA1933040 PA1933041 PA1933042 PA1933043 PA1933044 PA1933045 PA1933046 PA1933047 PA1933048 PA1934765 PA1934766 PA1934767 PA1934768 PA1934769 PA1934770 PA1934771 PA1934772 PA1934773 PA1934774 PA1934775 PA1934776 PA1934777 PA1934778 PA1934779 SD009524148GA SD009524154GA SD009524156GA SD009400078GA SD009400104GA SD009410365GA SD009410368GA SD009441092GA SD009441093GA SD009441096GA SD009441097GA SD009441098GA SD009441099GA SD009441100GA SD009441101GA SD009441104GA SD009441105GA SD009441106GA SD009441107GA SD009441108GA SD009441109GA SD009441111GA SD009441112GA SD009441113GA SD009441114GA SD009441115GA SD009441116GA SD009441117GA SD009441118GA SD009441119GA SD009441121GA SD009441284GA SD009441285GA SD009441286GA SD009441289GA PA1345571 PA1345572 PA1345573 PA1345574 PA1345587 PA1345589 PA1345591 PA1345592 PA1345607 PA1345608 PA1345609 PA1345610 PA1345620 PA1345853 PA1345855 PA1345859 PA1345861 PA1345864 PA1345865 PA1345866 PA1345869 PA1345870 PA1345871 PA1345872 PA1345873 PA1345874 PA1345876 PA1345884 PA1345887 PA1345917 PA1345921 PA1345946 PA1345951 PA1345955 SBG06026025GA SBG06026026GA SBG06026027GA SBG06026028GA SBG06056880GA SBG06056881GA SBG06056883GA SBG06056885GA SBG06056886GA SBG06056887GA SBG06056890GA SBG06056891GA SBG06056892GA SBG06056894GA SBG06056895GA SBG06141121GA SBG06141123GA SBG06182243GA SBG06182245GA SBG06182247GA SBG06182249GA SBG06182250GA SBG06182251GA SBG06182257GA SBG06182260GA SBG06182262GA SBG06182263GA SBG06182264GA SBG06182267GA SBG06182293GA SBG06182301GA SBG06223423GA SD008325557GA SD008325573GA SD008325577GA SD008325579GA PA1668167 PA1668168 PA1668169 SBG05504736GA SBG06026024GA SBG06141122GA SBG06182252GA SBG06182256GA SBG06182266GA SD008077122GA SBG06388157GA SBG06482962GA SBG06482967GA SBG06483161GA SBG06483196GA SBG06483197GA SBG06483199GA SBG06483229GA SBG06483266GA SBG06483267GA SBG06483268GA SBG06505110GA SBG06515574GA SD007504079GA SD007504084GA SD007504086GA SD007504087GA SD008462794GA SD008462795GA SD008462796GA SD008462797GA SD008462800GA SD008462802GA SD009316705GA SD009316706GA SD009420540GA SD009420541GA SD009420544GA SD009420648GA SBG06141111GA SBG06141113GA SBG06141114GA SBG06182302GA SBG06182382GA SBG06182383GA SBG06182385GA SBG06182386GA SBG06182387GA SD008294325GA SD008314961GA PA1709611 PA1709612 PA1835309 PA1835310 PA1835311 PA1835312 PA1835313 PA1835314 PA1835315 PA1835316 PA1835317 PA1835318 PA1835319 PA1835320 PA1835321 PA1835322 PA1835323 PA1835324 PA1835325 PA1835326 PA1651527 PA1651640 PA1651641 PA1651643 PA1656011 PA1656012 PA1656013 PA1656014 PA1656015 PA1656016 PA1656017 PA1656018 PA1656019 PA1656020 PA1656021 PA1656022 PA1656023 PA1656024 PA1656025 PA1656026 PA1656027 PA1656028 SD009078537GA SD009078539GA SD009078540GA SD009078541GA SD009119381GA SD009119383GA SD009119384GA SD009119385GA SD009119386GA SD009119388GA SD009119390GA SD009119391GA SD009119392GA SD009119393GA SD009119396GA SD009119621GA SD009119623GA SD009119625GA SD009119626GA SD009119719GA SD009119720GA SD009119721GA SD009119722GA SD009119723GA SD009129818GA SD009129819GA SD009129820GA SD009129822GA SD009129823GA SD009129826GA SD009129827GA SD009244556GA SD009255012GA SD010127748GA SD010138417GA SD010138419GA SD010138422GA SD010138424GA SD010138426GA SD010138439GA SD010138440GA SD010138446GA SD010138448GA SD010138449GA SD010148547GA SD010148552GA SD010261999GA SD010262000GA SD010262002GA SD010262003GA SD010262004GA SD010262005GA SD010262006GA SD010262009GA SD010262010GA SD010262011GA SD010262012GA SD010262013GA SD010262160GA SD010262161GA SD010262162GA SD010262168GA SD010262171GA SD010262172GA SD010262173GA SD010262174GA SD010148555GA SD010148557GA SD010148559GA SD010148671GA PA1414755 PA1414756 PA1414757 PA1414758 PA1414759 PA1414760 PA1414761 SBG07077673GR PA1932274 PA1932275 PA1932276 PA1932277 PA1932278 PA1932279 PA1932280 PA1932281 PA1932282 SD007210470GA SD007304034GA SD007314567GA SD007314574GA SD007314575GA SD007314576GA SD007314578GA SD007314580GA SD007314710GA SD007314715GA SD007314721GA SD007345832GA SD007345853GA SD007345854GA SD007356302GA SD007356303GA SD007356304GA SD007356305GA SD007356306GA SD007356308GA SD007356309GA SD007356310GA SD007356311GA SD007356312GA SD007356313GA SD007356314GA SD007356315GA SD007356316GA SD007366650GA SD007366652GA SD007377077GA SD007377079GA SD007377080GA SD007377088GA SD007377092GA SD007377113GA SD007377114GA SD007377115GA SD007377116GA SD007377117GA SD007377119GA SD007377121GA SD007377122GA SD007377123GA SD007377124GA SD007377125GA SD007377126GA SD007377127GA SD007377128GA SD007377129GA SD007377130GA SD007377131GA SD007377132GA SD007377133GA SD007377134GA SD007377137GA SD007377138GA SD007377139GA SD007377140GA SD007377141GA SD007377143GA SD007377144GA SD007377145GA SD007377146GA SD007377147GA SD007387338GA SD007387339GA SD007387340GA SD007387341GA SD007303958GA SD007303964GA SD007325116GA SD007325119GA SD007366648GA SD007366651GA SD007366659GA SD007366661GA SD007377078GA SD007377081GA SD007377082GA SD007377083GA SD007377084GA SD007377085GA SD007377086GA SD007377087GA SD007377089GA SD007377090GA SD007377091GA SD007377120GA SD007377135GA SD007377142GA SD007377150GA SD007377158GA SD008138359GA PA1348048 PA1348057 PA1348058 PA1348059 PA1348060 PA1348064 PA1348065 PA1348066 SD008304548GA SD008304565GA SBG06213175GA SBG06213176GA SBG06316061GA SD007366622GA PA1498775 PA1534444 PA1534448 PA1534449 PA1534451 SD008056175GA SBG06213123GA SBG06213127GA SBG06213132GA SBG06213135GA SBG06213187GA SBG06213192GA SBG06213199GA SBG06336584GR SD009234166GA SD009409687GA SD010127791GA SD010138479GA SD010313831GA SD010313832GA SD010313837GA SD010313862GA SD010313866GA SD010313871GA SD010314355GA SD010314356GA SD010314358GA SD010314363GA SD010314366GA SD010324438GA SD010376616GA SD010376623GA SD010376624GA SD010376625GA SD010376626GA SD010376627GA SD010376628GA SD010376629GA SD010376630GA SD010376633GA SD010376638GA SD010376639GA SD010376971GA SD010377026GA SD010377027GA SD010377030GA SD010377032GA SD010377045GA SD010376966GA SD010376967GA SD010376968GA SD010376969GA SD010376972GA SD010376973GA SD010376976GA SD010376977GA SD010377021GA SD010377023GA SD010377024GA SD010377025GA SD010377028GA SD010377029GA SD010377034GA SD010377035GA SD010377038GA SD010377039GA SD010377041GA SD010377044GA SD010377046GA SD010377051GA SD010377052GA SD010377054GA SD010377055GA SD010377056GA PA1652042 PA1652043 PA1652044 PA1652045 PA1652046 PA1652047 PA1741581 PA1741583 PA1741584 SBG06192589GA SBG06192590GA SBG06192595GA SBG06192597GA SBG06429549GR SD009057897GR SD008076873GA SD008077176GA SD008283660GA SD008283664GA SD008325578GA SD008388086GA SD008388088GA SD008388089GA SD010127661GA SD010138492GA SD010148700GA SD010148705GA SD010148708GA SD010148709GA SBG07026904GR SBG07088102GR SBG07088103GR SD008168773GA SD008189140GA SD008314964GA SD008367207GA SD009078507GA SD009078513GA SD009119466GA SD009119474GA SD009119479GA SD009223888GA SD009234168GA SD009234169GA PA1484893 PA1484894 SBG06316167GA SBG06316169GA SBG06505162GA SBG06525986GA SBG06525987GA SBG06526320GA SBG06316166GA SD009119478GA SD009119632GA SD009129870GA SD009141613GA SD009141623GA SD009223652GA SD009223654GA SD009223656GA SD009223765GA SD009223890GA SD009244719GA SBG06526383GA PA1464874 PA1464875 PA1464876 PA1464877 PA1464878 PA1464879 PA1464880 PA1464881 PA1464882 PA1464883 PA1464884 PA1464885 PA1927793 PA1927794 PA1927795 PA1927796 PA1927797 PA1927798 PA1927799 PA1927801 PA1934064 PA1934065 PA1934066 SD010127703GA SD010138468GA SD010138485GA SD010148542GA SD010148545GA SD010148556GA SD010148657GA SD010148675GA SD010148679GA SD010189685GA SD010189687GA SD010189701GA SD007439604GR SD007439608GR SD008442092GA SD008473362GA SD008473363GA SD008473364GA SBG06346972GA SBG06346974GA SBG06377775GA SBG06377776GA SBG06377777GA SBG06377778GA SBG06377779GA SBG06377780GA SBG06377781GA SBG06377782GA SBG06377824GA SBG06377827GA SBG06377828GA SBG06377830GA SBG06388014GA SBG06398319GA SBG06398320GA SBG06398321GA SBG06398322GA SBG06398323GA SBG06398324GA SBG06398325GA SBG06398515GA SBG06377821GA SBG06377822GA SD009119380GA SD009119389GA SD009119395GA SD009119619GA SD009119620GA PA1607390 PA1607392 SD008209737GA SD008211082GA SD008211090GA SD008325392GA SD008367241GA SD008367242GA SD008377924GA SD008377925GA SD008377926GA SD008377927GA SD008377928GA SD008377932GA SD008388024GA SD008388025GA SD008388026GA SD008388027GA SD008388028GA SD008388029GA SD008388031GA SD008388033GA SD008494530GA SBG06515438GA SBG06525831GA SBG06525833GA SBG06526118GA SBG06526314GA SBG06526315GA SBG06526316GA SBG06526317GA SBG06526318GA SBG06526321GA SBG06526322GA SBG06526323GA SBG06526324GA SBG06526325GA SBG06526326GA SBG06526327GA SBG06526329GA SBG06526330GA SBG06526337GA SBG06526338GA SBG06526379GA SBG06526382GA SBG06129622GA SBG06151692GA SD008494521GA SD008494525GA SD008494527GA SD008494531GA SD009513894GA SD008189079GA SD008209729GA SD008209731GA SD008209732GA SD008209735GA SD008209740GA SD008211087GA SD008219940GA SD008219942GA SD008219943GA SD007220843GA SD007231020GA SD007231021GA SD007252021GA SD007252022GA SD007252023GA SD007252025GA SD007252026GA SD007252027GA SD007252028GA SD007252029GA SD007252030GA SD007252033GA SD007252138GA SD007252141GA PA1743218 PA1743219 PA1743221 PA1743224 SBG07183564GA SBG07183565GA SBG07183567GA SD008077154GA SD008077163GA SD008077166GA SD008138360GA SD008138361GA SD008138363GA SBG06192736GA SBG06213361GA SBG06213365GA SBG06223370GA SBG06223391GA SBG06284768GA SBG06439719GA SBG06439725GA SBG06439729GA SBG06449917GA SBG06449918GA SBG06449919GA SBG06449920GA SBG06449921GA SBG06449922GA SBG06449923GA SBG06449924GA SBG06493800GA SBG06504954GA SBG06504955GA SBG06505105GA SBG06505106GA SBG06505107GA SBG06505108GA SBG06505109GA SBG06515497GA SBG06515498GA SBG07121295GA SBG07121301GA SD007282994GA SD007282996GA SD007293549GA SD007441100GA SD007441119GA PA1933613 PA1933822 PA1933823 PA1933824 PA1933825 PA1934062 PA1934063 PA1934067 PA1934069 PA1934073 PA1934075 PA1934076 PA1934078 PA1934079 SD010366208GA SD010366217GA SD010366218GA SD008219938GA SD008219939GA PA1743088 PA1743194 SBG06141096GA SBG06141097GA SBG06141098GA SBG06141099GA SBG06141100GA SBG06172163GA SBG06172164GA SBG06172166GA SBG06172167GA SBG06192754GA SBG06192755GA SBG06192756GA SBG06192757GA SBG06192758GA SBG06192759GA SBG06192760GA SBG06192761GA SBG06192762GA SBG06192763GA SBG06192765GA SBG06192766GA SBG06192768GA SBG06192769GA SBG06192772GA SBG06192773GA SBG06192774GA SBG06192776GA SBG06336573GA SBG06483019GA SBG06483023GA SD009223719GA PA1934068 SD009524160GA SD009524161GA SD009524162GA SD009524163GA SD009524164GA SD009524176GA SD009524179GA SD010127733GA SD008473937GA SD009203573GA SD009223925GA SD009223926GA SD009223927GA SD009223928GA SD009223929GA SD009223932GA SD009223933GA SD009223934GA SD009223935GA SD009223936GA SD009223937GA SD009223938GA SD009223939GA SD009223940GA SD009223941GA SD009223942GA SD009233995GA SD009233996GA SD009233997GA SD009233998GA SD009233999GA SD009234000GA SD009234002GA SD009234003GA SD009234004GA SD009234147GA SD009234149GA SD009244733GA SD009316902GA SD009316903GA SD009316904GA SD009316907GA SD009358127GA SBG06223382GA SBG06223384GA SBG06223418GA SBG06223420GA SD008463250GA SD008463252GA SD008473553GA SD008473554GA SD008473561GA SD008473562GA SD008473925GA SD008473931GA SD008473934GA SD008494699GA SD008494701GA SD009037203GA SD009037204GA SD009037207GA SD009119718GA SD009399475GA SD009410359GA SD009026875GA SD009027111GA PA1345909 PA1345910 PA1345916 PA1347661 PA1347662 PA1347663 PA1347664 PA1347665 PA1347666 PA1347667 PA1347668 PA1347669 PA1347670 PA1347671 PA1347672 PA1347673 SD009327364GR SD008077185GR SD009172305GA SD009223655GA SD009534341GA SBG06316164GA SBG06367350GA SBG06367351GA SBG06377800GA SBG06377801GA SBG06419411GA SBG06419481GA SBG06483262GA SBG06483263GA SBG07121335GA SBG07121336GA SBG07121338GA SBG07121339GA SBG07121340GA SBG07121342GA SBG07121343GA SD007252059GA SD007335441GA SD007335443GA SD007335445GA SD007408272GA SD007408273GA SD007439483GA SD007439484GA SD008127943GA SD008127945GA SD008168770GA SD008168776GA SD008188954GA SD008189077GA SD008189084GA SD008189141GA SD008505338GA SD008505339GA SD008515759GA SD008515760GA SD009472036GA SD009472037GA SD009472038GA SD009482528GA SD007220664GA SD007251981GA SD007251982GA SD007251985GA SD007252035GA SD007252037GA SD007252039GA SD007252041GA SD007252044GA SD007252045GA SD007252046GA SD007252049GA SD007252068GA SD007252071GA SD007252072GA SD007252075GA SD007283031GA SD007293365GA SD007293514GA SD007293573GA SD007304030GA SD007304056GA SD007314573GA SD007210420GA SD007220612GA SD007220615GA SD007220662GA SD007220663GA SD007220665GA SD007220666GA SD007220901GA SD007252060GA SD007397754GA SD007397755GA SD007397756GA SD007397757GA SD007397758GA SD007397760GA SD007397761GA SD007397762GA SD007397763GA SD007397765GA SD007397766GA SD007397767GA SD007397768GA SD007397770GA SD007397771GA SD007397772GA SD007397773GA SD007397774GA SD007397775GA SD007397776GA SD007397778GA SD007397779GA SD007397780GA SD007397781GA SD007397782GA SD007397783GA SD007397785GA SD007397786GA SD007397787GA SD007397788GA SBG06182384GA SBG06182394GA SBG06192588GA SBG06192591GA SBG06192592GA SBG06192593GA SBG06192594GA SBG06192596GA SBG06213129GA SBG06213130GA SBG06213131GA SBG06213134GA SBG06213136GA SBG06213189GA SBG06213191GA SBG06213194GA SBG06213362GA SBG06223389GA SBG06223409GA SBG06223412GA SBG06525832GA SBG06526363GA SBG07037072GA PA1654584 PA1654585 PA1654586 PA1742191 PA1742192 PA1742193 PA1742194 PA1742195 PA1742196 PA1742197 PA1742198 PA1742199 PA1742200 PA1742201 PA1742202 PA1742203 PA1742204 PA1742205 PA1742206 PA1742207 PA1742208 PA1742209 PA1742273 PA1742274 PA1742276 PA1742278 PA1742279 PA1742280 PA1742281 PA1742282 PA1742283 PA1742284 PA1742285 PA1742286 PA1742287 PA1742290 PA1742291 PA1742292 PA1742294 PA1742295 PA1770099 PA1770100 SD009037446GR PA1279385 PA1279386 PA1279387 PA1279388 PA1279389 PA1279390 PA1279391 PA1279392 PA1279393 PA1279394 PA1279395 PA1279396 PA1279397 PA1279398 PA1279399 PA1279400 PA1279401 PA1279402 PA1279403 PA1279404 PA1279405 PA1279406 PA1279407 PA1279408 PA1279409 PA1279410 PA1279411 PA1279412 PA1279413 PA1279414 PA1279415 PA1279416 PA1345575 PA1345576 PA1345577 PA1345578 PA1345579 PA1345580 PA1345581 PA1345582 PA1345583 PA1345584 PA1345585 PA1345586 PA1345588 PA1345590 PA1345593 PA1345594 PA1345595 PA1345596 PA1345597 PA1345598 PA1345599 PA1345600 PA1345601 PA1345602 PA1345603 PA1345604 PA1345605 PA1345606 PA1345611 PA1345612 PA1345613 PA1345614 PA1345615 PA1345616 PA1345617 PA1345618 PA1345619 PA1345913 PA1345914 PA1345918 PA1345919 PA1345920 PA1345922 PA1345923 PA1345924 PA1345925 PA1345926 PA1345927 PA1345928 PA1345929 PA1345930 PA1345931 PA1345932 PA1345933 PA1345936 PA1345937 PA1345938 PA1345939 PA1345940 PA1345941 PA1345943 PA1345944 PA1345945 PA1345947 PA1345948 PA1345949 PA1345950 PA1345952 SBG07098527GA SBG07193840GA SBG07193842GA SBG07193843GA SBG07193844GA SBG07193847GA SBG07193849GA SBG07193850GA SBG07193852GA SBG07193853GA SBG07193854GA SBG07193855GA SBG07193856GA SBG07193857GA SBG07193858GA SBG07193859GA SD007220616GA SD007220899GA SD007252036GA SD007252042GA SD007252043GA SD007252048GA SD007252069GA SD007252070GA SD007252073GA SD007252074GA SD007252077GA SD007252078GA SD007252079GA SD007252080GA SD007252334GA SD007252339GA SD007252344GA SD007441226GA SD007441229GA SD007483122GA SD007483123GA SD007483124GA SD007483125GA SD007483126GA SD007483127GA SD007483128GA SD007483129GA SD007514606GA SD007514607GA SD007514608GA SD007514609GA SD007514610GA SD007514611GA SD007514612GA SD007514615GA SD007514616GA SD007514618GA SD007514619GA SD007514620GA SD007514621GA SD007514622GA SD007514623GA SD007514624GA SD007514625GA SD007514626GA SD007514627GA SD007514630GA SD007514631GA SD007514633GA SD007514634GA SD007514635GA SD007514636GA SD007514637GA SD007514638GA SD007514639GA SD007514640GA SD007514641GA SD007514642GA SD007514643GA SD007514645GA SD007514651GA SD007514652GA SD007514653GA SD007514654GA SD007514655GA SD007514657GA SD007514658GA SD007514659GA SD007514660GA SD007514661GA SD007514662GA SD007514663GA SD007514664GA SD007514665GA SD007514666GA SD007514667GA SD007514668GA SD007514670GA SD007514671GA SD007514672GA SD007514673GA SD007514674GA SD007514675GA SD007514676GA SD007514677GA SD007514678GA SD007514679GA SD007514680GA SD007514681GA SD007514682GA SD007514683GA SD007514684GA SD007514685GA SD007514686GA SD007514687GA SD007514688GA SD007514689GA SD007514690GA SD007514691GA SD007514693GA SD007514694GA SD007514695GA SD007514696GA SD007514697GA SD007514698GA SD007514699GA SD007514700GA SD007514701GA SD007514704GA SD007514705GA SD007514707GA SD007524728GA SD007524731GA SD007524732GA SD007524733GA SD007524734GA SD007524735GA SD007524736GA SD007524737GA SD007524739GA SD007524740GA SD007524741GA SD007524742GA SD007524743GA SD007524744GA SD007524745GA SD007524746GA SD007524747GA SD007524748GA SD007524750GA SD007524752GA SD008035427GA SD008045898GA SD008077082GA SD008077084GA SD008077088GA SD008077144GA SD008077150GA SD008494540GA SD008505409GA SD008505410GA SD008516150GA SD008516154GA SD008516160GA SD008516162GA SD008516163GA SD008516164GA SD008516165GA SD008516167GA SD008516169GA SD008516170GA SD008516171GA SD008516172GA SD008516173GA SD008516174GA SD008516175GA SD008516176GA SD008516177GA SD008516178GA SD008516179GA SD008526455GA SD008526456GA SD008526457GA SD008526460GA SD008526461GA SD008526462GA SD009441103GA SD009441110GA SD009441120GA SD009441287GA SD009441288GA SD009461824GA SD009461826GA SD009461827GA SD009461828GA SD009461829GA SD009461830GA SD009461831GA SD009461832GA SD009472019GA SD009472020GA SD009472021GA SD009472022GA SD009472023GA SD009472024GA SD009472025GA SD009472027GA SD009472028GA SD009472029GA SD009472030GA SD009472032GA SD009472033GA SD009472034GA SD009472108GA SD009472109GA SD009482529GA SD009482530GA SD009482531GA SD009482532GA SD009482533GA SD009502935GA SD009502936GA SD009502937GA SD009502940GA SD010095483GA SD010127702GA SD010127778GA SD010127780GA SD010127789GA SD010127790GA SD010127799GA SD010138466GA SD007441105GA SD007441106GA SD007441113GA SD008526527GA SD008526528GA SD008526529GA SD008526530GA SD008526531GA SD008526532GA SD008526533GA SD008526534GA SD008526535GA SD008526536GA SD008526538GA SD008526539GA SD008526540GA SD008526541GA SBG06504312GA SBG06504314GA SBG06504315GA SBG06504318GA SD007472717GA SD007472718GA SD008076913GA SD008077135GA SD008077140GA SD008077159GA SD008077160GA SD008077171GA SD008242260GA SD008242262GA SD008242264GA SD009078552GA SD009099241GA SD009099252GA SD009119493GA SD009121190GA SD007366735GR SBG06192510GA SBG06192525GA SBG06192716GA SBG06213231GA SBG06326225GA SBG06326226GA SBG06493683GA PA1651308 PA1653461 PA1653462 PA1653463 PA1653464 PA1653465 PA1653466 PA1653467 PA1653468 PA1653469 PA1653470 PA1653471 PA1653472 PA1674684 PA1674685 PA1674686 PA1674689 PA1674690 PA1674691 PA1674692 PA1674693 PA1674694 PA1674695 PA1674696 PA1674697 PA1674698 PA1674699 PA1674700 PA1674701 PA1674703 PA1674704 PA1674705 PA1674706 PA1674707 SD008076880GA SD008077121GA SD008077123GA SD008077132GA SD008077133GA SD008077152GA SD008409089GA SD008409346GA SD008429796GA SD008429798GA PA1345908 PA1345912 PA1345915 PA1345934 PA1345942 PA1345953 PA1499693 PA1499694 PA1499695 PA1499696 PA1499697 PA1499699 SBG06441257GA SBG06515499GA SBG06525994GA PA1533317 PA1533318 PA1533319 PA1533320 PA1533321 PA1533322 PA1533323 PA1533324 PA1533325 PA1533326 PA1533327 PA1533328 PA1743333 PA1743974 PA1743975 PA1743976 PA1743977 PA1743983 PA1743986 PA1743987 PA1743988 PA1743989 PA1743990 PA1743993 PA1743994 PA1743995 PA1743996 SD008515762GA SD008516151GA SD008516152GA SD008516153GA SD008516156GA SD008516157GA SD008516158GA SD008516159GA SD008516161GA SBG06141112GA SBG06141115GA SBG06182306GA SBG06182311GA SBG06182312GA SBG06182320GA SBG06182388GA SBG06182389GA SBG06182390GA SBG06182391GA SBG06182395GA SD007472713GA SD007524922GA SD007524925GA SD007524940GA SD008409090GA SD008409091GA SD008409092GA SD008409347GA SD008283656GA SD008283658GA PA1652468 PA1652469 PA1652470 PA1652471 PA1652472 PA1652473 PA1652474 PA1652475 PA1652476 PA1652477 PA1652478 PA1652479 PA1652480 PA1652481 PA1652482 PA1652483 PA1652484 PA1652485 PA1652486 PA1652487 SD009420653GA SD009420655GA SBG06182248GA SBG06182368GA SBG06182398GA SBG06182399GA SBG06182400GA SBG06182401GA SBG06182402GA SBG06192618GA SBG06192619GA SBG06192620GA SBG06192621GA SBG06192622GA SBG06192623GA SBG06192624GA SBG06192625GA SBG06192626GA SBG06192628GA SBG06192729GA SBG06192730GA SBG06192731GA SBG06192732GA SBG06192733GA SBG06192735GA PA1533891 PA1533892 PA1533893 PA1533894 SD007231017GA SBG06451903GA SBG06451905GA SBG06462584GA SD007482853GA SD007482935GA SD007252047GA SD007304114GA SD007304117GA SD007314543GA SD007314544GA SD007314545GA SD007314547GA SD007314548GA SD007314549GA SD007335442GA SD007356142GA SD007356143GA SD007356144GA SD007356146GA SD007356147GA SD007356148GA SD007356149GA SD007356150GA SD007356151GA SD007356152GA SD007356153GA SD007356154GA SD007356155GA SD007356156GA SD007356157GA SD007356158GA SD007356159GA SD007356160GA SD007356161GA SD007356162GA SD007356163GA SD007356164GA SD007356165GA SD007356166GA SD007356167GA SD007356168GA SD007356169GA SD007356170GA SD007356171GA SD007356172GA SD007356173GA SD007356176GA SD007356177GA SD007356178GA SD007356179GA SD007356181GA SD007356182GA SD007356183GA SD007356184GA SD007356185GA SD007356186GA SD007356187GA SD007356188GA SD007356189GA SD007356190GA SD007356191GA SD007356192GA SD007356193GA SD007356194GA SD007356195GA SD007356196GA SD007356197GA SD007356198GA SD007356199GA SD007356200GA SD007356201GA SD007356202GA SD007356203GA SD007356204GA SD007356205GA SD007356226GA SD007356227GA SD007356228GA SD007356229GA SD007356230GA SD007356231GA SD007356232GA SD007356233GA SD007356234GA SD007356235GA SD008138381GA SD009327111GA SD009327112GA SD009358142GA SD009358143GA SD009358144GA SD009358146GA SBG06515495GA SBG06525996GA SBG06525998GA SBG06526001GA SBG06526002GA SBG06526003GA SD008209728GA SD008209739GA SD008219944GA SD008219946GA SD008219947GA SD008219948GA SD008219949GA SD008219950GA PA1332211 PA1332212 PA1332213 PA1332214 PA1332215 PA1332216 PA1332217 PA1332218 PA1332219 PA1332220 PA1332221 PA1332222 PA1332223 PA1332224 PA1332225 PA1332226 PA1332227 PA1332228 PA1332229 PA1394012 PA1394013 PA1394014 PA1394015 PA1394016 PA1394017 PA1394018 PA1394019 PA1394020 PA1394021 PA1394022 PA1394023 PA1394024 PA1394025 PA1394026 PA1394027 PA1394028 PA1394029 PA1394030 PA1651310 PA1673733 SBG05442701GA SBG05453014GA SBG05453015GA SBG05453016GA SBG05453017GA SBG05453018GA SBG05453019GA SBG05453021GA SBG05453022GA SBG05453023GA SBG05453024GA SBG05453025GA SBG05453026GA SBG05453027GA SBG05453028GA SBG05453029GA SBG05453030GA SBG05453031GA SBG05453032GA SBG06141124GA SBG06141125GA SBG06182244GA SBG06182253GA SBG06182254GA SBG06182295GA SBG06182363GA SBG06182365GA SBG06182369GA SD008419356GA SD008429835GA SD009233953GA SD009244337GA PA1347587 PA1347588 PA1347589 PA1347590 PA1347591 PA1347592 PA1347593 PA1347594 PA1347723 PA1347724 PA1347725 PA1347726 PA1347727 PA1347728 PA1347729 PA1347730 PA1347731 PA1347732 PA1347733 SD007252341GA SD007293596GA SD007293597GA SD007293598GA SD007293599GA SD007293600GA SBG06182303GA SBG06182307GA SBG06182309GA SBG06182310GA SBG06182315GA SBG06182316GA SBG06182317GA SBG06182318GA SBG06182319GA SBG06388156GA SBG06388158GA SBG07098528GA SBG07121289GA SBG07121292GA SD007439674GA SD007504083GA SD010292958GA SD010292959GA SD010292960GA SD010292961GA SD010292962GA SD008494541GR SD009037445GR PA1348735 PA1348736 PA1348737 PA1348738 PA1348739 PA1348740 PA1348741 PA1348742 PA1348743 PA1348744 PA1348745 PA1348746 PA1348747 PA1348748 PA1348749 PA1348750 PA1348751 PA1348752 PA1348753 PA1348754 PA1348755 PA1348756 PA1348757 PA1348758 PA1348759 PA1348760 PA1348761 PA1348762 PA1348763 PA1348764 PA1348765 PA1348766 SBG06243933GA SBG06243936GA SBG06388155GA SBG07088105GA SBG07098586GA SBG07098590GA SD010313855GA SD010314163GA SD010314266GA SD010324443GA SD009234195GA SD009265397GA SD009265400GA SD009441387GA SD010138431GA SD010138450GA SD010189716GA SD010189717GA SBG06316028GA SBG06316029GA SBG06316030GA SBG06316031GA SBG06316033GA SBG06316034GA SBG06316035GA SBG06316165GA SBG06377802GA SBG06377840GA SBG06388074GA SBG06388075GA SBG06388076GA SBG06398527GA SBG06398529GA SBG06398530GA SBG06398531GA SBG06398532GA SBG06398533GA SBG06398534GA SBG06398535GA SBG06398536GA SBG06398537GA SBG06398538GA SBG06398833GA SBG07111133GA SBG07121298GA SBG07142117GA SBG07142118GA SBG07193988GA SBG07193989GA SBG07193990GA SBG07193991GA SBG07193992GA SBG07193993GA SBG07193994GA SBG07193995GA SD007231018GA SD007252024GA SD007252031GA SD007252032GA SD007252137GA SD007252140GA SD007282993GA SD007282995GA SD007283022GA SD007283023GA SD007335444GA SD007346136GA SD007346138GA SD007346140GA SD007366629GA SD007397759GA SD007408254GA SD007408255GA SD007408256GA SD007408257GA SD007408258GA SD007408259GA SD007408260GA SD007408264GA SD007408265GA SD007408266GA SD007408267GA SD007408268GA SD007408269GA SD007408270GA SD007514598GA SD007514599GA SD007514600GA SD007514601GA SD007514605GA SD008388030GA SD008388280GA SD008388281GA SD008388282GA SD008388283GA SD008388284GA SD008388288GA SD008388289GA SD008494520GA SD008494532GA SD010189647GA SD010189648GA SD010189658GA SD010189659GA SD010189670GA SD010366297GA SD010366299GA SD010376703GA SD010376975GA SD010377037GA SD010377047GA SD010387328GA SD010387329GA SD010387330GA SD010387379GA SD010148544GA SD010148546GA SD010148550GA SD010313856GA SD010313875GA SD010313886GA SD010314256GA SD010376618GA SD010376636GA SD010376637GA PA1925447 PA1925448 PA1742296 PA1742297 PA1742298 PA1742299 PA1742300 PA1770097 PA1770101 SD009099239GA SD009099240GA SD009099243GA SD009099247GA SD007493731GA SD007514603GA SD007377157GR SD010189715GA SD010262008GA PA1651645 PA1651646 PA1743220 PA1743222 PA1597948 PA1597949 PA1651302 PA1651303 PA1651304 PA1651305 PA1651621 PA1651624 PA1651638 PA1651639 PA1741937 PA1741938 PA1741939 PA1741940 PA1741941 PA1741942 PA1741943 PA1741944 SD008505229GA SD008505231GA SD009121152GA SD009121153GA SD009121154GA SD009121155GA SD009121156GA SD009121161GA SD009121167GA SD009121168GA SD009121172GA SD009121173GA SD009234015GA SD010376660GA SD010376813GA SBG07098519GA SBG07098521GA SBG07098522GA SBG07098524GA SBG07098525GA SBG07098593GA SBG07098595GA PA1927579 PA1927821 PA1933049 SBG07098523GA SBG07121290GA SBG07121293GA SBG06141090GA SBG06141091GA SD009400084GR SD010127704GA SD010127705GA SD010127706GA SD010127709GA PA1673700 PA1673703 PA1673705 PA1673706 PA1673707 PA1673709 PA1673717 SD008242263GA PA1933050 SBG06482961GA SBG06482963GA SBG06482964GA SBG06482966GA SD008211007GA PA1933399 PA1933400 SD008066782GA SD008076849GA SD008076861GA SD008076863GA SD010148684GA SBG06213122GA SBG06213124GA SBG06213125GA SBG06213126GA SBG06213128GA SBG06213133GA SBG06213188GA SBG06213190GA SBG06213193GA SBG06213195GA SBG06213196GA SBG06213197GA SBG06213198GA SBG06213200GA PA1499309 PA1499310 PA1499311 PA1499312 PA1499313 PA1499314 PA1499315 PA1499316 PA1499317 PA1499318 PA1499319 PA1499320 PA1499321 PA1499322 PA1499324 PA1499384 PA1499385 PA1741582 PA1743189 PA1743190 PA1743191 PA1743248 PA1743249 SD008377931GA SD008388032GA SD008388285GA SD008388287GA SD009378738GA SD009399638GA SD009399641GA SD009399643GA SD009399644GA SD009399645GA SD009409709GA SD009409710GA SD009409724GA SD009409725GA SD008219945GA SD008263088GA SD008283919GA SD008325386GA SD008346356GA SD008346357GA SD008346358GA SD008346359GA SD008346360GA SD008346361GA SD008346362GA SD008346363GA SD008346364GA SD008346365GA SD008346367GA SD008346368GA SD008346369GA SD008346370GA SBG07142198GA SBG07142201GA SD007346046GA SD007397999GA SD007398140GA SD007398143GA SD007419043GA SD007439671GA SD007439780GA SD007439783GA SD007439784GA SD007439785GA SD007439786GA SD007439787GA SD007439905GA SD007439906GA SD007439907GA SD007439908GA SD010376801GA SD010376803GA SD010376804GA PA1709046 PA1709047 PA1709048 PA1709049 PA1656035 PA1656036 PA1656037 PA1656038 PA1656039 PA1656040 PA1656041 PA1656042 PA1656043 PA1656044 PA1656045 PA1656046 PA1656047 PA1656048 PA1656049 PA1656050 PA1656051 PA1656052 PA1656053 PA1656054 PA1656055 PA1656056 PA1656057 PA1656058 PA1656059 PA1656060 PA1656061 PA1656062 PA1656063 PA1656064 SD008209736GA SD008388286GA SD008442034GA SD008442093GA SD008452531GA SD009399637GA SD009441094GA SD009441095GA SD009524149GA SD009524150GA SD009524151GA SD009524152GA SD009524153GA SD009524155GA SD009524221GA SD009524222GA SD010127811GA SD010127813GA PA1348062 PA1348067 PA1348068 PA1348069 PA1348072 PA1348073 PA1348074 PA1348075 SBG06504316GA SD007472719GA SD007303898GA SD007439777GA SD007439862GA SD007439873GA SD007451701GA SD007451703GA SD007452003GA SD007452008GA SD007452021GA SD007524920GA SD007524921GA SD007524923GA SD007524924GA SD007524926GA SD007524927GA SD007524928GA SD007524929GA SD007524930GA SD007524931GA SD007524932GA SD007524935GA SD007524936GA SD007524937GA SD007524938GA SD007524939GA SD007524941GA SD007524942GA SD007524943GA SD007524944GA SD007524945GA PA1742055 PA1742056 PA1742057 PA1742058 PA1742059 PA1742060 PA1742061 PA1742062 PA1742063 PA1742064 PA1742065 PA1742066 PA1742067 PA1742068 PA1742069 PA1742070 PA1743024 PA1743025 PA1743029 PA1743034 PA1932303 PA1932305 PA1932306 PA1932307 PA1932308 PA1932310 PA1932313 PA1932314 PA1932315 PA1932316 PA1932320 PA1932321 PA1932322 PA1654565 PA1741667 PA1741668 PA1741669 PA1741670 PA1741671 PA1741672 PA1741673 PA1741674 PA1741675 PA1741676 PA1741677 PA1741678 PA1741679 PA1741680 PA1741681 PA1741682 PA1741683 PA1741684 PA1742243 PA1742244 PA1742245 PA1742246 PA1742247 PA1742248 PA1742249 PA1742250 PA1742251 PA1742252 PA1742253 PA1742254 PA1742255 PA1742256 PA1742257 PA1742258 PA1742259 PA1742260 PA1742261 PA1742262
PA1742263 PA1742264 PA1742265 PA1742266 PA1742267 PA1742268 PA1742270 PA1742271 PA1742272 PA1743074 PA1743078 PA1743079 PA1743080 PA1743085 PA1743087 PA1743089 PA1743192 PA1743193 PA1743246 PA1743247 SBG06141126GA SBG06141127GA SBG06141128GA SBG06141129GA SBG06141130GA SBG06182304GA SBG06182305GA SBG06182308GA SBG06182313GA SBG06182314GA SBG06182321GA SD007251983GA SD007252057GA SD007252058GA SBG06398540GA SBG06398541GA SBG06398542GA SBG06398584GA SBG06398586GA SBG06398608GA SBG06398609GA SBG06398610GA SBG06398611GA SBG06398612GA SBG06398613GA SBG06398624GA SBG06398625GA SBG06398831GA SBG06429547GA SD008077086GA SD009316912GA SD010127798GA SD008283655GA SD008283659GA SD008283661GA SD008283662GA SBG06213092GR SBG06213093GR SD007356141GA SD007356145GA SD007356174GA SD007356175GA SD007356180GA SD007366727GA SD007366728GA SD007366731GA SD007366732GA SD007366733GA SD007366734GA SD008494954GA SD009244340GA SD009420531GA SD010148712GA SD010303438GA SD009244336GA SD009244338GA SD009244341GA SD009244342GA SD009244344GA SD009244345GA SD009244824GA SD009244828GA SD010127659GA SD010127734GA SD010303450GA SD010387302GA PA1654569 PA1654570 PA1654571 PA1654572 PA1654573 PA1654574 PA1654575 PA1654576 PA1654577 PA1654578 PA1654579 PA1654580 PA1654581 PA1654582 PA1654583 PA1742481 PA1742482 PA1742483 PA1742484 PA1742485 PA1742486 PA1742488 PA1742489 PA1742490 PA1742494 PA1742496 PA1742498 PA1742502 PA1742503 PA1742510 PA1742511 PA1742512 PA1742515 PA1742517 PA1742518 SD008442035GA SD008442086GA SD008442088GA SD008442089GA SD008442090GA SD008442091GA SD008442094GA SD008452532GA SD008452534GA PA1743978 PA1743979 PA1743980 PA1743982 PA1743984 PA1743985 PA1743991 PA1743992 SBG06504874GA SBG06526116GA SBG06526360GA SBG06526361GA SBG06526362GA SBG07037073GA SBG07037074GA SBG07037075GA SBG07037076GA SBG07037077GA SD007493739GA SD008283918GA SD008283920GA SD008283921GA SD008283922GA SD008283923GA SD008283924GA SD008283926GA SD008283927GA SD008294314GA SD008294315GA SD008294316GA SD008294317GA SD008294318GA SD008294320GA SD008294321GA SD008294322GA SD008294324GA SD008294326GA SD008304537GA SD008304538GA SD008304540GA SD008304546GA SD008304547GA SD009244725GA SD009296285GA SD009296286GA SD009296287GA SD009296288GA SD009296290GA SD009296291GA SD009296292GA SD009296293GA SD009296294GA SD009316920GA SD009316923GA SD009317058GA SD009317059GA SD009317066GA SD009317072GA SBG06398585GA PA1741580 PA1743026 PA1743028 PA1743030 SD009524158GA SD009524165GA SD009524175GA SD009524177GA SD009524180GA SBG06451931GA SBG06451932GA SBG06451933GA SD007439782GA SD010054457GA SD010074815GA SD010095481GA SD010095482GA SD010095484GA SD010095487GA SD010095488GA SD010127700GA SD010127785GA SD010127816GA SD010127817GA SD009316905GA SD009316908GA SD009358129GA SD009358130GA SD009358131GA SD009358132GA SD009358133GA SD009358134GA SD009358136GA SD009358137GA SD009358138GA SD009358139GA SD009358140GA SD009399540GA SD009399541GA SD009399542GA SD009399543GA SD009399544GA SD009399548GA SD009409781GA SD009409782GA SD009409783GA PA1530541 PA1530558 PA1531879 PA1531880 PA1530554 PA1530555 PA1530557 SD010127637GA SD010127638GA SD010127639GA SD010127640GA PA1509241 PA1509243 PA1530553 PA1530556 SD007304110GA SD007304111GA SD007304113GA SD007304118GA SD007304119GA SD007314541GA SD007314542GA SD007314550GA SD010313826GA SD010313828GA SD010313878GA SD010314032GA SD010314259GA SD010314262GA SD010314360GA SD010324435GA PA1597941 PA1597942 PA1597943 PA1597944 PA1597945 PA1597946 PA1597947 PA1607391 PA1607393 PA1607394 PA1607395 PA1607396 SD008077083GA SD008077085GA SD008077089GA SD008138373GA SD008138374GA SD008138386GA SD008148501GA SD008148507GA SD008148510GA SD008189138GA SD008189139GA SD008209648GA SD008209649GA SD008209650GA SD008209651GA SD008209652GA SD008209707GA SD008209708GA SD008209709GA SD008209710GA SD008209711GA SD008209712GA SD008209713GA SD008209714GA SD008209715GA SD008209716GA SD008209717GA SD008209718GA SD008209719GA SD008209720GA SD008209721GA SD008209722GA SD008209723GA SD008209724GA SD008209725GA SD008209726GA SD008209727GA SD008209730GA SD008209733GA SD008219791GA SD008219792GA SD008219795GA SD008219797GA SD008219798GA SD008219799GA SD008219801GA SD008219802GA SD008219805GA SD008242114GA SD008242118GA SD008242119GA SD008242200GA SD008263087GA SD008263089GA PA1668027 PA1668028 PA1668029 SD008168771GA SD010127779GA SD007482897GA SD007493582GA SD007524842GA SD007524843GA SD007524845GA SD007524847GA SD007524852GA SD007524857GA SD007524858GA SD007524859GA SD007524860GA SD007524861GA SD007524864GA SD010148729GA SD010148737GA SD010148738GA SD010148741GA SD007524848GA SD007524849GA SD007524851GA SD007524853GA SD007524862GA SD007524863GA SD007439618GA SD007514613GA SD007514628GA SD007514629GA SD007514669GA SD007524751GA SD008045895GA SD008046130GA SD008077129GR SBG06493679GA SBG06493681GA SBG06398832GA SBG06419410GA SBG06451442GA SD007408271GA SD009316704GA SD010127651GA SD010272450GA SD010127782GA SD010127783GA SD010127784GA SD010127788GA PA1743225 PA1927565 SD009420529GA SD009420530GA SD009420532GA SD009420533GA SD009420654GA SD010303444GA SD009244339GA SD009244343GA SD009244821GA SD009244822GA SD009244823GA SD009244825GA SD009244826GA SD009244827GA SD009316888GA SD010313838GA SD010314365GA SD010324432GA SD010324436GA SD010324439GA SD010324441GA SD010324442GA SD010324444GA SBG07121334GA SBG07131949GA SBG07131952GA SBG07152561GA SBG07152564GA SD007493723GA SD007493742GA SD007514597GA SD007514656GA SD008045863GA SD008045864GA SD008045986GA SD008045987GA SD008077145GA SD008209706GA SD008209734GA SD008209738GA SD008211006GA SD008211008GA SD008211010GA SD008211084GA SD008211088GA SD008211089GA SD008219793GA SD008219794GA SD008219796GA SD008219800GA SD008219803GA SD008219804GA SD008219937GA SD008221091GA SD008231507GA SD008231508GA SD008231509GA SD008231511GA SD008231660GA SD008231661GA SD008231662GA SD008231664GA SD008388429GA SD008388430GA SD008388431GA SD008388432GA SD010148548GA SD010148661GA SD010148670GA SD010148678GA SD008442087GA SD008452530GA SD008452533GA SD008473360GA PA1741586 PA1743040 PA1743041 PA1744732 PA1650366 PA1650367 PA1650368 PA1650369 PA1650371 SD007190109GA SD007220927GA SD007220928GA SD007220929GA SD007220930GA SD007220931GA PA1932304 PA1933175 PA1933177 PA1933178 PA1933179 PA1933180 PA1933181 PA1933182 PA1933183 PA1933184 PA1933185 PA1933186 PA1933187 PA1933188 PA1933189 PA1933190 PA1933191 PA1933192 PA1933193 PA1933194 PA1933195 PA1933196 PA1933197 PA1933198 PA1933199 PA1933200 PA1933201 PA1933202 PA1933203 PA1933204 SD010127664GA SD010303431GA SD010303433GA SD010303434GA SD010303435GA SD010303436GA SD010303437GA SD010303439GA SD010303441GA SD010303442GA SD010303443GA SD010303445GA SD010303446GA SD010303448GA SD010303449GA SD010303451GA SD007398142GA SD007419044GA SD007419046GA SD007419047GA SD007419048GA SD007419049GA SD007419050GA SD007419051GA SD008377521GA SD008505227GA SD009119622GA SD009119670GA SD009119672GA SD009121158GA SD009121169GA SD009121170GA SD009121174GA SD009129825GA SD009151789GA SD009151790GA SD009151791GA SD009151792GA PA1653262 PA1653263 PA1653265 PA1653266 PA1654552 PA1654553 PA1654554 PA1654557 PA1654560 PA1654561 PA1654562 PA1654566 PA1743022 PA1743023 PA1743027 PA1743031 PA1743032 PA1743033 SD009524159GA SD010313812GA SD010313813GA SD010313814GA SD010313815GA SD010313816GA SD010313817GA SD010313818GA SD010313819GA SD010313820GA SD010313821GA SD010313822GA SD010313823GA SD010313824GA SD010313825GA SD010313827GA SD010313829GA SD010313830GA SD010313833GA SD010313834GA SD010313835GA SD010313841GA SD010313842GA SD010313843GA SD010313847GA SD010313849GA SD010313853GA SD010313857GA SD010313858GA SD010313859GA SD010313860GA SD010313861GA SD010313864GA SD010313865GA SD010313868GA SD010313869GA SD010313870GA SD010313873GA SD010313874GA SD010313877GA SD010313879GA SD010313880GA SD010313881GA SD010313882GA SD010313883GA SD010313884GA SD010313885GA SD010314022GA SD010314023GA SD010314024GA SD010314025GA SD010314026GA SD010314027GA SD010314028GA SD010314029GA SD010314031GA SD010314033GA SD010314034GA SD010314036GA SD010314037GA SD010314038GA SD010314039GA SD010314040GA SD010314041GA SD010314042GA SD010314043GA SD010314044GA SD010314045GA SD010314160GA SD010314162GA SD010314165GA SD010314257GA SD010314258GA SD010314261GA SD010314263GA SD010314264GA SD010314265GA SD010314267GA SD008377502GA SD008377503GA SD008377504GA SD008377505GA SD008377506GA SD008377507GA SD008377508GA SD008377509GA SD008377510GA SD008377511GA SD008377512GA SD008377513GA SD008377514GA SD008377515GA SD008377517GA SD008377518GA SD008377519GA SD008377520GA SD008377522GA SD008377523GA SD008377524GA SD008377525GA SD008377526GA SD008377637GA SD008377638GA SD008377639GA SD008377640GA SD008377641GA SD008377642GA SD008377643GA SD008377644GA SD008377645GA SD008377646GA SD008377647GA SD008377648GA SD008377649GA SD008377650GA SD008377651GA SD008377652GA SD008377653GA SD008377654GA SD008377655GA SBG06419480GA SBG06429548GA SD009223892GA SD009368556GA SD009368557GA SD009368558GA SD009368559GA SD009368561GA SD009368562GA SD009368565GA SD009368636GA SD009368637GA SD009368641GA SD009378914GA SD009388967GA SD009388968GA SD009502885GA SD009502886GA SD009502887GA SD008188953GA SD008188955GA SD008188956GA SD008188957GA SD008189078GA SD008189080GA SD008189081GA SD008189082GA SD008189083GA SD008189085GA SD008189086GA SD008189142GA SD010127699GA SD010127701GA SD010127795GA SD010127801GA SD008463249GA SD008463251GA SD008473555GA SD008473556GA SD008473557GA SD009399545GA SD009399546GA SBG06367409GA SBG06367410GA SBG06367411GA PA1932285 PA1932287 PA1932294 PA1932297 PA1932299 PA1345854 PA1345856 PA1345857 PA1345858 PA1345860 PA1345862 PA1345867 PA1345868 PA1345875 PA1345877 PA1345878 PA1345879 PA1345880 PA1345881 PA1345882 PA1345883 PA1345885 PA1345886 PA1345888 PA1345889 PA1345890 PA1345891 PA1345892 PA1345893 PA1345894 PA1345895 PA1345896 PA1345897 PA1345898 PA1345899 PA1345900 PA1347026 PA1347027 PA1347028 PA1347029 PA1347030 PA1347031 PA1347032 PA1347033 PA1347034 PA1347035 PA1347036 PA1347037 PA1347038 PA1347039 PA1347040 PA1347041 PA1741585 PA1743066 PA1743067 PA1743069 PA1743070 PA1743075 PA1743076 SD007220557GA SD007220558GA SD007220559GA SD007220561GA SD007356307GA SD007419033GA SD007419034GA SD007439472GA SD007439474GA SD008494677GA SD009316790GA SD009368571GA SD009368572GA SD009368574GA SD009368632GA SD009399476GA SD010074909GA SD010138305GA SD010138306GA SD010138310GA SD010138311GA SD010138317GA SD010138319GA SD010138411GA SD010148727GA SD010148730GA SD010148732GA SD010148735GA SD010148739GA SD010158821GA SD010158823GA SD010158824GA SD008409345GA SD008429797GA SD007252034GA SD007252038GA SD007252050GA SD007252076GA SD007252330GA SD007252331GA SD007252332GA SD007252333GA SD007252335GA SD007252336GA SD007252337GA SD007252338GA SD007252340GA SD007252342GA SD007252343GA SD008066608GA SD009378915GA SD009388958GA SD009388960GA SD009388961GA SD009388962GA SD009388966GA SD007241673GA SD007283244GA SD007283245GA SD007283246GA SD007283247GA SD007283248GA SD007293412GA SD007293413GA SD007293445GA SD007293446GA SD007293447GA SD007293448GA SD007293449GA SD007293450GA SD007293451GA SD007293452GA SD007293454GA PA1483771 PA1483772 PA1483773 PA1483774 PA1483775 PA1483776 PA1483777 PA1483778 PA1483779 PA1483780 SBG06483110GA SBG06483160GA SBG06483162GA SBG06483260GA SBG06483264GA SBG06483265GA SD007220842GA SD007220844GA SD007220845GA SD007220846GA SD007220847GA SD007220848GA SD007220849GA SD007220850GA SD007220851GA SD007231019GA PA1655767 PA1655768 PA1655769 PA1655770 PA1655771 PA1655776 PA1655778 PA1668030 SD007366662GA SD007377154GA SD007377155GA SD007377160GA SD007472731GA SD007493498GA SD007493499GA SD007493500GA SD007493501GA SD007493502GA SD007493503GA SD007493504GA SD007493505GA SD007493583GA SD007493584GA SD007493586GA SD010127677GA SD010127678GA SD010127679GA SD010127680GA SD010127682GA SD010127684GA SD010127686GA SD010127687GA SD008294313GA SD008304549GA SD008325632GA SD008325633GA SD008325634GA SD008325635GA SD008325636GA SD008325668GA SD008325670GA SD008325671GA SD008336034GA SBG06462148GA SBG06462149GA SBG06462150GA SBG06462151GA SBG06483106GA SBG06483107GA SBG06483108GA SBG06483109GA SBG06483111GA SBG06483261GA SD009119624GA SD009244561GA SD009255009GA SD009316792GA SD009316793GA SD009316795GA SD009316797GA SD009316798GA SD009316800GA SD009316802GA SD009316803GA SD009316804GA SD009316805GA SD009316806GA SD009316807GA SD009316808GA SD009316809GA SD009316810GA SBG06336383GA SBG06357109GA SBG06357221GA SBG06357222GA SBG06398539GA SBG06398626GA SBG06398627GA SBG06398628GA SD010127698GA SD010127781GA SD010127786GA SD010127787GA SD010127793GA SD010127796GA SD010127800GA SD010189714GA SD008494957GA SD008505375GA SD008494536GA SD008494539GA SD008494542GA SD008494543GA SD008494544GA SD008494545GA SD008494546GA SD008494547GA SD008494548GA SD008494549GA SD008494672GA SD008494823GA SD008494824GA SD008494825GA SD008494826GA SD008494827GA SD007493730GA SD008035670GA SD008035671GA SD008035673GA SD008046010GA SD008046011GA SD008046013GA SD008056366GA SD008056568GA SD008056569GA SD008056570GA SD008056571GA SD008076858GA SD008076867GA SD008076869GA SD007283070GR SD010313954GA SD010313956GA SD010313957GA SD010313958GA SD010313959GA SD010313960GA SD010313961GA SD010313962GA SD010313963GA SD010313965GA SD010313966GA SD010313967GA SD010313968GA SD010313969GA SD010313970GA SD010314157GA SD010314158GA PA1364460 PA1482447 SBG06439720GA SBG06439723GA SBG06439724GA SBG06439727GA SD007472726GA SD008283657GA SD009492663GA SD009492665GA PA1763797 PA1763798 SD008505230GA SD009121157GA SD009121160GA SD009121176GA SD008077158GA SD008242261GA SD008429836GA SD008429837GA SD008429838GA SD008429839GA SD008429840GA SD008429842GA PA1742269 PA1742302 PA1770096 PA1770098 SD008473916GA SD008526537GA SD009026871GA SD009026872GA SD009027108GA SD009121200GA SD009121201GA SD009121202GA SD009121203GA PA1277711 PA1277713 PA1277714 PA1277715 PA1277716 PA1277717 PA1279845 PA1279846 PA1279847 PA1279848 PA1279849 PA1279850 SD008148493GA SD008148494GA SD009027112GA SD009119714GA SBG06482968GA SBG06505159GA SBG06505160GA SBG06505161GA SD008314963GA PA1345911 PA1345935 PA1345954 PA1364459 SBG06182362GA SBG06182364GA SBG06182367GA SBG06336590GA SBG06377795GA SBG06377796GA SBG06377797GA SBG06377798GA SBG06377799GA SBG06388010GA SBG06388011GA SBG06388012GA SD007293414GA SD007293453GA SD007335404GA SD007387386GA SD007472542GA SD007472543GA SD007472544GA SD007472545GA SD007472546GA SD007472549GA SD007472550GA SD009451637GA SD009451639GA SD009492690GA SD009492691GA SD009492692GA SD009492693GA SD009451653GA SD010460297GA SD010460298GA SD010460299GA SD010460302GA SD010460303GA SD010460304GA SD010460305GA SD010470398GA SD010470399GA PA1463570 SD009513720GA SD009513721GA PA1345346 PA1345347 SD009389306GA SD008294177GA SD008294179GA SBG06462582GA SBG07087864GA SBG07087867GA SBG07087868GA SD008325405GA SD008409093GA SD008409344GA SD008419662GA SD008473361GA SD008473841GA SD008473843GA SD008473844GA SD008473845GA SD008494729GA SD008494730GA SD008494731GA SD008494732GA SD008494943GA SD008494944GA SD008494946GA SD008494947GA SD008494948GA SD008494949GA SD008494950GA SD008494951GA SD008494952GA SD008494953GA SD008494956GA SD008505371GA SD008505372GA SD008505373GA SD008505374GA SD009027109GA SD009037441GA SD009037449GA SD009037450GA SD009037464GA SD009037468GA SD009368564GA SD009368566GA SD009368567GA SD009368568GA SD009368569GA SD009368639GA PA1668170 PA1668171 PA1743068 PA1743071 PA1743072 PA1743073 SD007514361GA SD007514364GA SD007514382GA SD008014965GA SD008046109GA SD008046111GA SD008046113GA SD008046114GA SD008046115GA SD008046116GA SD008046118GA SD010127644GA SD010366286GA SD007335424GA PA1408586 PA1408587 PA1408588 PA1408589 PA1408590 PA1408591 PA1408592 PA1484039 PA1484040 PA1484041 PA1484042 PA1484043 PA1484044 PA1484045 PA1743981 SD009368633GA SD009244735GA SD009461833GA SBG07193841GA SBG07193848GA SD007304116GA SD007325111GA SBG06493799GA SD007514367GA SD007514370GA SD007304115GA SD007441250GA SD007514372GA SD009047740GA SD009047745GA SD009047746GA SD009047747GA SD009047749GA SD009047750GA SD009047751GA SD009047752GA SD009047753GA SD008419470GA SD008494670GA SD008494673GA SD008494674GA SD007482984GA SD007482985GA PA1925446 PA1925450 PA1925451 PA1925458 SD008442062GA SD008473842GA SBG07088074GA SBG07119475GA SBG07119797GA SBG07121392GA PA1463556 PA1463557 PA1463558 PA1463559 PA1463560 PA1463561 PA1463562 PA1463563 PA1463564 PA1463565 PA1463566 PA1463567 PA1463568 PA1463569 PA1463571 PA1463572 PA1463573 PA1463574 PA1463575 PA1463576 PA1463577 PA1463578 PA1463579 PA1463580 PA1463581 PA1463582 PA1463583 PA1463584 PA1463585 PA1463586 PA1463587 PA1463588 PA1463589 PA1463590 PA1463591 PA1463592 PA1463593 PA1463594 PA1463595 SBG07108973GA SBG07111117GA SBG07111125GA SBG07111126GA SBG07111127GA SBG07111128GA SBG07111130GA SBG07111131GA SBG07111132GA SBG07111135GA SBG07111136GA SBG07111137GA SBG07111138GA SBG07111139GA SBG07119787GA SBG07119789GA SBG07119790GA SBG07119791GA SBG07119800GA SBG07121390GA SBG07121403GA SBG07121404GA SBG07121406GA SBG07121407GA SBG07121408GA SBG07121427GA SD009099244GA SD009099250GA SD009099251GA SD009099254GA SD009099255GA SD009099256GA SD008014963GA SD008014964GA SD008025357GA SD008388085GA SD008388087GA SD008419354GA SD008419355GA SD008419359GA SD008419360GA SD009099236GA SD009119496GA SD009119497GA SD009129952GA SD009348061GA SBG06254219GA SBG06451904GA SBG06462583GA SBG06504892GA SD007293411GA SD007293415GA SD008138375GA PA1743226 PA1743227 PA1743228 PA1743229 PA1743230 PA1743231 PA1743232 PA1743233 PA1743234 PA1743235 PA1743321 PA1743322 PA1743323 PA1743324 PA1743325 PA1743326 PA1743327 PA1743328 PA1743329 PA1743330 PA1743331 PA1743332 SBG06305689GA SBG06483198GA SBG07088142GA SD010272452GA SD010449325GA SD010449326GA SD010449327GA SD010481139GA SD007514384GA SD009513722GA SD010127746GA SD010127764GA SD010262001GA SD010272403GA SD010272412GA SD010376808GA SD008505532GA SD009358145GA SD009368642GA SD009368643GA SD009368644GA SD009368645GA SD009368646GA SD009368647GA SD009378739GA SBG06483020GA SBG06493628GA SBG06493632GA SBG06493797GA SBG06493801GA SBG06493802GA SBG06504317GA SBG06504319GA SBG06504320GA SBG07026986GA SD009524223GA PA1742275 PA1742277 PA1742293 PA1652521 PA1655773 PA1655774 PA1655775 SD008325408GA PA1742288 PA1742289 SD008409348GA SD008462854GA SD008462857GA SD008462859GA SD008462860GA SD008462863GA SD009078554GA SD009078555GA SD009099068GA SD009099246GA SD009119484GA SD009119485GA SD009119486GA SD009119487GA SD009119491GA SD009119492GA SD009119494GA SD009492739GA SD008209569GA SD008209571GA SD008209572GA PA1650697 SD009037205GA SD009037206GA SD010138444GA SD007503959GA SD007503965GA SD007503966GA SD007503969GA SD007503970GA SD007514366GA SD007514377GA SD007514379GA SD008014956GA SD008014961GA SD008025358GA SD008025360GA SD007335418GA SD007439669GA SD008462862GA SD008463067GA SD008463068GA SD008494641GA SD008494642GA SD008494675GA SD008494676GA SD008494693GA SD008494694GA SD008494695GA SD008494697GA SD008494698GA SD008494700GA SD008494702GA SD008494703GA SD008494704GA SD008494705GA SD008494707GA SD008494708GA SD008494709GA SD008494710GA SD008494711GA SD008494712GA SD008494920GA SD008494922GA SD008494924GA SD008505159GA SD008505160GA SD008505161GA SD008505162GA SD008505163GA SD007241662GA SD007241669GA SD009420542GA SD009420645GA SD009420646GA SD009420647GA SD009441245GA SD009441246GA SD009441247GA SD009441248GA SD009441249GA SD009441250GA SD009441251GA SD009441252GA SD009441255GA SD009441256GA PA1349395 PA1349396 PA1349397 PA1484891 SBG06367355GA SBG06367357GA SBG06367358GA SBG06367359GA SBG06367360GA SBG06367362GA SBG06377725GA SBG07098534GA SBG07098538GA SBG07098541GA SBG07098542GA SBG07108993GA SBG07108995GA SBG07108996GA SBG07108997GA SD007220560GA SD008045856GA SD008045858GA SD008056350GA SD008066607GA SD008066609GA SD008077113GA SD008148499GA SD008148500GA SD008209570GA SD008294358GA SD008315222GA SD010138315GA SD010138407GA SD010148731GA SD010148733GA SD010158822GA SD008045910GA SD008045911GA SD008045912GA SD008045913GA SD008046009GA SD010148672GA SD010148680GA SD010148681GA SD010148683GA PA1484890 SBG06213233GA SBG06213243GA SBG06223396GA SBG06223397GA SBG06223424GA SBG06223425GA SBG06223427GA SD009513324GA SBG06036306GA SBG06141116GA SBG06141117GA SBG06141118GA SBG06141119GA SBG06141120GA SBG06182324GA SBG06182328GA SD010127819GA SD010158819GA SD010189652GA SD010189654GA SD010189661GA SD010189664GA SD010313964GA SD010314148GA SD010314151GA SD010314153GA SD010314154GA SD010314156GA SD010314295GA SD010314296GA SD010314298GA SD010314299GA SD010314300GA SD010324790GA SD010324791GA SD010324792GA SD010324793GA SD010324794GA PA1709119 PA1709120 PA1709121 PA1709122 PA1709123 PA1709124 PA1709125 PA1709126 PA1709127 PA1709128 PA1709129 PA1709130 PA1709131 PA1709132 PA1709133 PA1709134 PA1709135 PA1709136 PA1709137 PA1927820 PA1927823 SD010324798GA SD010366215GA SD010366219GA SD010366221GA SD010366282GA SD010376705GA SD010387332GA SD010387333GA SD010429097GA SD010429099GA SD010429100GA SD010429101GA SD010429102GA SD010429103GA SD010127652GA SD010387286GA SD010387287GA SD010387288GA SBG07098533GA SBG07098535GA SBG07098536GA SBG07098537GA SBG07098539GA SBG07098540GA SBG07108994GA SBG06429509GA SBG06429634GA SD007283159GA SD007303965GA SD007303968GA SD007303969GA SD007303971GA SD008045776GA SD008045901GA SD008045903GA SD008056385GA SD009420534GA SD010127647GA SD010127692GA SD010138414GA SD010251590GA SD010251592GA SD007387337GA SD007387342GA SD007387343GA SBG06305782GA SBG06305784GA SBG06305785GA SBG06305787GA SBG06305788GA SBG06305790GA SBG06305791GA SBG06336589GA SBG06462682GA SBG06462684GA SD008494526GA SD008494534GA SD008494758GA SD008494759GA SD008494760GA SD008494761GA SD008494762GA SD008505165GA SD008505166GA SD008505167GA SD008505168GA SD008505169GA SD007429260GA SD007439469GA SD007439471GA SD007439473GA SD008045893GA SD008076870GA SD008076876GA SD008076894GA SD008076906GA SD008076907GA SD009306449GA SD009316786GA SD009316832GA SD009410157GA SD009410158GA SD009410360GA SD010324796GA SD010324797GA SD010366209GA SD010366210GA SD010366211GA SD010366213GA SD010366214GA SD010366220GA PA1651642 PA1463646 PA1463647 PA1463648 PA1463649 PA1463650 PA1463651 PA1463652 PA1463653 SBG07087881GA SBG07098266GA SBG07108934GA SBG07108935GA PA1349394 PA1481025 PA1533365
PA1655777 PA1349392 PA1349393 SD007387388GA PA1532208 PA1532209 SD007210280GA SD007210468GA SD007210469GA SD007210471GA SD007293330GA SD007293500GA SD007293502GA SD007293503GA SD007293504GA SD007293510GA SD007293571GA SD007293575GA SD007293576GA SD007293578GA SD007304031GA SD007304033GA SD007314566GA SD007314568GA SD007314569GA SD007314570GA SD007314571GA SD007314572GA SD007314577GA SD007314579GA SD007314581GA SD007314582GA SD007314583GA SD007314584GA SD007314585GA SD007314711GA SD007314712GA SD007314713GA SD007314716GA SD007314717GA SD007314718GA SD007314719GA SD007314720GA SD007314722GA SD007314723GA SD007314724GA SD007377118GA PA1463644 PA1463645 PA1534109 PA1534110 PA1534111 PA1534112 PA1534113 PA1534114 PA1534115 PA1534116 PA1534117 PA1534118 PA1534119 PA1534120 PA1534121 PA1534122 PA1763796 PA1763799 PA1483781 PA1483782 SD008325560GA SD008325561GA SD008325562GA SD008325563GA SD008325566GA SD008325574GA SD008325575GA SD008325576GA SD008325580GA PA1461949 PA1461950 PA1461951 PA1462003 SBG06483018GA SBG06483021GA SBG06483022GA SD007283132GA SD007283134GA SD007283135GA SD007283141GA SD007429081GA PA1484892 SBG06141101GA SBG06141102GA SBG06141103GA SBG06141105GA SBG06213355GA SBG06213356GA SBG07108936GA SD008304620GA SD008304621GA SD008315219GA SD008315220GA SD008315221GA SD008315223GA SD008325411GA SD008463248GA SBG06316027GA SD007241696GA SD007303963GA SD008148491GA SD008148492GA SD008148495GA SD008283847GA SD008283850GA SD008283852GA SD008283853GA SD008283855GA SD008294178GA SD008294359GA SD008304619GA SD008315150GA SD010387331GA SD010459748GA SD010459749GA SD010459750GA SD010459761GA SD007283070GA SD009441253GA PA1348070 PA1348071 PA1349673 PA1481813 PA1483689 PA1483694 SD007210512GA SD007241686GA SD007241688GA SD007241690GA SD007241692GA SBG06482956GA SBG06482957GA SBG06504506GA SBG06505155GA SBG06505157GA SBG06505158GA PA1651637 PA1651644 PA1932300 PA1932301 PA1932302 SBG06482958GA SBG06482959GA SD008315101GA SD008315102GA SD008315103GA SD008315104GA SD008315105GA SD008335937GA SD008335938GA SD008335939GA SD008335940GA SD008335941GA SD008336011GA SD008356715GA SD008356716GA SD009099224GA SD009099225GA SD009099226GA SD009099227GA SD009099229GA SD009099230GA SD009099231GA SD009099232GA SD009099233GA SD009099234GA SD009099237GA SD009099238GA SD009119791GA SD009119792GA SD009119793GA SD009119794GA SD009119795GA SD009129954GA SD009129955GA SD009129956GA SD009316703GA SD009316707GA SD009410366GA SD009420538GA SD009420679GA SD009431089GA SD009431090GA SD009431091GA SD009451635GA SD009451636GA SD009451638GA SD009451640GA SD009482513GA SD009482514GA SD009482515GA SD009482516GA SD009482517GA SD009482518GA SD009492740GA SD009502947GA SD010054459GA SD010054460GA SD010054461GA SD010095740GA SD010127689GA SD010127690GA SD010127691GA SD010127693GA SD010127694GA SD010127695GA SD010127696GA SD010127697GA SD010138412GA SD010138413GA SD010251343GA SD010262007GA SD010262163GA SD010262164GA SD010262165GA SD010262166GA SD010262167GA SD010262169GA SD010262170GA SD010262175GA SD010262176GA SD010262177GA SD010272378GA SD010272380GA SD010272381GA SD010272382GA SD010272383GA SD010272384GA SD010272385GA SD010272386GA SD010272387GA SD010272388GA SD010272390GA SD010272391GA SD010272392GA SD010272394GA SD010272395GA SD010272396GA SD010272397GA SD010272398GA SD010272399GA SD010272400GA SD010272401GA SD010272402GA SD010272404GA SD010272405GA SD010272406GA SD010272408GA SD010272409GA SD010272410GA SD010272411GA SD010272413GA SD010272414GA SD010272415GA SD010272416GA SD010272417GA SD010272439GA SD010272440GA SD010272441GA SD010272442GA SD010272443GA SD010272445GA SD010272446GA SD010272447GA SD010272448GA SD010272449GA SD010272451GA SD010292956GA SD010292957GA SD010366207GA SD010366212GA SD010366216GA SD010366279GA SD010366280GA SD010366281GA SD010366283GA SD010366284GA SD010366285GA SD010366287GA SD010376704GA SD010376706GA SD010376707GA SD010376708GA SD008148508GA SD008231510GA SD008283788GA SD008294183GA SD008066605GA SBG06026029GA SBG06056888GA SBG06141089GA SBG06161721GA SBG06192509GA SBG06192718GA SBG06213174GA SBG06213352GA SBG06213364GA SBG06223386GA SBG06223390GA SBG06223398GA SBG06223399GA SBG06223401GA SBG06223410GA SBG06233733GA SBG06233734GA SBG06233735GA SBG06233736GA SBG06233737GA SBG06233738GA SBG06233739GA SBG06233740GA SBG06243924GA SBG06243925GA SBG06243926GA SBG06243927GA SBG06243928GA SBG06243929GA SBG06243930GA SBG06243931GA SBG06243932GA SBG06243937GA SBG06284776GA SBG06284778GA SBG06284780GA SBG06284781GA SBG06284782GA SBG06305819GA SBG06305823GA SBG06315852GA SBG06315853GA SBG06315856GA SBG06315857GA SBG06315859GA SBG06315860GA SBG06315861GA SBG06315863GA SBG06315864GA SBG06315865GA SBG06315866GA SBG06315867GA SBG06315868GA SBG06315869GA SBG06315870GA SBG06315871GA SBG06315872GA SBG06315873GA SD007293444GA SD007303973GA SD007324896GA SD007335401GA SD007335402GA SD007335403GA SD007419054GA SD007419057GA SD007419058GA SD007419061GA SD007429078GA SD007429079GA SD007429080GA SD007429130GA SD007429258GA SD007429261GA SD007472547GA SD007472548GA SD007472551GA SD007482852GA SD007482854GA SD007482855GA SD007482929GA SD007482930GA SD007482931GA SD007482932GA SD007482933GA SD007482936GA SD007482937GA SD007482938GA SD007503952GA SD007503953GA SD007503954GA SD007503955GA SD007503956GA SD007503962GA SD007503963GA SD007503964GA SD007503968GA SBG07077616GA SD007325117GA SD008056564GA SD010148718GA SD010189673GA SD010189674GA SD010189675GA SD010189679GA SD009502924GA SD009513896GA SD010116607GA SD010116611GA SD010148728GA SD010148736GA SD010148742GA SD010314030GA SD010366327GA PA1668019 PA1668020 PA1668021 PA1668022 PA1668023 PA1668024 PA1668025 PA1668026 SD008066786GA SBG06505150GA SBG06505151GA SBG07121337GA SBG07152562GA SBG07152563GA SD007397764GA SD007397777GA PA1533699 PA1533700 PA1533701 PA1533702 PA1533703 SBG06504313GA SBG06098371GA SBG06098374GA SBG06129616GA SBG06129617GA SBG06151691GA SBG06151696GA SBG06151698GA SBG06151700GA SBG07193846GA SBG07193851GA PA1464534 PA1464535 SBG07077738GA SBG07088091GA SBG07098520GA SBG07098587GA SBG07098588GA SBG07098589GA SBG07098591GA SBG07098592GA SBG06223369GA SBG06223387GA SBG06223394GA SBG06223416GA SBG07173173GA SBG07173222GA SD007210287GA SD007210288GA SD007241695GA SD007303957GA SD007303959GA SD007303961GA SD007325113GA SD007325115GA SD007419042GA SD007419045GA SD007439668GA SD007439675GA SD007439676GA SD007439677GA PA1743334 PA1932151 PA1932283 PA1932284 PA1932286 PA1932288 PA1932289 PA1932290 PA1932291 PA1932292 PA1932295 PA1932296 PA1932298 PA1933173 PA1933622 SD007325120GA SD007377151GA SD008056179GA SD008076905GA SD008077110GA SD008148496GA SD008148497GA SD008148498GA SD008325388GA SD008346366GA SD008356630GA SD008356634GA SD008367240GA SD008377516GA SD008377771GA SD008377866GA SD008377868GA SD008387989GA SD008387990GA SD008387991GA SD008387993GA SD008388290GA SD008388291GA SD008388293GA SD008429794GA SD008429795GA SD008463253GA SD008473488GA SD008473489GA SD008473490GA SD008473491GA SD008473492GA SD008473493GA SD008473494GA SD008473558GA SD008473559GA SD008473560GA SD008473908GA SD008473909GA SD008473910GA SD008473911GA SD008473912GA SD008473913GA SD008473915GA SD008473917GA SD008473918GA SD008473921GA SD008473922GA SD008473923GA SD008473924GA SD008473926GA SD008473927GA SD008473928GA SD008473929GA SD008473930GA SD008473932GA SD008494522GA SD008494523GA SD008494524GA SD008494529GA SD008505335GA SD008505336GA SD008515761GA SD008515763GA SD009057947GA SD009078538GA SD009078553GA SD009078556GA SD009088951GA SD009099065GA SD009099066GA SD009099067GA SD009099242GA SD009099253GA SD009099340GA SD009099341GA SD009119397GA SD009119488GA SD009129821GA SD009129824GA SD009172302GA SD009182686GA SD009203311GA SD009203312GA SD009203313GA SD009203314GA SD009203315GA SD009203316GA SD009203317GA SD009203318GA SD009203319GA SD009203320GA SD009223788GA SD009223790GA SD009223791GA SD009223792GA SD009223793GA SD009223794GA SD009223795GA SD009223796GA SD009223797GA SD009223798GA SD009223799GA SD009223800GA SD009223801GA SD009223802GA SD009223889GA SD009223901GA SD009234022GA SD009234167GA SD009234170GA SD009244726GA SD009316791GA SD009316794GA SD009316796GA SD009316801GA SD009327110GA SD009327373GA SD009420683GA SD009441102GA SD009441254GA SD009472018GA SD009472026GA SD009472035GA SD009472110GA SD009513891GA SD010054456GA SD010064535GA SD010064536GA SD010064537GA SD010064538GA SD010064539GA SD010127716GA SD010127719GA SD010138286GA SD010138307GA SD010138309GA SD010138314GA SD010138318GA SD010138321GA SD010138322GA SD010138408GA SD010138410GA SD010138489GA SD010148543GA SD010148734GA SD010148743GA SD010158813GA SD010158816GA SD010158827GA SD010189646GA SD010189665GA SD010189667GA SD010189681GA SD010189730GA SD010189732GA SD010189733GA SD010189735GA SD010251341GA SD010251342GA SD010251344GA SD010251345GA SD010251346GA SD010251377GA SD010251378GA SD010251379GA SD010303440GA SD010376664GA SD010376810GA SD010387299GA PA1498612 PA1498613 SD007282982GA SD007282983GA SD007282984GA SD007282985GA SD007282987GA SD007325114GA SD007335412GA SD007335422GA SD007335423GA SD007314714GA SD007366649GA SD007472588GA SD007472590GA PA1464520 PA1464521 PA1464522 PA1464523 PA1464524 PA1464525 PA1464526 PA1464527 PA1464528 PA1464529 SBG07077609GA SBG07077610GA SBG07077612GA SBG07077620GA SBG07152624GA SBG07152626GA SBG07152628GA SD008056563GA SD007483098GA SD007483099GA SD007483100GA SD007483101GA SD007493713GA SD007493714GA SD007493715GA SD007493716GA SD007493717GA SD007493718GA SD007493719GA SD007493720GA SD007493721GA SD007493722GA SD007493724GA SD007493725GA SD007493726GA SD007493727GA SD007493728GA SD007493729GA SD007493732GA SD007493733GA SD007493734GA SD007493735GA SD007493736GA SD007493737GA SD007493738GA SD007493740GA SD007493741GA SD007493743GA SD007493744GA SD007493745GA SD007493746GA SD007493747GA SD007366660GA SD007377156GA SD007377162GA SD007210465GA SD008077162GA SD008077165GA SD008077169GA SD008077173GA SBG07131946GA SBG07131947GA SBG07131948GA SBG07131950GA SBG07131951GA SBG06462551GA SBG06462552GA SBG06462681GA SBG06472760GA SD007441099GA SBG06441255GA SBG06441256GA SBG06441258GA SBG06441259GA SBG06254218GA SBG06254222GA SBG06357108GA SBG06377842GA SBG06377843GA SBG07077669GA SBG07077713GA SBG06305807GA SBG06305808GA SBG06305809GA SBG06305810GA SBG06305811GA SBG06305812GA SBG06305813GA SBG06305814GA SBG06305815GA SBG06305816GA SBG06305817GA SBG06305818GA SBG06305820GA SBG06305821GA SBG06305822GA SBG06305824GA SBG06305825GA SBG06305826GA SBG06315858GA PA1607387 PA1607389 SBG07173176GA SBG07173224GA SBG07173340GA SBG07173342GA SBG07173343GA SBG07193872GA SBG07193874GA PA1481691 PA1481692 PA1481693 PA1481694 PA1481695 SBG06483024GA SBG06483025GA SBG06493626GA SBG06493627GA SBG06493629GA SBG06493630GA SBG06493631GA SBG06493633GA SBG06493634GA SBG06493635GA SBG06493636GA SBG06493637GA SBG06493638GA SBG06493639GA SBG06493640GA SBG06493803GA SBG07026820GA SBG07026821GA SBG06377841GA SBG06377844GA SBG07173360GA SBG07193870GA SBG07193873GA SBG07193875GA SBG07193876GA SBG07193877GA SBG07193878GA SBG07193879GA SBG06462550GA SBG06462680GA SBG06462683GA SBG06472761GA SBG06493798GA PA1483732 PA1483733 SBG06462553GA SBG07087865GA SBG07152560GA SBG07173174GA SBG07193871GA PA1484889 SBG06141104GA SBG06213353GA SBG06213354GA SBG06243934GA SBG06243935GA SBG06243938GA SBG06462577GA SBG06462578GA SBG06462579GA SBG06462580GA SBG06462581GA SBG06493645GA SBG06493646GA SBG06493647GA SBG06493648GA SBG06493649GA SBG06493650GA SBG06493651GA SBG06493652GA SBG06504356GA SBG06504886GA SBG06504887GA SBG06504888GA SBG06504889GA SBG06504890GA SBG06504891GA SBG07026988GA SBG07026990GA SBG07026992GA SBG07077658GA SBG07077659GA SBG07077660GA SBG07087846GA SBG07087869GA SBG07087870GA SBG07087871GA SBG07087872GA SBG07087873GA SBG07087874GA SBG07087875GA SBG07087876GA SBG07087877GA SBG07087878GA SBG07087879GA SBG07087880GA SBG07087882GA SBG07087883GA SBG07087884GA SBG07087885GA SBG07087886GA SBG07087887GA SBG07087888GA SBG07131797GA PA1742475 PA1742476 PA1742478 PA1742479 PA1742480 PA1742513 PA1742514 PA1742516 PA1742520 PA1742521 PA1742522 PA1742523 PA1927566 PA1927567 PA1927568 PA1927570 PA1927571 PA1927572 SD008336010GA SD008429986GA SD008429987GA SD008429988GA SD008429989GA SD008429990GA SD008462798GA SD008462799GA SD008462801GA SD008462803GA SD008463182GA SD008463183GA SD008463202GA SD008463203GA SD008463205GA SD008505341GA SD008515749GA SD008515750GA SD009047739GA SD009047741GA SD009047743GA SD009129950GA SD009129953GA SD009129957GA SD009129958GA SD009203566GA SD009203567GA SD009203568GA SD009203569GA SD009203571GA SD009203572GA SD009203574GA SD009244566GA SD009244567GA SD009244568GA SD009244569GA SD009244570GA SD009285818GA SD009285819GA SD009285821GA SD009285823GA SD009316885GA SD009368630GA SD009368631GA SD009368634GA SD009420535GA SD009420536GA SD009420537GA SD009420539GA SD009420543GA SD009420545GA SD009420649GA SD009420650GA SD009482477GA SD009482478GA SD009482479GA SD010054458GA SD010095738GA SD010095739GA SD010127648GA SD010127649GA SD010189680GA SD010189682GA SD010189683GA SD010189684GA PA1476389 PA1509242 PA1509244 PA1509245 PA1509246 PA1509247 PA1530552 PA1532230 PA1532231 PA1532232 PA1533366 PA1533367 PA1533368 PA1533369 PA1533370 PA1533371 PA1607385 PA1607386 PA1607388 SBG07131799GA SBG07162972GA SBG07162974GA SD007210513GA SD007241625GA SD007241626GA SD007241627GA SD007241628GA SD007241629GA SD007241630GA SD007241631GA SD007241632GA SD007241633GA SD007241634GA SD007241664GA SD007241666GA SD007241667GA SD007241668GA SD007241689GA SD007241693GA SD007283157GA SD007303966GA SD007303967GA SD007303970GA SD007303972GA SD007303974GA SD007335405GA SD007418941GA SD007418942GA SD007418943GA SD007418944GA SD007418945GA SD007419029GA SD007419030GA SD007419031GA SD007419053GA SD007419055GA SD007419056GA SD007419059GA SD007429141GA SD007429142GA SD007429143GA SD007429144GA SD007429145GA SD007429146GA SD007429147GA SD007429148GA SD007429149GA SD007429150GA SD007429259GA SD007429262GA SD007482876GA SD007482877GA SD007482878GA SD007482879GA SD007482880GA SD007482881GA SD007482882GA SD007482883GA SD007482988GA SD007503951GA SD007503957GA SD007503958GA SD007503960GA SD007503961GA SD007503967GA SD007514356GA SD007514357GA SD007514358GA SD007514359GA SD007514360GA SD007514362GA SD007514363GA SD007514365GA SD007514368GA SD007514369GA SD007514371GA SD007514373GA SD007514374GA SD007514376GA SD007514378GA SD007514380GA SD007514381GA SD007514383GA SD007514385GA SD008045770GA SD008045771GA SD008045773GA SD008045774GA SD008045775GA SD008045777GA SD008045778GA SD008045779GA SD009368635GA SBG07121345GA SBG07121348GA SD007304112GA SD007366730GA SD007387296GA SD007439498GA PA1530467 PA1530468 PA1530539 PA1530540 SD007472587GA SD007472589GA SD008077111GA SD007241663GA SD007241665GA SD007241671GA SD007241674GA PA1763714 PA1651300 PA1651306 PA1656030 SD008283848GA PA1651613 PA1651614 PA1651615 PA1651616 PA1651617 PA1651618 PA1651619 PA1651620 PA1651622 PA1651623 SD007282988GA SD008077146GA SD008077147GA SD008211081GA SD008211083GA SD008211085GA SD008211086GA SD008219936GA SD010158812GA PA1481814 PA1481815 PA1481816 PA1481817 PA1481818 SD007419027GA SD007419028GA SD007419052GA SD007419060GA SD008377864GA SD008377865GA SD008377867GA SD008076851GA SD008463204GA SD008463206GA SD008494728GA SD008494945GA SD009057904GA SD009057946GA SD009068326GA SD009119473GA SD009119481GA SD009119483GA SD009119638GA SD009119694GA SD009119697GA SD009119698GA SD009119731GA SD009121187GA SD009121193GA SD009129869GA SD009151788GA SD009172303GA SD009172304GA SD009172306GA SD009172307GA SD009172308GA SD009172309GA SD009172310GA SD009172311GA SD009172312GA SD009223708GA SD009234020GA SD009254897GA SD009254898GA SD009255206GA SD009255207GA SD009358128GA SD009358135GA SD009358141GA SD009368575GA SD009368576GA SD009378781GA SD009378782GA SD009399547GA SD009399549GA SD009410159GA SD009410160GA SD009410161GA SD008127982GA SD007304107GA SD007366631GA SD009285814GA SD009285815GA SD009285816GA SD009285820GA SD009285822GA SD010303432GA SD010387298GA SD010387300GA SD010387301GA SD010387303GA SD010387304GA SD010387305GA SD010387306GA SD007524850GA SD008077107GA SD008077108GA SD008077109GA SD008077112GA SD008077114GA SD008077115GA SD008127978GA SD008127979GA SD008127980GA SD008127981GA SD008127983GA SBG07026985GA SD008056344GA PA1653257 PA1653258 PA1653259 PA1653260 PA1653261 SD008442063GA SD007429077GA SD007482986GA SD009265396GA SD009265398GA PA1673708 PA1673711 PA1673713 PA1673723 SD008377769GA SD008377770GA SD008377772GA SD008377773GA SD008387992GA SD008388292GA SD008388294GA SD007200248GA SD007200250GA SD007241698GA SD007241699GA PA1763741 PA1763742 PA1763743 PA1763744 PA1763745 PA1763746 PA1763747 SD009410358GA SD009410361GA SD009410362GA SD007303962GA SD007325118GA SD007366621GA SD007210286GA SD007482851GA SD007482934GA SBG06504882GA SD010158793GA PA1653254 PA1653255 PA1655772 SD008076856GA SD008077079GA SD008077186GA SD007441251GA SD008138401GA SD008283781GA SD008283783GA SD008283784GA SD008283785GA SD008283786GA SD008283787GA SD008283789GA SD008283790GA PA1651307 PA1674688 PA1348061 SD010189703GA SD010376663GA PA1499323 SBG06504878GA SBG06504879GA SBG06504881GA SBG06504883GA SBG06504884GA SBG06504885GA SBG06515439GA SBG07121344GA SD010138421GA SD010138427GA SD010138441GA SD010138445GA SD010138447GA SD010148682GA SD010148687GA SD007441242GA SD007472721GA SD007472730GA SD007472732GA SD007472734GA SD007482895GA SD007482896GA SD007482898GA SD007493390GA SD009244727GA SD009244731GA SD009492664GA SD009492666GA PA1530542 PA1530543 PA1530544 PA1530545 PA1530546 PA1530547 PA1530548 PA1530549 PA1530550 PA1530551 PA1709683 PA1709684 PA1743223 PA1763716 SD009420678GA SD009420680GA SD009461825GA SD009472031GA SD009088952GA SD008367316GA SD008377656GA SD008304543GA SD008314960GA SD008336033GA SD008336035GA SD009244510GA SD009244565GA PA1348052 PA1348054 PA1348055 PA1348056 SD008045907GA SD008045981GA SD008056177GA SD008056178GA SD008056180GA SD008377804GA SD008377805GA SD008377806GA SD008377807GA SD008377808GA SD008377809GA SD008377810GA SD008377811GA SD008377812GA SD008377813GA SD008377929GA SD008377930GA SD008377933GA SD008421279GA SD010189734GA SD008219941GA SD008494541GA SD009502906GA SD009502907GA SD009502908GA SD009502909GA SD009502910GA SD009502911GA SD009502912GA SD009502913GA SD009502914GA SD009502933GA SD009502934GA SD009502938GA SD009502939GA SD009502941GA SD009513382GA SD007293455GA SD009119696GA PA1464101 PA1464102 PA1464103 PA1651299 PA1651528 PA1651726 PA1651727 PA1651728 PA1651729 PA1651730 PA1651731 PA1651732 PA1651733 PA1651869 PA1651870 PA1656029 PA1656031 PA1656032 PA1656033 PA1743063 PA1743086 PA1743342 PA1763715 SBG06182246GA SBG06182296GA SBG06192745GA SBG06192750GA SBG06203051GA SBG06483163GA SBG06483164GA SBG06493754GA SBG06493755GA SBG06493756GA SBG06493757GA SBG06493758GA SBG06504507GA SBG06504508GA SD007241660GA SD007241661GA SD007241670GA SD007241672GA SD007252056GA SD007252345GA SD007252346GA SD007252347GA SD007252348GA SD007252349GA SD007252350GA SD007252351GA SD007252352GA SD007252353GA SD007252354GA SD007282992GA SD007293548GA SD007304105GA SD007304108GA SD007304109GA SD007397769GA SD007441217GA SD007441218GA SD007441219GA SD007441220GA SD007441221GA SD007441223GA SD007441224GA SD007441225GA SD007441227GA SD007441228GA SD007441230GA SD007441231GA SD007441232GA SD007441233GA SD007441234GA SD007441236GA SD007441238GA SD007441240GA SD007441241GA SD007441247GA SD007441248GA SD007493665GA SD008045994GA SD008138393GA SD008148505GA SD008231663GA SD008283925GA SD008294182GA SD008294319GA SD008294327GA SD008304533GA SD008304534GA SD008304535GA SD008304536GA SD008304539GA SD008304542GA SD008304544GA SD008304545GA SD008304551GA SD008304552GA SD008304563GA SD008304566GA SD008304567GA SD008315107GA SD008388433GA SD008419466GA SD008419468GA SD008419469GA SD008452732GA SD008452733GA SD008494533GA SD008494535GA SD008494537GA SD008494538GA SD008494671GA SD008505337GA SD008505340GA SD009037439GA SD009037443GA SD009037457GA SD009037529GA SD009037530GA SD009047742GA SD009047744GA SD009047748GA SD009057895GA SD009057898GA SD009057925GA SD009078505GA SD009078509GA SD009078510GA SD009078511GA SD009078514GA SD009099228GA SD009099235GA SD009099245GA SD009119464GA SD009119465GA SD009119467GA SD009119468GA SD009119469GA SD009119470GA SD009119471GA SD009119472GA SD009119476GA SD009119477GA SD009119480GA SD009119482GA SD009119489GA SD009119490GA SD009119495GA SD009119498GA SD009119634GA SD009119635GA SD009119640GA SD009119641GA SD009119662GA SD009119663GA SD009119665GA SD009119666GA SD009119667GA SD009119668GA SD009119673GA SD009119674GA SD009119675GA SD009119676GA SD009119695GA SD009119732GA SD009119733GA SD009121188GA SD009121189GA SD009121191GA SD009121192GA SD009121194GA SD009121197GA SD009121198GA SD009172194GA SD009172195GA SD009172196GA SD009223893GA SD009233946GA SD009233947GA SD009233948GA SD009233949GA SD009233950GA SD009233951GA SD009233952GA SD009233954GA SD009285825GA SD009285826GA SD009285827GA SD009285828GA SD009285829GA SD009285830GA SD009285831GA SD009285832GA SD009285833GA SD010095480GA PA1934072 PA1934074 SBG07077613GA SBG07077617GA SBG07077619GA SBG07077621GA SBG07077623GA SBG07077624GA SBG07077625GA SBG07077626GA SD007282986GA SD007282989GA SD007282990GA SD007282991GA SD007293516GA SD007293517GA SD007293518GA SD007293519GA SD007293520GA SD007335416GA SD007335419GA SD007335420GA SD008077064GA SD008077067GA SD008077069GA SD008077074GA SD008077099GA SD008077116GA SD008077180GA SD008077182GA SD008077184GA SD008077187GA SD008077188GA SD008077189GA SD008077190GA SD008077191GA SD008077192GA SD008077193GA SD008077195GA SD008077196GA SD008077197GA SD008077198GA SD008077199GA SD009223944GA SD009234148GA SD010127641GA |
| FEI Number |
2126677
|
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | 262-513-4122 |
Manufacturer Reason for Recall | GE Healthcare has recently become aware of an issue with the Dash 3000/4000/5000 patient monitors configured with the 802.11b wireless LAN option that may impact patient safety.
The Dash monitor may exhibit reboots on some networks when the 802.11b wireless LAN option is used to connect to the hospital wireless LAN infrastructure. This reboot results in a loss of patient monitoring for approxima |
FDA Determined Cause 2 | Software design |
| Action | GE Healthcare sent an URGENT MEDICAL DEVICE CORRECTION letter dated June 15, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. GE recommended actions that should be taken by a qualified service representative. For any questions regarding the medical device correction or identification of affected items contact Technical Support at 1-800-558-7044 or customers should contact their local Service Representative.
UPDATE:
A second letter dated 12/13/11 was sent on 1/20/12 to consignees. The revised letter provides an update to the safety instructions and product correction. |
| Quantity in Commerce | 8029 devices |
| Distribution | Worldwide Distribution - USA including AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VT, VA, WA, WV, WI and the countries of UNITED KINGDOM, TURKEY, THAILAND, TAIWAN, SWEDEN, SPAIN, SLOVENIA, SINGAPORE, SAUDI A ARABIA, RUSSIAN FEDERATION, REPUBLIC OF KOREA, PORTUGAL, POLAND, NORWAY, NEW ZEALAND, NETHERLANDS, MEXICO, KUWAIT, JAPAN, ITALY, ISRAEL, INDIA, GERMANY, FRANCE, FINLAND, ESTONIA, ECUADOR, DENMARK, CANDA, BELGIUM AUSTRIA, AUSTRALIA, and ALGERIA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|