  Class 2 Device Recall MX8000 Dual v. Exp CT XRay System see related information
Date Initiated by Firm June 03, 2011
Date Posted August 01, 2011
Recall Status1 Terminated 3 on August 07, 2013
Recall Number Z-2914-2011
Recall Event ID 59166
510(K)Number K010817  
Product Classification System, x-ray, tomography, computed - Product Code KPS
Product Vertical Brake Hubs component of the Philips MX8000 Series Dual v. Exp. Computed Tomography System, Serial Numbers: 8008-8413, 80001-800022, Model #728130, 510 (k) #K010817.

Product Usag: The MX 8000 series Dual v. Exp are Computed Tomography X-Ray system units intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Code Information The following Serial Numbers of the CT X-ray Scanners are subject to recall/correction: 8008-8413, 80001-800022.
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Road
Cleveland OH 44143-2131
For Additional Information Contact
440-483-7000
Manufacturer Reason
for Recall
Philips Medical Systems decided to conduct a recall/field correction of the patient support vertical brake hub component in all of their CT MX8000 Dual v. Exp Computed Tomography X-Ray System scanners due to a component defect which could cause an unexpected downward motion of the table (couch) while the patient is on the table. This event could occur because the short key connecting the vertical
FDA Determined
Cause 2
Device Design
Action Philips Medical Systems (Cleveland) Inc. issued an "URGENT - Medical Device Correction - Field Safety Notice" letter dated June 1, 2011 to all affected customers, via Certified Mail for tracking purposes. The notification identified the product, the problem involved, and the action(s) to be taken by the customers with regard to this device issue. The notification instructs the customers to monitor their units and if the recalled product malfunctions, to immediately notify their Phillips service representative. The letter also states that a Philips Field Service Engineer will visit each customer location and perform a brake repair or replacement on their system to resolve the issue. For further information or support, the customers are instructed to contact their local Philips representative or local Philips Healthcare office. For North America and Canada the customers are instructed to contact the Customer Care Solutions Center (1-800-722-9377, option 5: Enter Site ID or follow the prompt(s).
Quantity in Commerce 412 system units were distributed.
Distribution Worldwide Distribution -- USA (nationwide) including the states of: AR, AZ, CA, CO, FL, FA, HI, IL, IN, KY, MA, MI, MO, MS, NC, NV, NY, OH, OK, SC, SD, TN, TX, VA, VT, WA and Puerto Rico and the country of Canada.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = KPS and Original Applicant = PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
-
-