|
Class 2 Device Recall REVOLUTION XR/D DIGITAL RADIOGRAPHIC IMAGING SYSTEM WITH TOMOSYNTHESIS |
|
Date Initiated by Firm |
October 26, 2011 |
Date Posted |
February 21, 2012 |
Recall Status1 |
Terminated 3 on August 02, 2012 |
Recall Number |
Z-1065-2012 |
Recall Event ID |
60126 |
510(K)Number |
K051967
|
Product Classification |
System, x-ray, stationary - Product Code KPR
|
Product |
REVOLUTION XR/D DIGITAL RADIOGRAPHIC IMAGING SYSTEM WITH TOMOSYNTHESIS
The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic images of human anatomy. It is not intended for mammographic use. |
Code Information |
Serial Number 000000PROTOHL3 00000002400HL9 00000013894HL1 00000013895HL8 00000014468HL3 00000014643HL1 00000014646HL4 00000014647HL2 00000014648HL0 00000014649HL8 00000016626HL4 00000016629HL8 00000016630HL6 00000018158HL6 00000018159HL4 00000018160HL2 00000018163HL6 00000018164HL4 00000018649HL4 00000018733HL6 00000018769HL0 00000018804HL5 00000019142HL9 00000019224HL5 00000019387HL0 00000019550HL3 00000019551HL1 00000019712HL9 00000019748HL3 00000019749HL1 00000019750HL9 00000019751HL7 00000019752HL5 00000019754HL1 00000019855HL6 00000019858HL0 00000019861HL4 00000019925HL7 00000020047HL7 00000020048HL5 00000020049HL3 00000020050HL1 00000020051HL9 00000020053HL5 00000020275HL4 00000020276HL2 00000020277HL0 00000020279HL6 00000020280HL4 00000020281HL2 00000020283HL8 00000020287HL9 00000020289HL5 00000020290HL3 00000020295HL2 00000020297HL8 00000020298HL6 00000020299HL4 00000020300HL0 00000020301HL8 00000020302HL6 00000020304HL2 00000020305HL9 00000020306HL7 00000020307HL5 00000020309HL1 00000020311HL7 00000021375HL1 00000021376HL9 00000021377HL7 00000021916HL2 00000022300HL8 00000022356HL0 00000022359HL4 00000022361HL0 00000022364HL4 00000022365HL1 00000022494HL9 00000022495HL6 00000022499HL8 00000023345HL2 00000023346HL0 00000023356HL9 00000023568HL9 00000023582HLO 00000023582HL0 00000023640HL6 00000023642HL2 00000023643HL0 00000023644HL8 00000023645HL5 00000023706HL5 00000023707HL3 00000023708HL1 00000023932HL7 00000023934HL3 00000023935HL0 00000024598HL5 00000024600HL9 00000024601HL7 00000024602HL5 00000024605HL8 00000024686HL8 00000024695HL9 00000024696HL7 00000024930HL0 00000024932HL6 00000024933HL4 00000024934HL2 00000024935HL9 00000024936HL7 00000025333HL6 00000025395HL5 00000025396HL3 00000025463HL1 00000025464HL9 00000025465HL6 00000025599HL2 00000025601HL6 00000025602HL4 00000025603HL2 00000025604HL0 00000026100HL8 00000026101HL6 00000026102HL4 00000027827HL5 00000027828HL3 00000027830HL9 00000027831HL7 00000027832HL5 00000028087HL5 00000028088HL3 00000028268HL1 00000028392HL9 00000028393HL7 00000028723HL5 00000028764HL9 00000028765HL6 00000028767HL2 00000028893HL6 00000028894HL4 00000028895HL1 00000028896HL9 00000028897HL7 00000029009HL8 00000029011HL4 00000029042HL9 00000029043HL7 00000029200HL3 00000029300HL1 00000029541HL0 00000029542HL8 00000029598HL0 00000029715HL0 00000029716HL8 00000029717HL6 00000029788HL7 00000029789HL5 00000029791HL1 00000029832HL3 00000029833HL1 00000029855HL4 00000029856HL2 00000029894HL3 00000029895HL0 00000029896HL8 00000029897HL6 00000029898HL4 00000029983HL4 00000029984HL2 00000030099HL6 00000030100HL2 00000030154HL9 00000030155HL6 00000030300HL8 00000030478HL2 00000030521HL9 00000030522HL7 00000030524HL3 00000030527HL9 00000030657HL1 00000030872HL6 00000030934HL4 00000031030HL0 00000031058HL1 00000031059HL9 00000031196HL9 00000031197HL7 00000031244HL7 00000031413HL8 00000031474HL0 00000031518HL4 00000031520HL0 00000031521HL8 00000031522HL6 00000031643HL0 00000031647HL1 00000031723HL0 00000031726HL3 00000031827HL9 00000031829HL5 00000031830HL3 00000031831HL1 00000031832HL9 00000031833HL7 00000032058HL0 00000032059HL8 00000032060HL6 00000032119HL0 00000032121HL6 00000032196HL8 00000032197HL6 00000032198HL4 00000032199HL2 00000032245HL3 00000032246HL1 00000032247HL9 00000032344HL4 00000032345HL1 00000032416HL0 00000032417HL8 00000032444HL2 00000032448HL3 00000032449HL1 00000032476HL4 00000032478HL0 00000032539HL9 00000032540HL7 00000032542HL3 00000032612HL4 00000032613HL2 00000032684HL3 00000032685HL0 00000032767HL6 00000032768HL4 00000032975HL5 00000033053HL0 00000033054HL8 00000033055HL5 00000033110HL8 00000033113HL2 00000033334HL4 00000033335HL1 00000033337HL7 00000033338HL5 00000033404HL5 00000033405HL2 00000033406HL0 00000033487HL0 00000033517HL4 00000033518HL2 00000033519HL0 00000033520HL8 00000033521HL6 00000033592HL7 00000033593HL5 00000033594HL3 00000033595HL0 00000033687HL5 00000033688HL3 00000033689HL1 00000033690HL9 00000033692HL5 00000033770HL9 00000033814HL5 00000033842HL6 00000033843HL4 00000033844HL2 00000033890HL5 00000033891HL3 00000033892HL1 00000034011HL7 00000034014HL1 00000034056HL2 00000034057HLO 00000034162HL8 00000034180HL0 00000034181HL8 00000034182HL6 00000034265HL9 00000034329HL3 00000034331HL9 00000034391HL3 00000034392HL1 00000034406HL9 00000034409HL3 00000034529HL8 00000034530HL6 00000034531HL4 00000034532HL2 00000034533HL0 00000034587HL6 00000034590HL0 00000034742HL7 00000034743HL5 00000034791HL4 00000034793HL0 00000034830HL0 00000034831HL8 00000034878HL9 00000034879HL7 00000034880HL5 00000034881HL3 00000034915HL9 00000034916HL7 00000034976HL1 00000034977HL9 00000034978HL7 00000034979HL5 00000034980HL3 00000034981HL1 00000034982HL9 00000034984HL5 00000035029HL6 00000035064HL5 00000035065HL2 00000035066HL0 00000035085HL0 00000035086HL8 00000035087HL6 00000035209HL6 00000035232HL5 00000035323HL5 00000035324HL3 00000035326HL8 00000035524HL8 00000035592HL5 00000035593HL3 00000035594HL1 00000035595HL8 00000035596HL6 00000035597HL4 00000035635HL2 00000035843HL2 00000035844HL0 00000035845HL7 00000035921HL6 00000035922HL4 00000035925HL7 00000035993HL5 00000036001HL6 00000036014HL9 00000036015HL6 00000036047HL9 00000036049HL5 00000036050HL3 00000036051HL1 00000036142HL8 00000036195HL6 00000036198HL0 00000036300HL2 00000036325HL9 00000036369HL7 00000036371HL3 00000036413HL3 00000036414HL1 00000036415HL8 00000036416HL6 00000036417HL4 00000036505HL6 00000036506HL4 00000036507HL2 00000036509HL8 00000036592HL4 00000036594HL0 00000036595HL7 00000036775HL5 00000036776HL3 00000036777HL1 00000036813HL4 00000036815HL9 00000036893HL6 00000036896HL9 00000036898HL5 00000036899HL3 00000036900HL9 00000037040HL3 00000037041HL1 00000037042HL9 00000037132HL8 00000037134HL4 00000037135HL1 00000037274HL8 00000037275HL5 00000037276HL3 00000037382HL9 00000037383HL7 00000037384HL5 00000037385HL2 00000037546HL9 00000037549HL3 00000037603HL8 00000037724HL2 00000037725HL9 00000037728HL3 00000037897HL6 00000037898HL4 00000037900HL8 00000038023HL8 00000038024HL6 00000038025HL3 00000038089HL9 00000038090HL7 00000038191HL3 00000038192HL1 00000038193HL9 00000038364HL6 00000038365HL3 00000038435HL4 00000038485HL9 00000038487HL5 00000038775HL3 00000038776HL1 00000038937HL9 00000038938HL7 00000038939HL5 00000039213HL4 00000039282HL9 00000039283HL7 00000039970HL9 00000040017HL6 00000040018HL4 00000040019HL2 00000040111HL7 00000040125HL7 00000040277HL6 00000040278HL4 00000040279HL2 00000040388HL1 00000040418HL6 00000040419HL4 00000040580HL3 00000040581HL1 00000040676HL9 00000040802HL1 00000040868HL2 00000040926HL8 00000041018HL3 00000041123HL1 00000041272HL6 00000041349HL2 00000041351HL8 00000041353HL4 00000041432HL6 00000041433HL4 00000041435HL9 00000041436HL7 00000041437HL5 00000041439HL1 00000347434HL5 00000029170HL8 00000030527HL6 00000030933HL6 00000031057HL3 00000031247HL0 00000031409HL6 00000031412HL0 00000031523HL4 00000031646HL3 00000031724HL8 00000034163HL6 00000035170HL0 00000036145HL1 00000036596HL5 00000037547HL7 00000037727HL5 00000037899HL2 00000038021HL2 00000038022HL0 00000038091HL5 00000038774HL6 00000040420HL2 00000041273HL4 00000041352HL6 00000041434HL2 00000959507WK5 00000014646HL4 00000014468HL3 00000014587HL0 00000014642HL3 00000021376HL9 00000021377HL7 00000018770HL8 00000019223HL7 00000019145HL2 00000019924HL0 00000020278HL8 00000024685HL0 00000020298HL6 00000029790HL3 00000023357HL7 00000019750HL9 00000031723HL0 00000023580HL4 00000024599HL3 00000020310HL9 00000025396HL3 00000026102HL4 00000028894HL4 00000024929HL2 00000014641HL5 00000029300HL1 00000028548HL6 00000025602HL4 00000028723HL5 00000029009HL8 00000028549HL4 00000030100HL2 00000030099HL6 00000031833HL7 00000034742HL7 00000031831HL1 00000037134HL4 00000029543HL6 00000033814HL5 00000029789HL5 00000019142HL9 00000016629HL8 00000033406HL0 00000034181HL8 00000034057HL0 00000034056HL2 00000031643HL0 00000030872HL6 00000035170HL0 00000034984HL5 00000032614HL0 00000030300HL8 00000033053HL0 00000034532HL2 00000033518HL2 00000033519HL0 00000034977HL9 00000035326HL8 00000037133HL6 00000036593HL2 00000035921HL6 00000037546HL9 00000035524HL8 00000040278HL4 00000014644HL9 00000034057HLO 00000036775HL5 00000038775HL3 00000036812HL6 00000036893HL6 00000037605HL3 00000036592HL4 00000038939HL5 00000038025HL3 00000036414HL1 00000037725HL9 00000029983HL4 00000039282HL9 00000039283HL7 00000038090HL7 00000036592HL4 00000038485HL9 00000040802HL1 00000040581HL1 00000041353HL4 00000036049HL5 00000041437HL5 00000039970HL9 00000018803HL7 00000021917HL0 00000021919HL6 00000022360HL2 00000037272HL2 00000034792HL2 00000034741HL9 00000037136HL9 00000016625HL6 00000017074HL6 00000018156HL0 00000020286HL1 00000020296HL0 00000023358HL5 00000029714HL3 00000033812HL9 00000033815HL2 00000037726HL7 00000029003HL1 00000014645HL6 00000029012HL2 00000029168HL2 00000029169HL0 00000033927HL5 00000035923HL2 00000002666EN6 00000017073HL8 00000017075HL3 00000018154HL5 00000018162HL8 00000022363HL6 00000022497HL2 00000023583HL8 00000023931HL9 00000025600HL8 00000028766HL4 00000029599HL8 00000032770HL0 00000034410HL1 00000034589HL2 00000035171HL8 00000036196HL4 00000036814HL2 00000038088HL1 00000038773HL8 00000039085HL6 00000039086HL4 00000040801HL3 00000026099HL2 00000028898HL5 00000031201HL7 00000031246HL2 00000033112HL4 00000033402HL9 00000034688HL2 00000037606HL1 00000037899HL2 00000038363HL8 00000038486HL7 00000040175HL2 00000014650HL6 00000019859HL8 00000020308HL3 00000034914HL2 00000019856HL4 00000022362HL8 00000034393HL9 00000034407HL7 00000034740HL1 00000030299HL2 00000032541HL5 00000036197HL2 00000029301HL9 00000017068HL8 00000017072HL0 00000018647HL8 00000018650HL2 00000019141HL1 00000019711HL1 00000019755HL8 00000019854HL9 00000019926HL5 00000019927HL3 00000020046HL9 00000023569HL7 00000028722HL7 00000029857HL0 00000030525HL0 00000031121HL7 00000034832HL6 00000001914HL1 00000037605HL3 00000015863HL4 00000019143HL7 00000020274HL7 00000023359HL3 00000029201HL1 00000029303HL5 00000034183HL4 00000036370HL5 00000037548HL5 00000037604HL6 00000035169HL2 00000030658HL9 00000036048HL7 00000033811HL1 00000035325HL0 00000040924HL3 00000040925HL0 00000961628WK5 00000018155HL2 00000020284HL6 00000032683HL5 00000035167HL6 00000023581HL2 00000037273HL0 00000031642HL2 00000036144HL4 00000031828HL7 00000020291HL1 00000020294HL5 00000022301HL6 00000022496HL4 00000023709HL9 00000023930HL1 00000027829HL1 00000030153HL1 00000030156HL4 00000030383HL4 00000030526HL8 00000030528HL4 00000031198HL5 00000031411HL2 00000031519HL2 00000031644HL8 00000031645HL5 00000031648HL9 00000031725HL5 00000032120HL8 00000032445HL9 00000033691HL7 00000033813HL7 00000034687HL4 00000034689HL0 00000034690HL8 00000035208HL8 00000036508HL0 00000036897HL7 00000037043HL7 00000037381HL1 00000038284HL6 00000038285HL3 00000038286HL1 00000039778HL6 00000039971HL7 00000039972HL5 00000041350HL0 00000041438HL3 00000016627HL2 00000018157HL8 00000018161HL0 00000019713HL7 00000019857HL2 00000020285HL3 00000020288HL7 00000020293HL7 00000021378HL5 00000022302HL4 00000024688HL4 00000025334HL4 00000029004HL9 00000029171HL6 00000029302HL7 00000029540HL2 00000030479HL0 00000030480HL8 00000030659HL7 00000030873HL4 00000030932HL8 00000031122HL5 00000031199HL7 00000031200HL9 00000031245HL4 00000031475HL7 00000031721HL4 00000031722HL2 00000032118HL2 00000032244HL6 00000032477HL2 00000032976HL3 00000033109HL0 00000033336HL9 00000033403HL7 00000033407HL8 00000034009HL1 00000034010HL9 00000034012HL5 00000034013HL3 00000034983HL7 00000035068HL6 00000035322HL7 00000035327HL6 00000035523HL0 00000038484HL2 00000065910HL2 00000030660HL5 00000017070HL4 00000019386HL2 |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
262-513-4122
|
Manufacturer Reason for Recall |
GE Healthcare has become aware of a potential issue associated with the hand grip that is clamped to the table top of your Revolution XRD X-ray systems that may impact patient safety.
The hand grip accessory that is clamped to the table top may not be properly adjusted and can dislodge during use. Should this happen during patient loading or unloading, patient injury could occur.
|
FDA Determined Cause 2 |
Device Design |
Action |
GE Healthcare sent an Urgent Medical Device Correction letter dated October 24, 2011, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to remove the hand grip accessory from use until the product is corrected and to take extra precautions when loading or unloading patients from the table while the hand grips are not in use. A GE Healthcare service representative would contact the consignee to arrange for the correction.
For questions regarding this recall call 262-513-4122. |
Quantity in Commerce |
772 |
Distribution |
Worldwide Distribution - USA including AL, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WY, WI and the countries of VENEZUELA, UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, TAIWAN, SWITZERLAND, SPAIN, SLOVENIA, SAUDI ARABIA, REPUBLIC OF KOREA, NORWAY, NEW ZEALAND, NETHERLANDS, MALTA, KUWAIT, JORDAN, JAPAN, ITALY, ISRAEL, HONG KONG, GERMANY, FRANCE, ECUADOR, DENMARK, CZECH REPUBLIC, CHINA, CHILE, CANADA, and AUSTRALIA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KPR and Original Applicant = GE HEALTHCARE
|
|
|
|