 Class 2 Device Recall CDI 500 Monitorsee related information
Date Initiated by FirmFebruary 03, 2012
Date PostedApril 06, 2012
Recall Status1 Terminated 3 on August 29, 2013
Recall NumberZ-1378-2012
Recall Event ID 60152
510(K)NumberK972962 
Product Classification Monitor, blood-gas, on-line, cardiopulmonary bypass - Product Code DRY
ProductCDI Blood Parameter Monitoring System 500, 500AVHCT. Intended for use during cardiopulmonary bypass procedures.
Code Information 500AVHCT CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules and Hematocrit/ Saturation probe: 1001, 1005, 1007, 1011-1022, 1024, 1028, 1032, 1036, 1038, 1039, 1041-1044, 1046, 1047, 1049-1053, 1059, 1063, 1066-1071, 1075, 1096, 1097, 1101, 1103-1105, 1111-1113, 1115, 1130, 1133-1140, 1146-1152, 1166-1168, 1181-1185, 1216-1220, 1231-1235, 1244-1248, 1254, 1255, 1261-1264, 1270-1274, 1285, 1286-1289, 1295-1299, 1304-1309, 1312-1315, 1321, 1332-1334, 1340-1344, 1350-1354, 1357, 1358, 1379-1383, 1389-1393, 1399-1408, 1414-1418, 1421, 1422, 1424-1433, 1439-1443, 1449-1453, 1459-1463, 1466, 1479, 1480, 1494, 1525, 1554, 1597-1601, 1605, 1617-1631, 1637-1646, 1657-1661, 1672-1676, 1682-1686, 1697-1701, 1712-1716, 1733-1737, 1787-1789, 1805-1809, 1815-1819, 1830, 1844-1846, 1857-1860, 1886-1890, 1901-1905, 1916-1920, 1926-1930, 1936-1940, 1953-1974, 1976-1985, 2028-2034, 2036-2038, 2040, 2045, 2059-2063, 2065-2067, 2070-2077, 2086, 2097-2101, 2109-2111, 2113-2115, 2123-2131, 2138-2142, 2149-2155, 2177-2182, 2185-2187, 2189, 2195-2199, 2214, 2216-2219, 2224, 2228-2231, 2243-2245, 2249-2253, 2255-2258, 2260, 2263, 2265-2268, 2275-2277, 2279, 2285-2291, 2295-2298, 2313-2321, 2323-2329, 2338-2354, 2357, 2368, 2370-2375, 2377, 2392, 2403-2408, 2417-2426, 2432-2436, 2439-2444, 2453-2458, 2463-2467, 2470-2472, 2476-2483, 2487-2503, 2514-2517, 2521-2525, 2541-2556, 2565-2582, 2587-2603, 2629-2633, 2635-2642, 2646, 2647, 2656, 2661, 2662, 2664-2680, 2689-2691, 2695, 2696, 2701-2708, 2710-2716, 2722-2727, 2737, 2742-2747, 2757-2773, 2783, 2784, 2788, 2793-2795, 2802, 2805-2808, 2815-2820, 2837-2839, 2846-2849, 2865-2870, 2880-2889, 2892-2895, 2906-2908, 2913-2915, 2920, 2921, 2924-2926, 2937-2940, 2942-2947, 2952, 2958-2975, 2979-2983, 2990-2999, 3008-3010, 3022-3027, 3040, 3041, 3052-3055, 3073-3075, 3083-3087, 3096, 3115, 3116, 3139-3145, 3152, 3153, 3155-3162, 3169-3182, 3190-3196, 3214-3220, 3225-3228, 3232, 3238-3245, 3248, 3249, 3264-3268, 3287, 3288, 3291-3306, 3314-3323, 3329-3334, 3340-3349, 3354-3357, 3369-3373, 3383-3391, 3394-3397, 3402-3406, 3416, 3418-3427, 3434-3437, 3441-3447, 3449-3456, 3458-3462, 3469-3472, 3481, 3482, 3489, 3490, 3500-3503, 3509, 3510, 3518, 3534, 3535, 3541-3545, 3571-3573, 3590-3609, 3612-3619, 3624-3628, 3636-3639, 3643-3645, 3669-3681, 3698-3703, 3715, 3716, 3722-3726, 3738-3744, 3753-3757, 3769-3780, 3784-3787, 3799-3804, 3811-3816, 3827-3830, 3838-3841, 3852-3881, 3892-3899, 3902-3906, 3912-3921, 3924-3931, 3937-3952, 3962-3965, 3971-3974, 3986-3989, 3995-3998, 4006-4029, 4038-4043, 4050-4055, 4070-4075, 4082-4089, 4096-4099, 4105-4108, 4114-4131, 4143-4148, 4169-4185, 4203-4212, 4219-4224, 4230-4241, 4260-4280, 4290-4299, 4307-4310, 4318-4323, 4335-4344, 4351, 4352, 5001-5010, 5036-5039, 5048-5056, 5065-5077, 5082-5085, 5094-5101, 5118-5125, 5142-5149, 5158-5169, 5178, 5179, 5190, 5197, 5198, 5210-5213, 5222-5225, 5238-5245, 5250-5253, 5274-5277, 5282-5285, 5290-5293, 5298-5305, 5318-5321, 5334-5337, 5341-5344, 5354-5356, 5370-5374, 5383, 5385, 5386, 5516-5526, 5532-5541, 5557-5561, 5584-5587, and 5589-5592.
FEI Number 1828100
Recalling Firm/
Manufacturer
Terumo Cardiovascular Systems Corporation
6200 Jackson Road
Ann Arbor MI 48103-9586
For Additional Information Contact
734-741-6173
Manufacturer Reason
for Recall
Replacement of the SBC batteries (system batteries) should be managed as "routine maintenance" instead of as an in-house Service Procedure because some systems may be relying on batteries beyond their useful life.
FDA Determined
Cause 2
Labeling design
ActionOn 2/3/12 all consignees received an URGENT MEDICAL DEVICE RECALL letter via Federal Express. The letter described the reason for the recall, identified the affected product population, what to do in the event of a failure, the potential hazard, and correction. Terumo CVS will update the Operator's Manual for the CDI 500 with instructions to replace the battery within its useful life. Users are to review the notice and ensure that all users are made aware of the issue. In addition, the Customer Response Form should be completed and returned. Users can contact their local Terumo Cardiovascular Systems representative or call 1-800-521-2818 with questions or concerns, to check the last time system batteries were replaced, or to schedule maintenance for systems approaching 10 years.
Quantity in Commerce3,829 units
DistributionWorldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, WV, WA, and WI and the countries of Japan, Germany, Barbados, Saudi Arabia, Columbia, Uruguay, Ecuador, Trinidad, Costa Rica, Paraguay, Panama, Belgium, Jordan, Libya, Mexico, Brazil, Honduras, Canada, Korea, Dominica Republic, El Salvador, Guatemala, Vietnam, Chile, Hong Kong, United Arab Emirates, Singapore, Australia, Malaysia, Taiwan, Philippines, Mexico, Thailand, Vietnam, and South Africa.
Total Product Life CycleTPLC Device Report