• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall AngioDynamics Morpheus Smart PICC CT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall AngioDynamics Morpheus Smart PICC CT see related information
Date Initiated by Firm January 09, 2012
Date Posted January 26, 2012
Recall Status1 Terminated 3 on February 05, 2016
Recall Number Z-0910-2012
Recall Event ID 60909
510(K)Number K093406  
Product Classification Catheter,intravascular,therapeutic,long-term greater than 30 days - Product Code LJS
Product AngioDynamics Special Order Morpheus¿ Smart PICC CT 65cm 5F Dual Lumen with Reverse Taper --- Max CT Flow Rate: 4mL/sec --- Injector Pressure Limit --- Catalog No.: 52102618 --- Made in the USA --- [Manufacturer symbol] AngioDynamics, 603 Queensbury Avenue, Queensbury, NY 12804

indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injections of contrast media.
Code Information FG Lot Number: 554082, 554554
Recalling Firm/
Manufacturer
Angiodynamics Worldwide Headquarters
14 Plaza Drive
Latham NY 12110-3421
For Additional Information Contact David Greer
518-795-1676
Manufacturer Reason
for Recall
The red luer on the Morpheus CT PICC (Peripherally Inserted Central Catheter) has the potential to crack during use of the product.
FDA Determined
Cause 2
Nonconforming Material/Component
Action AngioDynamics, Inc. sent an Urgent - Medical Device Recall letter dated January 5, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to identify and segragate the recalled affected product. Complete the enclosed Morpheus Ct PICC Recall Reply Form and fax it to the attention of the Morpheus CT PICC Recall Coordinator at 518-798-1360. If customers do not have any affected product remaining in their possession they should fill out the attached form noting zero quantity to be returned and fax the form to 518-798-1360. If customers distributed the affected product, they should inform their customers of this recall action immediately and have them return the recalled units to them. For any questions regarding this recall call 1-800-772-6446 or e-mail customerservice@angiodynamics.com.
Quantity in Commerce 35 units
Distribution Nationwide
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = LJS and Original Applicant = ANGIODYNAMICS, INC.
-
-