• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall RF IgA Calibrator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall RF IgA Calibratorsee related information
Date Initiated by FirmMarch 29, 2012
Date PostedMay 14, 2012
Recall Status1 Terminated 3 on September 05, 2017
Recall NumberZ-1546-2012
Recall Event ID 61500
510(K)NumberK941398 
Product Classification System, test, rheumatoid factor - Product Code DHR
ProductRF IgA Calibrator, a component of the EL-RF/3 Rheumatoid Factor Test Kit; an in vitro diagnostic; TheraTest Laboratories, Lombard, IL 60148; component catalog #12833, kit catalog #303-305. The TheraTest EL-RF/3 is intended as an aid to the diagnosis of Rheumatoid Arthritis (RA).
Code Information RF IgA Calibrator: component catalog number 12833, lot number 11118586, exp. 28 DEC 2012; EL-RF/3: kit catalog number 303-305, lot number 11118583, exp. 28 NOV 2012.
FEI Number 1421346
Recalling Firm/
Manufacturer
Thera Test Laboratories, Inc.
1111 N Main St
Lombard IL 60148-1360
For Additional Information Contact
630-627-6069
Manufacturer Reason
for Recall
The RF IgA Calibrator component of the EL-RF/3 Rheumatoid Factor test system may be unstable, resulting in a drop in the RF IgA calibrator OD value which could invalidate an RF IgA test run.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionTheraTest sent "Urgent Device Recall" letters dated March 28, 2012 to customers that received the affected kit, requesting them to return or destroy the remainder of the product. TheraTest issued a replacement kit to all customers that had unused product as well as any material they consumed which resulted in a failed run. Customers were also instructed to complete the enclosed confirmation form and fax the completed form back to TheraTest. Any questions were directed to TheraTest at 800-441-0771.
Quantity in Commerce207 kits
DistributionNationwide Distribution (US)
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = DHR
-
-