Class 2 Device Recall Stryker MA204

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Stryker MA204see related information
Date Initiated by FirmMay 17, 2012
Date PostedJune 22, 2012
Recall Status1 Terminated 3 on May 12, 2014
Recall NumberZ-1862-2012
Recall Event ID 61712
Product Classification Bed, ac-powered adjustable hospital - Product Code FNL
ProductStryker MA204 Bed Model (FL25E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone
Code Information Model FL25E: QK02276, QK02274, QK01602, KK01602, K03922, K03921, K03920, K03919, K03918, K03917, K03916, K03915, K03914, K03913, K03912, K03911, K03910, K03909, K03908, K03907, K03906, K03905, K03904, K03903, K03902, K03901, K03900, K03899, K03898, K03897, K03896, K03895, K03894, K03893, K03892, K03891, K03890, K03889, K03888, K03887, K03886, K03885, K03884, K03883, K03882, K03881, K03880, K03879, K03878, K03877, K03876, K03875, K03874, K03873, K03872, K03871, K03870, K03869, K03868, K03867, K03866, K03865, K03864, K03863, K03862, K03861, K03860, K03859, K03858, K03857, K03856, K03855, K03854, K03853, K03852, K03851, K03850, K03849, K03848, K03847, K03846, K03845, K03844, K03843, K03842, K03841, K03840, K03839, K03838, K03837, K03836, K03835, K03834, K03833, K03832, K03831, K03830, K03829, K03828, K03827, K03826, K03825, K03824, K03823, K03822, K03821, K03820, K03819, K03818, K03817, K03816, K03815, K03814, K03813, K03812, K03811, K03810, K03809, K03808, K03807, K03806, K03805, K03804, K03803, K03802, K03801, K03800, K03799, K03798, K03797, K03796, K03795, K03794, K03793, K03792, K03791, K03790, K03789, K03788, K03787, K03786, K03785, K03784, K03783, K03782, K03781, K03780, K03779, K03778, K03777, K03776, K03775, K03774, K03773, K03772, K03771, K03770, K03769, K03768, K03767, K03766, K03765, K03764, K03763, K03762, K03761, K03760, K03759, K03758, K03757, K03756, K03755, K03754, K03753, K03752, K03751, K03750, K03749, K03748, K03747, K03746, K03745, K03744, K03743, K03742, K03741, K03740, K03739, K03738, K03737, K03736, K03735, K03734, K03733, K03732, K03731, K03730, K03729, K03728, K03727, K03726, K03725, K03724, K03723, K03722, K03721, K03720, K03719, K03718, K03717, K03716, K03715, K03714, K03713, K03712, K03711, K03710, K03709, K03708, K03707, K03706, K03705, K03704, K03703, K03702, K03701, K03700, K03699, K03698, K03697, K03696, K03695, K03694, K03693, K03692, K03691, K03690, K03689, K03688, K03687, K03686, K03685, K03684, K03683, K03682, K03681, K03680, K03679, K03678, K03677, K03676, K03675, K03674, K03673, K03672, K03671, K03670, K03669, K03668, K03667, K03666, K03665, K03664, K03663, K03662, K03661, K03660, K03659, K03658, K03657, K03656, K03655, K03654, K03653, K03652, K03651, K03650, K03649, K03648, K03647, K03646, K03645, K03644, K03643, K03642, K03641, K03640, K03639, K03638, K03637, K03636, K03635, K03634, K03633, K03632, K03631, K03630, K03629, K03628, K03627, K03626, K03625, K03624, K03623, K03622, K03621, K03620, K03619, K03618, K03617, K03616, K03615, K03614, K03613, K03612, K03611, K03610, K03609, K03608, K03607, K03606, K03605, K03604, K03603, K03602, K03601, K03600, K03599, K03598, K03597, K03596, K03595, K03594, K03593, K03592, K03591, K03590, K03589, K03588, K03587, K03586, K03585, K03584, K03583, K03582, K03581, K03580, K03579, K03578, K03577, K03576, K03575, K03574, K03573, K03572, K03571, K03570, K03569, K03568, K03567, K03566, K03565, K03564, K03563, K03562, K03561, K03560, K03559, K03558, K03557, K03556, K03555, K03554, K03553, K03552, K03551, K03550, K03549, K03548, K03547, K03546, K03545, K03544, K03543, K03542, K03541, K03540, K03539, K03538, K03537, K03536, K03535, K03534, K03533, K03532, K03531, K03530, K03529, K03528, K03527, K03526, K03525, K03524, K03523, K03522, K03521, K03520, K03519, K03518, K03517, K03516, K03515, K03514, K03513, K03512, K03511, K03510, K03509, K03508, K03507, K03506, K03505, K03504, K03503, K03502, K03501, K03500, K03499, K03498, K03497, K03496, K03495, K03494, K03493, K03492, K03491, K03490, K03489, K03488, K03487, K03486, K03485, K03484, K03483, K03482, K03481, K03480, K03479, K03478, K03477, K03476, K03475, K03474, K03473, K03472, K03471, K03470, K03469, K03468, K03467, K03466, K03465, K03464, K03463, K03462, K03461, K03460, K03459, K03458, K03457, K03456, K03455, K03454, K03453, K03452, K03451, K03450, K03449, K03448, K03447, K03446, K03445, K03444, K03443, K03442, K03441, K03440, K03439, K03438, K03437, K03436, K03435, K03434, K03433, K03432, K03431, K03430, K03429, K03428, K03427, K03426, K03425, K03424, K03423, K03422, K03421, K03420, K03419, K03418, K03417, K03416, K03415, K03414, K03413, K03412, K03411, K03410, K03409, K03408, K03407, K03406, K03405, K03404, K03403, K03402, K03401, K03400, K03399, K03398, K03397, K03396, K03395, K03394, K03393, K03392, K03391, K03390, K03389, K03388, K03387, K03386, K03385, K03384, K03383, K03382, K03381, K03380, K03379, K03378, K03377, K03376, K03375, K03374, K03373, K03372, K03371, K03370, K03369, K03368, K03367, K03366, K03365, K03364, K03363, K03362, K03361, K03360, K03359, K03358, K03357, K03356, K03355, K03354, K03353, K03352, K03351, K03350, K03349, K03348, K03347, K03346, K03345, K03344, K03343, K03342, K03341, K03340, K03339, K03338, K03337, K03336, K03335, K03334, K03333, K03332, K03331, K03330, K03329, K03328, K03327, K03326, K03325, K03324, K03323, K03322, K03321, K03320, K03319, K03318, K03317, K03316, K03315, K03314, K03313, K03312, K03311, K03310, K03309, K03308, K03307, K03306, K03305, K03304, K03303, K03302, K03301, K03300, K03299, K03298, K03297, K03296, K03295, K03294, K03293, K03292, K03291, K03290, K03289, K03288, K03287, K03286, K03285, K03284, K03283, K03282, K03281, K03280, K03279, K03278, K03277, K03276, K03275, K03274, K03273, K03272, K03271, K03270, K03269, K03268, K03267, K03266, K03265, K03264, K03263, K03262, K03261, K03260, K03259, K03258, K03257, K03256, K03255, K03254, K03253, K03252, K03251, K03250, K03249, K03248, K03247, K03246, K03245, K03244, K03243, K03242, K03241, K03240, K03239, K03238, K03237, K03236, K03235, K03234, K03233, K03232, K03231, K03230, K03229, K03228, K03227, K03226, K03225, K03224, K03223, K03222, K03221, K03220, K03219, K03218, K03217, K03216, K03215, K03214, K03213, K03212, K03211, K03210, K03209, K03208, K03207, K03206, K03205, K03204, K03203, K03202, K03201, K03200, K03199, K03198, K03197, K03196, K03195, K03194, K03193, K03192, K03191, K03190, K03189, K03188, K03187, K03186, K03185, K03184, K03183, K03182, K03181, K03180, K03179, K03178, K03177, K03176, K03175, K03174, K03173, K03172, K03171, K03170, K03169, K03168, K03167, K03166, K03165, K03164, K03163, K03162, K03161, K03160, K03159, K03158, K03157, K03156, K03155, K03154, K03153, K03152, K03151, K03150, K03149, K03148, K03147, K03146, K03145, K03144, K03143, K03142, K03141, K03140, K03139, K03138, K03137, K03136, K03135, K03134, K03133, K03132, K03131, K03130, K03129, K03128, K03127, K03126, K03125, K03124, K03123, K03122, K03121, K03120, K03119, K03118, K03117, K03116, K03115, K03114, K03113, K03112, K03111, K03110, K03109, K03108, K03107, K03106, K03105, K03104, K03103, K03102, K03101, K03100, K03099, K03098, K03097, K03096, K03095, K03094, K03093, K03092, K03091, K03090, K03089, K03088, K03087, K03086, K03085, K03084, K03083, K03082, K03081, K03080, K03079, K03078, K03077, K03076, K03075, K03074, K03073, K03072, K03071, K03070, K03069, K03068, K03067, K03066, K03065, K03064, K03063, K03062, K03061, K03060, K03059, K03058, K03057, K03056, K03055, K03054, K03053, K03052, K03051, K03050, K03049, K03048, K03047, K03046, K03045, K03044, K03043, K03042, K03041, K03040, K03039, K03038, K03037, K03036, K03035, K03034, K03033, K03032, K03031, K03030, K03029, K03028, K03027, K03026, K03025, K03024, K03023, K03022, K03021, K03020, K03019, K03018, K03017, K03016, K03015, K03014, K03013, K03012, K03011, K03010, K03009, K03008, K03007, K03006, K03005, K03004, K03003, K03002, K03001, K03000, K02999, K02998, K02997, K02996, K02995, K02994, K02993, K02992, K02991, K02990, K02989, K02988, K02987, K02986, K02985, K02984, K02983, K02982, K02981, K02980, K02979, K02978, K02977, K02976, K02975, K02974, K02973, K02972, K02971, K02970, K02969, K02968, K02967, K02966, K02965, K02964, K02963, K02962, K02961, K02960, K02959, K02958, K02957, K02956, K02955, K02954, K02953, K02952, K02951, K02950, K02949, K02948, K02947, K02946, K02945, K02944, K02943, K02942, K02941, K02940, K02939, K02938, K02937, K02936, K02935, K02934, K02933, K02932, K02931, K02930, K02929, K02928, K02927, K02926, K02925, K02924, K02923, K02922, K02921, K02920, K02919, K02918, K02917, K02916, K02915, K02914, K02913, K02912, K02911, K02910, K02909, K02908, K02907, K02906, K02905, K02904, K02903, K02902, K02901, K02900, K02899, K02898, K02897, K02896, K02895, K02894, K02893, K02892, K02891, K02890, K02889, K02888, K02887, K02886, K02885, K02884, K02883, K02882, K02881, K02880, K02879, K02878, K02877, K02876, K02875, K02874, K02873, K02872, K02871, K02870, K02869, K02868, K02867, K02866, K02865, K02864, K02863, K02862, K02861, K02860, K02859, K02858, K02857, K02856, K02855, K02854, K02853, K02852, K02851, K02850, K02849, K02848, K02847, K02846, K02845, K02844, K02843, K02842, K02841, K02840, K02839, K02838, K02837, K02836, K02835, K02834, K02833, K02832, K02831, K02830, K02829, K02828, K02827, K02826, K02825, K02824, K02823, K02822, K02821, K02820, K02819, K02818, K02817, K02816, K02815, K02814, K02813, K02812, K02811, K02810, K02809, K02808, K02807, K02806, K02805, K02804, K02803, K02802, K02801, K02800, K02799, K02798, K02797, K02796, K02795, K02794, K02793, K02792, K02791, K02790, K02789, K02788, K02787, K02786, K02785, K02784, K02783, K02782, K02781, K02780, K02779, K02778, K02777, K02776, K02775, K02774, K02773, K02772, K02771, K02770, K02769, K02768, K02767, K02766, K02765, K02764, K02763, K02762, K02761, K02760, K02759, K02758, K02757, K02756, K02755, K02754, K02753, K02752, K02751, K02750, K02749, K02748, K02747, K02746, K02745, K02744, K02743, K02742, K02741, K02740, K02739, K02738, K02737, K02736, K02735, K02734, K02733, K02732, K02731, K02730, K02729, K02728, K02727, K02726, K02725, K02724, K02723, K02722, K02721, K02720, K02719, K02718, K02717, K02716, K02715, K02714, K02713, K02712, K02711, K02710, K02709, K02708, K02707, K02706, K02705, K02704, K02703, K02702, K02701, K02700, K02699, K02698, K02697, K02696, K02695, K02694, K02693, K02692, K02691, K02690, K02689, K02688, K02687, K02686, K02685, K02684, K02683, K02682, K02681, K02680, K02679, K02678, K02677, K02676, K02675, K02674, K02673, K02672, K02671, K02670, K02669, K02668, K02667, K02666, K02665, K02664, K02663, K02662, K02661, K02660, K02659, K02658, K02657, K02656, K02655, K02654, K02653, K02652, K02651, K02650, K02649, K02648, K02647, K02646, K02645, K02644, K02643, K02642, K02641, K02640, K02639, K02638, K02637, K02636, K02635, K02634, K02633, K02632, K02631, K02630, K02629, K02628, K02627, K02626, K02625, K02624, K02623, K02622, K02621, K02620, K02619, K02618, K02617, K02616, K02615, K02614, K02613, K02612, K02611, K02610, K02609, K02608, K02607, K02606, K02605, K02604, K02603, K02602, K02601, K02600, K02599, K02598, K02597, K02596, K02595, K02594, K02593, K02592, K02591, K02590, K02589, K02588, K02587, K02586, K02585, K02584, K02583, K02582, K02581, K02580, K02579, K02578, K02577, K02576, K02575, K02574, K02573, K02572, K02571, K02570, K02569, K02568, K02567, K02566, K02565, K02564, K02563, K02562, K02561, K02560, K02559, K02558, K02557, K02556, K02555, K02554, K02553, K02552, K02551, K02550, K02549, K02548, K02547, K02546, K02545, K02544, K02543, K02542, K02541, K02540, K02539, K02538, K02537, K02536, K02535, K02534, K02533, K02532, K02531, K02530, K02529, K02528, K02527, K02526, K02525, K02524, K02523, K02522, K02521, K02520, K02519, K02518, K02517, K02516, K02515, K02514, K02513, K02512, K02511, K02510, K02509, K02508, K02507, K02506, K02505, K02504, K02503, K02502, K02501, K02500, K02499, K02498, K02497, K02496, K02495, K02494, K02493, K02492, K02491, K02490, K02489, K02488, K02487, K02486, K02485, K02484, K02483, K02482, K02481, K02480, K02479, K02478, K02477, K02476, K02475, K02474, K02473, K02472, K02471, K02470, K02469, K02468, K02467, K02466, K02465, K02464, K02463, K02462, K02461, K02460, K02459, K02458, K02457, K02456, K02455, K02454, K02453, K02452, K02451, K02450, K02449, K02448, K02447, K02446, K02445, K02444, K02443, K02442, K02441, K02440, K02439, K02438, K02437, K02436, K02433, K02432, K02431, K02430, K02429, K02428, K02427, K02426, K02425, K02424, K02423, K02422, K02421, K02420, K02419, K02418, K02417, K02416, K02415, K02414, K02413, K02412, K02411, K02410, K02409, K02408, K02407, K02406, K02405, K02403, K02402, K02401, K02400, K02399, K02398, K02397, K02396, K02395, K02394, K02393, K02392, K02391, K02390, K02389, K02388, K02387, K02386, K02385, K02384, K02383, K02382, K02381, K02380, K02379, K02378, K02377, K02376, K02375, K02374, K02373, K02372, K02371, K02370, K02369, K02368, K02367, K02366, K02365, K02364, K02330, K02329, K02328, K02327, K02326, K02325, K02323, K02322, K02321, K02320, K02319, K02318, K02317, K02316, K02315, K02313, K02312, K02311, K02309, K02308, K02307, K02306, K02305, K02304, K02303, K02302, K02301, K02300, K02299, K02298, K02297, K02296, K02295, K02294, K02292, K02290, K02289, K02288, K02287, K02286, K02285, K02284, K02282, K02281, K02279, K02277, K02275, K02273, K02272, K02271, K02270, K02269, K02268, K02267, K02266, K02265, K02264, K02263, K02262, K02261, K02260, K02259, K02258, K02257, K02256, K02255, K02254, K02253, K02252, K02251, K02250, K02249, K02222, K02221, K02220, K02219, K02218, K02217, K02216, K02215, K02214, K02213, K02212, K02211, K02210, K02209, K02208, K02207, K02206, K02205, K02204, K02203, K02202, K02201, K02200, K02199, K02198, K02197, K02196, K02195, K02194, K02193, K02192, K02191, K02190, K02188, K02187, K02186, K02185, K02184, K02183, K02182, K02181, K02180, K02179, K02178, K02177, K02176, K02175, K02174, K02173, K02172, K02171, K02170, K02169, K02168, K02167, K02166, K02165, K02164, K02163, K02162, K02161, K02160, K02159, K02158, K02157, K02156, K02155, K02154, K02153, K02152, K02151, K02150, K02149, K02148, K02147, K02146, K02145, K02144, K02143, K02142, K02141, K02140, K02139, K02138, K02137, K02136, K02135, K02134, K02133, K02132, K02131, K02130, K02129, K02128, K02127, K02126, K02125, K02124, K02123, K02122, K02121, K02120, K02119, K02118, K02117, K02116, K02115, K02114, K02011, K02010, K02009, K02008, K02007, K02006, K02005, K02004, K02003, K02002, K02001, K02000, K01999, K01998, K01997, K01996, K01995, K01994, K01993, K01992, K01991, K01990, K01989, K01988, K01987, K01986, K01985, K01984, K01983, K01982, K01981, K01980, K01979, K01973, K01937, K01936, K01935, K01934, K01933, K01931, K01930, K01929, K01928, K01927, K01926, K01925, K01924, K01923, K01922, K01921, K01920, K01919, K01918, K01917, K01916, K01915, K01914, K01913, K01912, K01911, K01910, K01909, K01908, K01907, K01906, K01905, K01904, K01903, K01902, K01901, K01900, K01899, K01898, K01897, K01896, K01895, K01894, K01893, K01892, K01859, K01858, K01857, K01856, K01855, K01854, K01853, K01852, K01814, K01813, K01812, K01811, K01810, K01809, K01808, K01807, K01806, K01805, K01804, K01803, K01802, K01801, K01800, K01799, K01798, K01797, K01796, K01795, K01794, K01793, K01792, K01791, K01790, K01789, K01788, K01787, K01786, K01785, K01784, K01783, K01782, K01781, K01780, K01779, K01778, K01777, K01776, K01775, K01774, K01773, K01772, K01771, K01770, K01769, K01768, K01767, K01766, K01765, K01738, K01737, K01736, K01735, K01734, K01709, K01708, K01707, K01706, K01705, K01704, K01703, K01702, K01701, K01700, K01699, K01698, K01697, K01696, K01695, K01694, K01693, K01692, K01691, K01690, K01689, K01688, K01687, K01686, K01685, K01684, K01683, K01682, K01681, K01680, K01679, K01678, K01677, K01676, K01675, K01674, K01673, K01672, K01671, K01670, K01669, K01668, K01667, K01666, K01665, K01664, K01663, K01662, K01661, K01660, K01659, K01658, K01657, K01656, K01655, K01654, K01653, K01652, K01651, K01650, K01649, K01648, K01647, K01646, K01645, K01644, K01643, K01642, K01641, K01640, K01639, K01638, K01613, K01612, K01611, K01610, K01609, K01608, K01607, K01606, K01605, K01604, K01603, K01601, K01600, K01599, K01598, K01597, K01596, K01595, K01594, K01593, K01592, K01591, K01590, K01589, K01588, K01587, K01586, K01585, K01584, K01583, K01582, K01581, K01580, K01579, K01578, K01577, K01576, K01575, K01574, K01573, K01572, K01571, K01570, K01569, K01568, K01567, K01566, K01565, K01564, K01563, K01531, K01530, K01529, K01528, K01527, K01526, K01525, K01524, K01523, K01522, K01521, K01520, K01519, K01518, K01517, K01516, K01515, K01514, K01513, K01512, K01511, K01510, K01509, K01508, K01507, K01506, K01505, K01504, K01503, K01502, K01501, K01500, K01499, K01498, K01497, K01496, K01495, K01494, K01493, K01492, K01491, K01490, K01486, K01485, K01484, K01483, K01482, K01481, K01480, K01479, K01478, K01477, K01476, K01475, K01474, K01473, K01472, K01471, K01470, K01469, K01468, K01467, K01466, K01465, K01464, K01463, K01462, K01461, K01460, K01459, K01458, K01457, K01456, K01455, K01454, K01453, K01452, K01451, K01450, K01449, K01448, K01447, K01445, K01443, K01442, K01441, K01440, K01439, K01438, K01437, K01436, K01435, K01434, K01433, K01432, K01431, K01430, K01429, K01428, K01427, K01426, K01425, K01424, K01423, K01422, K01421, K01420, K01419, K01418, K01417, K01416, K01415, K01414, K01413, K01412, K01411, K01410, K01409, K01408, K01407, K01406, K01405, K01404, K01403, K01402, K01401, K01400, K01399, K01398, K01397, K01396, K01395, K01394, K01393, K01392, K01391, K01390, K01389, K01388, K01387, K01385, K01384, K01383, K01382, K01381, K01380, K01379, K01378, K01377, K01376, K01375, K01374, K01373, K01372, K01371, K01370, K01369, K01368, K01367, K01366, K01365, K01364, K01363, K01362, K01361, K01360, K01359, K01358, K01357, K01356, K01355, K01354, K01353, K01352, K01351, K01350, K01349, K01348, K01347, K01346, K01345, K01344, K01343, K01342, K01341, K01340, K01339, K01338, K01337, K01336, K01335, K01334, K01333, K01332, K01331, K01330, K01329, K01328, K01327, K01326, K01325, K01324, K01323, K01322, K01321, K01320, K01318, K01317, K01316, K01315, K01314, K01313, K01312, K01311, K01310, K01309, K01308, K01307, K01306, K01305, K01304, K01303, K01302, K01301, K01300, K01299, K01298, K01297, K01296, K01295, K01294, K01293, K01292, K01291, K01290, K01289, K01288, K01287, K01286, K01285, K01284, K01283, K01282, K01281, K01280, K01279, K01278, K01277, K01276, K01275, K01274, K01273, K01272, K01271, K01270, K01269, K01268, K01267, K01266, K01265, K01264, K01263, K01262, K01261, K01260, K01259, K01258, K01257, K01256, K01255, K01254, K01253, K01252, K01251, K01250, K01249, K01248, K01247, K01246, K01245, K01244, K01243, K01242, K01241, K01240, K01239, K01238, K01237, K01236, K01235, K01234, K01233, K01232, K01231, K01230, K01229, K01228, K01227, K01226, K01225, K01224, K01223, K01222, K01221, K01220, K01219, K01218, K01217, K01216, K01215, K01214, K01213, K01212, K01211, K01210, K01209, K01208, K01207, K01206, K01205, K01204, K01203, K01201, K01197, K01196, K01195, K01194, K01193, K01192, K01191, K01190, K01189, K01188, K01187, K01186, K01185, K01184, K01183, K01182, K01181, K01180, K01179, K01178, K01177, K01176, K01175, K01174, K01173, K01172, K01171, K01170, K01169, K01168, K01167, K01166, K01165, K01164, K01163, K01162, K01161, K01160, K01159, K01158, K01157, K01156, K01155, K01154, K01153, K01152, K01151, K01150, K01149, K01148, K01147, K01146, K01145, K01144, K01143, K01142, K01141, K01140, K01139, K01138, K01137, K01136, K01135, K01134, K01133, K01132, K01131, K01130, K01129, K01128, K01127, K01126, K01125, K01124, K01123, K01122, K01121, K01120, K01119, K01118, K01117, K01116, K01115, K01114, K01113, K01112, K01111, K01110, K01109, K01108, K01107, K01106, K01105, K01104, K01103, K01102, K01101, K01100, K01099, K01098, K01097, K01096, K01095, K01094, K01093, K01092, K01091, K01090, K01089, K01088, K01087, K01086, K01085, K01084, K01083, K01082, K01081, K01080, K01079, K01078, K01077, K01076, K01075, K01074, K01073, K01072, K01071, K01070, K01069, K01068, K01067, K01066, K01065, K01064, K01063, K01062, K01061, K01060, K01059, K01058, K01057, K01056, K01055, K01054, K01053, K01052, K01051, K01050, K01049, K01048, K01047, K01046, K01045, K01044, K01043, K01042, K01041, K01040, K01039, K01038, K01037, K01036, K01035, K01034, K01033, K01032, K01031, K01030, K01029, K01028, K01027, K01026, K01025, K01024, K01022, K01021, K01020, K01019, K01018, K01017, K01016, K01013, K00064, K00063, K00062, K00061, K00060, K00058, K00056, K00055, K00054, K00053, K00052, K00051, K00050, K00049, K00048, K00047, K00046, K00045, K00044, K00043, K00042, K00041, K00040, K00039, K00038, K00037, K00036, K00035, K00034, K00033, K00032, K00031, K00030, K00029, K00005, K00004, and K00000.
FEI Number 1831750
Recalling Firm/
Manufacturer
Stryker Medical Division of Stryker Corporation
3800 East Centre Ave.
Portage MI 49002
For Additional Information ContactRenata Sila
269-324-6609
Manufacturer Reason
for Recall
An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac
FDA Determined
Cause 2
Component design/selection
ActionThe firm, Stryker Medical, sent an "URGENT- Medical Device Correction" letter dated May 17, 2012 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to observe the beds for any malfunctions according to Preventive Maintenance Checklists in maintenance manual; report scale or bed exit malfunctions to Stryker (1-800-327-0770, option 2) to arrange for field service or to order service parts, at no charge. The customers were also instructed to forward this notice and advise Stryker of any subaccounts if the product was further distributed. A postage paid return postcard is to be filled for tracking and to make arrangements for repair. If you have any urgent questions or concerns, call 269-389-6466, Monday-Friday 8a.m-5p.m. (ET).
Quantity in Commerce2546
DistributionWorldwide distribution:USA (nationwide) and countries including: Argentina, Bertec, Brasil, Brazil, Canada, Chile, China, India, Latin America, Mexico,Osteonics SA, Polska, Portugal, Puerto Rico, and Singapore.