• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall UniCel DxH Slidemaker Stainer Stainer Floor Cabinet

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall UniCel DxH Slidemaker Stainer Stainer Floor Cabinetsee related information
Date Initiated by FirmDecember 13, 2011
Date PostedAugust 20, 2012
Recall Status1 Terminated 3 on August 20, 2012
Recall NumberZ-2226-2012
Recall Event ID 62755
Product Classification Slide stainer, automated - Product Code KPA
ProductUniCel DxH Slidemaker Stainer Stainer Floor Cabinet, PN A96152. The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole blood-sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear. The DxH Slidemaker Stainer allows for adaptation of the smear appearance and stain methodology according to user preferences. Blood smears produced by the Slidemaker portion of the DxH Slidemaker Stainer are moved to baskets for transfer to the Stainer portion by a robot arm. Microscopic examination of the stained blood smears can be used to help determine the hematologic status of a patient. A stained blood film: " Allows for the differentiation of white blood cells " Facilitates the characterization of red blood cells and platelets " Aids in the identification of blood components and cellular abnormalities The optional floor stand has four drawers for storing the reagent and waste containers currently in use in the DxH Slidemaker Stainer.
Code Information Part Number(s) 775222 and A96152; Serial #s-MP010, MP023, MP026, AS23001, AS23002, AS25003, AS25004, AS25005, AS33008, AS33009, AS33010, AS33011, AS37013, AS38014, AS38015, AS39016, AS39017, AS40018, AS44012, AT04001, AT04002, AT04003, AT04004, AT04005, AT22010, AT22011, AT22012, AT35013, AT35014, AT35015, AT35016, AT35017, AT35018, AT40019, AT40020, AT40021, AT40022, AT40023, AT40024, AT40025, AT40026, AT40027, AT40028, AT40029, AT40030, AT45031, AT45032, AT45033, AT45034, AT45035 
FEI Number 2050012
Recalling Firm/
Manufacturer
Beckman Coulter Inc.
250 S Kraemer Blvd
Brea CA 92821-6232
For Additional Information ContactClair K. O'Donovan, Ph.D.
714-961-4483
Manufacturer Reason
for Recall
The recall was initiated because Beckman Coulter has confirmed that the DxH Slidemaker Stainer Floor Stand shipped with Slidemaker Stainer instruments were missing vinyl caps, exposing sharpt edges on the DxH Slidemaker Stainer Reagent Cabinet (Floor Stand) drawer slides. The sharp edges on the drawer slides may expose operators to physical injury when replacing reagents and waste containers whic
FDA Determined
Cause 2
Process change control
ActionBeckman Coulter initiated a "PRODUCT CORRECTIVE ACTION (PCA)" which was communicated via phone calls starting the week of December 19, 2011 to all customers who purchased UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System. The phone calls provided the customers with an explanation of the problem identified, potential impact should the problem occur and an action to be taken. Customers were instructed to share the PCA information with their laboratory staff and if they had any questions, they should contact Beckman Coulter at (800) 526-7694.
Quantity in Commerce41 units (15 units in the US)
DistributionWorldwide Distribution-USA (nationwide) including the states GA, NY, IL, TX, WA, PA, IN, MS, NC, and MO and the countries of Belgium, France, Hong Kong, Italy, Spain, and Switzerland.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-