| Class 1 Device Recall NIM TriVantage EMG Endotracheal Tube | |
Date Initiated by Firm | March 04, 2013 |
Date Posted | June 04, 2013 |
Recall Status1 |
Terminated 3 on January 30, 2014 |
Recall Number | Z-1420-2013 |
Recall Event ID |
64631 |
510(K)Number | K112686 |
Product Classification |
Stimulator, nerve - Product Code ETN
|
Product | REF 8229705 x1 NIM TriVantage EMG Endotracheal Tube 5.0 mm I.D. x 6.5 mm O.D. Medtronic Xomed, Inc., 6743 Southpoint Drive North, Jacksonville, FL 32216-0980 USA STERILE Rx Only
The electrodes are designed to make contact with the patient's vocal cords to facilitate electromyographic (EMG) monitoring of the laryngeal musculature during surgery when connected to a multi-channel EMG neuromonitoring device. |
Code Information |
Product Catalog Numbers: REF 8229705, 8229706, 8229707, 8229708, 8229709, 8229735, 8229736, 8229737, 8229738, 8229739. Lot Numbers: 205830052 to 206486732. 0206516104, 0206516105, 0206516106, 0206516108, 0206520224, 0206520225, 0206520226, 0206520227, 0206520228, 0206520358, 0206542163, 0206545356, 0206545502. |
Recalling Firm/ Manufacturer |
Medtronic Xomed, Inc. 6743 Southpoint Dr N Jacksonville FL 32216-6218
|
For Additional Information Contact | Gabriela Anchondo 904-279-7550 |
Manufacturer Reason for Recall | In March 2013, Medtronic issued a recall of the NIM TriVantage EMG Endotracheal Tube due to cuff leakage in the intraoperative cuff deflation which then requires re-inflation or replacement of the deflated tube in order to ensure the continued ventilation of the patient. |
FDA Determined Cause 2 | Use error |
Action | The firm, Medtronic Xomed, Inc. sent an "URGENT Product Recall Notification" letter dated March 14, 2013, to its U.S. and International customers. The letter described the product, problem and actions to be taken. The customers were instructed to return any affected devices still in their possession per the following procedure: 1. Check inventories for the affected products listed on the enclosed checklist. 2. Fill-in the "quantity on-hand" column on the checklist. 3. Fax checklist to Medtronic ENT at 1-904-296-2386. 4. If you have any of the subject product, contact Medtronic ENT Customer Service at 1-800-874-5797 to arrange for returns and replacement or credit. 5.When returning products, clearly mark the outside of the container Returned Goods Authorization (RGA) number.
If you have any questions regarding the subject action or the content of the letter, contact Gabriela Anchondo at 904-279-7550. |
Quantity in Commerce | 2,591 units |
Distribution | Worldwide distribution: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MA, MD, MI, MN, MT, NC, NH, NJ, NY, OH, PA, RI, SC, TX, UT, VA, and WA; and countries of : Australia, Canada, Germany, Italy, Norway, Poland and United Kingdom. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|