• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall LIGACLIP Endoscopic Rotating Multiple Clip Applier

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall LIGACLIP Endoscopic Rotating Multiple Clip Appliersee related information
Date Initiated by FirmApril 12, 2013
Date PostedJuly 19, 2013
Recall Status1 Terminated 3 on May 01, 2014
Recall NumberZ-1775-2013
Recall Event ID 65141
510(K)NumberK812291 K830503 K864102 
Product Classification Applier, surgical, clip - Product Code GDO
ProductLIGACLIP Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L Product Usage: The LIGACLIP Endoscopic Rotating Multiple Clip Applier is a sterile, single patient use instrument designed to provide a means of ligation through ENDOPATH surgical trocars. The instrument delivers 20 titanium clips that individually advance after each firing. The shaft is made of a low glare material that minimizes reflective distortion. It is designed to rotate 360 in either direction. The rotating knob is located to allow for a one-handed technique.
Code Information H44Z7W, 37ER320, 01/03/12 to 11/15/16; J4A016, 37ER320, 01/05/12 to 12/15/16; J4A11G, 37ER320, 01/10/12 to 12/15/16; J4A124, 37ER320, 01/10/12 to 12/15/16; J4A145 37ER320, 01/11/12 to 12/15/16  J4A182, 37ER320, 01/13/12 to 12/15/16; J4A21T, 37ER320, 01/27/12 to 12/15/16; J4A28D, 37ER320, 01/19/12 to 12/15/16; J4A28U, 37ER320, 01/19/12 to 12/15/16; J4A335, 37ER320, 01/23/12 to 12/15/16; J4A38L, 37ER320, 01/25/12 to 12/15/16; J4A39D, 37ER320, 01/26/12 to 12/15/16; J4A406, 37ER320, 01/26/12 to 12/15/16; J4A40P, 37ER320, 02/07/12 to 12/15/16; J4A45T, 37ER320, 01/29/12 to 12/15/16; J4A461, 37ER320, 01/30/12 to 12/15/16; J4A46T, 37ER320, 01/30/12 to 12/15/16; J4A48U,), 37ER320, 02/02/12 to 12/15/16; J4A49L, 37ER320, 02/01/12 to 12/15/16; J4A525, 37ER320, 02/02/12 to 12/15/16; J4A535, 37ER320, 02/02/12 to 12/15/16; J4A559, 41XBB48B 02/02/12 to 07/15/16; J4A60K, 37ER320, 02/08/12 to 01/15/17; J4A60W, 37ER320, 02/08/12 to 01/15/17; J4A61Z, 37ER320, 02/08/12 to 01/15/17; J4A62E, 37ER320, 02/08/12 to 01/15/17; J4A62U, 37ER320, 02/09/12 to 01/15/17; J4A71W, 37ER320, 02/15/12 to 01/15/17; J4A729, 37ER320, 02/15/12 to 01/15/17; J4A73H, 37ER320, 02/15/12 to 01/15/17; J4A743, 37ER320, 02/16/12 to 01/15/17; J4A74T, 37ER320, 02/16/12 to 01/15/17; J4A755, 37ER320, 02/16/12 to 01/15/17; J4A75P, 37ER320, 02/16/12 to 01/15/17; J4A86E, 37ER320, 02/24/12 to 01/15/17; J4A871, 37ER320, 02/25/12 to 01/15/17; J4A87L, 37ER320, 02/25/12 to 01/15/17; J4A889, 37ER320, 02/25/12 to 01/15/17; J4A92X, 41XBB48B 02/28/12 to 08/15/16; J4A97H, 37ER320, 03/02/12 to 01/15/17; J4A98T, 37ER320, 03/02/12 to 01/15/17; J4A997, 37ER320, 03/03/12 to 01/15/17; J4A99V, 37ER320, 03/03/12 to 02/15/17; J4AAOM, 37ER320, 03/03/12 to 02/15/17; J4AA1H, 37ER320, 03/05/12 to 02/15/17; J4AA71, 37ER320, 03/07/12 to 02/15/17; J4AA7V, 37ER320, 03/08/12 to 02/15/17; J4AA84, 37ER320, 03/08/12 to 02/15/17; J4AA9D, 37ER320, 03/08/12 to 02/15/17; J4AC2X, 37ER320, 03/10/12 to 02/15/17; J4AD09, 37ER320, 03/15/12 to 02/15/17; J4AD1A, 37ER320, 03/15/12 to 02/15/17; J4AD2G, 37ER320, 03/16/12 to 02/15/17; J4AE3W, 37ER320, 03/22/12 to 02/15/17; J4AE49, 37ER320, 03/22/12 to 02/15/17; J4AE57, 37ER320, 03/23/12 to 02/15/17; J4AE63, 37ER320, 03/23/12 to 02/15/17; J4AE6R, 37ER320, 04/13/12 to 02/15/17; J4AE7E, 37ER320, 04/11/12 to 02/15/17; J4AE86, 37ER320, 03/24/12 to 02/15/17; J4AE8L, 37ER320, 03/24/12 to 02/15/17; J4AF09, 41XBB48B, 03/27/12 to 10/15/16 84; J4AF15, 37ER320, 03/28/12 to 02/15/17; J4AF1N, 37ER320, 03/28/12 to 02/15/17; J4AF2G, 37ER320, 03/28/12 to 02/15/17; J4AF3K, 37ER320, 03/28/12" 02/15/17; J4AF48, 37ER320, 03/29/12 to 02/15/17; J4AF4J, 37ER320, 03/29/12 to 02/15/17; J4AGOV, 37ER320, 04/03/12 to 02/15/17; J4AG4V, 37ER320, 04/05/12 to 03/15/17; J4AG91, 37ER320, 04/11/12 to 03/15/17; J4AG9G, 37ER320, 04/18/12 to 03/15/17; J4AHOV, 37ER320, 04/12/12 to 03/15/17; J4AH8N, 37ER320, 04/15/12 to 03/15/17; J4AH96, 37ER320, 04/16/12 to 03/15/17; J4AH9C, 37ER320, 04/16/12 to 03/15/17; J4AH9R, 37ER320, 04/16/12 to 03/15/17; J4AJ3A, 37ER320, 04/18/12 to 03/15/17; J4AJ3T, 37ER320, 04/18/12 to 03/15/17; J4AJ6P, 37ER320, 04/20/12 to 03/15/17; J4AJ7M, 37ER320, 04/20/12 to 03/15/17; J4AKOP, 37ER320, 04/21/12 to 03/15/17; J4AK14, 37ER320, 04/22/12 to 03/15/17; J4AK1H, 37ER320, 04/23/12 to 03/15/17; J4AK1R, 37ER320, 04/23/12 to 03/15/17; J4AK2V, 37ER320, 04/24/12 to 03/15/17; J4AK35, 37ER320, 04/24/12 to 03/15/17; J4AK40, 37ER320, 04/24/12 to 03/15/17; J4AK5U, 37ER320, 04/25/12 to 03/15/17; J4AK67, 37ER320, 04/25/12 to 03/15/17; 2136; J4AK7D, 37ER320, 04/26/12 to 03/15/17; J4AK8C, 37ER320, 04/26/12 to 03/15/17; 1152; 4AK9R, 37ER320, 04/28/12 to 03/15/17; J4ALOU, 37ER320, 04/27/12 to 03/15/17; J4AL65, 37ER320, 05/02/12 to 03/15/17; J4AL6P, 37ER320, 05/02/12 to 03/15/17; 2172; J4AL87, 37ER320, 05/11/12 to 03/15/17; 2244; J4AL90, 37ER320, 05/03/12 to 04/15/17; J4AL9H, 37ER320, 05/03/12 to 04/15/17; J4AM11, 37ER320, 05/03/12 to 04/15/17; J4AM1K, 37ER320, 05/04/12 to 04/15/17; J4AM2E, 37ER320, 05/04/12 to 04/15/17; J4AM32, 37ER320, 05/04/12 to 04/15/17; J4AM3G, 37ER320, 05/05/12 to 04/15/17; J4AM45, 37ER320, 05/05/12 to 04/15/17; J4AM7R, 37ER320, 05/08/12 to 04/15/17; J4AM7W, 37ER320, 05/08/12 to 04/15/17; J4AM8H, 37ER320, 05/09/12 to 04/15/17; J4AM9J, 37ER320, 05/09/12 to 04/15/17; J4ANOC, 37ER320, 05/09/12 to 04/15/17; J4AN14, 37ER320, 05/09/12 to 04/15/17; J4AN1P, 37ER320, 05/10/12 to 04/15/17; J4AN1X, 37ER320, 05/10/12 to 04/15/17; J4AN2K, 37ER320, 05/10/12 to 04/15/17; J4AN35, 37ER320, 05/10/12 to 04/15/17; J4AR7E, 41XBB48B 05/24/12 to 01/15/17; J4AT2G, 37ER320, 06/02/12 to 04/15/17; 1152; J4AT2H, 37ER320, 05/31/12 to 04/15/17; J4AUOH, 37ER320, 06/04/12 to 04/15/17; J4AU23, 37ER320, 06/04/12 to 04/15/17; J4AU29, 37ER320, 06/05/12 to 05/15/17; J4AU2V, 37ER320, 06/05/12 to 05/15/17; J4AV5J, 37ER320, 06/12/12 to 05/15/17; J4AV5T, 37ER320, 06/12/12 to 05/15/17; J4AW02, 37ER320, 06/13/12 to 05/15/17; J4AWOG, 37ER320, 06/14/12 to 05/15/17; J4AW7D, 37ER320, 06/18/12 to 05/15/17; J4AW7N, 37ER320, 06/18/12 to 05/15/17; J4AW86, 37ER320, 06/19/12 to 05/15/17; J4AX78, 37ER320, 06/23/12 to 05/15/17; J4AX7M, 37ER320, 06/23/12 to 05/15/17; J4AX84, 37ER320, 06/24/12 to 05/15/17; J4AX8M, 37ER320, 06/24/12 to 05/15/17; J4AY22, 37ER320, 06/26/12 to 05/15/17; J4AY23, 37ER320, 06/27/12 to 05/15/17; J4AY24, 37ER320, 06/27/12 to 05/15/17;  J4AY26, 37ER320, 06/27/12 to 05/15/17; J4AY7E, 37ER320, 06/29/12 to 05/15/17; J4AY8R, 37ER320, 06/29/12 to 05/15/17; J4AY9D, 37ER320, 07/09/12 to 05/15/17; J4AZOL, 37ER320, 07/02/12 to 05/15/17; J4AZ1X, 37ER320, 07/11/12 to 05/15/17; J4AZ2H, 37ER320, 07/03/12 to 06/15/17; J4AZ6R, 37ER320, 07/06/12 to 06/15/17; J4AZ7K, 37ER320, 07/06/12 to 06/15/17; J4C04N, 37ER320, 07/11/12 to 06/15/17; J4C059, 37ER320, 07/12/12 to 06/15/17; J4C08U, 37ER320, 07/13/12 to 06/15/17; J4C09E, 37ER320, 07/14/12 to 06/15/17; J4C12Y, 37ER320, 07/16/12 to 06/15/17; J4C14F, 37ER320, 07/18/12 to 06/15/17; J4C16V, 37ER320, 07/18/12 to 06/15/17; J4C216, 37ER320, 07/23/12 to 06/15/17; J4C21H, 37ER320, 07/24/12 to 06/15/17; J4C21N, 37ER320, 07/24/12 to 06/15/17; J4C22M, 37ER320, 07/24/12 to 06/15/17; J4C30T, 37ER320, 07/27/12 to 06/15/17; J4C31J, 37ER320, 07/28/12 to 06/15/17; J4C32M, 37ER320, 08/21/12 to 06/15/17; J4C332, 37ER320, 07/30/12 to 06/15/17; J4C399, 37ER320, 08/03/12 to 06/15/17; J4C41U, 37ER320, 08/06/12 to 07/15/17; J4C42C, 37ER320, 08/04/12 to 07/15/17; J4C42X, 37ER320, 08/03/12 to 07/15/17; J4C45J, 37ER320, 08/07/12 to 07/15/17; J4C46Y, 37ER320, 08/07/12 to 07/15/17; J4C47A, 37ER320, 08/07/12 to 07/15/17; J4C489, 37ER320, 08/08/12 to 07/15/17; J4C49Y, 37ER320, 08/08/12 to 07/15/17; J4C508, 37ER320, 08/09/12 to 07/15/17; J4C510, 37ER320, 08/09/12 to 07/15/17; J4C57X, 37ER320, 08/14/12 to 07/15/17; J4C59E, 37ER320, 08/14/12 to 07/15/17; J4C606, 37ER320, 08/15/12 to 07/15/17; J4C65Z, 37ER320, 08/18/12 to 07/15/17; J4C66P, 37ER320, 08/20/12 to 07/15/17; J4C678, 37ER320, 08/21/12 to 07/15/17; J4C67K, 37ER320, 08/21/12 to 07/15/17; J4C762, 37ER320, 08/24/12 to 07/15/17; J4C765, 37ER320, 08/24/12 to 07/15/17; J4C76T, 37ER320, 08/24/12 to 07/15/17; J4C84K, 37ER320, 08/30/12 to 07/15/17; J4C84Z, 37ER320, 08/30/12 to 07/15/17; J4C90P, 37ER320, 09/05/12 to 07/15/17; J4C91H, 37ER320, 09/02/12 to 07/15/17; J4C91W, 37ER320, 09/05/12 to 07/15/17; J4C92P, 37ER320, 09/05/12 to 07/15/17; J4C96L, 37ER320, 09/06/12 to 08/15/17; J4C97D, 37ER320, 09/06/12 to 08/15/17; J4C97Z, 37ER320, 09/07/12 to 08/15/17; J4C98X, 37ER320, 09/07/12 to 08/15/17; J4CA3A, 37ER320, 09/11/12 to 08/15/17; J4CA4L, 37ER320, 09/11/12 to 08/15/17; J4CA62, 37ER320, 09/12/12 to 08/15/17; J4CA6K, 37ER320, 09/12/12 to 08/15/17; J4CA7F, 37ER320, 09/13/12 to 08/15/17; J4CA7J, 37ER320, 09/13/12 to 08/15/17; J4CA96, 37ER320, 09/13/12 to 08/15/17; J4CA9V, 37ER320, 09/13/12 to 08/15/17; J4CC63, 37ER320, 09/19/12 to 08/15/17; J4CD25, 42XBB48B 09/21/12 to 02/15/17; J4CD4R, 37ER320, 09/21/12 to 08/15/17; J4CD5H, 37ER320, 09/22/12 to 08/15/17; J4CD6D, 42XBB48B 09/25/12 to 02/15/17; J4CD6R, 37ER320, 09/25/12 to 08/15/17; J4CD8F, 42XBB48B 09/27/12 to 02/15/17; J4CE05, 37ER320, 09/26/12 to 08/15/17; J4CEOA, 37ER320, 09/26/12 to 08/15/17; J4CE1U, 37ER320, 09/26/12 to 08/15/17; J4CF6M, 37ER320, 10/05/12 to 09/15/17; J4CF6P, 37ER320, 10/05/12 to 09/15/17; J4CF7E, 37ER320, 10/05/12 to 09/15/17; J4CF7R, 37ER320, 10/06/12 to 09/15/17; J4CF80, 37ER320, 10/06/12 to 09/15/17; J4CF8X, 37ER320, 10/08/12 to 09/15/17; J4CGOP, 37ER320, 10/09/12 to 09/15/17; J4CG17, 37ER320, 10/10/12 to 09/15/17; J4CG1U, 37ER320, 10/10/12 to 09/15/17; J4CG2U, 37ER320, 10/10/12 to 09/15/17; J4CG37, 37ER320, 10/10/12 to 09/15/17; J4CG4L, 37ER320, 10/11/12 to 09/15/17; J4CG66, 37ER320, 10/11/12 to 09/15/17; J4CG8R, 37ER320, 10/16/12 to 09/15/17; J4CG8X, 37ER320, 10/16/12 to 09/15/17; J4CH10, 37ER320, 10/15/12 to 09/15/17; J4CH4U, 37ER320, 10/17;/12 to 09/15/17; J4CH4V, 37ER320, 11/16/12 to 09/15/17; J4CH97, 37ER320, 10/20/12 to 09/15/17; J4CH9R, 37ER320, 10/20/12 to 09/15/17; J4CJ17, 37ER320, 10/20/12 to 09/15/17; J4CJ61, 37ER320, 10/25/12 to 09/15/17; J4CJ6L, 37ER320, 10/25/12 to 09/15/17; J4CKOV, 37ER320, 10/27/12 to 09/15/17; J4CK16, 37ER320, 10/29/12 to 09/15/17; J4CK3L, 37ER320, 10/30/12 to 09/15/17; J4CK9D, 37ER320, 11/02/12 to 09/15/17; J4CK9J, 37ER320, 11/02/12 to 09/15/17; J4CL23, 37ER320, 11/03/12 to 10/15/17; J4CL3F, 37ER320, 11/06/12 to 10/15/17; J4CL3U, 37ER320, 11/06/12 to 10/15/17; J4CL49, 37ER320, 11/06/12 to 10/15/17; J4CL8D, 37ER320, 11/08/12 to 10/15/17; J4CL9A, 37ER320, 11/08/12 to 10/15/17; J4CL9Z, 37ER320, 11/17;/12 to 10/15/17; J4CN1C, 37ER320, 11/15/12 to 10/15/17; J4CN9T, 37ER320, 11/20/12 to 10/15/17; J4CP1P, 37ER320, 11/27/12 to 10/15/17; J4CP1Y, 37ER320, 11/27/12 to 10/15/17; J4CP4C, 37ER320, 11/27/12 to 10/15/17; J4CP4L, 37ER320, 11/27/12 to 10/15/17; J4CP8U, 37ER320, 11/29/12 to 10/15/17; J4CR1C, 37ER320, 11/30/12 to 10/15/17; J4CR3L, 37ER320, 12/04/12 to 10/15/17; J4CR44, 37ER320, 12/04/12 to 10/15/17; J4CR6Y, 37ER320, 12/04/12 to 11/15/17; J4CR7X, 37ER320, 12/07/12 to 11/15/17; J4CT1U, 37ER320, 12/07/12 to 11/15/17; J4CT27, 37ER320, 12/09/12 to 11/15/17; J4CTSR, 37ER320, 12/11/12 to 11/15/17; J4CU1U, 37ER320, 12/12/12 to 11/15/17; J4CU27, 37ER320, 12/12/12 to 11/15/17; J4CU3K, 37ER320, 12/14/12 to 11/15/17; J4CU68, 37ER320, 12/15/12 to 11/15/17; J4CV2F, 37ER320, 12/20/12 to 11/15/17; J4CV48, 37ER320, 12/20/12 to 11/15/17; J4CV4J, 37ER320, 12/21/12 to 11/15/17; J4CV4U, 37ER320, 12/21/12 to 11/15/17; J4CV58, 37ER320, 12/21/12 to 11/15/17; J9002K, 41XCB57S 01/03/12 to 12/15/16; J9002L, 41XCD52S 01/03/12 to 12/15/16; J9003P, 02CK017 01/05/12 to 12/31/99; J90058, 41XCC51S 01/06/12 to 12/15/16; J90078, 41XCDS15 01/05/12 to 12/15/16; J9012T, 41KNC61XL 01/06/12 to 12/15/16; J9012U, 41XCD50S 01/06/12 to 12/15/16; J9012V, 41FNC42XL, 01/06/12 to 12/15/16; J90143, 41XCB52S 01/09/12 to 12/15/16; J90145, 41XCB57S 01/09/12 to 12/15/16; J90146, 41FNC41XL 01/09/12 to 12/15/16; J9014N, 02CK017 01/06/12 to 12/31/99; J9016G, 41XCB52S 01/09/12 to 12/15/16; J9016H, 41FNC43XL 01/09/12 to 12/15/16; J9016J 41TNC49XL 01/10/12 to 12/15/16; J9016K 41KNC60XL 01/10/12 to 12/15/16; J9020X 41XCD51S 01/10/12 to 12/15/16; J9022W 02CK017 01/10/12 to 12/31/99; J90249 41KBC17XL 01/12/12 to 12/15/16; J9026C 41XCD51S 01/11/12 to 12/15/16; J9026D 41XCC51S 01/12/12 to 12/15/16; J9026H 02CK017 01/11/12 to 12/31/99; J9028G 41XCB57S 01/13/12 to 12/15/16; J9032A 41XCD50S 01/13/12 to 12/15/16; J9032D 41KBC17XL 01/13/12 to 12/15/16; J9033P 02CK017 01/13/12 to 12/31/99; J90379 41XCC51S 01/16/12 to 12/15/16; J9037A 41FNC42XL, 01/16/12 to 12/15/16; J9038A 41KBC17XL 01/17;/12 to 12/15/16; J90390 41KNC61XL 01/17;/12 to 12/15/16; J9042C 02CK017 01/17;/12 to 12/31/99; J9044R 41XCD51S 01/18/12 to 12/15/16; J9044T 41FNC42XL, 01/18/12 to 12/15/16; J9047G 41XCD51S 01/19/12 to 12/15/16; J9047H 41FNC41XL 01/19/12 to 12/15/16; J9051P 41XCD51S 01/20/12 to 12/15/16; J9051R 41XCC51S 01/20/12 to 12/15/16; J9051T 41XCC50S 01/20/12 to 12/15/16; J9051U 41FNC41XLD 01/20/12 to 12/15/16; J90574, 41XCB57S 01/23/12 to 12/15/16; J90575, 41TNC49XL 01/24/12 to 12/15/16; J90598, 41FNC42XL, 01/24/12 to 12/15/16; J9059A, 41XCD51S 01/25/12 to 12/15/16; J90643, 41XCD52S 01/26/12 to 12/15/16; J90644, 41KNC61XL 01/26/12 to 12/15/16; J90646, 41XCC53S 01/27/12 to 12/15/16; J9066Z, 41XCD51S 01/27/12 to 12/15/16; J90671, 41KNC60XL 01/27/12 to 12/15/16; J9068K, 02CK017 01/26/12 to 12/31/99; J90743, 41XCD51S 01/30/12 to 12/15/16; J9074T, 41TNC69XL 01/31/12 to 12/15/16; J9074X, 41FNC41XL 02/01/12 to 12/15/16; J90771, 41XCB52S, 01/31/12 to 12/15/16; J9080P, 41XCD51S, 02/01/12 to 12/15/16; J9080R, 41FNC42XL, 02/01/12 to 12/15/16; J90823, 41XCD52S, 02/01/12 to 12/15/16; J90824, 41XCC51S, 02/01/12 to 12/15/16; J90825, 41XCB57S, 02/01/12 to 12/15/16; J9082G, 02CK017, 02/01/12 to 12/31/99; J9084F, 41FNC43XL, 02/02/12 to 12/15/16; J90867, 41FNC41XL, 02/03/12 to 12/15/16; J90881, 41KNC09XL, 02/03/12 to 12/15/16; J90883, 41KNC60XL, 02/03/12 to 12/15/16; J9090A, 41XCD50S, 02/03/12 to 12/15/16; J9091Z, 02CK017, 02/03/12 to 12/31/99; J90931, 41XCD51S, 02/07/12 to 01/15/17; J90948, 41XCC53S, 02/07/12 to 01/15/17; J90949, 41TNC49XL, 02/07/12 to 01/15/17; J90968, 41FNC41XLD, 02/08/12 to 01/15/17; J90969, 41KNC07XL, 02/09/12 to 01/15/17; J9098C, 41TNC36XL, 02/09/12 to 01/15/17; J90993, 02CK017, 02/08/12 to 12/31/99; J90C1W, 41XCC51S, 02/14/12 to 01/15/17; J90C1X, 41XCD51S, 02/14/12 to 01/15/17; J90C1Y, 41XCD52S, 02/14/12 to 01/15/17; J90C1Z, 41XCC50S, 02/14/12 to 01/15/17; J90C67, 41XCD51S, 02/15/12 to 01/15/17; J90C8C, 02CK017, 02/15/12 to 12/31/99; J90DOL, 41XCD51S, 02/16/12 to 01/15/17; J9003N, 02CK017, 02/17/12 to 12/31/99; J9004Z, 41FNC43XL, 02/17/12 to 01/15/17; J90050, 41XCC51S, 02/17/12 to 01/15/17; J90051, 41FNC41XL, 02/17/12 to 01/15/17; J9007A, 41XCD51S, 02/17/12 to 01/15/17; J9008J, 41XCD51S, 02/21/12 to 01/15/17; J9008K, 41TNC36XL, 02/21/12 to 01/15/17; J9008L, 41KNC60XL, 02/21/12 to 01/15/17; J9009H, 41FNC41XLD, 02/20/12 to 01/15/17; J9009K, 02CK017, 02/20/12 to 12/31/99; J90E2R, 41XCD51S, 02/21/12 to 01/15/17; J90E2T, 41XCB57S, 02/21/12 to 01/15/17; J90E40, 41FNC42XL, 02/22/12 to 01/15/17; J90E5P, 41XCD51S, 02/22/12 to 01/15/17; J90E5T, 41XCB52S, 02/22/12 to 01/15/17; J90E7G, 41XCC53S, 02/22/12 to 01/15/17; J90E7R, 02CK017, 02/22/12 to 12/31/99; J90F03, 41XCD51S, 02/23/12 to 01/15/17; J90F06, 41XCD50S, 02/23/12 to 01/15/17; J90F2P, 41XCD515, 02/24/12 to 01/15/17; J90F2R, 41KBC17XL, 02/24/12 to 01/15/17; J90F2T, 41TNC69XL, 02/24/12 to 01/15/17; J90F2U, 41XCD525, 02/27/12 to 01/15/17; J90F4T, 41XCC515, 02/27/12 to 01/15/17; J90F68, 41XCD515, 02/27/12 to 01/15/17; J90F6A, 41KNC07XL, 02/27/12 to 01/15/17; J90F6C, 41FNC42XL, 02/27/12 to 01/15/17; J90F6U, 41FNC41XL, 02/27/12 to 01/15/17; J90F90, 41FNC41XLD, 02/28/12 to 01/15/17; J90F91, 41XCD515, 02/28/12 to 01/15/17; J90F9Z, 41KNC60XL, 02/28/12 to 01/15/17; J90GOO, 41KNC61XL, 02/28/12 to 01/15/17; J90GOF, 02CK017, 02/28/12 to 12/31/99; J90G23, 41XCD515, 02/29/12 to 01/15/17; J90G25, 41TNC36XL, 02/29/12 to 01/15/17; J90G26, 41XCC515, 02/29/12 to 01/15/17; J90G3A, 41KNC60XL, 02/29/12 to 01/15/17; J90G5F, 41XCD515, 03/01/12 to 01/15/17; J90G5H, 41FNC42XL, 03/01/12 to 01/15/17; J90G5J, 41XCD525, 03/01/12 to 01/15/17; J90G9G, 41XCB515, 03/02/12 to 01/15/17; J90G9H, 41KBC17XL, 03/02/12 to 01/15/17; J90G9J, 41XCC515, 03/02/12 to 01/15/17; J90G9K, 41TNC36XL, 03/02/12 to 01/15/17; J90HOP, 41TNC69XL, 03/05/12 to 01/15/17; J90H1R, 41XCD51S, 03/05/12 to 02/15/17; J90H6E, 41FNC42XL, 03/07/12 to 02/15/17; J90H6K, 41FNC41XL, 03/07/12 to 02/15/17; J90H6L, 41KBC17XL, 03/07/12 to 02/15/17; J90H7J, 41KNC09XL, 03/07/12 to 02/15/17; J90H7K, 41XCD525, 03/07/12 to 02/15/17; J90H7L, 41XCB575, 03/07/12 to 02/15/17; J90J1A, 02CK017, 03/07/12 to 12/31/99; J90J36, 41XCD51S, 03/09/12 to 02/15/17; J90J7X, 41XCC515, 03/09/12 to 02/15/17; J90J7Y, 41XCD525, 03/09/12 to 02/15/17; J90J7Z, 41FNC42XL, 03/09/12 to 02/15/17; J90J8P, 41FNC43XL, 03/12/12 to 02/15/17; J90KOH, 41XCD515, 03/12/12 to 02/15/17; J90KOJ, 41TNC20XL, 03/12/12 to 02/15/17; J90KOK, 41XCB575, 03/12/12 to 02/15/17; J90K1V, 41KNC60XL, 03/13/12 to 02/15/17; J90K4E, 41XCC505, 03/13/12 to 02/15/17; J90K4G, 41FNC41XL, 03/13/12 to 02/15/17; J90K4H, 41TNC49XL, 03/13/12 to 02/15/17; J90K4J, 41KNC07XL, 03/13/12 to 02/15/17; J90K5E, 41KNC61XL, 03/13/12 to 02/15/17; J90K5F, 41KNC09XL, 03/13/12 to 02/15/17; J90K6K, 02CK017, 03/13/12 to 12/31/99; J90K7N, 41FNC41XLD, 04/13/12 to 02/15/17; J90K7P, 41KNC09XL, 03/14/12 to 02/15/17; J90K7R, 41TNC49XL, 03/14/12 to 02/15/17; J90K7T, 41KNC07XL, 03/14/12 to 02/15/17; J90K9D, 41XCC51S, 03/14/12 to 02/15/17; J90L1A, 41FNC42XL, 03/15/12 to 02/15/17; J90L1D, 41XCD50S, 03/15/12 to 02/15/17; J90L1E, 41XCD52S, 03/15/12 to 02/15/17; J90L2L, 41FNC41XL, 03/15/12 to 02/15/17; J90L4E, 41KNC03XL, 03/16/12 to 02/15/17; J90L4F, 41KNC60XL, 03/16/12 to 02/15/17; J90L62, 41FNC43XL, 03/16/12 to 02/15/17; J90L6G, 02CK017, 03/16/12 to 12/31/99; J90M16, 02CK017, 03/19/12 to 12/31/99; J90M1T, 41FNC43XLD, 03/20/12 to 02/15/17; J90M1U, 41XCC51S, 03/20/12 to 02/15/17; J90M3G, 41FNC41XLD, 03/21/12 to 02/15/17; J90M3H, 41KNC19XL, 03/22/12 to 02/15/17; J90M3K, 41FNC42XL, 03/22/12 to 02/15/17; J90M4H, 41FNC41XL, 03/23/12 to 02/15/17; J90M4J, 41KNC60XL, 03/23/12 to 02/15/17; J90M50, 02CK017, 03/20/12 to 12/31/99; J90NOJ, 41KNC03XL, 03/23/12 to 02/15/17; J90NOK, 41FNC42XL, 03/23/12 to 02/15/17; J90N33, 41XCC51S 03/22/12 to 02/15/17; J90N4Y, 41XCC51S 03/27/12 to 02/15/17; J90P15, 41XCD51S 03/27/12 to 02/15/17; J90P16, 41FNC43XLD 03/27/12 to 02/15/17; J90P44, 41XCC51S 03/28/12 to 02/15/17; J90P5U, 02CK017 03/27/12 to 12/31/99; J90P6G, 02CK017 03/28/12 to 12/31/99; J90P7T, 41FNC42XL, 03/28/12 to 02/15/17; J90R22, 41KNC60XL 03/29/12 to 02/15/17; J90R23, 41FNC43XL 03/29/12 to 02/15/17; J90R24, 41XCC51S 03/29/12 to 02/15/17; J90R25, 41XCB52S 03/29/12 to 02/15/17; J90R3H, 41KNC07XL 03/29/12 to 02/15/17; J90R3J, 41FNC41XL 03/29/12 to 02/15/17; J90R45, 02CK017 03/29/12 to 12/31/99; J90T34, 41XCC51S 04/02/12 to 02/15/17; J90T35, 41KNC61XL 04/03/12 to 02/15/17; J90T3Z, 41XCB57S 04/02/12 to 02/15/17; J90T5D, 41XCC51S 04/04/12 to 03/15/17; J90T5E, 41XCB57S 04/04/12 to 03/15/17; J90T5F, 41FNC42XL, 04/04/12 to 03/15/17; J90T5G, 41FNC41XL 04/04/12 to 03/15/17; J90T90, 41KNC07XL 04/09/12 to 03/15/17; J90T91, 41XCB52S 04/09/12 to 03/15/17; J90T92, 41FNC43XL 04/09/12 to 03/15/17; J90T93, 41KNC61XL 04/09/12 to 03/15/17; J90UOT, 02CK017 04/04/12 to 12/31/99; J90U41, 41FNC41XLD 04/10/12 to 03/15/17; J90U50, 41FNC42XL, 04/10/12 to 03/15/17; J90U51, 41XCC51S 04/10/12 to 03/15/17; J90U52, 41KNC07XL 04/10/12 to 03/15/17; J90U53, 41FNC43XL 04/10/12 to 03/15/17; J90U54, 41TNC20XL 04/10/12 to 03/15/17; J90W98, 02CK017 04/10/12 to 12/31/99; J90X3A, 02CK017 04/11/12 to 12/31/99; J90X6K, 02CK017 04/12/12 to 12/31/99; J90X6X, 41FNC42XL, 04/12/12 to 03/15/17; J90X6Y, 41XCC51S 04/12/12 to 03/15/17; J90X7D, 41FNC41XL 04/12/12 to 03/15/17; J90X7K, 41KNC07XL 04/13/12 to 03/15/17; J90X9A, 41FNC43XL 04/13/12 to 03/15/17; J90VOO, 02CK017 04/13/12 to 12/31/99; J90V2U, 41XCD51S 04/17;/12 to 03/15/17; J90V41, 02CK017 04/16/12 to 12/31/99; J90V82, to 02CK017 04/17;/12 to 12/31/99; J90Z32, 41FNC42XL, 04/18/12 to 03/15/17; J90Z33, 41XCC51S 04/18/12 to 03/15/17; J90Z43, 41KBC17XL 04/19/12 to 03/15/17; J90Z44, 41XCD52S 04/18/12 to 03/15/17; J90Z5Z, 41FNC41XL, 04/19/12 to 03/15/17; J91005, 02CK017, 04/20/12 to 12/31/99; J9107X, 41FNC42XL, 04/24/12 to 03/15/17; J9110D, 41XCC51S, 04/25/12 to 03/15/17; J9110E, 41KBC17XL, 04/25/12 to 03/15/17; J9112K, 02CK017, 04/25/12 to 12/31/99; J91167, 41TNC49XL, 04/27/12 to 03/15/17; J91168, 41FNC41XL, 04/27/12 to 03/15/17; J91179, 02CK017, 04/26/12 to 12/31/99; J91195, 41XCD52S, 04/27/12 to 03/15/17; J9125D, 02CK017, 04/28/12 to 12/31/99; J9128G, 41FNC42XL, 05/02/12 to 03/15/17; J9128H, 41XCC51S, 05/02/12 to 03/15/17; J91306, 02CK017, 05/02/12 to 12/31/99; J9133G, 02CK017, 05/03/12 to 12/31/99; J9140G, 02CK017, 05/07/12 to 12/31/99; J91484, 02CK017, 05/09/12 to 12/31/99; J9167F, 02CK017, 05/17;/12 to 12/31/99; J91685, 41FNC41XL, 05/17/12 to 04/15/17; J9171F, 02CK017, 05/18/12 to 12/31/99; J9175D, 41XCC51S, 05/21/12 to 04/15/17; J9175F, 41KBC17XL, 05/21/12 to 04/15/17; J9180F, 41FNC42XL, 05/23/12 to 04/15/17; J9180G, 41KNC07XL, 05/23/12 to 04/15/17; J9181K, 02CK017, 05/22/12 to 12/31/99; J9183T, 41XCB52S, 05/23/12 to 04/15/17; J9183U, 41FNC43XL, 05/23/12 to 04/15/17; J9184V, 02CK017, 05/23/12 to 12/31/99; J9187R, 41KNC60XL, 05/25/12 to 04/15/17; J9196V, 41FNC43XLD, 05/28/12 to 04/15/17; J9196W, 41XCC51S, 06/15/12 to 04/15/17; J9196X, 41KNC07XL, 05/28/12 to 04/15/17; J9199X, 02CK017, 05/28/12 to 12/31/99; J91A1C, 41KBC17XL, 05/30/12 to 04/15/17; J91A1D, 41FNC42XL, 05/30/12 to 04/15/17; J91A1E, 41FNC43XL, 05/29/12 to 04/15/17; J91A47, 41FNC41XL, 05/30/12 to 04/15/17; J91A57, 41XCB52S, 05/30/12 to 04/15/17; J91A58, 41FNC41XLD, 05/30/12 to 04/15/17; J91A59, 41XCB57S, 05/30/12 to 04/15/17; J91A62, 41XCD52S, 05/30/12 -04/15/17; J91A63, 41TNC49XL, 05/31/12 to 04/15/17; J91A6V, 02CK017, 05/30/12 to 12/31/99; J91A9P, 02CK017, 05/31/12 to 12/31/99; J91C2H, 41KNC07XL, 06/01/12 to 04/15/17; J91C2Z, 41XCC51S, 06/04/12 to 04/15/17; J91C30, 41XCC51S, 06/01/12 to 04/15/17; J91C31, 41FNC42XL, 06/01/12 to 04/15/17; J91C3J, 41FNC43XL, 06/04/12 to 04/15/17; J91C3K, 41KBC17XL, 06/04/12 to 04/15/17; J91C60, 41FNC41XL, 06/05/12 to 04/15/17; J91C6Z, 02CK017, 06/04/12 to 12/31/99; J91D05, 41XCD51S, 06/05/12 to 05/15/17; J91D1D, 02CK017

06/05/12 to 12/31/99; J91D3V, 41KNC07XL, 06/06/12 to 05/15/17; J91D7X, 41XCB57S, 06/07/12 to 05/15/17; J91D7Y, 41FNC42XL, 06/07/12 to 05/15/17; J91D7Z, 41KNC07XL, 06/07/12 to 05/15/17; J91D8H, 41XCC51S, 06/08/12 to 05/15/17; J91D8J, 41TNC69XL, 06/07/12 to 05/15/17; J91E4M, 41XCD51S, 06/12/12 to 05/15/17; J91E4N, 41FNC41XLD, 06/12/12 to 05/15/17; J91E9R, 41FNC43XL, 06/13/12 to 05/15/17; J91E9T, 41FNC42XL, 06/12/12 to 05/15/17; J91E9V, 41KNC07XL, 06/13/12 to 05/15/17; J91F23, 02CK017

06/12/12 to 12/31/99; J91F45, 41XCB52S, 06/14/12 to 05/15/17; J91F47, 41TNC49XL, 06/14/12 to 05/15/17; J91F77, 41XBB79S, 06/14/12 to 05/15/17; J91F9F, 02CK017

06/14/12 to 12/31/99; J91G24, 41XCD51S, 06/14/12 to 05/15/17; J91G2D, 41KNC61XL, 06/20/12 to 05/15/17; J91G2Z, 02CK017

06/15/12 to 12/31/99; J91G5K, 41XCD51S, 06/15/12 to 05/15/17; J91G70, 41XCC51S, 06/18/12 to 05/15/17; J91H03, 41FNC42XL, 06/19/12 to 05/15/17; J91H04, 41KNC07XL, 06/21/12 to 05/15/17; J91H05, 41KBC17XL, 06/22/12 to 05/15/17; J91H15, 02CK017

06/19/12 to 12/31/99; J91H35, 41FNC41XL, 06/19/12 to 05/15/17; J91H36, 41XCD51S, 06/21/12 to 05/15/17; J91H9N, 02CK017

06/21/12 to 12/31/99; J91J39, 41KNC60XL, 06/22/12 to 05/15/17; J91K3F, 41XBB79S, 06/26/12 to 05/15/17; J91K5G, 41XBB79S, 06/26/12 to 05/15/17; J91K6V, 41XCC51S, 06/28/12 to 05/15/17; J91K6Z, 41KNC61XL, 07/04/12 to 05/15/17; J91K71, 41FNC42XL, 06/28/12 to 05/15/17; J91K73, 41XCC50S, 06/28/12 to 05/15/17; J91K75, 41KBC17XL, 06/29/12 to 05/15/17; J91L08, 41KNC61XL, 06/29/12 to 05/15/17; J91L09, 41FNC41XL, 06/29/12 to 05/15/17; J91LOA, 41XCB52S, 06/29/12 to 05/15/17; J91LOD, 41FNC43XL, 07/04/12 to 05/15/17; J91L68, 02CK017

06/29/12 to 12/31/99; J91MOH, 02CK017

07/04/12 to 12/31/99; J91M5Z, 41XBB79S, 07/05/12 to 06/15/17; J91M62, 41FNC43XLD, 07/05/12 to 06/15/17; J91M63, 41KBC17XL, 07/05/12 to 06/15/17; J91M6U, 02CK017

07/05/12 to 12/31/99; J91M9U, 41TNC49XL, 07/06/12 to 06/15/17; J91M9W, 41FNC41XL, 07/06/12 to 06/15/17; J91N3A, 41FNC41XLD, 07/09/12 to 06/15/17; J91N3C, 41XCB57S, 07/09/12 to 06/15/17; J91N3D, 41XCD52S, 07/09/12 to 06/15/17; J91N4K, 02CK017

07/09/12 to 12/31/99; J91N9L, 41FNC43XL, 07/10/12 to 06/15/17; J91N9M, 41XCB52S, 07/10/12 to 06/15/17;J91N9N, 41XCD51S, 07/10/12 to 06/15/17; J91P1K, 02CK017

07/10/12 to 12/31/99; J91P2X, 41KBC17Xl, 07/11/12 to 06/15/17; J91P2Y, 41XCC50S, 07/11/12 to 06/15/17; J91P2Z, 41XCC51S, 07/12/12 to 06/15/17; J91P47, 02CK017

07/11/12 to 12/31/99; J91P79, 41XCB52S, 07/12/12 to 06/15/17; J91P7A, 41FNC41Xl, 07/12/12 to 06/15/17; J91ROH, 41TNC36Xl, 07/13/12 to 06/15/17; J91ROK, 41FNC42Xl, 07/13/12 to 06/15/17; J91R4P, 41XCC53S, 07/17;/12 to 06/15/17; J91R4R, 41XCD51S, 07/17;/12 to 06/15/17; J91R9C, 41TNC49Xl, 07/17;/12 to 06/15/17; J91R9D, 41XCB57S, 07/18/12 to 06/15/17; J91R9F, 41FNC41Xl, 07/17;/12 to 06/15/17; J91T22, to 02CK017

07/17;/12 to 12/31/99; J91T42, 41XCD52S, 07/18/12 to 06/15/17; J91T67, 02CK017

07/18/12 to 12/31/99; J91T98, 41XCD50S, 07/19/12 to 06/15/17; J91T99, 41FNC42Xl, 07/19/12 to 06/15/17; J91T9A, 41XCC51S, 07/19/12 to 06/15/17; J91U23, 41XCD51S, 07/19/12 to 06/15/17; J91U36, 41KNC60Xl, 07/23/12 to 06/15/17; J91U37, 41KBC17Xl, 07/20/12 to 06/15/17; J91U38, 41XCC51S, 07/23/12 to 06/15/17; J91U46, 41TNC36Xl, 07/20/12 to 06/15/17; J91U7U, 41KNC09Xl, 07/24/12 to 06/15/17; J91U7V, 41XCC53S, 07/24/12 to 06/15/17; J91V2E, 41KNC03Xl, 07/25/12 to 06/15/17; J91V2F, 41XCD51S, 07/25/12 to 06/15/17; J91V30, 02CK017

07/24/12 to 12/31/99; J91V60, 41KBC17Xl, 07/25/12 to 06/15/17; J91V6U, 41XCC51S, 07/25/12 to 06/15/17; J91V6Z, 02CK017

07/25/12 to 12/31/99; J91W06, 41XCD51S, 07/26/12 to 06/15/17; J91W07, 41KNC60Xl, 07/27/12 to 06/15/17; J91W42, 41FNC43Xl, 07/27/12 to 06/15/17; J91W43, 41FNC41Xl, 07/27/12 to 06/15/17; J91W44, 41XCD52S, 07/27/12 to 06/15/17; J91W45, 41XCD51S, 07/27/12 to 06/15/17; J91W7J, 41XCC51S, 07/31/12 to 06/15/17; J91W7l, 41KBC17Xl, 07/30/12 to 06/15/17; J91W7M, 41KNC03Xl, 08/01/12 to 06/15/17; J91X1l, 02CK017

07/31/12 to 12/31/99; J91X3U, 41FNC42Xl, 08/01/12 to 06/15/17; J91X3V, 41FNC43XL, 08/01/12 to 06/15/17; J91X3W, 41FNC41XL, 08/02/12 to 06/15/17; J91X3X, 41FNC41XLD, 08/02/12 to 06/15/17; J91X7K, 41KNC60XL, 08/02/12 to 06/15/17; J91Y10, 41XCD51S, 08/03/12 to 07/15/17; J91Y11, 41XCC50S, 08/06/12 to 07/15/17; J91Y49, 41XCD51S, 08/06/12 to 07/15/17; J91Y5G, 02CK017

08/06/12 to 12/31/99; J91Y8A, 41TNC49XL, 08/07/12 to 07/15/17; J91Y8C, 41FNC42XL, 08/07/12 to 07/15/17; J91Y9C, 02CK017

08/08/12 to 12/31/99; J91Z1H, 41XCB57S, 08/08/12 to 07/15/17; J91Z1J, 41FNC41XL, 08/08/12 to 07/15/17; J91Z26, 41KBC17XL, 08/08/12 to 07/15/17; J91Z5J, 41FNC43XL, 08/09/12 to 07/15/17; J91Z9E, 41XCD51S, 08/13/12 to 07/15/17; J91Z9F, 41XCC51S, 08/14/12 to 07/15/17; J9206A, 41XCD51S, 08/15/12 to 07/15/17; J92079, 41XCD51S, 08/15/12 to 07/15/17; J9208E, 02CK017

08/15/12 to 12/31/99; J9210Y, 41XCC50S, 08/16/12 to 07/15/17; J9210Z, 41XCD50S, 08/16/12 to 07/15/17; J92110, 41XCD51S, 08/16/12 to 07/15/17; J9214G, 41XCC51S, 08/17;/12 to 07/15/17; J9214H, 41FNC41XL, 08/17;/12 to 07/15/17; J9214J, 41FNC42XL, 08/17;/12 to 07/15/17; J9214K, 41XCD51S, 08/17;/12 to 07/15/17; J9219P, 41FNC41XLD, 08/20/12 to 07/15/17; J92239, 41FNC43XL, 08/21/12 to 07/15/17; J9223A, 41KBC17XL, 08/21/12 to 07/15/17; J9223C, 41FNC43XLD, 08/21/12 to 07/15/17; J9223D, 41XCB52S, 08/21/12 to 07/15/17; J92245, 02CK017

08/21/12 to 12/31/99; J92273, 41XCD51S, 08/23/12 to 07/15/17; J92274, 41XCC51S, 08/22/12 to 07/15/17; J92275, 41FNC42XL, 08/23/12 to 07/15/17; J9230G, 41XCD51S, 08/24/12 to 07/15/17; J9231U, 02CK017

08/23/12 to 12/31/99; J9236P, 02CK017

08/24/12 to 12/31/99; J9237F, 41XCD51S, 08/27/12 to 07/15/17; J9243Y, 41FNC43XL, 08/28/12 to 07/15/17; J9243Z, 41FNC41XL, 08/28/12 to 07/15/17; J92440, 41XCC51S, 08/28/12 to 07/15/17; J92441, 41FNC41XLD, 08/28/12 to 07/15/17; J9244Y, 02CK017

08/28/12 to 12/31/99; J9247V, 41XCB55S, 08/29/12 to 07/15/17; J9247W, 41KBC17XL, 08/29/12 to 07/15/17; J9247X, 41XCC51S, 08/29/12 to 07/15/17; J9247Y, 41FNC41XL, 08/29/12 to 07/15/17; J9247Z, 41FNC43XL, 08/29/12 to 07/15/17; J9248W, 02CK017

08/30/12 to 12/31/99; J92578, 02CK017

08/31/12 to 12/31/99; J9260L, 02CK017

09/03/12 to 12/31/99; J9262H, 41FNC41XLD, 09/05/12 to 08/15/17; J9262J, 41XCC53S, 09/05/12 to 08/15/17; J9262K, 41FNC42XL, 09/04/12 to 08/15/17; J92639, 41XCC51S, 09/05/12 to 08/15/17; J9263A, 41XCB57S, 09/04/12 to 08/15/17; J92700, 41KBC17XL, 09/06/12 to 08/15/17; J92701, 41FNC43XL, 09/06/12 to 08/15/17; J92702, 41FNC42XL, 09/06/12 to 08/15/17; J9276X, 02CK017

09/07/12 to 12/31/99; J9277T, 41XCC51S, 09/10/12 to 08/15/17; J9277V, 41FNC43XL, 09/10/12 to 08/15/17; J9284L, 02CK017

09/11/12 to 12/31/99; J9284M, 41KNC60XL, 09/11/12 to 08/15/17; J92866, 41XCC50S, 09/12/12 to 08/15/17; J92868, 41FNC41XL, 09/12/12 to 08/15/17; J9289Y, 41TNC49XL, 09/13/12 to 08/15/17; J9289Z, 41FNC42XL, 09/13/12 to 08/15/17; J92900, 41FNC43XL, 09/13/12 to 08/15/17; J92902, 41XCC51S, 09/14/12 to 08/15/17; J9297F, 41XCC51S, 09/18/12 to 08/15/17; J9297H, 41XCB52S, 09/18/12 to 08/15/17; J92A2C, 41FNC41XL, 09/19/12 to 08/15/17; J92A2D, 41XCB52S, 09/19/12 to 08/15/17; J92A2E, 41XCC51S, 09/19/12 to 08/15/17; J92A2N, 02CK017

09/19/12 to 12/31/99; J92A50, 41FNC43XL, 09/20/12 to 08/15/17; J92A51, 41XCC53S, 09/20/12 to 08/15/17; J92A52, 41FNC42XL, 09/20/12 to 08/15/17; J92A53, 41FNC41XLD, 09/20/12 to 08/15/17; J92A7N, 02CK017

09/20/12 to 12/31/99; J92C1E, 02CK017

09/21/12 to 12/31/99; J92C5E, 41FNC43XLD, 09/27/12 to 08/15/17; J92C75, 02CK017

09/25/12 to 12/31/99; J92D7A, 41XCD52S, 09/28/12 to 08/15/17; J92EOO, 02CK017

09/28/12 to 12/31/99; J92E15, 41KBC17XL, 10/01/12 to 08/15/17; J92E42, to 02CK017

10/01/12 to 12/31/99; J92E5N, 41KNC60XL, 10/02/12 to 08/15/17; J92E5P, 41KNC61XL, 10/02/12 to 08/15/17; J92E5R, 41XCD51S, 10/02/12 to 08/15/17; J92E7J, 02CK017

10/02/12 to 12/31/99; J92F04, 41XCB57S, 10/04/12 to 09/15/17; J92F05, 41FNC43XL, 10/03/12 to 09/15/17; J92F06, 41FNC42XL, 10/04/12 to 09/15/17; J92F3G, 41KNC09XL, 10/04/12 to 09/15/17; J92F3H, 41FNC41XLD, 10/04/12 to 09/15/17; J92F73, 41FNC40XL, 10/08/12 to 09/15/17; J92F74, 41XCB57S, 10/08/12 to 09/15/17; J92F76, 41XCD51S, 10/08/12 to 09/15/17; J92GOO, 02CK017

10/05/12 to 12/31/99; J92G48, 02CK017

10/08/12 to 12/31/99; J92G5T, 41XCC50S, 10/09/12 to 09/15/17; J92G6G, 41KNC60XL, 10/09/12 to 09/15/17; J92G90, 41XCD51S, 10/12/12 to 09/15/17; J92G92, 41TNC49XL, 10/10/12 to 09/15/17; J92H03, 41FNC43XLD, 10/11/12 to 09/15/17; J92H20, 02CK017

10/10/12 to 12/31/99; J92H3C, 41XCD52S, 10/10/12 to 09/15/17; J92H3E, 41XCD50S, 10/12/12 to 09/15/17; J92H4E, 42KNC07XL, 10/11/12 to 09/15/17; J92H4Z, 42KNC35XL, 10/11/12 to 09/15/17; J92H50, 42XCB55S, 11/05/12 to 09/15/17; J92H6E, 02CK017

10/11/12 to 12/31/99; J92H6L, 41XCD50S, 10/11/12 to 09/15/17; J92H8J, 42XCB51S, 10/15/12 to 09/15/17; J92H8N, 41KBC17XL, 10/15/12 to 09/15/17; J92H8P, 41FNC41XL, 10/12/12 to 09/15/17; J92H8R, 41FNC42XL, 10/15/12 to 09/15/17; J92J2D, 41FNC41XLD, 10/16/12 to 09/15/17; J92J2F, 41TNC49XL, 10/16/12 to 09/15/17; J92J2H, 41FNC43XLD, 10/16/12 to 09/15/17; J92J32, 42KNC03XL, 10/16/12 to 09/15/17; J92J6N, 41XCD50S, 10/16/12 to 09/15/17; J92J6P, 41FNC41XL, 10/17;/12 to 09/15/17; J92J8W, 02CK017

10/16/12 to 12/31/99; J92K9A, 42KNC07XL, 11/05/12 to 09/15/17; J92K9C, 42XCB55S, 11/05/12 to 09/15/17; J92L9N, 02CK017

10/23/12 to 12/31/99; J92M19, 41XCC53S, 10/25/12 to 09/15/17; J92M4E, 02CK017

10/24/12 to 12/31/99; J92M6C, 41FNC41XLD, 10/25/12 to 09/15/17; J92M8K, 02CK017

10/25/12 to 12/31/99; J92N4R, 41FNC43XL, 10/30/12 to 09/15/17; J92N7Y, 42KNC35XL, 11/02/12 to 09/15/17; J92N7Z, 42KNC03XL, 11/05/12 to 09/15/17; J92N8Z, 02CK017

10/30/12 to 12/31/99; J92R3Y, 02CK017

11/02/12 to 12/31/99; J92T1A, 41FNC42XL, 11/06/12 to 10/15/17; J92T1C, 41KBC17XL, 11/06/12 to 10/15/17; J92T1D, 42KNC35XL, 11/06/12 to 10/15/17; J92T3A, 02CK017

11/06/12 to 12/31/99; J92T4G, 42KNC07XL, 11/07/12 to 10/15/17; J92T6M, 02CK017

11/07/12 to 12/31/99; J92T7E, 41XCC51S, 11/08/12 to 10/15/17; J92T7F, 41XCD52S, 11/08/12 to 10/15/17; J92T7G, 41XCB57S, 11/08/12 to 10/15/17; J92U06, 43TNC20XL, 11/12/12 to 10/15/17; J92U07, 43TNC69XL, 11/12/12 to 10/15/17; J92U26, 41KNC61XL, 11/14/12 to 10/15/17; J92U62, 42XCB55S, 11/14/12 to 10/15/17; J92U63, 42KNC03XL, 11/14/12 to 10/15/17; J92U88, 41KNC60XL, 11/14/12 to 10/15/17; J92V5U, 02CK017

11/14/12 to 12/31/99; J92V63, 41XCC51S, 11/15/12 to 10/15/17; J92V64, 41FNC41XLD, 11/15/12 to 10/15/17; J92V7H, 41FNC43XL, 11/14/12 to 10/15/17; J92V7K, 42XCB51S, 11/15/12 to 10/15/17; J92V8M, 43TNC20XL, 11/15/12 to 10/15/17; J92V8N, 43TNC69XL, 11/16/12 to 10/15/17; J92WOT, 02CK017

11/15/12 to 12/31/99; J92W2E, 43TNC69XL, 11/21/12 to 10/15/17; J92W2F, 43TNC20XL, 11/16/12 to 10/15/17; J92W2G, 43KNC03XL, 11/16/12 to 10/15/17; J92X07, 42KNC07XL, 11/21/12 to 10/15/17; J92XOA, 41FNC41XL, 11/22/12 to 10/15/17; J92XOC, 41XCC50S, 11/22/12 to 10/15/17; J92XOZ, 41FNC41XL, 11/21/12 to 10/15/17; J92X3P, 02CK017

11/22/12 to 12/31/99; J92X8C, 02CK017

11/24/12 to 12/31/99; J92X8D, 42KNC35XL, 11/23/12 to 10/15/17; J92X8E, 42XCB55S, 11/23/12 to 10/15/17; J92X9R, 41FNC43XL, 11/23/12 to 10/15/17; J92X9T, 41FNC42XL, 11/23/12 to 10/15/17; J92Y2P, 41KBC17XL, 11/26/12 to 10/15/17; J92Y2R, 41XCC50S, 11/26/12 to 10/15/17; J92Y2T, 41KNC61XL, 11/26/12 to 10/15/17; J92Y2U, 41XCC51S, 11/26/12 to 10/15/17; J92Y6R, 41XCD51S, 11/27/12 to 10/15/17; J92Y71, 41XCC51S, 11/28/12 to 10/15/17; J92ZOJ, 41FNC43XLD, 11/28/12 to 10/15/17; J92ZOK, 41KNC09XL, 11/28/12 to 10/15/17; J92Z14, 41XCD52S, 11/28/12 to 10/15/17; J92Z3Z, 02CK017

11/28/12 to 12/31/99; J92Z4R, 41XCB57S, 11/29/12 to 10/15/17; J92Z4T, 42XCB55S, 11/29/12 to 10/15/17; J92Z4U, 41XCC53S, 11/29/12 to 10/15/17; J92Z92, 41FNC41XL, 11/30/12 to 10/15/17; J92Z93, 41XCD51S, 12/01/12 to 10/15/17; J9302F, 41FNC42XL, 12/03/12 to 10/15/17; J9302G, 41KBC17XL, 12/03/12 to 10/15/17; J9302H, 41FNC43XL, 12/03/12 to 10/15/17; J9303L, 41XCC50S, 12/03/12 to 10/15/17; J93105, 41XCD51S, 12/05/12 to 11/15/17; J9313T, 41XCD52S, 12/05/12 to 11/15/17; J9313Z, 41XCC51S, 12/05/12 to 11/15/17; J93140, 41XCD50S, 12/07/12 to 11/15/17; J93168, 41XCD51S, 12/07/12 to 11/15/17; J9316C, 41FNC43XL, 12/10/12 to 11/15/17; J9316D, 41FNC42XL, 12/10/12 to 11/15/17; J9318Z, 02CK017

12/07/12 to 12/31/99; J93231, 02CK017

12/10/12 to 12/31/99; J9324T, 41KBC17XL, 12/10/12 to 11/15/17; J9324U, 41FNC40XL, 12/10/12 to 11/15/17; J9324V, 41XCC50S, 12/11/12 to 11/15/17; J9325M, 41FNC41XL, 12/11/12 to 11/15/17; K4C03G, 37ER320, 01/04/13 12/15/17; K4C03N, 37ER320, 01/04/13 12/15/17; K4C07P, 37ER320, 01/08/13 12/15/17; K4C09V, 37ER320, 01/09/13 12/15/17; K4C136, 37ER320, 01/10/13 12/15/17; K4C15P, 37ER320, 01/12/13 12/15/17; K4C16E, 37ER320, 01/12/13 12/15/17; K4C185, 37ER320, 01/13/13 12/15/17; K4C18P, 37ER320, 01/14/13 12/15/17; K4C18Y, 37ER320, 01/13/13 12/15/17; K4C19W, 37ER320, 01/15/13 12/15/17; K4C206, 37ER320, 01/16/13 12/15/17; K4C22M, 37ER320, 01/17;/13 12/15/17; K4C26W, 37ER320, 01/18/13 12/15/17; K4C27N, 37ER320, 01/21/13 12/15/17; K4C28T, 37ER320, 01/22/13 12/15/17; K4C293, 37ER320, 01/22/13 12/15/17; K4C31W, 37ER320, 01/22/13 12/15/17; K4C37N, 37ER320, 01/26/13 12/15/17; K4C384, 37ER320, 01/26/13 12/15/17; K4C39J, 37ER320, 01/25/13 12/15/17; K4C412, 37ER320, 01/26/13 12/15/17; K4C41Z, 37ER320, 01/26/13 12/15/17; K4C42N, 37ER320, 01/26/13 12/15/17; K4C504, 37ER320, 02/01/13 12/15/17; K4C51A, 37ER320, 02/02/13 12/15/17; K4C523, 37ER320, 02/02/13 12/15/17; K4C573, 37ER320, 02/06/13 01/15/18; K4C57K, 37ER320, 02/05/13 01/15/18; K4C581, 37ER320, 02/06/13 01/15/18; K4C58G, 37ER320, 02/06/13 01/15/18; K4C591, 37ER320, 02/06/13 01/15/18; K4C63E, 37ER320, 02/08/13 01/15/18; K4C645, 37ER320, 02/09/13 01/15/18; K4C64P, 37ER320, 02/12/13 01/15/18; K4C69N, 37ER320, 02/13/13 01/15/18; K4C709, 37ER320, 02/13/13 01/15/18; K4C71H, 37ER320, 02/13/13 01/15/18; K4C74T, 37ER320, 02/15/13 01/15/18; K4C791, 37ER320, 02/19/13 01/15/18; K4C82W, 37ER320, 02/20/13 01/15/18; K4C833, 37ER320, 02/20/13 01/15/18; K4C848, 37ER320, 02/21/13 01/15/18; K4C88J, 37ER320, 02/23/13 01/15/18; K4C88Y, 37ER320, 02/24/13 01/15/18; K4C89U, 37ER320, 02/24/13 01/15/18; K4C98D, 37ER320, 03/02/13 01/15/18; K4C98K, 37ER320, 03/01/13 01/15/18; K4CA38, 37ER320, 03/11/13 01/15/18; K4CA3V, 37ER320, 03/08/13 02/15/18; K4CC3F, 37ER320, 03/08/13 02/15/18; K4CC3H, 37ER320, 03/09/13 02/15/18; K4CC4R, 37ER320, 03/09/13 02/15/18; K4CC5J, 37ER320, 03/09/13 02/15/18; K4CC5K, 37ER320, 03/11/13 02/15/18; K4CD16, 37ER320, 03/13/13 02/15/18; K4CD1Y, 37ER320, 03/14/13 02/15/18; K4CD24, 37ER320, 03/14/13 02/15/18; K4CD2M, 37ER320, 03/14/13 02/15/18; K4CD3F, 37ER320, 03/15/13 02/15/18; K4CD3Y, 37ER320, 03/15/13 02/15/18; K4CD45, 37ER320, 03/15/13 02/15/18; K4CD7W, 37ER320, 03/18/13 02/15/18; K4CEOW, 37ER320, 03/20/13 02/15/18; K4CE2N, 37ER320, 03/20/13 02/15/18; K4CE9K, 37ER320, 03/23/13 02/15/18; K4CF07, 37ER320, 03/23/13 02/15/18; K4CFOJ, 37ER320, 03/26/13 02/15/18; K4CF43, 37ER320, 03/27/13 02/15/18; K4CF49, 37ER320, 03/27/13 02/15/18; K4CF60, 37ER320, 03/28/13 02/15/18; K4CF72, 37ER320, 03/28/13 02/15/18; K4CF7R, 37ER320, 03/28/13 02/15/18; K4CF8Z, 37ER320, 04/03/13 02/15/18; K4CG2H, 37ER320, 04/04/13 03/15/18; K4CG3H, 37ER320, 04/04/13 03/15/18; K90393, 41TNC49XL, 01/17;/13 12/15/17; K90394, 41FNC43XLD, 01/17;/13 12/15/17; K90395, 41FNC43XL, 01/17;/13 12/15/17; K9042A, 02CK017

01/18/13 12/31/99; K90441, 41FNC41XL, 01/21/13 12/15/17; K90442, 41KBC17XL, 01/21/13 12/15/17; K90443, 41FNC41XLD, 01/21/13 12/15/17; K90486, 41XCC51S, 01/22/13 12/15/17; K90487, 41XCD51S, 01/23/13 12/15/17; K90488, 41XCD52S, 01/23/13 12/15/17; K90489, 41XCB57S, 01/23/13 12/15/17; K90519, 02CK017

01/23/13 12/31/99; K9057D, 41TNC49XL, 02/01/13 12/15/17; K9059K, 02CK017

01/24/13 12/31/99; K9063J, 41FNC41XLD, 01/28/13 12/15/17; K9063K, 41FNC41XL, 01/28/13 12/15/17; K9083C, 41FNC43XL, 02/05/13 12/15/17; K9083D, 41XCC51S, 02/05/13 12/15/17; K9083E, 02CK017

02/01/13 12/31/99; K9084H, 41FNC42XL, 02/05/13 12/15/17; K9084L, 41XCB57S, 02/06/13 12/15/17; K9088P, 41FNC40XL 02/05/13 01/15/18; K9088R 41XCD52S 02/06/13 01/15/18; K90894 41XCB52S 02/06/13 01/15/18; K90895 41KBC17XL 02/06/13 01/15/18; K9093F 41XCD51S 02/07/13 01/15/18; K90947 43TNC69XL 02/08/13 01/15/18; K9094P 41FNC43XLD 02/11/13 01/15/18; K9097G 41XCD50S 02/12/13 01/15/18; K9097H 41XCD51S 02/11/13 01/15/18; K90A5Z, 43TNC69XL 02/13/13 01/15/18; K90A9V, 41FNC41XLD 02/13/13 01/15/18; K90A9W, 41FNC41XL, 02/13/13 01/15/18; K90A9X, 41XCC51S, 02/13/13 01/15/18; K90C2J, 02CK017

02/14/13 12/31/99; K90C3A, 41FNC43XL, 02/13/13 01/15/18; K90C3K, 41XCB57S, 02/14/13 01/15/18; K90C5C, 41XCD51S, 02/15/13 01/15/18; K90C89, 41XCD51S, 02/16/13 01/15/18; K90C8L, 41FNC43XLD, 02/15/13 01/15/18; K90C8N, 41XCD52S, 02/18/13 01/15/18; K90C9W, 02CK017

02/15/13 12/31/99; K90D1X, 41FNC42XL, 02/19/13 01/15/18; K90D1Y, 41KBC17XL, 02/19/13 01/15/18; K90D4U, 02CK01,7 02/18/13 12/31/99; K90D7N, 41XCC50S, 02/19/13 01/15/18; K90E1A, 41XCB57S, 02/20/13 01/15/18; K90E1D, 43TNC69XL, 02/20/13 01/15/18; K90E29, 41XCD50S, 02/20/13 01/15/18; K90E2A, 41XCB52S, 02/20/13 01/15/18; K90E4G, 02CK017

02/20/13 12/31/99; K90E5Y, 41XCD51S, 02/21/13 01/15/18; K90F02, 41FNC43XL, 02/22/13 01/15/18; K90F04, 41XCC51S, 02/22/13 01/15/18; K90F30, 02CK017, 02/23/13 to 12/31/99; K90F3X, 41XCD51S 02/22/13 01/15/18; K90F4E, 41XCD51S 02/25/13 01/15/18; K90F4H, 41FNC42XL, 02/25/13 01/15/18; K90F90 41FNC43XL 02/26/13 01/15/18; K90F91 41XCC51S 02/26/13 01/15/18; K90G1G 02CK017, 02/26/13 12/31/99; K90G2V 41FNC43XL 02/27/13 01/15/18; K90G2W 41FNC42XL, 02/27/13 01/15/18; K90G2X 41XCC51S 02/27/13 01/15/18; K90G4X 02CK017, 02/27/13 12/31/99; K90H1P 41KBC17XL 03/04/13 01/15/18; K90H1R 41XCD52S 03/04/13 01/15/18; K90H1T 41XCD51S 03/01/13 01/15/18; K90H5V 41FNC41XLD 03/04/13 01/15/18; K90JOZ 41KNC60XL 03/11/13 02/15/18; K90J10 41KNC61XL 03/07/13 02/15/18; K90J11 41TNC36XL 03/07/13 02/15/18; K90J93, 41XCC53S 03/07/13 02/15/18; K90K3, A 02CK017, 03/08/13 12/31/99; K90K55, 41KNC09XL 03/13/13 02/15/18; K90K56, 41FNC43XLD 03/12/13 02/15/18; K90K57, 41XCD51S 03/12/13 02/15/18; K90L08, 41FNC43XL 03/14/13 02/15/18; K90L09, 41KNC60XL 03/14/13 02/15/18; K90L17, 41FNC41XL 03/13/13 02/15/18; K90L1V, 02CK017, 03/12/13 12/31/99; K90L5A, 41KBC17XL 03/14/13 02/15/18; K90M20, 41XCD51S 03/14/13 02/15/18; K90M22, 41XCC53S 03/22/13 02/15/18; K90M3N, 41XCC50S 03/15/13 02/15/18; K90M3P, 41XCD50S 03/15/13 02/15/18; K90M6M, 41XCC51S 03/18/13 02/15/18; K90M6N, 41FNC43XL 03/18/13 02/15/18; K90M6P, 41XCD51S 03/18/13 02/15/18; K90N1Z, 41TNC49XL 03/19/13 02/15/18; K90N6A, 41XCB52S 03/20/13 02/15/18; K90N6C, 41FNC41XL 03/20/13 02/15/18; K90N9F, 41XCC51S 03/21/13 02/15/18; K90N9G, 41FNC43XL 03/21/13 02/15/18; K90N9H, 41FNC42XL, 03/21/13 02/15/18; K90P3G, 41TNC49XL 03/22/13 02/15/18; K90P3H, 41FNC41XLD 03/22/13 02/15/18; K90P4F, 41KNC61XL 03/22/13 02/15/18; K90R2P, 02CK017, 03/26/13 12/31/99; K90R8R, 41KBC17XL 04/01/13 02/15/18; K90R8T, 41FNC41XLD 04/01/13 02/15/18; K90U1H, 41TNC36XL 04/04/13 03/15/18; and K90U1J, 41KNC09XL 04/04/13 03/15/18. 
FEI Number 1527736
Recalling Firm/
Manufacturer
Ethicon Endo-Surgery Inc
4545 Creek Rd
Cincinnati OH 45242-2803
For Additional Information ContactThomas A. Morris
513-337-3419
Manufacturer Reason
for Recall
Ethicon Endo-Surgery is initiating a voluntary recall for LIGACLIP 10mm M/L Endoscopic Rotating Multiple Clip Applier (ER320) due to potential clip formation and feeding issues which may result in improper clip formation, vessel damage and/or insufficient occlusion of the vessel or other structure
FDA Determined
Cause 2
Under Investigation by firm
ActionEthicon sent an Urgent : Medical Device Recall letter dated April 30, 2013 to all primary and secondary consignees via overnight UPS to allow tracking of the receipt. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory immediately and return affected products for credit. This credit is intended to facilitate replacement purchase. For questions contact your local representative or call Ethicon Customer Support Center at 1-877-384-4266 option 6.
Quantity in Commerce738,356
DistributionWorldwide Distribution - USA Nationwide to all fifty (50) states, and internationally including the following countries: ARGENTINA, ARGENTINA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CYPRUS, ECUADOR, EGYPT, EL SALVADOR, GREECE, HONDURAS, INDIA, INDONESIA, ISRAEL, JAMAICA, JAPAN, JORDAN, KOREA, REPUBLIC OF, KUWAIT, LEBANON, MEXICO, NETHERLANDS ANTILLES, NICARAGUA, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, PUERTO RICO, QATAR, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, TAIWAN, PROVINCE OF CHINA, THAILAND, TRINIDAD AND TOBAGO, TURKEY, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA and YEMEN.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = GDO
-
-