U.S. Department of Health and Human Services

Class 2 Device Recall Medical Linear Accelerator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Medical Linear Accelerator see related information
Date Initiated by Firm May 28, 2013
Date Posted March 07, 2014
Recall Status1 Terminated 3 on September 15, 2021
Recall Number Z-1065-2014
Recall Event ID 65451
510(K)Number K051932  
Product Classification Accelerator, linear, medical - Product Code IYE
Product All Elekta digital accelerators with electrons and beam modulator, MLCi/MLCi2, Agility or Asymmetric Heads


Product Usage: Deliver of radiation to defined target volumes
Code Information 151855, 151587, 152246, 151694, 152019, 105864, 151729, 105932, 151574, 152397, 152559, 152248, 151875, 153036, 153072, 151965, 151076, 151363, 152448, 105782, 151110, 105841, 105731, 105989, 152154, 151093, 151369, 105727, 151705, 151954, 152502, 152454, 152262, 151504, 105941, 151418, 105990, 151674, 105555, 151399, 105809, 151802, 105610, 151134, 151421, 105716, 105717, 152427, 105829, 153035, 105910, 152117, 152803, 151566, 151033, 152271, 151736, 151700, 151866, 151259, 151698, 151699, 151539, 151517, 151636, 151811, 105584, 151568, 152452, 151390, 152960, 151258, 151159, 152051, 153045, 151708, 152425, 151267, 152939, 152491, 105590, 151167, 105853, 105854, 152356, 152959, 151584, 151989, 105894, 151391, 151317, 151522, 105925, 151055, 151423, 151981, 151808, 152801, 152970, 151367, 151769, 151969, 153019, 153021, 151977, 151809, 152205, 151763, 151695, 151507, 151623, 151627, 105883, 105914, 151238, 152703, 151640, 152163, 151380, 151479, 152200, 152192, 151933, 152126, 152475, 152115, 152116, 152965, 151422, 152802, 152927, 152123, 105992, 105950, 151955, 151340, 153061, 151014, 151722, 152222, 151025, 151724, 151779, 152652, 152659, 152715, 152091, 105559, 105844, 151178, 152585, 105960, 151810, 151467, 151678, 151791, 152522, 151514, 152176, 151051, 151848, 151849, 151411, 153030, 151216, 151430, 151173, 151931, 151696, 152373, 151077, 152388, 151616, 151867, 151008, 151595, 152948, 151141, 151228, 152365, 151433, 151611, 152008, 105728, 152732, 151151, 152833, 151714, 105646, 152815, 152023, 151095, 105773, 105587, 105890, 152433, 152954, 151653, 152223, 152303, 152767, 152358, 151892, 151785, 105846, 152352, 152353, 152359, 152362, 152881, 151515, 105828, 151360, 151704, 151765, 151325, 152618, 105591, 151035, 151059, 151861, 152946, 105630, 151675, 152305, 152468, 152787, 151538, 152031, 152469, 152481, 152601, 153114, 105644, 151709, 151029, 105561, 151886, 151956, 105929, 151024, 151860, 151523, 151521, 105568, 105936, 151803, 153050, 105881, 105806, 151885, 105868, 151260, 151519, 151432, 151783, 152175, 152232, 152177, 152604, 151952, 152257, 152798, 151410, 153032, 105593, 151371, 151084, 151098, 152792, 105969, 151030, 151774, 152220, 151016, 152217, 105926, 105676, 105621, 151435, 151319, 151387, 151978, 151734, 152446, 105872, 151530, 105636, 151328, 152673, 151176, 152371, 151080, 152370, 151689, 105856, 105918, 151759, 151626, 151842, 151942, 151567, 151598, 151601, 151958, 151959, 151960, 152364, 151820, 152717, 105666, 151760, 105902, 105998, 105637, 151253, 151269, 151396, 151104, 151770, 151864, 105678, 105879, 151782, 152098, 151099, 151557, 151558, 151789, 152210, 152211, 105649, 151309, 152103, 152301, 105754, 105982, 151160, 151634, 152155, 152599, 152600, 152963, 151388, 151394, 151793, 105859, 105759, 151305, 105774, 151398, 151540, 151632, 151015, 152214, 152520, 152955, 105580, 151446, 151143, 152158, 151301, 151962, 152392, 105837, 151847, 105647, 105699, 105705, 105898, 151684, 152784, 105660, 151038, 152530, 151229, 152593, 151441, 152483, 152536, 105564, 105565, 153046, 105886, 151256, 152678, 153037, 151154, 151168, 151839, 151537, 151155, 151156, 151157, 151491, 151579, 105585, 152043, 151801, 105744, 105984, 151412, 105567, 151728, 153071, 151053, 151359, 105624, 152079, 152080, 105949, 152783, 152307, 152308, 105563, 105648, 105655, 151585, 105882, 151326, 151327, 151443, 152331, 151130, 151250, 151828, 151881, 151882, 151245, 105796, 105812, 152504, 105586, 151683, 151764, 105612, 151416, 105566, 152064, 151961, 152916, 151664, 151953, 105657, 151751, 152036, 151266, 152828, 152432, 151812, 151672, 151074, 152174, 105770, 105772, 105776, 151352, 151298, 105257, 105803, 151545, 151628, 152207, 152817, 152818, 105723, 151362, 151073, 153068, 152549, 151710     
Recalling Firm/
Manufacturer
Elekta, Inc.
400 Perimeter Center Ter Ne
Ste 50
Atlanta GA 30346-1227
For Additional Information Contact Linda Wetsel
404-300-9725
Manufacturer Reason
for Recall
Customers are resetting the default values outside recommended factory settings.
FDA Determined
Cause 2
Device Design
Action Elekta sent an Important Field Safety Notice dated May 2013 to all affected customers. The Notice identified the affected product, problem, clinical impact and actions to be taken. Customers were instructed to adhere to the instructions and advice given in manuals and system dialogues. For question contact your local Elekta office.
Quantity in Commerce 472 units
Distribution USA Nationwide Distribution in the states of: AL, AK, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, TN, TX, UT, VT, VA, WA, WV, WI, WY, DC and including Puerto Rico.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = IYE and Original Applicant = ELEKTA, INC
-
-