| | Class 2 Device Recall VITROS Chemistry Products CK Slides |  |
| Date Initiated by Firm | June 12, 2013 |
| Date Posted | November 15, 2013 |
| Recall Status1 |
Terminated 3 on April 11, 2017 |
| Recall Number | Z-0305-2014 |
| Recall Event ID |
65572 |
| 510(K)Number | K922072 |
| Product Classification |
Differential rate kinetic method, cpk or isoenzymes - Product Code JHS
|
| Product | VITROS Chemistry Products CK Slides packaged as 300 slides/pack catalog number 8479396 and packaged as 90 slides/pack catalog number 8478034
VITROS CK Slides quantitatively measure creatine kinase (CK) activity in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. |
| Code Information |
Multiple lots affected, with expiry dates through 22-Mar-2015 |
| FEI Number |
1000136573
|
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics 100 Indigo Creek Dr Rochester NY 14626-5101
|
Manufacturer Reason for Recall | Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues. |
FDA Determined Cause 2 | Equipment maintenance |
| Action | Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 12, 2013, to all affected customers via FedEx regarding Slide Dispense issues with various VITROS Chemistry Product Slides. VITROS ALT, CL-, GGT, LDH, K+, Na+ and TP slides use a different ABP, they are not affected by this issue. OCD provided instructions that may allow their customers to reposition the ABP if a jam occurs. OCD will provide you with a credit for any cartridge where the ABP becomes misaligned within the cartridge body in such a way that you are unable to reposition the ABP successfully to resolve the slide dispense issue. In addition to the Confirmation of Receipt form, OCD also enclosed three (3) copies of a Request for Credit form for you to complete when you require credit for affected cartridges. Complete and return the Confirmation of Receipt by June 21, 2013. If you experience slide dispense condition codes more than once a week, call our Customer Technical Support at 1-800-421-3311.
For questions regarding this recall call 585-453-4224. |
| Quantity in Commerce | 86,503 units (30,378 units domestically & 56,125 units internationally) |
| Distribution | Worldwide Distribution - USA (nationwide) including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY including Puerto Rico and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, England, Germany, France, Italy, and Spain. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|