 Class 2 Device Recall VITROS Chemistry Products UPRO Slidessee related information
Date Initiated by FirmJune 12, 2013
Date PostedNovember 15, 2013
Recall Status1 Terminated 3 on April 11, 2017
Recall NumberZ-0327-2014
Recall Event ID 65572
Product Classification Lowry (colorimetric), total protein - Product Code JGP
ProductVITROS Chemistry Products UPRO Slides packaged as 90 slides/pack catalog number 6800120 VITROS UPRO Slides quantitatively measure protein (UPRO) concentration in urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Code Information Multiple lots affected, with expiry dates through 22-Mar-2015
FEI Number 1000136573
Recalling Firm/
Manufacturer
Ortho-Clinical Diagnostics
100 Indigo Creek Dr
Rochester NY 14626-5101
Manufacturer Reason
for Recall
Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.
FDA Determined
Cause 2
Equipment maintenance
ActionOrtho Clinical Diagnostics sent an Important Product Correction Notification letter dated June 12, 2013, to all affected customers via FedEx regarding Slide Dispense issues with various VITROS Chemistry Product Slides. VITROS ALT, CL-, GGT, LDH, K+, Na+ and TP slides use a different ABP, they are not affected by this issue. OCD provided instructions that may allow their customers to reposition the ABP if a jam occurs. OCD will provide you with a credit for any cartridge where the ABP becomes misaligned within the cartridge body in such a way that you are unable to reposition the ABP successfully to resolve the slide dispense issue. In addition to the Confirmation of Receipt form, OCD also enclosed three (3) copies of a Request for Credit form for you to complete when you require credit for affected cartridges. Complete and return the Confirmation of Receipt by June 21, 2013. If you experience slide dispense condition codes more than once a week, call our Customer Technical Support at 1-800-421-3311. For questions regarding this recall call 585-453-4224.
Quantity in Commerce31,395 units (10,864 units domestically & 20,531 units internationally)
DistributionWorldwide Distribution - USA (nationwide) including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY including Puerto Rico and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, England, Germany, France, Italy, and Spain.