Date Initiated by Firm |
October 07, 2013 |
Date Posted |
January 28, 2014 |
Recall Status1 |
Terminated 3 on August 27, 2014 |
Recall Number |
Z-0851-2014 |
Recall Event ID |
66585 |
510(K)Number |
K051087
|
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product |
User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System with software version 3.5.1 or lower
User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. |
Code Information |
Material numbers 10284473, 10488224, 10444801, and 10444802. |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 500 Gbc Dr Ms 514 PO BOX 6101 Newark DE 19702-2466
|
For Additional Information Contact |
Customer Support 800-441-9250
|
Manufacturer Reason for Recall |
When utilizing the Routine Inventory screen to enter a User Defined Method (EMPTY) Flex the system may assign the User Defined Method Flex to a different Flex that is currently in inventory on the system, and then use the incorrect Flex Cartridge to process the user defined method.
|
FDA Determined Cause 2 |
Software design |
Action |
Siemens sent a Urgent Medical Device Correction letter dated October 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Please use the following actions if using Vista SW Versions 3.5.1 and below to prevent the issue from occurring:
1. When assigning a User Defined (EMPTY) Flex, use the Advanced Inventory Screen per the Operators Guide.
2. Verify that EMPTY Flex is no longer listed in the inventory before proceeding with any testing.
3. If the EMPTY Flex is still listed in the inventory, contact your Siemens Customer Care Center - Technical Solutions for assistance.
Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Discuss the content of this letter with your Medical Director.
We apologize for the inconvenience this situation has caused. If you have any questions, please contact your Siemens Customer Care Center - Technical Solutions at (800-441-9250) or your local Siemens technical support representative. |
Quantity in Commerce |
2038 |
Distribution |
USA (nationwide) Distribution including Puerto Rico. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JJE and Original Applicant = DADE BEHRING, INC.
|