| | Class 2 Device Recall Ziehm Imaging GmbH |  |
| Date Initiated by Firm | November 21, 2013 |
| Date Posted | December 24, 2013 |
| Recall Status1 |
Terminated 3 on June 23, 2014 |
| Recall Number | Z-0565-2014 |
| Recall Event ID |
66976 |
| 510(K)Number | K073346 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Ziehm Vision(2) C-Arm Interventional Fluoroscopic X-Ray System.
Mobile Fluoroscopic Imaging |
| Code Information |
9760 9937 90067 90816 9763 9945 90102 90817 9828 9946 90103 90819 9840 9983 90142 90823 9841 9985 90636 90838 9848 9986 90699 90842 9849 90024 90715 90863 9855 90031 90750 90864 9856 90032 90758 90865 9894 90033 90773 90899 9896 90034 90774 90909 9923 90054 90813 90919 9924 90065 90814 9936 90066 90815 |
| FEI Number |
2027299
|
Recalling Firm/ Manufacturer |
Ziehm Imaging Inc
6280 Hazeltine National Dr
Suite 100
Orlando
FL
32822-5114
|
| For Additional Information Contact | 47-615-8560 Ext. 140 |
Manufacturer Reason for Recall | Engineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices. |
FDA Determined Cause 2 | Device Design |
| Action | Ziehm Imaging sent an "URGENT RECALL NOTICE" letter dated November 22, 2013, to all affected customers. The letter identified the products the problem and the action needed to be taken by the customers.
Customers were advised and instructed to:
Your Model and Serial number can be found on the product label attached to the side of the mobile C-arm Stand. Our records indicate that your facility has one or more of the products listed above that are being used in your facility.
Recall of the product was initiated because according to our manufacturing documentation, the referenced device installed at your facility was not equipped with an emergency stopping device for vertical travel of the c-arm imaging assembly. To fully comply with applicable standards the referenced device must be retrofitted with an emergency stopping device identification and label.
The issue:
According to our manufacturing documentation, the referenced device installed at
your facility was not equipped with an emergency stopping device.
To fully comply with applicable standards the referenced device must be retrofitted with an emergency stopping device. The manufacturer Ziehm Imaging GmbH advised customers they may continue to operate the affected device with caution until the correction has been carried out on-site. During this time Ziehm Imaging GmbH advises the operator to be aware of the following safety
measures which may resolve the possibility of potentially hazardous situations by:
" Operating the power-off switch located either on the C-arm Stand top panel or on the monitor cart
" Removing the power cable from the wall outlet or,
" Activate the push-buttons for reverse vertical movement and keeping the pushbutton pressed until patient /operator is no longer in the risk zone or the device stops
Please alert all staff members working with the device to these potential hazardous situations, and familiarize them with all possibilities to resolve the hazardous situation |
| Quantity in Commerce | 54 devices |
| Distribution | USA (Nationwide) Distribution including the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MT, MA, MD, MI, MN, MN, MO,MS, NC ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Hawaii. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|
|
|
|