Class 2 Device Recall ADVIA Centaur Systems Phenytoin Calibrator N ( PHTN) 2pack |
|
Date Initiated by Firm |
December 18, 2013 |
Date Posted |
February 04, 2014 |
Recall Status1 |
Terminated 3 on May 18, 2016 |
Recall Number |
Z-0929-2014 |
Recall Event ID |
67021 |
510(K)Number |
K011393
|
Product Classification |
Enzyme immunoassay, diphenylhydantoin - Product Code DIP
|
Product |
ADVIA Centaur¿ Systems Phenytoin Calibrator N ( PHTN) 2-pack SMN 10311391
For in vitro diagnostic use in the quantitative determination of phenytoin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems |
Code Information |
Lot # Expiry Date 19002A58 06/09/2014; 19088A58 06/09/2014; 21376A58 06/09/2014; 25657A59 09/29/2014; 26830A59 09/29/2014; 28447A59 09/29/2014. |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc 333 Coney St East Walpole MA 02032-1516
|
For Additional Information Contact |
508-668-5000
|
Manufacturer Reason for Recall |
ADVIA Centaur¿ Systems Phenytoin Calibrator N,
lots CN58 and higher, change in recovery observed with the ADVIA Centaur Systems Phenytoin assay compared to recoveries using previous Calibrator N lots.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Siemens Healthcare Diagnostics Customer Notification notice was sent to affected Siemens Healthcare Diagnostic customers in the United States on December 18, 2013.. The notifications inform customers that Siemens Healthcare Diagnostics received customer feedback regarding the performance of the ADVIA Centara¿ Systems Phenytoin assay, indicating that the assay was recovering higher relative to the performance of other assays in proficiency studies. Siemens internal investigation confirmed the positive bias for Calibrator N lot CN 57 in the ADVIA Centaur Systems Phenytoin assay. This lot has expired and is no longer in use. Expected performance has been restored starting with Calibrator N lot CN58. Siemens is notifying customers that have received Calibrator N lots CN58 and higher that there will be a change in recovery observed with the ADVIA Centaur Systems Phenytoin assay compared to recoveries using previous Calibrator N lots. Siemens is providing data to inform customers of the expected difference between Calibrator lots CN58 and higher versus prior Calibrator N lots.
Customers were instructed to retain the letter with their laboratory records, and forward to those who may have received the affected product.
Customers with questions were instructed to contact their Siemens Customers Care Center or their local Siemens technical support representative.
For questions regarding this recall call 508-668-5000. |
Quantity in Commerce |
1294 (296 units US) |
Distribution |
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia,Brazil,Canada, Columbia, China, Ecuador, Egypt, India, Japan, Korea, Mexico, Saudi Arabia,Singapore, Thailand, and Uruguay. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = DIP and Original Applicant = RANDOX LABORATORIES, LTD.
|
|
|
|