|
Class 2 Device Recall IntelliFill i.v. Pharmacy Compounding Device (PCD) |
|
Date Initiated by Firm |
November 08, 2013 |
Date Posted |
December 29, 2013 |
Recall Status1 |
Terminated 3 on March 23, 2015 |
Recall Number |
Z-0607-2014 |
Recall Event ID |
66979 |
Product Classification |
System/device, pharmacy compounding - Product Code NEP
|
Product |
IntelliFill i.v. Pharmacy Compounding Device (PCD).
Intended to fill syringes from 0.5mL up to 11.5mL using mixing instructions from its formulary to fill the orders. |
Code Information |
Lot # 123, 125R, 127, 129, 145, 146R, 147R, 148, 215R, 500, 501, 502, 503, 504, 505, 506, 508, 510, 511R, 512, 514, 515, 516R, 517, 518, 519, and 520. |
Recalling Firm/ Manufacturer |
Baxter Healthcare Corp. 1 Baxter Pkwy Deerfield IL 60015-4625
|
For Additional Information Contact |
Center for One Baxter 224-948-2000
|
Manufacturer Reason for Recall |
Following a retrospective review for Baxter Daytona Complaint CR-954, it was identified that the IntelliFill i.v. Pharmacy Compounding Device (PCD) (product code: IFIV) has the potential to continue to run and fill syringes with the HEPA filtration disabled.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
A Service Bulletin was sent to all IntelliFill i. v. customers February 22, 20 I 0 by USPS mail.
The Service Bulletin informs customers that IntelliFill i.v. operators are to be trained to listen for the running HEPA filtration fan motor and observe/feel for airflow coming from the automation deck.
The field service engineer dispatched to the site replaced the damaged sensor on 11-Feb-2010 and a Service Bulletin was sent to all customers regarding the function and use of the HEPA filter on 22-Feb-2010.
A new HEPA sensor was designed and installed into all current IntelliFill i.v. devices between 21-July-2011 to 9-Feb-2012. |
Quantity in Commerce |
28 devices |
Distribution |
Nationwide Distribution including the states of CA, CO, CT, FL, IL, LA, MA, MD, MI, NC, NE, NY, OH, OR, PA, TN, TX, and WV. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|